A standardized validity assessment protocol for physiological signals from wearable technology: Methodological underpinnings and an application to the E4 biosensor

Wearable physiological measurement devices for ambulatory research with novel sensing technology are introduced with ever increasing frequency, requiring fast, standardized, and rigorous validation of the physiological signals measured by these devices and their derived parameters. At present, there...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Behavior Research Methods 2020-04, Vol.52 (2), p.607-629
Hauptverfasser: van Lier, Hendrika G., Pieterse, Marcel E., Garde, Ainara, Postel, Marloes G., de Haan, Hein A., Vollenbroek-Hutten, Miriam M. R., Schraagen, Jan Maarten, Noordzij, Matthijs L.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 629
container_issue 2
container_start_page 607
container_title Behavior Research Methods
container_volume 52
creator van Lier, Hendrika G.
Pieterse, Marcel E.
Garde, Ainara
Postel, Marloes G.
de Haan, Hein A.
Vollenbroek-Hutten, Miriam M. R.
Schraagen, Jan Maarten
Noordzij, Matthijs L.
description Wearable physiological measurement devices for ambulatory research with novel sensing technology are introduced with ever increasing frequency, requiring fast, standardized, and rigorous validation of the physiological signals measured by these devices and their derived parameters. At present, there is a lack of consensus on a standardized protocol or framework with which to test the validity of this new technology, leading to the use of various (often unfit) methods. This study introduces a comprehensive validity assessment protocol for physiological signals (electrodermal activity and cardiovascular activity) and investigates the validity of the E4 wearable (an example of such a new device) on the three levels proposed by the protocol: (1) the signal level, with a cross-correlation; (2) the parameter level, with Bland–Altman plots; and (3) the event level, with the detection of physiological changes due to external stressor levels via event difference plots. The results of the protocol show that the E4 wearable is valid for heart rate, RMSSD, and SD at the parameter and event levels, and for the total amplitude of skin conductance responses at the event level when studying strong sustained stressors. These findings are in line with the prior literature and demonstrate the applicability of the protocol. The validity assessment protocol proposed in this study provides a comprehensive, standardized, and feasible method for assessment of the quality of physiological data coming from new wearable (sensor) technology aimed at ambulatory research.
doi_str_mv 10.3758/s13428-019-01263-9
format Article
fullrecord <record><control><sourceid>gale_pubme</sourceid><recordid>TN_cdi_pubmed_primary_31290128</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><galeid>A714174755</galeid><sourcerecordid>A714174755</sourcerecordid><originalsourceid>FETCH-LOGICAL-c541t-4e3629334d35f31a840f7e9495b7f25eb450f0abd97868665cda01d934022ef23</originalsourceid><addsrcrecordid>eNqNktuK1TAUhosozjj6Al5IwBtBOubYpl4MDJvxACPeKHgX0nalO0Ob1CSdYfs6vqjZ03F7uBAJISH5_rWStf6ieErwKauFfBUJ41SWmDR50oqVzb3imAjBSyaovH_Yky9HxaMYrzBmkhL-sDhihDZZIo-L7-coJu16HXr7DXp0rUfb27RDOkaIcQKX0Bx88p0fkfEBzdtdtH70g-30iKIdnB4jMsFP6AZ00O0IKEG3dXtm9xp9gLT1_UGwuB7CbJ2zbogoJ84T6Xke822y3qHkUdoCuuCotT6Ciz48Lh6YnASe3K0nxec3F58278rLj2_fb84vy05wkkoOrKINY7xnwjCiJcemhoY3oq0NFdBygQ3Wbd_UspJVJbpeY9I3jGNKwVB2UpytceelnaDv8t-DHtUc7KTDTnlt1Z83zm7V4K9VTbikch_gxV2A4L8uEJOabOxgHLUDv0RFaW5IJXIPMvr8L_TKL2FfS0WZlIzIhrJMna7UoEdQ1hmf83Z59DDZzjswNp-f5_yk5rUQWUBXQRd8jAHM4fUEq71p1GoalU2jbk2jmix69vu_D5KfLsmAXIEbaL2JnQXXwQHDGAuaq1xVeUfIxqbbTm784lKWvvx_aabZSsdMuAHCr6r84_0_AKbw8eI</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2388318923</pqid></control><display><type>article</type><title>A