Contactless Conductivity Detection for Capillary Electrophoresis

A contactless capacitively coupled conductivity detector for capillary electrophoresis is introduced. The detector consists of two electrodes which are placed cylindrically around the outer polyimide coating of the fused-silica capillary with a detection gap of 2 mm. The electrodes form a cylindrica...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytical chemistry (Washington) 1998-02, Vol.70 (3), p.563-567
Hauptverfasser: Zemann, Andreas J, Schnell, Erhard, Volgger, Dietmar, Bonn, Günther K
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A contactless capacitively coupled conductivity detector for capillary electrophoresis is introduced. The detector consists of two electrodes which are placed cylindrically around the outer polyimide coating of the fused-silica capillary with a detection gap of 2 mm. The electrodes form a cylindrical capacitor, and the electric conductivity of the solution in the gap between the electrodes is measured. A high audio or low ultrasonic frequency for coupling of the ac voltage is used in order to minimize the influence of reactance of the liquid. For an improved version of the detector, two syringe cannulas are used as the electrodes and the capillary is simply assembled into the tubing. This allows an easy placement of the detector on various positions along the capillary. The limit of detection of inorganic cations and anions is 200 ppb, as determined for sodium and chloride, respectively.
ISSN:0003-2700
1520-6882
DOI:10.1021/ac9707592