Nonuniform Transmission in Brain SPECT Using 201Tl, 153Gd, and 99mTc Static Line Sources: Anthropomorphic Dosimetry Studies and Influence on Brain Quantification

Nonuniform attenuation correction in brain SPECT can be done routinely by means of additional gamma transmission CT (TCT) measurements, using different commercially available line-source isotopes, 201Tl, 153Gd, and 99mTc are among the most commonly used isotopes, depending on practical and cost-effe...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of nuclear medicine (1978) 2000-12, Vol.41 (12), p.2051-2062
Hauptverfasser: Van Laere, Koen, Koole, Michel, Kauppinen, Tomi, Monsieurs, Myriam, Bouwens, Luc, Dierck, Rudi
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 2062
container_issue 12
container_start_page 2051
container_title The Journal of nuclear medicine (1978)
container_volume 41
creator Van Laere, Koen
Koole, Michel
Kauppinen, Tomi
Monsieurs, Myriam
Bouwens, Luc
Dierck, Rudi
description Nonuniform attenuation correction in brain SPECT can be done routinely by means of additional gamma transmission CT (TCT) measurements, using different commercially available line-source isotopes, 201Tl, 153Gd, and 99mTc are among the most commonly used isotopes, depending on practical and cost-effectiveness issues. We have measured additional radiation burden from static uncollimated brain SPECT transmission sources for these isotopes. The influence of the transmission isotope on brain quantification was also measured and compared with uniform attenuation correction for phantom and human data. Full iterative transmission and emission reconstruction were compared with filtered backprojection techniques. Rod sources with 201Tl, 153Gd, and 99mTc were used on a triple-head gamma camera. Dosimetry was performed using LiF TLD-100 pellets and an anthropomorphic RANDO phantom. Effective dose equivalents were calculated on the basis of measured and extrapolated absorbed doses. For brain activity measurements, a Hoffman phantom was used. Images were corrected for scatter (triple-energy window) and were reconstructed by Chang attenuation correction and filtered backprojection as well as full iterative reconstruction (ordered-subsets expectation maximization [OSEM]). To study the effect of inhomogeneous bone attenuation, realistic measurements were performed on 10 young, healthy volunteers with 153Gd TCT. After stereotactic image realignment, a volume-of-interest analysis normalized to total counts was performed. Brain SPECT-TCT using 201Tl, 153Gd, and 99mTc produced total effective dose-rate equivalents of 50.3 +/- 11.2, 32.0 +/- 2.7, and 71.1 +/- 7.1 microSv/GBq x h, respectively, representing dose equivalents of 18.6, 11.9, and 26.3 microSv for a typical 20-min brain SPECT scan at maximal used source strength. Standardized quantification resulted in insignificant differences between the isotopes and methods (Chang versus OSEM) used for nonuniform correction. Iterative reconstruction enhanced image contrast and provided more accurate gray-to-white matter ratios. Between nonuniform and uniform attenuation with an optimized attenuation coefficient, slight central discrepancies were found for volunteer studies. Significantly lower intersubject variation was found for nonuniform corrected values in infratentorial and posterior brain regions. Brain transmission scanning using 201Tl, 153Gd, or 99mTc results in limited effective radiation dose equivalents compared with th
format Article
fullrecord <record><control><sourceid>pubmed_pasca</sourceid><recordid>TN_cdi_pubmed_primary_11138692</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>11138692</sourcerecordid><originalsourceid>FETCH-LOGICAL-h1152-49bb6b45009770c7f1ba7e37b02c6391e103ba2c630461b6ce141d595434d5fa3</originalsourceid><addsrcrecordid>eNo9kF1LwzAUhosobk7_ggS88WKFnDZJW-_mnHMw_GDddUnTdM1ok5G0yH6O_9TObd6cc-A8PC-8F94QaEh9ylh06Q0xMPApxXTg3Ti3xRizOI6vvQEAhDFLgqH38250p1VpbINSy7VrlHPKaKQ0era8n6vP2TRFa6f0BgUY0nqM-ox5MUZcFyhJmlSgVctbJdBSaYlWprNCuic00W1lzc40xu6q_vtinGpka_c93hVKuj_BQpd1J7WQyJwTvzquW1Uq0UuNvvWuSl47eXfaI2_9Okunb_7yY76YTpZ-BUADnyR5znJCMU6iCIuohJxHMoxyHAgWJiABhzk_3JgwyJmQQKCgCSUhKWjJw5F3f_TuuryRRbazquF2n5276oGHE8Cd4HXZtyWU--diEpAk7qnHI1WpTfWtrMx0J2rJ7cG51Q2BDIIswBTCX1SmgH0</addsrcrecordid><sourcetype>Index Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Nonuniform