Exchange of an obstructed loop nephrostomy catheter: technical note
Straightening of the distal loop of an obstructed self-retaining "Cope" style nephrostomy catheter may be impossible when the retaining thread is fixed by concretions. We describe a transcatheter technique using an 18 gauge Turner biopsy needle to place a 0.038 inch exchange guidewire for...
Gespeichert in:
Veröffentlicht in: | Cardiovascular and interventional radiology 1990-10, Vol.13 (5), p.327-328 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Straightening of the distal loop of an obstructed self-retaining "Cope" style nephrostomy catheter may be impossible when the retaining thread is fixed by concretions. We describe a transcatheter technique using an 18 gauge Turner biopsy needle to place a 0.038 inch exchange guidewire for retaining access to the collecting system. |
---|---|