A standardized validity assessment protocol for physiological signals from wearable technology: Methodological underpinnings and an application to the E4 biosensor
Wearable physiological measurement devices for ambulatory research with novel sensing technology are introduced with ever increasing frequency, requiring fast, standardized, and rigorous validation of the physiological signals measured by these devices and their derived parameters. At present, there...
Gespeichert in:
Veröffentlicht in: | Behavior Research Methods 2020-04, Vol.52 (2), p.607-629 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 629 |
---|---|
container_issue | 2 |
container_start_page | 607 |
container_title | Behavior Research Methods |
container_volume | 52 |
creator | van Lier, Hendrika G. Pieterse, Marcel E. Garde, Ainara Postel, Marloes G. de Haan, Hein A. Vollenbroek-Hutten, Miriam M. R. Schraagen, Jan Maarten Noordzij, Matthijs L. |
description | Wearable physiological measurement devices for ambulatory research with novel sensing technology are introduced with ever increasing frequency, requiring fast, standardized, and rigorous validation of the physiological signals measured by these devices and their derived parameters. At present, there is a lack of consensus on a standardized protocol or framework with which to test the validity of this new technology, leading to the use of various (often unfit) methods. This study introduces a comprehensive validity assessment protocol for physiological signals (electrodermal activity and cardiovascular activity) and investigates the validity of the E4 wearable (an example of such a new device) on the three levels proposed by the protocol: (1) the signal level, with a cross-correlation; (2) the parameter level, with Bland–Altman plots; and (3) the event level, with the detection of physiological changes due to external stressor levels via event difference plots. The results of the protocol show that the E4 wearable is valid for heart rate, RMSSD, and
SD
at the parameter and event levels, and for the total amplitude of skin conductance responses at the event level when studying strong sustained stressors. These findings are in line with the prior literature and demonstrate the applicability of the protocol. The validity assessment protocol proposed in this study provides a comprehensive, standardized, and feasible method for assessment of the quality of physiological data coming from new wearable (sensor) technology aimed at ambulatory research. |
doi_str_mv | 10.3758/s13428-019-01263-9 |
format | Article |
fullrecord | <record><control><sourceid>gale_pubme</sourceid><recordid>TN_cdi_pubmed_primary_31290128</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><galeid>A714174755</galeid><sourcerecordid>A714174755</sourcerecordid><originalsourceid>FETCH-LOGICAL-c541t-4e3629334d35f31a840f7e9495b7f25eb450f0abd97868665cda01d934022ef23</originalsourceid><addsrcrecordid>eNqNktuK1TAUhosozjj6Al5IwBtBOubYpl4MDJvxACPeKHgX0nalO0Ob1CSdYfs6vqjZ03F7uBAJISH5_rWStf6ieErwKauFfBUJ41SWmDR50oqVzb3imAjBSyaovH_Yky9HxaMYrzBmkhL-sDhihDZZIo-L7-coJu16HXr7DXp0rUfb27RDOkaIcQKX0Bx88p0fkfEBzdtdtH70g-30iKIdnB4jMsFP6AZ00O0IKEG3dXtm9xp9gLT1_UGwuB7CbJ2zbogoJ84T6Xke822y3qHkUdoCuuCotT6Ciz48Lh6YnASe3K0nxec3F58278rLj2_fb84vy05wkkoOrKINY7xnwjCiJcemhoY3oq0NFdBygQ3Wbd_UspJVJbpeY9I3jGNKwVB2UpytceelnaDv8t-DHtUc7KTDTnlt1Z83zm7V4K9VTbikch_gxV2A4L8uEJOabOxgHLUDv0RFaW5IJXIPMvr8L_TKL2FfS0WZlIzIhrJMna7UoEdQ1hmf83Z59DDZzjswNp-f5_yk5rUQWUBXQRd8jAHM4fUEq71p1GoalU2jbk2jmix69vu_D5KfLsmAXIEbaL2JnQXXwQHDGAuaq1xVeUfIxqbbTm784lKWvvx_aabZSsdMuAHCr6r84_0_AKbw8eI</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2388318923</pqid></control><display><type>article</type><title>A standardized validity assessment protocol for physiological signals from wearable technology: Methodological underpinnings and an application to the E4 biosensor</title><source>MEDLINE</source><source>SpringerNature Journals</source><source>Web of Science - Social Sciences Citation Index – 2020<img src="https://exlibris-pub.s3.amazonaws.com/fromwos-v2.jpg" /></source><creator>van Lier, Hendrika G. ; Pieterse, Marcel E. ; Garde, Ainara ; Postel, Marloes G. ; de Haan, Hein A. ; Vollenbroek-Hutten, Miriam M. R. ; Schraagen, Jan Maarten ; Noordzij, Matthijs L.