Straight tapered titanium stem with alumina bearing in cementless primary total hip arthroplasty: a minimum 5-year follow-up
The aim of this study was to assess the midterm results of primary cementless total hip arthroplasty using a tapered stem and alumina bearing couple in active patients. After a minimum of 5 years of follow-up, 78 arthroplasties in 72 patients were reviewed retrospectively. The mean Harris hip score...
Gespeichert in:
Veröffentlicht in: | The Journal of arthroplasty 2011-12, Vol.26 (8), p.1310-1317 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The aim of this study was to assess the midterm results of primary cementless total hip arthroplasty using a tapered stem and alumina bearing couple in active patients. After a minimum of 5 years of follow-up, 78 arthroplasties in 72 patients were reviewed retrospectively. The mean Harris hip score was 94 points, and 2 hips had thigh pain. All components radiographically demonstrated stable fixation by bone ingrowth and mild stress shielding of the proximal femur were noted in 14% of hips. There was no significant osteolysis or aseptic loosening. There was a ceramic head fracture in 1 hip and audible sounds in 2 hips. The results of total hip arthroplasty with a straight, tapered, proximally porous-coated stem and alumina-on-alumina bearing were encouraging for active patients. |
---|---|
ISSN: | 0883-5403 1532-8406 |
DOI: | 10.1016/j.arth.2010.12.015 |