Home uroflowmetry by means of the Da Capo home uroflowmeter

The Da Capo home flowmeter was tested versus the Urodyn 1000 flowmeter. The two flowmeters are based on different principles. The Da Capo is a portable, battery powered flowmeter designed to record all voidings during a period of time (e.g. 24 h) for a single patient. The flowmeters were tested with...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:European urology 1998, Vol.33 (1), p.64-68
Hauptverfasser: Jørgensen, J B, Jacobsen, H L, Bagi, P, Hvarnes, H, Colstrup, H
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 68
container_issue 1
container_start_page 64
container_title European urology
container_volume 33
creator Jørgensen, J B
Jacobsen, H L
Bagi, P
Hvarnes, H
Colstrup, H
description The Da Capo home flowmeter was tested versus the Urodyn 1000 flowmeter. The two flowmeters are based on different principles. The Da Capo is a portable, battery powered flowmeter designed to record all voidings during a period of time (e.g. 24 h) for a single patient. The flowmeters were tested with regard to accuracy of measurement of the voided volume and maximum flow (Qmax). Further, the Da Capo was tested by 10 healthy male volunteers, median age 47 years, range 21-57. Both flowmeters were very accurate measuring Qmax and voided volume. A few artifacts arose, i.e. extremely high Qmax values were recorded. All test persons found the flowmeter easy to handle. The weight transducer based Da Capo home flowmeter proved as accurate as the stationary flowmeters. It is easy to handle and it provides all-day monitoring of uroflow and voided volume.
doi_str_mv 10.1159/000019534
format Article
fullrecord <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_proquest_miscellaneous_79690883</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>79690883</sourcerecordid><originalsourceid>FETCH-LOGICAL-p206t-ac61735f2b8d1c5161124d113f2a00cd9260fb3d68b599bc579e4ca54c3049783</originalsourceid><addsrcrecordid>eNpVj71OwzAYRT2ASikMPACSJ7aAP__FFhMKhSJVYoE5sh1bLUrqYCdCeXsikYW7nOEeXekidAPkHkDoBzIHtGD8DK0JI7SgiqkLdJnz19wwodkKrTQvgXC6Ro-72Hk8phja-NP5IU3YTrjz5pRxDHg4ePxscGX6iA__TZ-u0HkwbfbXCzfo82X7Ue2K_fvrW_W0L3pK5FAYJ6FkIlCrGnACJADlDQAL1BDiGk0lCZY1UlmhtXWi1J47I7hjhOtSsQ26-9vtU_wefR7q7pidb1tz8nHMdamlJkqxWbxdxNF2vqn7dOxMmurlLfsFZzlRwg</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>79690883</pqid></control><display><type>article</type><title>Home uroflowmetry by means of the Da Capo home uroflowmeter</title><source>Karger Journals</source><source>MEDLINE</source><creator>Jørgensen, J B ; Jacobsen, H L ; Bagi, P ; Hvarnes, H ; Colstrup, H</creator><creatorcontrib>Jørgensen, J B ; Jacobsen, H L ; Bagi, P ; Hvarnes, H ; Colstrup, H</creatorcontrib><description>The Da Capo home flowmeter was tested versus the Urodyn 1000 flowmeter. The two flowmeters are based on different principles. The Da Capo is a portable, battery powered flowmeter designed to record all voidings during a period of time (e.g. 24 h) for a single patient. The flowmeters were tested with regard to accuracy of measurement of the voided volume and maximum flow (Qmax). Further, the Da Capo was tested by 10 healthy male volunteers, median age 47 years, range 21-57. Both flowmeters were very accurate measuring Qmax and voided volume. A few artifacts arose, i.e. extremely high Qmax values were recorded. All test persons found the flowmeter easy to handle. The weight transducer based Da Capo home flowmeter proved as accurate as the stationary flowmeters. It is easy to handle and it provides all-day monitoring of uroflow and voided volume.</description><identifier>ISSN: 0302-2838</identifier><identifier>DOI: 10.1159/000019534</identifier><identifier>PMID: 9471042</identifier><language>eng</language><publisher>Switzerland</publisher><subject>Adult ; Humans ; Male ; Middle Aged ; Reproducibility of Results ; Rheology - instrumentation ; Self Care ; Urination - physiology ; Urodynamics</subject><ispartof>European urology, 1998, Vol.33 (1), p.64-68</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,776,780,4010,27900,27901,27902</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/9471042$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Jørgensen, J B</creatorcontrib><creatorcontrib>Jacobsen, H L</creatorcontrib><creatorcontrib>Bagi, P</creatorcontrib><creatorcontrib>Hvarnes, H</creatorcontrib><creatorcontrib>Colstrup, H</creatorcontrib><title>Home uroflowmetry by means of the Da Capo home uroflowmeter</title><title>European urology</title><addtitle>Eur Urol</addtitle><description>The Da Capo home flowmeter was tested versus the Urodyn 1000 flowmeter. The two flowmeters are based on different principles. The Da Capo is a portable, battery powered flowmeter designed to record all voidings during a period of time (e.g. 24 h) for a single patient. The flowmeters were tested with regard to accuracy of measurement of the voided volume and maximum flow (Qmax). Further, the Da Capo was tested by 10 healthy male volunteers, median age 47 years, range 21-57. Both flowmeters were very accurate measuring Qmax and voided volume. A few artifacts arose, i.e. extremely high Qmax values were recorded. All test persons found the flowmeter easy to handle. The weight transducer based Da Capo home flowmeter proved as accurate as the stationary flowmeters. It is easy to handle and it provides all-day monitoring of uroflow and voided volume.