</creator><creatorcontrib>van Lier, Hendrika G. ; Pieterse, Marcel E. ; Garde, Ainara ; Postel, Marloes G. ; de Haan, Hein A. ; Vollenbroek-Hutten, Miriam M. R. ; Schraagen, Jan Maarten ; Noordzij, Matthijs L.</creatorcontrib><description>Wearable physiological measurement devices for ambulatory research with novel sensing technology are introduced with ever increasing frequency, requiring fast, standardized, and rigorous validation of the physiological signals measured by these devices and their derived parameters. At present, there is a lack of consensus on a standardized protocol or framework with which to test the validity of this new technology, leading to the use of various (often unfit) methods. This study introduces a comprehensive validity assessment protocol for physiological signals (electrodermal activity and cardiovascular activity) and investigates the validity of the E4 wearable (an example of such a new device) on the three levels proposed by the protocol: (1) the signal level, with a cross-correlation; (2) the parameter level, with Bland–Altman plots; and (3) the event level, with the detection of physiological changes due to external stressor levels via event difference plots. The results of the protocol show that the E4 wearable is valid for heart rate, RMSSD, and
SD
at the parameter and event levels, and for the total amplitude of skin conductance responses at the event level when studying strong sustained stressors. These findings are in line with the prior literature and demonstrate the applicability of the protocol. The validity assessment protocol proposed in this study provides a comprehensive, standardized, and feasible method for assessment of the quality of physiological data coming from new wearable (sensor) technology aimed at ambulatory research.</description><identifier>ISSN: 1554-351X</identifier><identifier>ISSN: 1554-3528</identifier><identifier>EISSN: 1554-3528</identifier><identifier>DOI: 10.3758/s13428-019-01263-9</identifier><identifier>PMID: 31290128</identifier><language>eng</language><publisher>New York: Springer US</publisher><subject>Behavioral Science and Psychology ; Biosensing Techniques ; Biosensors ; Cognitive Psychology ; Conductance ; Detectors ; Heart beat ; Heart Rate ; Innovations ; Physiology ; Psychology ; Psychology, Experimental ; Psychology, Mathematical ; Social Sciences ; Validity ; Wearable computers ; Wearable Electronic Devices</subject><ispartof>Behavior Research Methods, 2020-04, Vol.52 (2), p.607-629</ispartof><rights>The Author(s) 2019</rights><rights>COPYRIGHT 2020 Springer</rights><rights>The Author(s) 2019. This work is published under https://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>true</woscitedreferencessubscribed><woscitedreferencescount>76</woscitedreferencescount><woscitedreferencesoriginalsourcerecordid>wos000524956600011</woscitedreferencesoriginalsourcerecordid><citedby>FETCH-LOGICAL-c541t-4e3629334d35f31a840f7e9495b7f25eb450f0abd97868665cda01d934022ef23</citedby><cites>FETCH-LOGICAL-c541t-4e3629334d35f31a840f7e9495b7f25eb450f0abd97868665cda01d934022ef23</cites><orcidid>0000-0001-8730-1487</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://link.springer.com/content/pdf/10.3758/s13428-019-01263-9$$EPDF$$P50$$Gspringer$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://link.springer.com/10.3758/s13428-019-01263-9$$EHTML$$P50$$Gspringer$$Hfree_for_read</linktohtml><link.rule.ids>230,315,782,786,887,27931,27932,28256,41495,42564,51326</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/31290128$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>van Lier, Hendrika G.