A new method to combine contrast-enhanced magnetic resonance imaging during live ultrasound of the breast using volume navigation technique: A study for evaluating feasibility, accuracy and reproducibility in healthy volunteers
Abstract Objectives To evaluate feasibility, accuracy and reproducibility of combined US-MR of the breast using volume navigation technique. Subjects and methods Five healthy females underwent bilateral contrast-enhanced MR (CE-MR) of the breast in supine position, after positioning three couples of...
Gespeichert in:
Veröffentlicht in: | European journal of radiology 2012-03, Vol.81 (3), p.e332-e337 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | e337 |
---|---|
container_issue | 3 |
container_start_page | e332 |
container_title | European journal of radiology |
container_volume | 81 |
creator | Fausto, Alfonso Rizzatto, Giorgio Preziosa, Alessandro Gaburro, Luca Washburn, Michael J Rubello, Domenico Volterrani, Luca |
description | Abstract Objectives To evaluate feasibility, accuracy and reproducibility of combined US-MR of the breast using volume navigation technique. Subjects and methods Five healthy females underwent bilateral contrast-enhanced MR (CE-MR) of the breast in supine position, after positioning three couples of markers on the breast. After CE-MR data uploading in the ultrasound (US) database, manual co-registration was obtained during live US of the breast by means of an electromagnetic transmitter positioned near the subject under examination and two electromagnetic sensors were mounted on the transducer bracket. Transmitter and sensors were connected to a position-sensing unit embedded in the US equipment allowing to track probe position and orientation within the electromagnetic field. Live US image were co-registered to the previously loaded breast CE-MR volume by coupling markers. For each subject, two independent radiologists recorded the examination time and verified twice image alignment using five fixed checkpoints. Pair t Student test and Wilcoxon test were used for statistical analysis. Results In all subjects US and CE-MR images were successfully combined. The examination time was 10 ± 2 vs. 9 ± 4 min, respectively ( p = 0.642; NS). A total of one hundred measurements of images misalignment were performed: the measurements recorded between the two operators were 0.42 ± 0.32 cm and 0.58 ± 0.41 cm ( p = 0.161; NS), and 0.50 ± 0.32 cm and 0.56 ± 0.52 cm ( p = 0.928; NS), respectively. Discussion In our preliminary experience, volume navigation technique appears to be a accurate and reproducible method to combine CE-MR image during unilateral US of the breast. |
doi_str_mv | 10.1016/j.ejrad.2011.11.001 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_921143967</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>1_s2_0_S0720048X11007698</els_id><sourcerecordid>921143967</sourcerecordid><originalsourceid>FETCH-LOGICAL-c479t-b0051ffb5820615ec2a392fa917931979fd291bf1f9547cc367039da2182df9a3</originalsourceid><addsrcrecordid>eNqFUk2LFDEQbURxZ1d_gSC5ebHHJN0z6QgKw7J-wIIHFbyFdFKZztidrPkY6d_rHzG9M3rwIhQUVN6rl6pXVfWM4DXBZPvqsIZDkHpNMSHrEhiTB9WKdIzWjFH2sFphRnGN2-7bRXUZ4wFjvGk5fVxdUEpwi7tuVf3aIQc_0QRp8Bolj5SfeuugZJeCjKkGN0inQKNJ7h0kq1CA6N1SQ7bUrNsjncOSRnsElMeF57PTyBuUBkB9gNII5bhgjn7MEyAnj3Yvk_UOJVCDsz8yvEY7FFPWMzI-IDjKMRdE4ZjCt70dbZpfIqlUDlLNSBaFAHfB66zOr8g6NIAc0zDfC7kEEOKT6pGRY4Sn53xVfX138-X6Q3376f3H691trVrGU92X9RBj-k1H8ZZsQFHZcGokJ4w3hDNuNOWkN8TwTcuUarYMN1xLSjqqDZfNVfXi1Lf8qYwTk5hsVDCO0oHPUXBKSNvwLSvI5oRUwccYwIi7UHYZZkGwWMwVB3FvrljMFSWKuYX1_Nw_9xPov5w_bhbAmxMAypRHC0FEZWExzwZQSWhv_yPw9h--Gq2zSo7fYYZ48Dm4skBBRKQCi8_LfS3nRQjGbMu75jcum9JE</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>921143967</pqid></control><display><type>article</type><title>A new method to combine contrast-enhanced magnetic resonance imaging during live ultrasound of the breast using volume navigation technique: A study for evaluating feasibility, accuracy and reproducibility in healthy volunteers</title><source>MEDLINE</source><source>ScienceDirect Journals (5 years ago - present)</source><creator>Fausto, Alfonso ; Rizzatto, Giorgio ; Preziosa, Alessandro ; Gaburro, Luca ; Washburn, Michael J ; Rubello, Domenico ; Volterrani, Luca</creator><creatorcontrib>Fausto, Alfonso ; Rizzatto, Giorgio ; Preziosa, Alessandro ; Gaburro, Luca ; Washburn, Michael J ; Rubello, Domenico ; Volterrani, Luca</creatorcontrib><description>Abstract Objectives To evaluate feasibility, accuracy and reproducibility of combined US-MR