</description><subject>Adult</subject><subject>Humans</subject><subject>Male</subject><subject>Middle Aged</subject><subject>Reproducibility of Results</subject><subject>Rheology - instrumentation</subject><subject>Self Care</subject><subject>Urination - physiology</subject><subject>Urodynamics</subject><issn>0302-2838</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>1998</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNpVj71OwzAYRT2ASikMPACSJ7aAP__FFhMKhSJVYoE5sh1bLUrqYCdCeXsikYW7nOEeXekidAPkHkDoBzIHtGD8DK0JI7SgiqkLdJnz19wwodkKrTQvgXC6Ro-72Hk8phja-NP5IU3YTrjz5pRxDHg4ePxscGX6iA__TZ-u0HkwbfbXCzfo82X7Ue2K_fvrW_W0L3pK5FAYJ6FkIlCrGnACJADlDQAL1BDiGk0lCZY1UlmhtXWi1J47I7hjhOtSsQ26-9vtU_wefR7q7pidb1tz8nHMdamlJkqxWbxdxNF2vqn7dOxMmurlLfsFZzlRwg</recordid><startdate>1998</startdate><enddate>1998</enddate><creator>Jørgensen, J B</creator><creator>Jacobsen, H L</creator><creator>Bagi, P</creator><creator>Hvarnes, H</creator><creator>Colstrup, H</creator><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>7X8</scope></search><sort><creationdate>1998</creationdate><title>Home uroflowmetry by means of the Da Capo home uroflowmeter</title><author>Jørgensen, J B ; Jacobsen, H L ; Bagi, P ; Hvarnes, H ; Colstrup, H</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-p206t-ac61735f2b8d1c5161124d113f2a00cd9260fb3d68b599bc579e4ca54c3049783</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>1998</creationdate><topic>Adult</topic><topic>Humans</topic><topic>Male</topic><topic>Middle Aged</topic><topic>Reproducibility of Results</topic><topic>Rheology - instrumentation</topic><topic>Self Care</topic><topic>Urination - physiology</topic><topic>Urodynamics</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Jørgensen, J B</creatorcontrib><creatorcontrib>Jacobsen, H L</creatorcontrib><creatorcontrib>Bagi, P</creatorcontrib><creatorcontrib>Hvarnes, H</creatorcontrib><creatorcontrib>Colstrup, H</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>MEDLINE - Academic</collection><jtitle>European urology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Jørgensen, J B</au><au>Jacobsen, H L</au><au>Bagi, P</au><au>Hvarnes, H</au><au>Colstrup, H</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Home uroflowmetry by means of the Da Capo home uroflowmeter</atitle><jtitle>European urology</jtitle><addtitle>Eur Urol</addtitle><date>1998</date><risdate>1998</risdate><volume>33</volume><issue>1</issue><spage>64</spage><epage>68</epage><pages>64-68</pages><issn>0302-2838</issn><abstract>The Da Capo home flowmeter was tested versus the Urodyn 1000 flowmeter. The two flowmeters are based on different principles. The Da Capo is a portable, battery powered flowmeter designed to record all voidings during a period of time (e.g. 24 h) for a single patient. The flowmeters were tested with regard to accuracy of measurement of the voided volume and maximum flow (Qmax). Further, the Da Capo was tested by 10 healthy male volunteers, median age 47 years, range 21-57. Both flowmeters were very accurate measuring Qmax and voided volume. A few artifacts arose, i.e. extremely high Qmax values were recorded. All test persons found the flowmeter easy to handle. The weight transducer based Da Capo home flowmeter proved as accurate as the stationary flowmeters. It is easy to handle and it provides all-day monitoring of uroflow and voided volume.</abstract><cop>Switzerland</cop><pmid>9471042</pmid><doi>10.1159/000019534</doi><tpages>5</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0302-2838
ispartof European urology, 1998, Vol.33 (1), p.64-68
issn 0302-2838
language eng
recordid cdi_proquest_miscellaneous_79690883
source Karger Journals; MEDLINE
subjects Adult
Humans
Male
Middle Aged
Reproducibility of Results
Rheology - instrumentation
Self Care
Urination - physiology
Urodynamics
title Home uroflowmetry by means of the Da Capo home uroflowmeter
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-10T12%3A25%3A50IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Home%20uroflowmetry%20by%20means%20of%20the%20Da%20Capo%20home%20uroflowmeter&rft.jtitle=European%20urology&rft.au=J%C3%B8rgensen,%20J%20B&rft.date=1998&rft.volume=33&rft.issue=1&rft.spage=64&rft.epage=68&rft.pages=64-68&rft.issn=0302-2838&rft_id=info:doi/10.1159/000019534&rft_dat=%3Cproquest_pubme%3E79690883%3C/proquest_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=79690883&rft_id=info:pmid/9471042&rfr_iscdi=true