</creatorcontrib><creatorcontrib>Pieterse, Marcel E.</creatorcontrib><creatorcontrib>Garde, Ainara</creatorcontrib><creatorcontrib>Postel, Marloes G.</creatorcontrib><creatorcontrib>de Haan, Hein A.</creatorcontrib><creatorcontrib>Vollenbroek-Hutten, Miriam M. R.</creatorcontrib><creatorcontrib>Schraagen, Jan Maarten</creatorcontrib><creatorcontrib>Noordzij, Matthijs L.</creatorcontrib><title>A standardized validity assessment protocol for physiological signals from wearable technology: Methodological underpinnings and an application to the E4 biosensor</title><title>Behavior Research Methods</title><addtitle>Behav Res</addtitle><addtitle>BEHAV RES METHODS</addtitle><addtitle>Behav Res Methods</addtitle><description>Wearable physiological measurement devices for ambulatory research with novel sensing technology are introduced with ever increasing frequency, requiring fast, standardized, and rigorous validation of the physiological signals measured by these devices and their derived parameters. At present, there is a lack of consensus on a standardized protocol or framework with which to test the validity of this new technology, leading to the use of various (often unfit) methods. This study introduces a comprehensive validity assessment protocol for physiological signals (electrodermal activity and cardiovascular activity) and investigates the validity of the E4 wearable (an example of such a new device) on the three levels proposed by the protocol: (1) the signal level, with a cross-correlation; (2) the parameter level, with Bland–Altman plots; and (3) the event level, with the detection of physiological changes due to external stressor levels via event difference plots. The results of the protocol show that the E4 wearable is valid for heart rate, RMSSD, and
SD
at the parameter and event levels, and for the total amplitude of skin conductance responses at the event level when studying strong sustained stressors. These findings are in line with the prior literature and demonstrate the applicability of the protocol. The validity assessment protocol proposed in this study provides a comprehensive, standardized, and feasible method for assessment of the quality of physiological data coming from new wearable (sensor) technology aimed at ambulatory research.</description><subject>Behavioral Science and Psychology</subject><subject>Biosensing Techniques</subject><subject>Biosensors</subject><subject>Cognitive Psychology</subject><subject>Conductance</subject><subject>Detectors</subject><subject>Heart beat</subject><subject>Heart Rate</subject><subject>Innovations</subject><subject>Physiology</subject><subject>Psychology</subject><subject>Psychology, Experimental</subject><subject>Psychology, Mathematical</subject><subject>Social Sciences</subject><subject>Validity</subject><subject>Wearable computers</subject><subject>Wearable Electronic Devices</subject><issn>1554-351X</issn><issn>1554-3528</issn><issn>1554-3528</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2020</creationdate><recordtype>article</recordtype><sourceid>C6C</sourceid><sourceid>ARHDP</sourceid><sourceid>EIF</sourceid><recordid>eNqNktuK1TAUhosozjj6Al5IwBtBOubYpl4MDJvxACPeKHgX0nalO0Ob1CSdYfs6vqjZ03F7uBAJISH5_rWStf6ieErwKauFfBUJ41SWmDR50oqVzb3imAjBSyaovH_Yky9HxaMYrzBmkhL-sDhihDZZIo-L7-coJu16HXr7DXp0rUfb27RDOkaIcQKX0Bx88p0fkfEBzdtdtH70g-30iKIdnB4jMsFP6AZ00O0IKEG3dXtm9xp9gLT1_UGwuB7CbJ2zbogoJ84T6Xke822y3qHkUdoCuuCotT6Ciz48Lh6YnASe3K0nxec3F58278rLj2_fb84vy05wkkoOrKINY7xnwjCiJcemhoY3oq0NFdBygQ3Wbd_UspJVJbpeY9I3jGNKwVB2UpytceelnaDv8t-DHtUc7KTDTnlt1Z83zm7V4K9VTbikch_gxV2A4L8uEJOabOxgHLUDv0RFaW5IJXIPMvr8L_TKL2FfS0WZlIzIhrJMna7UoEdQ1hmf83Z59DDZzjswNp-f5_yk5rUQWUBXQRd8jAHM4fUEq71p1GoalU2jbk2jmix69vu_D5KfLsmAXIEbaL2JnQXXwQHDGAuaq1xVeUfIxqbbTm784lKWvvx_aabZSsdMuAHCr6r84_0_AKbw8eI</recordid><startdate>20200401</startdate><enddate>20200401</enddate><creator>van Lier, Hendrika G.