Investigation of adhesion of modern wound dressings: a comparative analysis of 56 different wound dressings

Introduction  In the process of chronic wound care, adhesive wound dressings may cause pain and injury in the wound environment during dressing changes. At present, no standardized test procedures are available for the investigation of adhesion of wound dressings. Therefore, our study aimed to test...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of the European Academy of Dermatology and Venereology 2011-08, Vol.25 (8), p.933-939
Hauptverfasser: Klode, J., Schöttler, L., Stoffels, I., Körber, A., Schadendorf, D., Dissemond, J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Introduction  In the process of chronic wound care, adhesive wound dressings may cause pain and injury in the wound environment during dressing changes. At present, no standardized test procedures are available for the investigation of adhesion of wound dressings. Therefore, our study aimed to test the adhesion of different wound dressings on steel as well as on healthy skin. Methods  Within an open, comparative study, the adhesive areas of 56 wound dressings were investigated. The adhesives were categorized into acrylate (n = 23), silicone (n = 9), hydrocolloid (n = 17) and polyurethane groups (n = 7). Using an especially modified testing machine, the adhesion of the wound dressings was measured on steel as well as on the skin of healthy study participants, in compliance with the European EN 1939:2003 standard. Results  The energy required to remove the wound dressings from human skin, was measured in Newton (N) and the following median values were obtained: hydrocolloid (2.25 N) > acrylate (1.14 N) > polyurethane (0.9 N) > silicone (0.7 N). The subjective pain intensity during the removal of the wound dressings was recorded using the visual analogue scale (VAS) with values ranging from 0 to 10. For hydrocolloid, it was 6.8, for acrylate 4.9, for polyurethane 3.1 and for silicone 2.5 points VAS. In comparison with human skin, the adhesion of wound dressings was significantly higher on steel (P 
ISSN:0926-9959
1468-3083
DOI:10.1111/j.1468-3083.2010.03886.x