Magnetic resonance artifact induced by the electrode Activa 3389: an in vitro and in vivo study
The electrode Activa 3389 is widely implanted for deep brain stimulation (DBS) and MRI is often used to control the position of the electrode. However, induced distorsion artifacts may result in imprecise localization and may lead to misinterpretations of the clinical effects and mechanisms of DBS....
Gespeichert in:
Veröffentlicht in: | Acta neurochirurgica 2004-02, Vol.146 (2), p.161-164 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 164 |
---|---|
container_issue | 2 |
container_start_page | 161 |
container_title | Acta neurochirurgica |
container_volume | 146 |
creator | Pollo, C Villemure, J-G Vingerhoets, F Ghika, J Maeder, P Meuli, R |
description | The electrode Activa 3389 is widely implanted for deep brain stimulation (DBS) and MRI is often used to control the position of the electrode. However, induced distorsion artifacts may result in imprecise localization and may lead to misinterpretations of the clinical effects and mechanisms of DBS.
In vitro 3D MR study: the proximal and distal contacts of one electrode were spotted by two localizers. The maximal artifact height (MAH) and width (MAW: measured on distal contact), and the distances between the artifact and the localizers (proximal, distal and lateral) were measured on 2 transverse and sagittal MR sequences with 90 degrees rotation of frequency-encoded gradient and phase direction. In vivo 3D MR study: coronal and sagittal reconstructions along the main axis of the electrode were performed on 10 postoperative MR (20 electrodes) to measure MAH and MAW. A Student t test was used to compare in vitro and in vivo measurements.
In vitro study: A MAH of 10.35 mm (+/-0.23) and MAW of 3.6 mm (+/-0.2) were found. We measured symmetrical extensions of the artifact over the distal contact. In vivo study: A MAH of 10.36 mm (+/-0.44) and MAW of 3.56 mm (+/-0.30) were obtained. No significant different artifact dimensions were measured between in vitro and in vivo studies (p |
doi_str_mv | 10.1007/s00701-003-0181-4 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_80160212</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2147572111</sourcerecordid><originalsourceid>FETCH-LOGICAL-c356t-fcb4021e3cb13a9bd54017fe18765d6bcf275613be0a564a97cf1d91f4a362a13</originalsourceid><addsrcrecordid>eNp9kU1rGzEQQEVoiVO3PyCXIHpIT5vMrD52N7dg0jbgkkt7FlppNllj7zqS1uB_HxkbAjn0Mpph3swgHmOXCDcIUN3GHAALAFEA1ljIM3YBjSyLHOBTziF3danrGfsS4ypXZSXFOZuhbLSoZHPBzB_7PFDqHQ8Ux8EOjrgNqe-sS7wf_OTI83bP0wtxWpNLYfTE713qd5YLUTd33A4Z5Ls-t3Luj8Vu5DFNfv-Vfe7sOtK30ztn_34-_F38LpZPvx4X98vCCaVT0blWQokkXIvCNq1XErDqCOtKK69b15WV0ihaAqu0tE3lOvQNdtIKXVoUc3Z93LsN4-tEMZlNHx2t13agcYqmBtT5QJnBH_8FKyWVEiBFJr9_IFfjFIb8iwwpQNGUh7t4hFwYYwzUmW3oNzbsDYI5SDJHSSZLMgdJRuaZq9Piqd2Qf584WRFvQ4yK0g</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>755013921</pqid></control><display><type>article</type><title>Magnetic resonance artifact induced by the electrode Activa 3389: an in vitro and in vivo study</title><source>MEDLINE</source><source>SpringerLink Journals - AutoHoldings</source><creator>Pollo, C ; Villemure, J-G ; Vingerhoets, F ; Ghika, J ; Maeder, P ; Meuli, R</creator><creatorcontrib>Pollo, C ; Villemure, J-G ; Vingerhoets, F ; Ghika, J ; Maeder, P ; Meuli, R</creatorcontrib><description>The electrode Activa 3389 is widely implanted for deep brain stimulation (DBS) and MRI is often used to control the position of the electrode. However, induced distorsion artifacts may result in imprecise localization and may lead to misinterpretations of the clinical effects and mechanisms of DBS.