</creator><creator>Pieterse, Marcel E.</creator><creator>Garde, Ainara</creator><creator>Postel, Marloes G.</creator><creator>de Haan, Hein A.</creator><creator>Vollenbroek-Hutten, Miriam M. R.</creator><creator>Schraagen, Jan Maarten</creator><creator>Noordzij, Matthijs L.</creator><general>Springer US</general><general>Springer Nature</general><general>Springer</general><general>Springer Nature B.V</general><scope>C6C</scope><scope>17B</scope><scope>ARHDP</scope><scope>BLEPL</scope><scope>DVR</scope><scope>EGQ</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>IAO</scope><scope>4T-</scope><scope>7TK</scope><scope>K9.</scope><scope>7X8</scope><scope>5PM</scope><orcidid>https://orcid.org/0000-0001-8730-1487</orcidid></search><sort><creationdate>20200401</creationdate><title>A standardized validity assessment protocol for physiological signals from wearable technology: Methodological underpinnings and an application to the E4 biosensor</title><author>van Lier, Hendrika G. ; Pieterse, Marcel E. ; Garde, Ainara ; Postel, Marloes G. ; de Haan, Hein A. ; Vollenbroek-Hutten, Miriam M. R. ; Schraagen, Jan Maarten ; Noordzij, Matthijs L.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c541t-4e3629334d35f31a840f7e9495b7f25eb450f0abd97868665cda01d934022ef23</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2020</creationdate><topic>Behavioral Science and Psychology</topic><topic>Biosensing Techniques</topic><topic>Biosensors</topic><topic>Cognitive Psychology</topic><topic>Conductance</topic><topic>Detectors</topic><topic>Heart beat</topic><topic>Heart Rate</topic><topic>Innovations</topic><topic>Physiology</topic><topic>Psychology</topic><topic>Psychology, Experimental</topic><topic>Psychology, Mathematical</topic><topic>Social Sciences</topic><topic>Validity</topic><topic>Wearable computers</topic><topic>Wearable Electronic Devices</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>van Lier, Hendrika G.</creatorcontrib><creatorcontrib>Pieterse, Marcel E.</creatorcontrib><creatorcontrib>Garde, Ainara</creatorcontrib><creatorcontrib>Postel, Marloes G.</creatorcontrib><creatorcontrib>de Haan, Hein A.</creatorcontrib><creatorcontrib>Vollenbroek-Hutten, Miriam M. R.</creatorcontrib><creatorcontrib>Schraagen, Jan Maarten</creatorcontrib><creatorcontrib>Noordzij, Matthijs L.</creatorcontrib><collection>Springer Nature OA/Free Journals</collection><collection>Web of Knowledge</collection><collection>Web of Science - Social Sciences Citation Index – 2020</collection><collection>Web of Science Core Collection</collection><collection>Social Sciences Citation Index</collection><collection>Web of Science Primary (SCIE, SSCI & AHCI)</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>Gale Academic OneFile</collection><collection>Docstoc</collection><collection>Neurosciences Abstracts</collection><collection>ProQuest Health & Medical Complete (Alumni)</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><jtitle>Behavior Research Methods</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>van Lier, Hendrika G.</au><au>Pieterse, Marcel E.</au><au>Garde, Ainara</au><au>Postel, Marloes G.</au><au>de Haan, Hein A.</au><au>Vollenbroek-Hutten, Miriam M. R.</au><au>Schraagen, Jan Maarten</au><au>Noordzij, Matthijs L.