of the breast using volume navigation technique. Subjects and methods Five healthy females underwent bilateral contrast-enhanced MR (CE-MR) of the breast in supine position, after positioning three couples of markers on the breast. After CE-MR data uploading in the ultrasound (US) database, manual co-registration was obtained during live US of the breast by means of an electromagnetic transmitter positioned near the subject under examination and two electromagnetic sensors were mounted on the transducer bracket. Transmitter and sensors were connected to a position-sensing unit embedded in the US equipment allowing to track probe position and orientation within the electromagnetic field. Live US image were co-registered to the previously loaded breast CE-MR volume by coupling markers. For each subject, two independent radiologists recorded the examination time and verified twice image alignment using five fixed checkpoints. Pair t Student test and Wilcoxon test were used for statistical analysis. Results In all subjects US and CE-MR images were successfully combined. The examination time was 10 ± 2 vs. 9 ± 4 min, respectively ( p = 0.642; NS). A total of one hundred measurements of images misalignment were performed: the measurements recorded between the two operators were 0.42 ± 0.32 cm and 0.58 ± 0.41 cm ( p = 0.161; NS), and 0.50 ± 0.32 cm and 0.56 ± 0.52 cm ( p = 0.928; NS), respectively. Discussion In our preliminary experience, volume navigation technique appears to be a accurate and reproducible method to combine CE-MR image during unilateral US of the breast.</description><identifier>ISSN: 0720-048X</identifier><identifier>EISSN: 1872-7727</identifier><identifier>DOI: 10.1016/j.ejrad.2011.11.001</identifier><identifier>PMID: 22104088</identifier><language>eng</language><publisher>Ireland: Elsevier Ireland Ltd</publisher><subject>Adult ; Aged ; Contrast Media ; Feasibility Studies ; Female ; Humans ; Image Enhancement - methods ; Intra and inter-observer variation ; Magnetic Resonance Imaging ; Magnetic Resonance Imaging - methods ; Meglumine - analogs & derivatives ; Middle Aged ; Organometallic Compounds ; Radiology ; Reproducibility of Results ; Statistics, Nonparametric ; Supine Position ; Ultrasonography, Mammary - methods ; Ultrasound, Mammary ; Volume navigation technique</subject><ispartof>European journal of radiology, 2012-03, Vol.81 (3), p.e332-e337</ispartof><rights>Elsevier Ireland Ltd</rights><rights>2011 Elsevier Ireland Ltd</rights><rights>Copyright © 2011 Elsevier Ireland Ltd. All rights reserved.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c479t-b0051ffb5820615ec2a392fa917931979fd291bf1f9547cc367039da2182df9a3</citedby><cites>FETCH-LOGICAL-c479t-b0051ffb5820615ec2a392fa917931979fd291bf1f9547cc367039da2182df9a3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://dx.doi.org/10.1016/j.ejrad.2011.11.001$$EHTML$$P50$$Gelsevier$$H</linktohtml><link.rule.ids>314,780,784,3550,27924,27925,45995</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/22104088$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Fausto, Alfonso</creatorcontrib><creatorcontrib>Rizzatto, Giorgio</creatorcontrib><creatorcontrib>Preziosa, Alessandro</creatorcontrib><creatorcontrib>Gaburro, Luca</creatorcontrib><creatorcontrib>Washburn, Michael J</creatorcontrib><creatorcontrib>Rubello, Domenico</creatorcontrib><creatorcontrib>Volterrani, Luca</creatorcontrib><title>A new method to combine contrast-enhanced magnetic resonance imaging during live ultrasound of the breast using volume navigation technique: A study for evaluating feasibility, accuracy and reproducibility in healthy volunteers</title><title>European journal of radiology</title><addtitle>Eur J Radiol</addtitle><description>Abstract Objectives To evaluate feasibility, accuracy and reproducibility of combined US-MR of the breast using volume navigation technique. Subjects and methods Five healthy females underwent bilateral contrast-enhanced MR (CE-MR) of the breast in supine position, after positioning three couples of markers on the breast. After CE-MR data uploading in the ultrasound (US) database, manual co-registration was obtained during live US of the breast by means of an electromagnetic transmitter positioned near the subject under examination and two electromagnetic sensors were mounted on the transducer bracket. Transmitter and sensors were connected to a position-sensing unit embedded in the US equipment allowing to track probe position and orientation within the