Performance assessment of four 64-slice computed tomographic devices for a typical clinical protocol

To compare the performances of four 64-slice CT devices, as regards radiation dose and image quality. Effective dose was measured with thermoluminescent dosimeters in an Alderson Rando phantom (Alderson Research Laboratories, New York, NY). Quantitative image quality was evaluated in a Catphan 600 p...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of computer assisted tomography 2011-01, Vol.35 (1), p.57-64
Hauptverfasser: Colli, Vittoria, Mangini, Monica, Strocchi, Sabina, Lumia, Domenico, Canì, Andrea, Boffano, Carlo, Cesarani, Federico, Rizzi, Elisa, Ferrara, Rosa Maria, Andreis, Ivo Andrea Bergamo, Bruno, Jolanda, Gozzi, Gino, Cacciatori, Marco, Ostinelli, Angelo, Vite, Cristina, Conte, Leopoldo, Fugazzola, Carlo
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To compare the performances of four 64-slice CT devices, as regards radiation dose and image quality. Effective dose was measured with thermoluminescent dosimeters in an Alderson Rando phantom (Alderson Research Laboratories, New York, NY). Quantitative image quality was evaluated in a Catphan 600 phantom (The Phantom Laboratory, New York, NY) using 3 parameters (modulation transfer function, contrast-to-noise ratio, and figure of merit). Effective and main organ doses were measured in 40 patients (10 for each CT device), who underwent an abdominal study with the same standard protocol; moreover, in these patients, clinical image quality, using a 5-grade quality score, was assessed. In-phantom measured doses and quantitative image quality showed some differences among the 4 devices. On the contrary, effective and organ doses provided to the patients were similar; no statistically significant differences were found also for clinical image quality. Some differences were found among the 4 devices from a physical point of view; on the other hand, the patient data were similar.
ISSN:0363-8715
1532-3145
DOI:10.1097/RCT.0b013e3181f124fe