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>A standardized validity assessment protocol for physiological signals from wearable technology: Methodological underpinnings and an application to the E4 biosensor</atitle><jtitle>Behavior Research Methods</jtitle><stitle>Behav Res</stitle><stitle>BEHAV RES METHODS</stitle><addtitle>Behav Res Methods</addtitle><date>2020-04-01</date><risdate>2020</risdate><volume>52</volume><issue>2</issue><spage>607</spage><epage>629</epage><pages>607-629</pages><issn>1554-351X</issn><issn>1554-3528</issn><eissn>1554-3528</eissn><abstract>Wearable physiological measurement devices for ambulatory research with novel sensing technology are introduced with ever increasing frequency, requiring fast, standardized, and rigorous validation of the physiological signals measured by these devices and their derived parameters. At present, there is a lack of consensus on a standardized protocol or framework with which to test the validity of this new technology, leading to the use of various (often unfit) methods. This study introduces a comprehensive validity assessment protocol for physiological signals (electrodermal activity and cardiovascular activity) and investigates the validity of the E4 wearable (an example of such a new device) on the three levels proposed by the protocol: (1) the signal level, with a cross-correlation; (2) the parameter level, with Bland–Altman plots; and (3) the event level, with the detection of physiological changes due to external stressor levels via event difference plots. The results of the protocol show that the E4 wearable is valid for heart rate, RMSSD, and
SD
at the parameter and event levels, and for the total amplitude of skin conductance responses at the event level when studying strong sustained stressors. These findings are in line with the prior literature and demonstrate the applicability of the protocol. The validity assessment protocol proposed in this study provides a comprehensive, standardized, and feasible method for assessment of the quality of physiological data coming from new wearable (sensor) technology aimed at ambulatory research.</abstract><cop>New York</cop><pub>Springer US</pub><pmid>31290128</pmid><doi>10.3758/s13428-019-01263-9</doi><tpages>23</tpages><orcidid>https://orcid.org/0000-0001-8730-1487</orcidid><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1554-351X |
ispartof | Behavior Research Methods, 2020-04, Vol.52 (2), p.607-629 |
issn | 1554-351X 1554-3528 1554-3528 |
language | eng |
recordid | cdi_pubmed_primary_31290128 |
source | MEDLINE; SpringerNature Journals; Web of Science - Social Sciences Citation Index – 2020<img src="https://exlibris-pub.s3.amazonaws.com/fromwos-v2.jpg" /> |
subjects | Behavioral Science and Psychology Biosensing Techniques Biosensors Cognitive Psychology Conductance Detectors Heart beat Heart Rate Innovations Physiology Psychology Psychology, Experimental Psychology, Mathematical Social Sciences Validity Wearable computers Wearable Electronic Devices |
title | A standardized validity assessment protocol for physiological signals from wearable technology: Methodological underpinnings and an application to the E4 biosensor |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-05T14%3A28%3A45IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-gale_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=A%20standardized%20validity%20assessment%20protocol%20for%20physiological%20signals%20from%20wearable%20technology:%20Methodological%20underpinnings%20and%20an%20application%20to%20the%20E4%20biosensor&rft.jtitle=Behavior%20Research%20Methods&rft.au=van%20Lier,%20Hendrika%20G.&rft.date=2020-04-01&rft.volume=52&rft.issue=2&rft.spage=607&rft.epage=629&rft.pages=607-629&rft.issn=1554-351X&rft.eissn=1554-3528&rft_id=info:doi/10.3758/s13428-019-01263-9&rft_dat=%3Cgale_pubme%3EA714174755%3C/gale_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2388318923&rft_id=info:pmid/31290128&rft_galeid=A714174755&rfr_iscdi=true |