standardized validity assessment protocol for physiological signals from wearable technology: Methodological underpinnings and an application to the E4 biosensor</title><source>MEDLINE</source><source>SpringerNature Journals</source><source>Web of Science - Social Sciences Citation Index – 2020&lt;img src="https://exlibris-pub.s3.amazonaws.com/fromwos-v2.jpg" /&gt;</source><creator>van Lier, Hendrika G. ; Pieterse, Marcel E. ; Garde, Ainara ; Postel, Marloes G. ; de Haan, Hein A. ; Vollenbroek-Hutten, Miriam M. R. ; Schraagen, Jan Maarten ; Noordzij, Matthijs L.</creator><creatorcontrib>van Lier, Hendrika G. ; Pieterse, Marcel E. ; Garde, Ainara ; Postel, Marloes G. ; de Haan, Hein A. ; Vollenbroek-Hutten, Miriam M. R. ; Schraagen, Jan Maarten ; Noordzij, Matthijs L.</creatorcontrib><description>Wearable physiological measurement devices for ambulatory research with novel sensing technology are introduced with ever increasing frequency, requiring fast, standardized, and rigorous validation of the physiological signals measured by these devices and their derived parameters. At present, there is a lack of consensus on a standardized protocol or framework with which to test the validity of this new technology, leading to the use of various (often unfit) methods. This study introduces a comprehensive validity assessment protocol for physiological signals (electrodermal activity and cardiovascular activity) and investigates the validity of the E4 wearable (an example of such a new device) on the three levels proposed by the protocol: (1) the signal level, with a cross-correlation; (2) the parameter level, with Bland–Altman plots; and (3) the event level, with the detection of physiological changes due to external stressor levels via event difference plots. The results of the protocol show that the E4 wearable is valid for heart rate, RMSSD, and SD at the parameter and event levels, and for the total amplitude of skin conductance responses at the event level when studying strong sustained stressors. These findings are in line with the prior literature and demonstrate the applicability of the protocol. The validity assessment protocol proposed in this study provides a comprehensive, standardized, and feasible method for assessment of the quality of physiological data coming from new wearable (sensor) technology aimed at ambulatory research.</description><identifier>ISSN: 1554-351X</identifier><identifier>ISSN: 1554-3528</identifier><identifier>EISSN: 1554-3528</identifier><identifier>DOI: 10.3758/s13428-019-01263-9</identifier><identifier>PMID: 31290128</identifier><language>eng</language><publisher>New York: Springer US</publisher><subject>Behavioral Science and Psychology ; Biosensing Techniques ; Biosensors ; Cognitive Psychology ; Conductance ; Detectors ; Heart beat ; Heart Rate ; Innovations ; Physiology ; Psychology ; Psychology, Experimental ; Psychology, Mathematical ; Social Sciences ; Validity ; Wearable computers ; Wearable Electronic Devices</subject><ispartof>Behavior Research Methods, 2020-04, Vol.52 (2), p.607-629</ispartof><rights>The Author(s) 2019</rights><rights>COPYRIGHT 2020 Springer</rights><rights>The Author(s) 2019. This work is published under https://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>true</woscitedreferencessubscribed><woscitedreferencescount>76</woscitedreferencescount><woscitedreferencesoriginalsourcerecordid>wos000524956600011</woscitedreferencesoriginalsourcerecordid><citedby>FETCH-LOGICAL-c541t-4e3629334d35f31a840f7e9495b7f25eb450f0abd97868665cda01d934022ef23</citedby><cites>FETCH-LOGICAL-c541t-4e3629334d35f31a840f7e9495b7f25eb450f0abd97868665cda01d934022ef23</cites><orcidid>0000-0001-8730-1487</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://link.springer.com/content/pdf/10.3758/s13428-019-01263-9$$EPDF$$P50$$Gspringer$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://link.springer.com/10.3758/s13428-019-01263-9$$EHTML$$P50$$Gspringer$$Hfree_for_read</linktohtml><link.rule.ids>230,315,782,786,887,27931,27932,28256,41495,42564,51326</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/31290128$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>van Lier, Hendrika G.