Biopsy needle tips with markers—MR compatible needles for high-precision needle tip positioning

Needle tip visualization is of high importance in magnetic resonance imaging (MRI) guided interventional procedures, for example for taking biopsies from suspicious lesions in the liver or kidney. The exact position of the needle tip is often obscured by image artifacts arising from the magnetic pro...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical physics (Lancaster) 2008-06, Vol.35 (6), p.2273-2278
Hauptverfasser: Müller-Bierl, Bernd M., Martirosian, Petros, Graf, Hansjörg, Boss, Andreas, König, Claudius, Pereira, Philippe L., Schick, Fritz
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 2278
container_issue 6
container_start_page 2273
container_title Medical physics (Lancaster)
container_volume 35
creator Müller-Bierl, Bernd M.
Martirosian, Petros
Graf, Hansjörg
Boss, Andreas
König, Claudius
Pereira, Philippe L.
Schick, Fritz
description Needle tip visualization is of high importance in magnetic resonance imaging (MRI) guided interventional procedures, for example for taking biopsies from suspicious lesions in the liver or kidney. The exact position of the needle tip is often obscured by image artifacts arising from the magnetic properties of the needle. The authors investigated two special biopsy needle tip designs using diamagnetic coatings. For common interventional MR sequences, the needle tip can be identified in the MR image by several equidistant dark spots arranged along a straight line. A dotted instead of a solid line allows for an improved control of the movement of the needle, not only if the needle is tilted toward the imaging plane, but also if the needle leaves an empty canal with signal extinction, which cannot be distinguished from the needle material itself. With the proposed design the position of the needle tip can be estimated with a precision of approximately 1 mm using conventional FLASH, FISP, and TSE sequences, as used for interventional MR. Furthermore, the size of the biopsy probe can be estimated from the artifact. In using needles with a properly designed tip coating, taking biopsies under MR control is beginning to be greatly simplified. The approach to design artifacts using diamagnetic material in combination with paramagnetic material paves the way toward new instruments and implants, suitably tailored to the needs of the interventional radiologist.
doi_str_mv 10.1118/1.2919082
format Article
fullrecord <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_pubmed_primary_18649458</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>69338136</sourcerecordid><originalsourceid>FETCH-LOGICAL-c3912-1647fff6d30360bfaf3f16548d6b9636a5d8a96e845ae5d790b375ca6f287d883</originalsourceid><addsrcrecordid>eNp90N9KwzAUBvAgipvTC19AeiUodCZNmiaXOvwHG4rodUjbZIu2S006x-58CJ_QJ7GjRQeiV4HD73ycfAAcIjhECLEzNIw44pBFW6AfkQSHJIJ8G_Qh5CSMCIx7YM_7ZwghxTHcBT3EKOEkZn0gL4yt_CqYK5UXKqhN5YOlqWdBKd2Lcv7z_WPyEGS2rGRt0ka00AfaumBmprOwcioz3tj5RkZQWW_qZmbm032wo2Xh1UH3DsDT1eXj6CYc313fjs7HYYY5ikJESaK1pjmGmMJUS401ojFhOU05xVTGOZOcKkZiqeI84TDFSZxJqiOW5IzhAThucytnXxfK16I0PlNFIefKLrygHGOGMG3gSQszZ713SovKmea7K4GgWPcpkOj6bOxRF7pIS5X_yK7ABoQtWJpCrf5OEpP7LvC09T4ztVxX9L3zZt2Gr3L9H_596hd3Cps0</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>69338136</pqid></control><display><type>article</type><title>Biopsy needle tips with markers—MR compatible needles for high-precision needle tip positioning</title><source>MEDLINE</source><source>Wiley Online Library Journals Frontfile Complete</source><source>Alma/SFX Local Collection</source><creator>Müller-Bierl, Bernd M. ; Martirosian, Petros ; Graf, Hansjörg ; Boss, Andreas ; König, Claudius ; Pereira, Philippe L. ; Schick, Fritz</creator><creatorcontrib>Müller-Bierl, Bernd M. ; Martirosian, Petros ; Graf, Hansjörg ; Boss, Andreas ; König, Claudius ; Pereira, Philippe L. ; Schick, Fritz</creatorcontrib><description>Needle tip visualization is of high importance in magnetic resonance imaging (MRI) guided interventional procedures, for example for taking biopsies from suspicious lesions in the liver or kidney. The exact position of the needle tip is often obscured by image artifacts arising from the magnetic