In vitro 3D MR study: the proximal and distal contacts of one electrode were spotted by two localizers. The maximal artifact height (MAH) and width (MAW: measured on distal contact), and the distances between the artifact and the localizers (proximal, distal and lateral) were measured on 2 transverse and sagittal MR sequences with 90 degrees rotation of frequency-encoded gradient and phase direction. In vivo 3D MR study: coronal and sagittal reconstructions along the main axis of the electrode were performed on 10 postoperative MR (20 electrodes) to measure MAH and MAW. A Student t test was used to compare in vitro and in vivo measurements.
In vitro study: A MAH of 10.35 mm (+/-0.23) and MAW of 3.6 mm (+/-0.2) were found. We measured symmetrical extensions of the artifact over the distal contact. In vivo study: A MAH of 10.36 mm (+/-0.44) and MAW of 3.56 mm (+/-0.30) were obtained. No significant different artifact dimensions were measured between in vitro and in vivo studies (p<0.0001).
Precise 3D localization of the electrode in implanted patients is provided by MR identification of the limits of the distal contact artifact. The position of the other contacts is deduced given the size of the contacts and the intercontact distance.</description><identifier>ISSN: 0001-6268</identifier><identifier>EISSN: 0942-0940</identifier><identifier>DOI: 10.1007/s00701-003-0181-4</identifier><identifier>PMID: 14963749</identifier><language>eng</language><publisher>Austria: Springer Nature B.V</publisher><subject>Brain - pathology ; Electric Stimulation Therapy - instrumentation ; Electrodes ; Electrodes, Implanted ; Equipment Design ; Humans ; Image Processing, Computer-Assisted ; Imaging, Three-Dimensional ; In Vitro Techniques ; Magnetic Resonance Imaging ; Mathematical Computing ; Movement Disorders - pathology ; Movement Disorders - therapy ; Phantoms, Imaging</subject><ispartof>Acta neurochirurgica, 2004-02, Vol.146 (2), p.161-164</ispartof><rights>Springer-Verlag/Wien 2004</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c356t-fcb4021e3cb13a9bd54017fe18765d6bcf275613be0a564a97cf1d91f4a362a13</citedby></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,776,780,27901,27902</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/14963749$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Pollo, C</creatorcontrib><creatorcontrib>Villemure, J-G</creatorcontrib><creatorcontrib>Vingerhoets, F</creatorcontrib><creatorcontrib>Ghika, J</creatorcontrib><creatorcontrib>Maeder, P</creatorcontrib><creatorcontrib>Meuli, R</creatorcontrib><title>Magnetic resonance artifact induced by the electrode Activa 3389: an in vitro and in vivo study</title><title>Acta neurochirurgica</title><addtitle>Acta Neurochir (Wien)</addtitle><description>The electrode Activa 3389 is widely implanted for deep brain stimulation (DBS) and MRI is often used to control the position of the electrode. However, induced distorsion artifacts may result in imprecise localization and may lead to misinterpretations of the clinical effects and mechanisms of DBS.
In vitro 3D MR study: the proximal and distal contacts of one electrode were spotted by two localizers. The maximal artifact height (MAH) and width (MAW: measured on distal contact), and the distances between the artifact and the localizers (proximal, distal and lateral) were measured on 2 transverse and sagittal MR sequences with 90 degrees rotation of frequency-encoded gradient and phase direction. In vivo 3D MR study: coronal and sagittal reconstructions along the main axis of the electrode were performed on 10 postoperative MR (20 electrodes) to measure MAH and MAW. A Student t test was used to compare in vitro and in vivo measurements.
In vitro study: A MAH of 10.35 mm (+/-0.23) and MAW of 3.6 mm (+/-0.2) were found. We measured symmetrical extensions of the artifact over the distal contact. In vivo study: A MAH of 10.36 mm (+/-0.44) and MAW of 3.56 mm (+/-0.30) were obtained. No significant different artifact dimensions were measured between in vitro and in vivo studies (p<0.0001).