Transmission in Brain SPECT Using 201Tl, 153Gd, and 99mTc Static Line Sources: Anthropomorphic Dosimetry Studies and Influence on Brain Quantification</title><source>Electronic Journals Library</source><source>MEDLINE</source><creator>Van Laere, Koen ; Koole, Michel ; Kauppinen, Tomi ; Monsieurs, Myriam ; Bouwens, Luc ; Dierck, Rudi</creator><creatorcontrib>Van Laere, Koen ; Koole, Michel ; Kauppinen, Tomi ; Monsieurs, Myriam ; Bouwens, Luc ; Dierck, Rudi</creatorcontrib><description>Nonuniform attenuation correction in brain SPECT can be done routinely by means of additional gamma transmission CT (TCT) measurements, using different commercially available line-source isotopes, 201Tl, 153Gd, and 99mTc are among the most commonly used isotopes, depending on practical and cost-effectiveness issues. We have measured additional radiation burden from static uncollimated brain SPECT transmission sources for these isotopes. The influence of the transmission isotope on brain quantification was also measured and compared with uniform attenuation correction for phantom and human data. Full iterative transmission and emission reconstruction were compared with filtered backprojection techniques. Rod sources with 201Tl, 153Gd, and 99mTc were used on a triple-head gamma camera. Dosimetry was performed using LiF TLD-100 pellets and an anthropomorphic RANDO phantom. Effective dose equivalents were calculated on the basis of measured and extrapolated absorbed doses. For brain activity measurements, a Hoffman phantom was used. Images were corrected for scatter (triple-energy window) and were reconstructed by Chang attenuation correction and filtered backprojection as well as full iterative reconstruction (ordered-subsets expectation maximization [OSEM]). To study the effect of inhomogeneous bone attenuation, realistic measurements were performed on 10 young, healthy volunteers with 153Gd TCT. After stereotactic image realignment, a volume-of-interest analysis normalized to total counts was performed. Brain SPECT-TCT using 201Tl, 153Gd, and 99mTc produced total effective dose-rate equivalents of 50.3 +/- 11.2, 32.0 +/- 2.7, and 71.1 +/- 7.1 microSv/GBq x h, respectively, representing dose equivalents of 18.6, 11.9, and 26.3 microSv for a typical 20-min brain SPECT scan at maximal used source strength. Standardized quantification resulted in insignificant differences between the isotopes and methods (Chang versus OSEM) used for nonuniform correction. Iterative reconstruction enhanced image contrast and provided more accurate gray-to-white matter ratios. Between nonuniform and uniform attenuation with an optimized attenuation coefficient, slight central discrepancies were found for volunteer studies. Significantly lower intersubject variation was found for nonuniform corrected values in infratentorial and posterior brain regions. Brain transmission scanning using 201Tl, 153Gd, or 99mTc results in limited effective radiation dose equivalents compared with the typical radiation burden. Relative brain perfusion quantification is not significantly different for the various nonuniform TCT isotopes. Iterative reconstruction improves gray-to-white contrasts but has no significant influence on brain perfusion semiquantification. Nonuniform attenuation correction decreases intersubject variability in the posterior brain regions that were compared, which may lead to improved sensitivity toward clinical applications.