properties of the needle. The authors investigated two special biopsy needle tip designs using diamagnetic coatings. For common interventional MR sequences, the needle tip can be identified in the MR image by several equidistant dark spots arranged along a straight line. A dotted instead of a solid line allows for an improved control of the movement of the needle, not only if the needle is tilted toward the imaging plane, but also if the needle leaves an empty canal with signal extinction, which cannot be distinguished from the needle material itself. With the proposed design the position of the needle tip can be estimated with a precision of approximately 1 mm using conventional FLASH, FISP, and TSE sequences, as used for interventional MR. Furthermore, the size of the biopsy probe can be estimated from the artifact. In using needles with a properly designed tip coating, taking biopsies under MR control is beginning to be greatly simplified. The approach to design artifacts using diamagnetic material in combination with paramagnetic material paves the way toward new instruments and implants, suitably tailored to the needs of the interventional radiologist.</description><identifier>ISSN: 0094-2405</identifier><identifier>EISSN: 2473-4209</identifier><identifier>DOI: 10.1118/1.2919082</identifier><identifier>PMID: 18649458</identifier><identifier>CODEN: MPHYA6</identifier><language>eng</language><publisher>United States: American Association of Physicists in Medicine</publisher><subject>biomedical equipment ; Biomedical imaging ; biomedical MRI ; biopsy needles ; Biopsy, Needle ; Bismuth ; diamagnetic materials ; Diamagnetism ; Image scanners ; image sequences ; Instrumentation ; interventional MR imaging ; kidney ; liver ; Magnetic resonance imaging ; Magnetic Resonance Imaging - instrumentation ; Magnetics ; Medical image artifacts ; Medical imaging ; Medical magnetic resonance imaging ; MR artifacts ; MR compatibility ; Needles ; paramagnetic materials ; Prion diseases ; Sensitivity and Specificity ; Sequence analysis ; Titanium</subject><ispartof>Medical physics (Lancaster), 2008-06, Vol.35 (6), p.2273-2278</ispartof><rights>American Association of Physicists in Medicine</rights><rights>2008 American Association of Physicists in Medicine</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c3912-1647fff6d30360bfaf3f16548d6b9636a5d8a96e845ae5d790b375ca6f287d883</citedby><cites>FETCH-LOGICAL-c3912-1647fff6d30360bfaf3f16548d6b9636a5d8a96e845ae5d790b375ca6f287d883</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://onlinelibrary.wiley.com/doi/pdf/10.1118%2F1.2919082$$EPDF$$P50$$Gwiley$$H</linktopdf><linktohtml>$$Uhttps://onlinelibrary.wiley.com/doi/full/10.1118%2F1.2919082$$EHTML$$P50$$Gwiley$$H</linktohtml><link.rule.ids>314,776,780,1411,27901,27902,45550,45551</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/18649458$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Müller-Bierl, Bernd M.</creatorcontrib><creatorcontrib>Martirosian, Petros</creatorcontrib><creatorcontrib>Graf, Hansjörg</creatorcontrib><creatorcontrib>Boss, Andreas</creatorcontrib><creatorcontrib>König, Claudius</creatorcontrib><creatorcontrib>Pereira, Philippe L.</creatorcontrib><creatorcontrib>Schick, Fritz</creatorcontrib><title>Biopsy needle tips with markers—MR compatible needles for high-precision needle tip positioning</title><title>Medical physics (Lancaster)</title><addtitle>Med Phys</addtitle><description>Needle tip visualization is of high importance in magnetic resonance imaging (MRI) guided interventional procedures, for example for taking biopsies from suspicious lesions in the liver or kidney. The exact position of the needle tip is often obscured by image artifacts arising from the magnetic properties of the needle. The authors investigated two special biopsy needle tip designs using diamagnetic coatings. For common interventional MR sequences, the needle tip can be identified in the MR image by several equidistant dark spots arranged along a straight line. A dotted instead of a solid line allows for an improved control of the movement of the needle, not only if the needle is tilted toward the imaging plane, but also if the needle leaves an empty