</creatorcontrib><creatorcontrib>Pieterse, Marcel E.</creatorcontrib><creatorcontrib>Garde, Ainara</creatorcontrib><creatorcontrib>Postel, Marloes G.</creatorcontrib><creatorcontrib>de Haan, Hein A.</creatorcontrib><creatorcontrib>Vollenbroek-Hutten, Miriam M. R.</creatorcontrib><creatorcontrib>Schraagen, Jan Maarten</creatorcontrib><creatorcontrib>Noordzij, Matthijs L.</creatorcontrib><title>A standardized validity assessment protocol for physiological signals from wearable technology: Methodological underpinnings and an application to the E4 biosensor</title><title>Behavior Research Methods</title><addtitle>Behav Res</addtitle><addtitle>BEHAV RES METHODS</addtitle><addtitle>Behav Res Methods</addtitle><description>Wearable physiological measurement devices for ambulatory research with novel sensing technology are introduced with ever increasing frequency, requiring fast, standardized, and rigorous validation of the physiological signals measured by these devices and their derived parameters. At present, there is a lack of consensus on a standardized protocol or framework with which to test the validity of this new technology, leading to the use of various (often unfit) methods. This study introduces a comprehensive validity assessment protocol for physiological signals (electrodermal activity and cardiovascular activity) and investigates the validity of the E4 wearable (an example of such a new device) on the three levels proposed by the protocol: (1) the signal level, with a cross-correlation; (2) the parameter level, with Bland–Altman plots; and (3) the event level, with the detection of physiological changes due to external stressor levels via event difference plots. The results of the protocol show that the E4 wearable is valid for heart rate, RMSSD, and SD at the parameter and event levels, and for the total amplitude of skin conductance responses at the event level when studying strong sustained stressors. These findings are in line with the prior literature and demonstrate the applicability of the protocol. The validity assessment protocol proposed in this study provides a comprehensive, standardized, and feasible method for assessment of the quality of physiological data coming from new wearable (sensor) technology aimed at ambulatory research.</description><subject>Behavioral Science and Psychology</subject><subject>Biosensing Techniques</subject><subject>Biosensors</subject><subject>Cognitive Psychology</subject><subject>Conductance</subject><subject>Detectors</subject><subject>Heart beat</subject><subject>Heart Rate</subject><subject>Innovations</subject><subject>Physiology</subject><subject>Psychology</subject><subject>Psychology, Experimental</subject><subject>Psychology, Mathematical</subject><subject>Social Sciences</subject><subject>Validity</subject><subject>Wearable computers</subject><subject>Wearable Electronic Devices</subject><issn>1554-351X</issn><issn>1554-3528</issn><issn>1554-3528</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2020</creationdate><recordtype>article</recordtype><sourceid>C6C</sourceid><sourceid>ARHDP</sourceid><sourceid>EIF</sourceid><recordid>eNqNktuK1TAUhosozjj6Al5IwBtBOubYpl4MDJvxACPeKHgX0nalO0Ob1CSdYfs6vqjZ03F7uBAJISH5_rWStf6ieErwKauFfBUJ41SWmDR50oqVzb3imAjBSyaovH_Yky9HxaMYrzBmkhL-sDhihDZZIo-L7-coJu16HXr7DXp0rUfb27RDOkaIcQKX0Bx88p0fkfEBzdtdtH70g-30iKIdnB4jMsFP6AZ00O0IKEG3dXtm9xp9gLT1_UGwuB7CbJ2zbogoJ84T6Xke822y3qHkUdoCuuCotT6Ciz48Lh6YnASe3K0nxec3F58278rLj2_fb84vy05wkkoOrKINY7xnwjCiJcemhoY3oq0NFdBygQ3Wbd_UspJVJbpeY9I3jGNKwVB2UpytceelnaDv8t-DHtUc7KTDTnlt1Z83zm7V4K9VTbikch_gxV2A4L8uEJOabOxgHLUDv0RFaW5IJXIPMvr8L_TKL2FfS0WZlIzIhrJMna7UoEdQ1hmf83Z59DDZzjswNp-f5_yk5rUQWUBXQRd8jAHM4fUEq71p1GoalU2jbk2jmix69vu_D5KfLsmAXIEbaL2JnQXXwQHDGAuaq1xVeUfIxqbbTm784lKWvvx_aabZSsdMuAHCr6r84_0_AKbw8eI</recordid><startdate>20200401</startdate><enddate>20200401</enddate><creator>van Lier, Hendrika G.