electromagnetic field. Live US image were co-registered to the previously loaded breast CE-MR volume by coupling markers. For each subject, two independent radiologists recorded the examination time and verified twice image alignment using five fixed checkpoints. Pair t Student test and Wilcoxon test were used for statistical analysis. Results In all subjects US and CE-MR images were successfully combined. The examination time was 10 ± 2 vs. 9 ± 4 min, respectively ( p = 0.642; NS). A total of one hundred measurements of images misalignment were performed: the measurements recorded between the two operators were 0.42 ± 0.32 cm and 0.58 ± 0.41 cm ( p = 0.161; NS), and 0.50 ± 0.32 cm and 0.56 ± 0.52 cm ( p = 0.928; NS), respectively. Discussion In our preliminary experience, volume navigation technique appears to be a accurate and reproducible method to combine CE-MR image during unilateral US of the breast.</description><subject>Adult</subject><subject>Aged</subject><subject>Contrast Media</subject><subject>Feasibility Studies</subject><subject>Female</subject><subject>Humans</subject><subject>Image Enhancement - methods</subject><subject>Intra and inter-observer variation</subject><subject>Magnetic Resonance Imaging</subject><subject>Magnetic Resonance Imaging - methods</subject><subject>Meglumine - analogs & derivatives</subject><subject>Middle Aged</subject><subject>Organometallic Compounds</subject><subject>Radiology</subject><subject>Reproducibility of Results</subject><subject>Statistics, Nonparametric</subject><subject>Supine Position</subject><subject>Ultrasonography, Mammary - methods</subject><subject>Ultrasound, Mammary</subject><subject>Volume navigation technique</subject><issn>0720-048X</issn><issn>1872-7727</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2012</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNqFUk2LFDEQbURxZ1d_gSC5ebHHJN0z6QgKw7J-wIIHFbyFdFKZztidrPkY6d_rHzG9M3rwIhQUVN6rl6pXVfWM4DXBZPvqsIZDkHpNMSHrEhiTB9WKdIzWjFH2sFphRnGN2-7bRXUZ4wFjvGk5fVxdUEpwi7tuVf3aIQc_0QRp8Bolj5SfeuugZJeCjKkGN0inQKNJ7h0kq1CA6N1SQ7bUrNsjncOSRnsElMeF57PTyBuUBkB9gNII5bhgjn7MEyAnj3Yvk_UOJVCDsz8yvEY7FFPWMzI-IDjKMRdE4ZjCt70dbZpfIqlUDlLNSBaFAHfB66zOr8g6NIAc0zDfC7kEEOKT6pGRY4Sn53xVfX138-X6Q3376f3H691trVrGU92X9RBj-k1H8ZZsQFHZcGokJ4w3hDNuNOWkN8TwTcuUarYMN1xLSjqqDZfNVfXi1Lf8qYwTk5hsVDCO0oHPUXBKSNvwLSvI5oRUwccYwIi7UHYZZkGwWMwVB3FvrljMFSWKuYX1_Nw_9xPov5w_bhbAmxMAypRHC0FEZWExzwZQSWhv_yPw9h--Gq2zSo7fYYZ48Dm4skBBRKQCi8_LfS3nRQjGbMu75jcum9JE</recordid><startdate>20120301</startdate><enddate>20120301</enddate><creator>Fausto, Alfonso</creator><creator>Rizzatto, Giorgio</creator><creator>Preziosa, Alessandro</creator><creator>Gaburro, Luca</creator><creator>Washburn, Michael J</creator><creator>Rubello, Domenico</creator><creator>Volterrani, Luca</creator><general>Elsevier Ireland Ltd</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>20120301</creationdate><title>A new method to combine contrast-enhanced magnetic resonance imaging during live ultrasound of the breast using volume navigation technique: A study for evaluating feasibility, accuracy and reproducibility in healthy volunteers</title><author>Fausto, Alfonso ; Rizzatto, Giorgio ; Preziosa, Alessandro ; Gaburro, Luca ; Washburn, Michael J ; Rubello, Domenico ; Volterrani, Luca</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c479t-b0051ffb5820615ec2a392fa917931979fd291bf1f9547cc367039da2182df9a3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2012</creationdate><topic>Adult</topic><topic>Aged</topic><topic>Contrast Media</topic><topic>Feasibility Studies</topic><topic>Female</topic><topic>Humans</topic><topic>Image Enhancement - methods</topic><topic>Intra and inter-observer variation</topic><topic>Magnetic Resonance Imaging</topic><topic>Magnetic Resonance Imaging - methods</topic><topic>Meglumine - analogs & derivatives</topic><topic>Middle Aged</topic><topic>Organometallic Compounds</topic><topic>Radiology</topic><topic>Reproducibility of Results</topic><topic>Statistics, Nonparametric</topic><topic>Supine Position</topic><topic>Ultrasonography, Mammary - methods</topic><topic>Ultrasound, Mammary</topic><topic>Volume navigation technique</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Fausto, Alfonso</creatorcontrib><creatorcontrib>Rizzatto, Giorgio</creatorcontrib><creatorcontrib>Preziosa, Alessandro</creatorcontrib><creatorcontrib>Gaburro, Luca</creatorcontrib><creatorcontrib>Washburn, Michael