</description><identifier>ISSN: 0161-5505</identifier><identifier>EISSN: 1535-5667</identifier><identifier>PMID: 11138692</identifier><language>eng</language><publisher>Reston, VA: Soc Nuclear Med</publisher><subject>Adult ; Biological and medical sciences ; Brain - diagnostic imaging ; Gadolinium ; Humans ; Image Processing, Computer-Assisted ; Investigative techniques, diagnostic techniques (general aspects) ; Medical sciences ; Nervous system ; Phantoms, Imaging ; Radiation Dosage ; Radioisotopes ; Radionuclide investigations ; Technetium ; Thallium Radioisotopes ; Tomography, Emission-Computed, Single-Photon</subject><ispartof>The Journal of nuclear medicine (1978), 2000-12, Vol.41 (12), p.2051-2062</ispartof><rights>2001 INIST-CNRS</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&amp;idt=842498$$DView record in Pascal Francis$$Hfree_for_read</backlink><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/11138692$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Van Laere, Koen</creatorcontrib><creatorcontrib>Koole, Michel</creatorcontrib><creatorcontrib>Kauppinen, Tomi</creatorcontrib><creatorcontrib>Monsieurs, Myriam</creatorcontrib><creatorcontrib>Bouwens, Luc</creatorcontrib><creatorcontrib>Dierck, Rudi</creatorcontrib><title>Nonuniform Transmission in Brain SPECT Using 201Tl, 153Gd, and 99mTc Static Line Sources: Anthropomorphic Dosimetry Studies and Influence on Brain Quantification</title><title>The Journal of nuclear medicine (1978)</title><addtitle>J Nucl Med</addtitle><description>Nonuniform attenuation correction in brain SPECT can be done routinely by means of additional gamma transmission CT (TCT) measurements, using different commercially available line-source isotopes, 201Tl, 153Gd, and 99mTc are among the most commonly used isotopes, depending on practical and cost-effectiveness issues. We have measured additional radiation burden from static uncollimated brain SPECT transmission sources for these isotopes. The influence of the transmission isotope on brain quantification was also measured and compared with uniform attenuation correction for phantom and human data. Full iterative transmission and emission reconstruction were compared with filtered backprojection techniques. Rod sources with 201Tl, 153Gd, and 99mTc were used on a triple-head gamma camera. Dosimetry was performed using LiF TLD-100 pellets and an anthropomorphic RANDO phantom. Effective dose equivalents were calculated on the basis of measured and extrapolated absorbed doses. For brain activity measurements, a Hoffman phantom was used. Images were corrected for scatter (triple-energy window) and were reconstructed by Chang attenuation correction and filtered backprojection as well as full iterative reconstruction (ordered-subsets expectation maximization [OSEM]). To study the effect of inhomogeneous bone attenuation, realistic measurements were performed on 10 young, healthy volunteers with 153Gd TCT. After stereotactic image realignment, a volume-of-interest analysis normalized to total counts was performed. Brain SPECT-TCT using 201Tl, 153Gd, and 99mTc produced total effective dose-rate equivalents of 50.3 +/- 11.2, 32.0 +/- 2.7, and 71.1 +/- 7.1 microSv/GBq x h, respectively, representing dose equivalents of 18.6, 11.9, and 26.3 microSv for a typical 20-min brain SPECT scan at maximal used source strength. Standardized quantification resulted in insignificant differences between the isotopes and methods (Chang versus OSEM) used for nonuniform correction. Iterative reconstruction enhanced image contrast and provided more accurate gray-to-white matter ratios. Between nonuniform and uniform attenuation with an optimized attenuation coefficient, slight central discrepancies were found for volunteer studies. Significantly lower intersubject variation was found for nonuniform corrected values in infratentorial and posterior brain regions. Brain transmission scanning using 201Tl, 153Gd, or 99mTc results in limited effective radiation dose equivalents compared with the typical radiation burden. Relative brain perfusion quantification is not significantly different for the various nonuniform TCT isotopes. Iterative reconstruction improves gray-to-white contrasts but has no significant influence on brain perfusion semiquantification. Nonuniform attenuation correction decreases intersubject variability in the posterior brain regions that were compared, which may lead to improved sensitivity toward clinical applications.