</creator><creator>Pieterse, Marcel E.</creator><creator>Garde, Ainara</creator><creator>Postel, Marloes G.</creator><creator>de Haan, Hein A.</creator><creator>Vollenbroek-Hutten, Miriam M. R.</creator><creator>Schraagen, Jan Maarten</creator><creator>Noordzij, Matthijs L.</creator><general>Springer US</general><general>Springer Nature</general><general>Springer</general><general>Springer Nature B.V</general><scope>C6C</scope><scope>17B</scope><scope>ARHDP</scope><scope>BLEPL</scope><scope>DVR</scope><scope>EGQ</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>IAO</scope><scope>4T-</scope><scope>7TK</scope><scope>K9.</scope><scope>7X8</scope><scope>5PM</scope><orcidid>https://orcid.org/0000-0001-8730-1487</orcidid></search><sort><creationdate>20200401</creationdate><title>A standardized validity assessment protocol for physiological signals from wearable technology: Methodological underpinnings and an application to the E4 biosensor</title><author>van Lier, Hendrika G. ; Pieterse, Marcel E. ; Garde, Ainara ; Postel, Marloes G. ; de Haan, Hein A. ; Vollenbroek-Hutten, Miriam M. R. ; Schraagen, Jan Maarten ; Noordzij, Matthijs L.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c541t-4e3629334d35f31a840f7e9495b7f25eb450f0abd97868665cda01d934022ef23</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2020</creationdate><topic>Behavioral Science and Psychology</topic><topic>Biosensing Techniques</topic><topic>Biosensors</topic><topic>Cognitive Psychology</topic><topic>Conductance</topic><topic>Detectors</topic><topic>Heart beat</topic><topic>Heart Rate</topic><topic>Innovations</topic><topic>Physiology</topic><topic>Psychology</topic><topic>Psychology, Experimental</topic><topic>Psychology, Mathematical</topic><topic>Social Sciences</topic><topic>Validity</topic><topic>Wearable computers</topic><topic>Wearable Electronic Devices</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>van Lier, Hendrika G.</creatorcontrib><creatorcontrib>Pieterse, Marcel E.</creatorcontrib><creatorcontrib>Garde, Ainara</creatorcontrib><creatorcontrib>Postel, Marloes G.</creatorcontrib><creatorcontrib>de Haan, Hein A.</creatorcontrib><creatorcontrib>Vollenbroek-Hutten, Miriam M. R.</creatorcontrib><creatorcontrib>Schraagen, Jan Maarten</creatorcontrib><creatorcontrib>Noordzij, Matthijs L.</creatorcontrib><collection>Springer Nature OA/Free Journals</collection><collection>Web of Knowledge</collection><collection>Web of Science - Social Sciences Citation Index – 2020</collection><collection>Web of Science Core Collection</collection><collection>Social Sciences Citation Index</collection><collection>Web of Science Primary (SCIE, SSCI &amp; AHCI)</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>Gale Academic OneFile</collection><collection>Docstoc</collection><collection>Neurosciences Abstracts</collection><collection>ProQuest Health &amp; Medical Complete (Alumni)</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><jtitle>Behavior Research Methods</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>van Lier, Hendrika G.</au><au>Pieterse, Marcel E.</au><au>Garde, Ainara</au><au>Postel, Marloes G.</au><au>de Haan, Hein A.</au><au>Vollenbroek-Hutten, Miriam M. R.</au><au>Schraagen, Jan Maarten</au><au>Noordzij, Matthijs L.