Graft force-setting technique in reconstruction of the anterior cruciate ligament
The objective of this work was to present a method for controlling the mechanical state of the knee during graft fixation in an ACL reconstruction; that is, adjusting the unloaded length of the graft at the time of fixation so that the force-carrying characteristics of the graft match that of the in...
Gespeichert in:
Veröffentlicht in: | The American journal of sports medicine 1990-01, Vol.18 (1), p.12-19 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The objective of this work was to present a method for controlling the mechanical state of the knee during graft fixation
in an ACL reconstruction; that is, adjusting the unloaded length of the graft at the time of fixation so that the force-carrying
characteristics of the graft match that of the intact ACL, when the normal and recon structed knees are subjected to the same
external load. To demonstrate the new method, an experimental knee testing system was used to measure ligament and graft forces
with buckle transducers, as external loads were applied to five fresh knee specimens with a pneumatic load apparatus. ACL
and collateral ligament force data were collected as anteriorly directed loads of 90 N (20 pounds) were applied at 0°, 30°,
60°, and 90° of flexion in the normal intact knee and the same knee following ACL excision. A composite graft was prepared,
con sisting of distally based semitendinosus and gracilis autogenous tissue augmented with the Kennedy Liga ment Augmentation
Device. Femoral fixation was ob tained using a fixation plug that was inserted through the lateral femoral condyle. With a
90 N anterior tibial load applied at 30° of flexion and a single buckle transducer measuring the total composite graft force,
the graft length was varied using the fixation plug until the force matched that seen by the intact ACL under the same loading
condition. The graft was then clamped in the fixation plug and the anterior load sequence at the four flexion angles was applied
to this reconstructed knee state. Ligament and graft forces were measured and compared to the normal intact knee and cut ACL
force data. The proposed force-setting method of ACL reconstruction resulted in graft forces that were closer to normal than
when using current reconstruction meth ods based on isometry and pretension, and allowed better reproducibility of total graft
forces among a series of knees. |
---|---|
ISSN: | 0363-5465 1552-3365 |
DOI: | 10.1177/036354659001800102 |