Precise 3D localization of the electrode in implanted patients is provided by MR identification of the limits of the distal contact artifact. The position of the other contacts is deduced given the size of the contacts and the intercontact distance.</description><subject>Brain - pathology</subject><subject>Electric Stimulation Therapy - instrumentation</subject><subject>Electrodes</subject><subject>Electrodes, Implanted</subject><subject>Equipment Design</subject><subject>Humans</subject><subject>Image Processing, Computer-Assisted</subject><subject>Imaging, Three-Dimensional</subject><subject>In Vitro Techniques</subject><subject>Magnetic Resonance Imaging</subject><subject>Mathematical Computing</subject><subject>Movement Disorders - pathology</subject><subject>Movement Disorders - therapy</subject><subject>Phantoms, Imaging</subject><issn>0001-6268</issn><issn>0942-0940</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2004</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><sourceid>BENPR</sourceid><recordid>eNp9kU1rGzEQQEVoiVO3PyCXIHpIT5vMrD52N7dg0jbgkkt7FlppNllj7zqS1uB_HxkbAjn0Mpph3swgHmOXCDcIUN3GHAALAFEA1ljIM3YBjSyLHOBTziF3danrGfsS4ypXZSXFOZuhbLSoZHPBzB_7PFDqHQ8Ux8EOjrgNqe-sS7wf_OTI83bP0wtxWpNLYfTE713qd5YLUTd33A4Z5Ls-t3Luj8Vu5DFNfv-Vfe7sOtK30ztn_34-_F38LpZPvx4X98vCCaVT0blWQokkXIvCNq1XErDqCOtKK69b15WV0ihaAqu0tE3lOvQNdtIKXVoUc3Z93LsN4-tEMZlNHx2t13agcYqmBtT5QJnBH_8FKyWVEiBFJr9_IFfjFIb8iwwpQNGUh7t4hFwYYwzUmW3oNzbsDYI5SDJHSSZLMgdJRuaZq9Piqd2Qf584WRFvQ4yK0g</recordid><startdate>200402</startdate><enddate>200402</enddate><creator>Pollo, C</creator><creator>Villemure, J-G</creator><creator>Vingerhoets, F</creator><creator>Ghika, J</creator><creator>Maeder, P</creator><creator>Meuli, R</creator><general>Springer Nature B.V</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>3V.</scope><scope>7TK</scope><scope>7U9</scope><scope>7X7</scope><scope>7XB</scope><scope>88E</scope><scope>8AO</scope><scope>8FI</scope><scope>8FJ</scope><scope>8FK</scope><scope>ABUWG</scope><scope>AFKRA</scope><scope>BENPR</scope><scope>CCPQU</scope><scope>FYUFA</scope><scope>GHDGH</scope><scope>H94</scope><scope>K9.</scope><scope>M0S</scope><scope>M1P</scope><scope>M7N</scope><scope>PQEST</scope><scope>PQQKQ</scope><scope>PQUKI</scope><scope>PRINS</scope><scope>7X8</scope></search><sort><creationdate>200402</creationdate><title>Magnetic resonance artifact induced by the electrode Activa 3389: an in vitro and in vivo study</title><author>Pollo, C ; Villemure, J-G ; Vingerhoets, F ; Ghika, J ; Maeder, P ; Meuli, R</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c356t-fcb4021e3cb13a9bd54017fe18765d6bcf275613be0a564a97cf1d91f4a362a13</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2004</creationdate><topic>Brain - pathology</topic><topic>Electric Stimulation Therapy - instrumentation</topic><topic>Electrodes</topic><topic>Electrodes, Implanted</topic><topic>Equipment Design</topic><topic>Humans</topic><topic>Image Processing, Computer-Assisted</topic><topic>Imaging, Three-Dimensional</topic><topic>In Vitro Techniques</topic><topic>Magnetic Resonance Imaging</topic><topic>Mathematical Computing</topic><topic>Movement Disorders - pathology</topic><topic>Movement Disorders - therapy</topic><topic>Phantoms, Imaging</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Pollo, C</creatorcontrib><creatorcontrib>Villemure, J-G</creatorcontrib><creatorcontrib>Vingerhoets, F</creatorcontrib><creatorcontrib>Ghika, J</creatorcontrib><creatorcontrib>Maeder, P</creatorcontrib><creatorcontrib>Meuli, R</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>ProQuest Central (Corporate)</collection><collection>Neurosciences Abstracts</collection><collection>Virology and AIDS Abstracts</collection><collection>Health & Medical Collection</collection><collection>ProQuest Central (purchase pre-March 2016)</collection><collection>Medical Database (Alumni Edition)</collection><collection>ProQuest Pharma Collection</collection><collection>Hospital Premium Collection</collection><collection>Hospital Premium Collection (Alumni Edition)</collection><collection>ProQuest Central (Alumni) (purchase pre-March 2016)</collection><collection>ProQuest Central (Alumni Edition)</collection><collection>ProQuest Central UK/Ireland</collection><collection>ProQuest Central</collection><collection>ProQuest One Community College</collection><collection>Health