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>A standardized validity assessment protocol for physiological signals from wearable technology: Methodological underpinnings and an application to the E4 biosensor</atitle><jtitle>Behavior Research Methods</jtitle><stitle>Behav Res</stitle><stitle>BEHAV RES METHODS</stitle><addtitle>Behav Res Methods</addtitle><date>2020-04-01</date><risdate>2020</risdate><volume>52</volume><issue>2</issue><spage>607</spage><epage>629</epage><pages>607-629</pages><issn>1554-351X</issn><issn>1554-3528</issn><eissn>1554-3528</eissn><abstract>Wearable physiological measurement devices for ambulatory research with novel sensing technology are introduced with ever increasing frequency, requiring fast, standardized, and rigorous validation of the physiological signals measured by these devices and their derived parameters. At present, there is a lack of consensus on a standardized protocol or framework with which to test the validity of this new technology, leading to the use of various (often unfit) methods. This study introduces a comprehensive validity assessment protocol for physiological signals (electrodermal activity and cardiovascular activity) and investigates the validity of the E4 wearable (an example of such a new device) on the three levels proposed by the protocol: (1) the signal level, with a cross-correlation; (2) the parameter level, with Bland–Altman plots; and (3) the event level, with the detection of physiological changes due to external stressor levels via event difference plots. The results of the protocol show that the E4 wearable is valid for heart rate, RMSSD, and SD at the parameter and event levels, and for the total amplitude of skin conductance responses at the event level when studying strong sustained stressors. These findings are in line with the prior literature and demonstrate the applicability of the protocol. The validity assessment protocol proposed in this study provides a comprehensive, standardized, and feasible method for assessment of the quality of physiological data coming from new wearable (sensor) technology aimed at ambulatory research.</abstract><cop>New York</cop><pub>Springer US</pub><pmid>31290128</pmid><doi>10.3758/s13428-019-01263-9</doi><tpages>23</tpages><orcidid>https://orcid.org/0000-0001-8730-1487</orcidid><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 1554-351X
ispartof Behavior Research Methods, 2020-04, Vol.52 (2), p.607-629
issn 1554-351X
1554-3528
1554-3528
language eng
recordid cdi_pubmed_primary_31290128
source MEDLINE; SpringerNature Journals; Web of Science - Social Sciences Citation Index – 2020<img src="https://exlibris-pub.s3.amazonaws.com/fromwos-v2.jpg" />
subjects Behavioral Science and Psychology
Biosensing Techniques
Biosensors
Cognitive Psychology
Conductance
Detectors
Heart beat
Heart Rate
Innovations
Physiology
Psychology
Psychology, Experimental
Psychology, Mathematical
Social Sciences
Validity
Wearable computers
Wearable Electronic Devices
title A standardized validity assessment protocol for physiological signals from wearable technology: Methodological underpinnings and an application to the E4 biosensor
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-05T14%3A28%3A45IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-gale_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=A%20standardized%20validity%20assessment%20protocol%20for%20physiological%20signals%20from%20wearable%20technology:%20Methodological%20underpinnings%20and%20an%20application%20to%20the%20E4%20biosensor&rft.jtitle=Behavior%20Research%20Methods&rft.au=van%20Lier,%20Hendrika%20G.&rft.date=2020-04-01&rft.volume=52&rft.issue=2&rft.spage=607&rft.epage=629&rft.pages=607-629&rft.issn=1554-351X&rft.eissn=1554-3528&rft_id=info:doi/10.3758/s13428-019-01263-9&rft_dat=%3Cgale_pubme%3EA714174755%3C/gale_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2388318923&rft_id=info:pmid/31290128&rft_galeid=A714174755&rfr_iscdi=true