</description><subject>Adult</subject><subject>Biological and medical sciences</subject><subject>Brain - diagnostic imaging</subject><subject>Gadolinium</subject><subject>Humans</subject><subject>Image Processing, Computer-Assisted</subject><subject>Investigative techniques, diagnostic techniques (general aspects)</subject><subject>Medical sciences</subject><subject>Nervous system</subject><subject>Phantoms, Imaging</subject><subject>Radiation Dosage</subject><subject>Radioisotopes</subject><subject>Radionuclide investigations</subject><subject>Technetium</subject><subject>Thallium Radioisotopes</subject><subject>Tomography, Emission-Computed, Single-Photon</subject><issn>0161-5505</issn><issn>1535-5667</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2000</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNo9kF1LwzAUhosobk7_ggS88WKFnDZJW-_mnHMw_GDddUnTdM1ok5G0yH6O_9TObd6cc-A8PC-8F94QaEh9ylh06Q0xMPApxXTg3Ti3xRizOI6vvQEAhDFLgqH38250p1VpbINSy7VrlHPKaKQ0era8n6vP2TRFa6f0BgUY0nqM-ox5MUZcFyhJmlSgVctbJdBSaYlWprNCuic00W1lzc40xu6q_vtinGpka_c93hVKuj_BQpd1J7WQyJwTvzquW1Uq0UuNvvWuSl47eXfaI2_9Okunb_7yY76YTpZ-BUADnyR5znJCMU6iCIuohJxHMoxyHAgWJiABhzk_3JgwyJmQQKCgCSUhKWjJw5F3f_TuuryRRbazquF2n5276oGHE8Cd4HXZtyWU--diEpAk7qnHI1WpTfWtrMx0J2rJ7cG51Q2BDIIswBTCX1SmgH0</recordid><startdate>200012</startdate><enddate>200012</enddate><creator>Van Laere, Koen</creator><creator>Koole, Michel</creator><creator>Kauppinen, Tomi</creator><creator>Monsieurs, Myriam</creator><creator>Bouwens, Luc</creator><creator>Dierck, Rudi</creator><general>Soc Nuclear Med</general><general>Society of Nuclear Medicine</general><scope>IQODW</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope></search><sort><creationdate>200012</creationdate><title>Nonuniform Transmission in Brain SPECT Using 201Tl, 153Gd, and 99mTc Static Line Sources: Anthropomorphic Dosimetry Studies and Influence on Brain Quantification</title><author>Van Laere, Koen ; Koole, Michel ; Kauppinen, Tomi ; Monsieurs, Myriam ; Bouwens, Luc ; Dierck, Rudi</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-h1152-49bb6b45009770c7f1ba7e37b02c6391e103ba2c630461b6ce141d595434d5fa3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2000</creationdate><topic>Adult</topic><topic>Biological and medical sciences</topic><topic>Brain - diagnostic imaging</topic><topic>Gadolinium</topic><topic>Humans</topic><topic>Image Processing, Computer-Assisted</topic><topic>Investigative techniques, diagnostic techniques (general aspects)</topic><topic>Medical sciences</topic><topic>Nervous system</topic><topic>Phantoms, Imaging</topic><topic>Radiation Dosage</topic><topic>Radioisotopes</topic><topic>Radionuclide investigations</topic><topic>Technetium</topic><topic>Thallium Radioisotopes</topic><topic>Tomography, Emission-Computed, Single-Photon</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Van Laere, Koen</creatorcontrib><creatorcontrib>Koole, Michel</creatorcontrib><creatorcontrib>Kauppinen, Tomi</creatorcontrib><creatorcontrib>Monsieurs, Myriam</creatorcontrib><creatorcontrib>Bouwens, Luc</creatorcontrib><creatorcontrib>Dierck, Rudi</creatorcontrib><collection>Pascal-Francis</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><jtitle>The Journal of nuclear medicine (1978)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Van Laere, Koen</au><au>Koole, Michel</au><au>Kauppinen, Tomi</au><au>Monsieurs, Myriam</au><au>Bouwens, Luc</au><au>Dierck, Rudi</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Nonuniform Transmission in Brain SPECT Using 201Tl, 153Gd, and 99mTc Static Line Sources: Anthropomorphic Dosimetry Studies and Influence on Brain Quantification</atitle><jtitle>The Journal of nuclear medicine (1978)</jtitle><addtitle>J Nucl Med</addtitle><date>2000-12</date><risdate>2000</risdate><volume>41</volume><issue>12</issue><spage>2051</spage><epage>2062</epage><pages>2051-2062</pages><issn>0161-5505</issn><eissn>1535-5667</eissn><abstract>Nonuniform attenuation correction in brain SPECT can be done routinely by means of additional gamma transmission CT (TCT) measurements, using different commercially available line-source isotopes, 201Tl, 153Gd, and 99mTc are among the most commonly used isotopes, depending on practical and cost-effectiveness issues. We have measured additional radiation burden from static uncollimated brain SPECT transmission sources for these isotopes. The