canal with signal extinction, which cannot be distinguished from the needle material itself. With the proposed design the position of the needle tip can be estimated with a precision of approximately 1 mm using conventional FLASH, FISP, and TSE sequences, as used for interventional MR. Furthermore, the size of the biopsy probe can be estimated from the artifact. In using needles with a properly designed tip coating, taking biopsies under MR control is beginning to be greatly simplified. The approach to design artifacts using diamagnetic material in combination with paramagnetic material paves the way toward new instruments and implants, suitably tailored to the needs of the interventional radiologist.</description><subject>biomedical equipment</subject><subject>Biomedical imaging</subject><subject>biomedical MRI</subject><subject>biopsy needles</subject><subject>Biopsy, Needle</subject><subject>Bismuth</subject><subject>diamagnetic materials</subject><subject>Diamagnetism</subject><subject>Image scanners</subject><subject>image sequences</subject><subject>Instrumentation</subject><subject>interventional MR imaging</subject><subject>kidney</subject><subject>liver</subject><subject>Magnetic resonance imaging</subject><subject>Magnetic Resonance Imaging - instrumentation</subject><subject>Magnetics</subject><subject>Medical image artifacts</subject><subject>Medical imaging</subject><subject>Medical magnetic resonance imaging</subject><subject>MR artifacts</subject><subject>MR compatibility</subject><subject>Needles</subject><subject>paramagnetic materials</subject><subject>Prion diseases</subject><subject>Sensitivity and Specificity</subject><subject>Sequence analysis</subject><subject>Titanium</subject><issn>0094-2405</issn><issn>2473-4209</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2008</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp90N9KwzAUBvAgipvTC19AeiUodCZNmiaXOvwHG4rodUjbZIu2S006x-58CJ_QJ7GjRQeiV4HD73ycfAAcIjhECLEzNIw44pBFW6AfkQSHJIJ8G_Qh5CSMCIx7YM_7ZwghxTHcBT3EKOEkZn0gL4yt_CqYK5UXKqhN5YOlqWdBKd2Lcv7z_WPyEGS2rGRt0ka00AfaumBmprOwcioz3tj5RkZQWW_qZmbm032wo2Xh1UH3DsDT1eXj6CYc313fjs7HYYY5ikJESaK1pjmGmMJUS401ojFhOU05xVTGOZOcKkZiqeI84TDFSZxJqiOW5IzhAThucytnXxfK16I0PlNFIefKLrygHGOGMG3gSQszZ713SovKmea7K4GgWPcpkOj6bOxRF7pIS5X_yK7ABoQtWJpCrf5OEpP7LvC09T4ztVxX9L3zZt2Gr3L9H_596hd3Cps0</recordid><startdate>200806</startdate><enddate>200806</enddate><creator>Müller-Bierl, Bernd M.</creator><creator>Martirosian, Petros</creator><creator>Graf, Hansjörg</creator><creator>Boss, Andreas</creator><creator>König, Claudius</creator><creator>Pereira, Philippe L.</creator><creator>Schick, Fritz</creator><general>American Association of Physicists in Medicine</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>200806</creationdate><title>Biopsy needle tips with markers—MR compatible needles for high-precision needle tip positioning</title><author>Müller-Bierl, Bernd M. ; Martirosian, Petros ; Graf, Hansjörg ; Boss, Andreas ; König, Claudius ; Pereira, Philippe L. ; Schick, Fritz</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c3912-1647fff6d30360bfaf3f16548d6b9636a5d8a96e845ae5d790b375ca6f287d883</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2008</creationdate><topic>biomedical equipment</topic><topic>Biomedical imaging</topic><topic>biomedical MRI</topic><topic>biopsy needles</topic><topic>Biopsy, Needle</topic><topic>Bismuth</topic><topic>diamagnetic materials</topic><topic>Diamagnetism</topic><topic>Image scanners</topic><topic>image sequences</topic><topic>Instrumentation</topic><topic>interventional MR imaging</topic><topic>kidney</topic><topic>liver</topic><topic>Magnetic resonance imaging</topic><topic>Magnetic Resonance Imaging - instrumentation</topic><topic>Magnetics</topic><topic>Medical image artifacts</topic><topic>Medical imaging</topic><topic>Medical magnetic resonance imaging</topic><topic>MR artifacts</topic><topic>MR compatibility</topic><topic>Needles</topic><topic>paramagnetic materials</topic><topic>Prion diseases</topic><topic>Sensitivity and Specificity</topic><topic>Sequence analysis</topic><topic>Titanium</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Müller-Bierl, Bernd M.