J</creatorcontrib><creatorcontrib>Rubello, Domenico</creatorcontrib><creatorcontrib>Volterrani, Luca</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>European journal of radiology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Fausto, Alfonso</au><au>Rizzatto, Giorgio</au><au>Preziosa, Alessandro</au><au>Gaburro, Luca</au><au>Washburn, Michael J</au><au>Rubello, Domenico</au><au>Volterrani, Luca</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>A new method to combine contrast-enhanced magnetic resonance imaging during live ultrasound of the breast using volume navigation technique: A study for evaluating feasibility, accuracy and reproducibility in healthy volunteers</atitle><jtitle>European journal of radiology</jtitle><addtitle>Eur J Radiol</addtitle><date>2012-03-01</date><risdate>2012</risdate><volume>81</volume><issue>3</issue><spage>e332</spage><epage>e337</epage><pages>e332-e337</pages><issn>0720-048X</issn><eissn>1872-7727</eissn><abstract>Abstract Objectives To evaluate feasibility, accuracy and reproducibility of combined US-MR of the breast using volume navigation technique. Subjects and methods Five healthy females underwent bilateral contrast-enhanced MR (CE-MR) of the breast in supine position, after positioning three couples of markers on the breast. After CE-MR data uploading in the ultrasound (US) database, manual co-registration was obtained during live US of the breast by means of an electromagnetic transmitter positioned near the subject under examination and two electromagnetic sensors were mounted on the transducer bracket. Transmitter and sensors were connected to a position-sensing unit embedded in the US equipment allowing to track probe position and orientation within the electromagnetic field. Live US image were co-registered to the previously loaded breast CE-MR volume by coupling markers. For each subject, two independent radiologists recorded the examination time and verified twice image alignment using five fixed checkpoints. Pair t Student test and Wilcoxon test were used for statistical analysis. Results In all subjects US and CE-MR images were successfully combined. The examination time was 10 ± 2 vs. 9 ± 4 min, respectively ( p = 0.642; NS). A total of one hundred measurements of images misalignment were performed: the measurements recorded between the two operators were 0.42 ± 0.32 cm and 0.58 ± 0.41 cm ( p = 0.161; NS), and 0.50 ± 0.32 cm and 0.56 ± 0.52 cm ( p = 0.928; NS), respectively. Discussion In our preliminary experience, volume navigation technique appears to be a accurate and reproducible method to combine CE-MR image during unilateral US of the breast.</abstract><cop>Ireland</cop><pub>Elsevier Ireland Ltd</pub><pmid>22104088</pmid><doi>10.1016/j.ejrad.2011.11.001</doi></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0720-048X |
ispartof | European journal of radiology, 2012-03, Vol.81 (3), p.e332-e337 |
issn | 0720-048X 1872-7727 |
language | eng |
recordid | cdi_proquest_miscellaneous_921143967 |
source | MEDLINE; ScienceDirect Journals (5 years ago - present) |
subjects | Adult Aged Contrast Media Feasibility Studies Female Humans Image Enhancement - methods Intra and inter-observer variation Magnetic Resonance Imaging Magnetic Resonance Imaging - methods Meglumine - analogs & derivatives Middle Aged Organometallic Compounds Radiology Reproducibility of Results Statistics, Nonparametric Supine Position Ultrasonography, Mammary - methods Ultrasound, Mammary Volume navigation technique |
title | A new method to combine contrast-enhanced magnetic resonance imaging during live ultrasound of the breast using volume navigation technique: A study for evaluating feasibility, accuracy and reproducibility in healthy volunteers |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-07T05%3A55%3A47IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=A%20new%20method%20to%20combine%20contrast-enhanced%20magnetic%20resonance%20imaging%20during%20live%20ultrasound%20of%20the%20breast%20using%20volume%20navigation%20technique:%20A%20study%20for%20evaluating%20feasibility,%20accuracy%20and%20reproducibility%20in%20healthy%20volunteers&rft.jtitle=European%20journal%20of%20radiology&rft.au=Fausto,%20Alfonso&rft.date=2012-03-01&rft.volume=81&rft.issue=3&rft.spage=e332&rft.epage=e337&rft.pages=e332-e337&rft.issn=0720-048X&rft.eissn=1872-7727&rft_id=info:doi/10.1016/j.ejrad.2011.11.001&rft_dat=%3Cproquest_cross%3E921143967%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=921143967&rft_id=info:pmid/22104088&rft_els_id=1_s2_0_S0720048X11007698&rfr_iscdi=true |