The development of a new direct digital extra-oral radiographic system prototype using a thin-film transistor panel
To develop a new method of direct digital image acquisition for extra-oral radiography. A selenium detector was combined with a thin-film transistor sensor and integrated with current technology for computed radiography in a system named Direct Radiography (DR). The effective image area of the proto...
Gespeichert in:
Veröffentlicht in: | Dento-maxillo-facial radiology 1998-05, Vol.27 (3), p.172-177 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To develop a new method of direct digital image acquisition for extra-oral radiography.
A selenium detector was combined with a thin-film transistor sensor and integrated with current technology for computed radiography in a system named Direct Radiography (DR). The effective image area of the prototype is 178 mm x 213.5 mm, and consists of a 1280 x 1536 pixel matrix with a pixel pitch of 139 microns x 139 microns. The electrical characteristics were evaluated and image quality was tested with a dried human mandible and half-skull phantom. The resultant images were subjectively reviewed by seven oral and maxillofacial radiologists.
The prototype had a signal-to-noise ratio of 400:1 when the exposure was 250 x 10(-8) C/Kg. Spatial resolution was equal to the Nyquist frequency of 3.55 cycles/mm. The images of the phantoms were considered to be of equivalent quality to conventional screen-film images and demonstrated the normal anatomical structures used for clinical evaluation. Three to four images per minute were generated with this system.
The DR system has an X-ray sensitivity comparable with a 400-speed film-screen system and appears suitable for clinical radiography. Further evaluation is therefore planned. |
---|---|