An automated measurement system for characterization of RF and gradient coil parameters

A fully automated laboratory‐based measurement system for characterization of coil system parameters is presented. This method uses an inexpensive personal computer (PC)‐controlled stepper motor positioning system in conjunction with a network/spectrum analyzer and an analog‐to‐digital converter (A/...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of magnetic resonance imaging 1998-05, Vol.8 (3), p.740-747
Hauptverfasser: Boyer, Jeffrey S., Wright, Steven M., Porter, Jay R.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A fully automated laboratory‐based measurement system for characterization of coil system parameters is presented. This method uses an inexpensive personal computer (PC)‐controlled stepper motor positioning system in conjunction with a network/spectrum analyzer and an analog‐to‐digital converter (A/D) board that allows high resolution data acquisition in an unattended manner. A graphical interface was created for complete control of stepper motor movement, measurement, and data acquisition. The system is capable of performing a wide range of measurements that can, either individually or combined, characterize radiofre‐quency (RF) and gradient coils used in MRI. Measurement methods, theory, and results for conductor and shield current distributions, mutual impedance, and magnetic fields are given. Comparisons with theoretical calculations are included to validate the accuracy and utility of the system.
ISSN:1053-1807
1522-2586
DOI:10.1002/jmri.1880080334