influence of the transmission isotope on brain quantification was also measured and compared with uniform attenuation correction for phantom and human data. Full iterative transmission and emission reconstruction were compared with filtered backprojection techniques. Rod sources with 201Tl, 153Gd, and 99mTc were used on a triple-head gamma camera. Dosimetry was performed using LiF TLD-100 pellets and an anthropomorphic RANDO phantom. Effective dose equivalents were calculated on the basis of measured and extrapolated absorbed doses. For brain activity measurements, a Hoffman phantom was used. Images were corrected for scatter (triple-energy window) and were reconstructed by Chang attenuation correction and filtered backprojection as well as full iterative reconstruction (ordered-subsets expectation maximization [OSEM]). To study the effect of inhomogeneous bone attenuation, realistic measurements were performed on 10 young, healthy volunteers with 153Gd TCT. After stereotactic image realignment, a volume-of-interest analysis normalized to total counts was performed. Brain SPECT-TCT using 201Tl, 153Gd, and 99mTc produced total effective dose-rate equivalents of 50.3 +/- 11.2, 32.0 +/- 2.7, and 71.1 +/- 7.1 microSv/GBq x h, respectively, representing dose equivalents of 18.6, 11.9, and 26.3 microSv for a typical 20-min brain SPECT scan at maximal used source strength. Standardized quantification resulted in insignificant differences between the isotopes and methods (Chang versus OSEM) used for nonuniform correction. Iterative reconstruction enhanced image contrast and provided more accurate gray-to-white matter ratios. Between nonuniform and uniform attenuation with an optimized attenuation coefficient, slight central discrepancies were found for volunteer studies. Significantly lower intersubject variation was found for nonuniform corrected values in infratentorial and posterior brain regions. Brain transmission scanning using 201Tl, 153Gd, or 99mTc results in limited effective radiation dose equivalents compared with the typical radiation burden. Relative brain perfusion quantification is not significantly different for the various nonuniform TCT isotopes. Iterative reconstruction improves gray-to-white contrasts but has no significant influence on brain perfusion semiquantification. Nonuniform attenuation correction decreases intersubject variability in the posterior brain regions that were compared, which may lead to improved sensitivity toward clinical applications.</abstract><cop>Reston, VA</cop><pub>Soc Nuclear Med</pub><pmid>11138692</pmid><tpages>12</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0161-5505
ispartof The Journal of nuclear medicine (1978), 2000-12, Vol.41 (12), p.2051-2062
issn 0161-5505
1535-5667
language eng
recordid cdi_pubmed_primary_11138692
source Electronic Journals Library; MEDLINE
subjects Adult
Biological and medical sciences
Brain - diagnostic imaging
Gadolinium
Humans
Image Processing, Computer-Assisted
Investigative techniques, diagnostic techniques (general aspects)
Medical sciences
Nervous system
Phantoms, Imaging
Radiation Dosage
Radioisotopes
Radionuclide investigations
Technetium
Thallium Radioisotopes
Tomography, Emission-Computed, Single-Photon
title Nonuniform Transmission in Brain SPECT Using 201Tl, 153Gd, and 99mTc Static Line Sources: Anthropomorphic Dosimetry Studies and Influence on Brain Quantification
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-01T08%3A30%3A48IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-pubmed_pasca&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Nonuniform%20Transmission%20in%20Brain%20SPECT%20Using%20201Tl,%20153Gd,%20and%2099mTc%20Static%20Line%20Sources:%20Anthropomorphic%20Dosimetry%20Studies%20and%20Influence%20on%20Brain%20Quantification&rft.jtitle=The%20Journal%20of%20nuclear%20medicine%20(1978)&rft.au=Van%20Laere,%20Koen&rft.date=2000-12&rft.volume=41&rft.issue=12&rft.spage=2051&rft.epage=2062&rft.pages=2051-2062&rft.issn=0161-5505&rft.eissn=1535-5667&rft_id=info:doi/&rft_dat=%3Cpubmed_pasca%3E11138692%3C/pubmed_pasca%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/11138692&rfr_iscdi=true