Research Premium Collection</collection><collection>Health Research Premium Collection (Alumni)</collection><collection>AIDS and Cancer Research Abstracts</collection><collection>ProQuest Health & Medical Complete (Alumni)</collection><collection>Health & Medical Collection (Alumni Edition)</collection><collection>Medical Database</collection><collection>Algology Mycology and Protozoology Abstracts (Microbiology C)</collection><collection>ProQuest One Academic Eastern Edition (DO NOT USE)</collection><collection>ProQuest One Academic</collection><collection>ProQuest One Academic UKI Edition</collection><collection>ProQuest Central China</collection><collection>MEDLINE - Academic</collection><jtitle>Acta neurochirurgica</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Pollo, C</au><au>Villemure, J-G</au><au>Vingerhoets, F</au><au>Ghika, J</au><au>Maeder, P</au><au>Meuli, R</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Magnetic resonance artifact induced by the electrode Activa 3389: an in vitro and in vivo study</atitle><jtitle>Acta neurochirurgica</jtitle><addtitle>Acta Neurochir (Wien)</addtitle><date>2004-02</date><risdate>2004</risdate><volume>146</volume><issue>2</issue><spage>161</spage><epage>164</epage><pages>161-164</pages><issn>0001-6268</issn><eissn>0942-0940</eissn><abstract>The electrode Activa 3389 is widely implanted for deep brain stimulation (DBS) and MRI is often used to control the position of the electrode. However, induced distorsion artifacts may result in imprecise localization and may lead to misinterpretations of the clinical effects and mechanisms of DBS.
In vitro 3D MR study: the proximal and distal contacts of one electrode were spotted by two localizers. The maximal artifact height (MAH) and width (MAW: measured on distal contact), and the distances between the artifact and the localizers (proximal, distal and lateral) were measured on 2 transverse and sagittal MR sequences with 90 degrees rotation of frequency-encoded gradient and phase direction. In vivo 3D MR study: coronal and sagittal reconstructions along the main axis of the electrode were performed on 10 postoperative MR (20 electrodes) to measure MAH and MAW. A Student t test was used to compare in vitro and in vivo measurements.
In vitro study: A MAH of 10.35 mm (+/-0.23) and MAW of 3.6 mm (+/-0.2) were found. We measured symmetrical extensions of the artifact over the distal contact. In vivo study: A MAH of 10.36 mm (+/-0.44) and MAW of 3.56 mm (+/-0.30) were obtained. No significant different artifact dimensions were measured between in vitro and in vivo studies (p<0.0001).
Precise 3D localization of the electrode in implanted patients is provided by MR identification of the limits of the distal contact artifact. The position of the other contacts is deduced given the size of the contacts and the intercontact distance.</abstract><cop>Austria</cop><pub>Springer Nature B.V</pub><pmid>14963749</pmid><doi>10.1007/s00701-003-0181-4</doi><tpages>4</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0001-6268 |
ispartof | Acta neurochirurgica, 2004-02, Vol.146 (2), p.161-164 |
issn | 0001-6268 0942-0940 |
language | eng |
recordid | cdi_proquest_miscellaneous_80160212 |
source | MEDLINE; SpringerLink Journals - AutoHoldings |
subjects | Brain - pathology Electric Stimulation Therapy - instrumentation Electrodes Electrodes, Implanted Equipment Design Humans Image Processing, Computer-Assisted Imaging, Three-Dimensional In Vitro Techniques Magnetic Resonance Imaging Mathematical Computing Movement Disorders - pathology Movement Disorders - therapy Phantoms, Imaging |
title | Magnetic resonance artifact induced by the electrode Activa 3389: an in vitro and in vivo study |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-08T08%3A43%3A21IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Magnetic%20resonance%20artifact%20induced%20by%20the%20electrode%20Activa%203389:%20an%20in%20vitro%20and%20in%20vivo%20study&rft.jtitle=Acta%20neurochirurgica&rft.au=Pollo,%20C&rft.date=2004-02&rft.volume=146&rft.issue=2&rft.spage=161&rft.epage=164&rft.pages=161-164&rft.issn=0001-6268&rft.eissn=0942-0940&rft_id=info:doi/10.1007/s00701-003-0181-4&rft_dat=%3Cproquest_cross%3E2147572111%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=755013921&rft_id=info:pmid/14963749&rfr_iscdi=true |