Instrument to test patency of lacrimal drainage system

The lacrimal drainage system drains fluid that exists normally on the surface of the eye into the nasal cavity. This paper describes a hand-held, personal-computer-based instrument to quantitatively estimate the resistance to fluid flow through the lacrimal drainage system. This resistance measureme...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of clinical engineering 1991-01, Vol.16 (1), p.57-60
Hauptverfasser: Stanley, C F, Masri, F A
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The lacrimal drainage system drains fluid that exists normally on the surface of the eye into the nasal cavity. This paper describes a hand-held, personal-computer-based instrument to quantitatively estimate the resistance to fluid flow through the lacrimal drainage system. This resistance measurement is proposed as a possible means to diagnose blockage of the system and to verify the degree of change in resistance after clinical treatment. The instrument system consists of a hand-held syringe attached to pressure and flow transducers. Resistance is calculated as the ratio of differential pressure to flow rate and is displayed in real time on a computer monitor.
ISSN:0363-8855
DOI:10.1097/00004669-199101000-00014