</creatorcontrib><creatorcontrib>Martirosian, Petros</creatorcontrib><creatorcontrib>Graf, Hansjörg</creatorcontrib><creatorcontrib>Boss, Andreas</creatorcontrib><creatorcontrib>König, Claudius</creatorcontrib><creatorcontrib>Pereira, Philippe L.</creatorcontrib><creatorcontrib>Schick, Fritz</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Medical physics (Lancaster)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Müller-Bierl, Bernd M.</au><au>Martirosian, Petros</au><au>Graf, Hansjörg</au><au>Boss, Andreas</au><au>König, Claudius</au><au>Pereira, Philippe L.</au><au>Schick, Fritz</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Biopsy needle tips with markers—MR compatible needles for high-precision needle tip positioning</atitle><jtitle>Medical physics (Lancaster)</jtitle><addtitle>Med Phys</addtitle><date>2008-06</date><risdate>2008</risdate><volume>35</volume><issue>6</issue><spage>2273</spage><epage>2278</epage><pages>2273-2278</pages><issn>0094-2405</issn><eissn>2473-4209</eissn><coden>MPHYA6</coden><abstract>Needle tip visualization is of high importance in magnetic resonance imaging (MRI) guided interventional procedures, for example for taking biopsies from suspicious lesions in the liver or kidney. The exact position of the needle tip is often obscured by image artifacts arising from the magnetic properties of the needle. The authors investigated two special biopsy needle tip designs using diamagnetic coatings. For common interventional MR sequences, the needle tip can be identified in the MR image by several equidistant dark spots arranged along a straight line. A dotted instead of a solid line allows for an improved control of the movement of the needle, not only if the needle is tilted toward the imaging plane, but also if the needle leaves an empty canal with signal extinction, which cannot be distinguished from the needle material itself. With the proposed design the position of the needle tip can be estimated with a precision of approximately 1 mm using conventional FLASH, FISP, and TSE sequences, as used for interventional MR. Furthermore, the size of the biopsy probe can be estimated from the artifact. In using needles with a properly designed tip coating, taking biopsies under MR control is beginning to be greatly simplified. The approach to design artifacts using diamagnetic material in combination with paramagnetic material paves the way toward new instruments and implants, suitably tailored to the needs of the interventional radiologist.</abstract><cop>United States</cop><pub>American Association of Physicists in Medicine</pub><pmid>18649458</pmid><doi>10.1118/1.2919082</doi><tpages>6</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0094-2405
ispartof Medical physics (Lancaster), 2008-06, Vol.35 (6), p.2273-2278
issn 0094-2405
2473-4209
language eng
recordid cdi_pubmed_primary_18649458
source MEDLINE; Wiley Online Library Journals Frontfile Complete; Alma/SFX Local Collection
subjects biomedical equipment
Biomedical imaging
biomedical MRI
biopsy needles
Biopsy, Needle
Bismuth
diamagnetic materials
Diamagnetism
Image scanners
image sequences
Instrumentation
interventional MR imaging
kidney
liver
Magnetic resonance imaging
Magnetic Resonance Imaging - instrumentation
Magnetics
Medical image artifacts
Medical imaging
Medical magnetic resonance imaging
MR artifacts
MR compatibility
Needles
paramagnetic materials
Prion diseases
Sensitivity and Specificity
Sequence analysis
Titanium
title Biopsy needle tips with markers—MR compatible needles for high-precision needle tip positioning
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-07T22%3A48%3A40IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Biopsy%20needle%20tips%20with%20markers%E2%80%94MR%20compatible%20needles%20for%20high-precision%20needle%20tip%20positioning&rft.jtitle=Medical%20physics%20(Lancaster)&rft.au=M%C3%BCller-Bierl,%20Bernd%20M.&rft.date=2008-06&rft.volume=35&rft.issue=6&rft.spage=2273&rft.epage=2278&rft.pages=2273-2278&rft.issn=0094-2405&rft.eissn=2473-4209&rft.coden=MPHYA6&rft_id=info:doi/10.1118/1.2919082&rft_dat=%3Cproquest_pubme%3E69338136%3C/proquest_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=69338136&rft_id=info:pmid/18649458&rfr_iscdi=true