Improved Pattern of Rate Responsiveness with Dynamic Slope Setting for the QT Sensing Pacemaker

We have recently described the electrophysiological basis of a new algorithm for the QT (TX) sensing rate responsive pacemaker. By using the new software program running on the standard programmer it has been possible to simulate the new algorithm in ten patients with complete heart block (seven pat...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 1989-02, Vol.12 (2), p.311-320
Hauptverfasser: BAIG, M.W., WILSON, J., BOUTE, W., BEGEMANN, M.J.S., COBBOLD, J.P., PERRINS, E.J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 320
container_issue 2
container_start_page 311
container_title Pacing and clinical electrophysiology
container_volume 12
creator BAIG, M.W.
WILSON, J.
BOUTE, W.
BEGEMANN, M.J.S.
COBBOLD, J.P.
PERRINS, E.J.
description We have recently described the electrophysiological basis of a new algorithm for the QT (TX) sensing rate responsive pacemaker. By using the new software program running on the standard programmer it has been possible to simulate the new algorithm in ten patients with complete heart block (seven patients had implanted TX units and three were paced with an external TX pacemaker) during routine exercise testing. In this way a single‐blind, intra‐patient comparison of the pattern of pacing rote change using both the existing and new algorithms was possible. In nine out of the ten cases the time taken to increase the pacing rate from 70 to 80 bpm was reduced significantly when the new algorithm was used (P = 0.037). Additionally, the correlation between the atrial and ventricular rates in those patients with normal sinus node function (seven patients) was determined. In all cases we have observed a significantly improved correlation between the atrial and ventricular paced rates during exercise with the new algorithm (P < 0.001).
doi_str_mv 10.1111/j.1540-8159.1989.tb02664.x
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_78936858</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>78936858</sourcerecordid><originalsourceid>FETCH-LOGICAL-c4081-c86e7fbe280cbee60b58aa9405799ea5d77003f6e431930b911ab26655ed95693</originalsourceid><addsrcrecordid>eNqVkE1PGzEQhq2qFU1pf0Ili0Nvu7Xjby4IBUiRUBsgFdws72YWNuxHsB1I_n29SpR75zLSvPO-o3kQOqEkp6l-LnMqOMk0FSanRps8FmQsJc83H9DoIH1EI0K5yjTT5jP6EsKSECIJF0foaMylppyOkL1uV75_gwWeuRjBd7iv8J2LgO8grPou1G_QQQj4vY7P-GLbubYu8X3TrwDfQ4x194Sr3uP4DPh2nkbJkUYzV0LrXsB_RZ8q1wT4tu_H6O_V5XzyK7v5M72enN9kJSeaZqWWoKoCxpqUBYAkhdDOGU6EMgacWChFCKskcEYNI4Wh1BXpZSFgYYQ07Bj92OWmb17XEKJt61BC07gO-nWwShsmtdBp8XS3WPo-BA-VXfm6dX5rKbEDXbu0A0I7ILQDXbunazfJ_H1_ZV20sDhY9ziTfrbT3-sGtv-RbGfnk0tGh4Rsl1CHCJtDgvMvViqmhH34PbXm4lbNH6czK9k_itKZhg</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>78936858</pqid></control><display><type>article</type><title>Improved Pattern of Rate Responsiveness with Dynamic Slope Setting for the QT Sensing Pacemaker</title><source>MEDLINE</source><source>Wiley Online Library Journals Frontfile Complete</source><creator>BAIG, M.W. ; WILSON, J. ; BOUTE, W. ; BEGEMANN, M.J.S. ; COBBOLD, J.P. ; PERRINS, E.J.</creator><creatorcontrib>BAIG, M.W. ; WILSON, J. ; BOUTE, W. ; BEGEMANN, M.J.S. ; COBBOLD, J.P. ; PERRINS, E.J.</creatorcontrib><description>We have recently described the electrophysiological basis of a new algorithm for the QT (TX) sensing rate responsive pacemaker. By using the new software program running on the standard programmer it has been possible to simulate the new algorithm in ten patients with complete heart block (seven patients had implanted TX units and three were paced with an external TX pacemaker) during routine exercise testing. In this way a single‐blind, intra‐patient comparison of the pattern of pacing rote change using both the existing and new algorithms was possible. In nine out of the ten cases the time taken to increase the pacing rate from 70 to 80 bpm was reduced significantly when the new algorithm was used (P = 0.037). Additionally, the correlation between the atrial and ventricular rates in those patients with normal sinus node function (seven patients) was determined. In all cases we have observed a significantly improved correlation between the atrial and ventricular paced rates during exercise with the new algorithm (P &lt; 0.001).</description><identifier>ISSN: 0147-8389</identifier><identifier>EISSN: 1540-8159</identifier><identifier>DOI: 10.1111/j.1540-8159.1989.tb02664.x</identifier><identifier>PMID: 2468141</identifier><language>eng</language><publisher>Oxford, UK: Blackwell Publishing Ltd</publisher><subject>Adult ; Aged ; algorithm ; Computer Simulation ; dynamic slope ; Electrocardiography ; Exercise Test ; Female ; Heart Block - therapy ; Heart Rate ; Humans ; Male ; Middle Aged ; Models, Cardiovascular ; Pacemaker, Artificial ; QT sensing pacemaker</subject><ispartof>Pacing and clinical electrophysiology, 1989-02, Vol.12 (2), p.311-320</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c4081-c86e7fbe280cbee60b58aa9405799ea5d77003f6e431930b911ab26655ed95693</citedby><cites>FETCH-LOGICAL-c4081-c86e7fbe280cbee60b58aa9405799ea5d77003f6e431930b911ab26655ed95693</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://onlinelibrary.wiley.com/doi/pdf/10.1111%2Fj.1540-8159.1989.tb02664.x$$EPDF$$P50$$Gwiley$$H</linktopdf><linktohtml>$$Uhttps://onlinelibrary.wiley.com/doi/full/10.1111%2Fj.1540-8159.1989.tb02664.x$$EHTML$$P50$$Gwiley$$H</linktohtml><link.rule.ids>314,776,780,1411,27903,27904,45553,45554</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/2468141$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>BAIG, M.W.</creatorcontrib><creatorcontrib>WILSON, J.</creatorcontrib><creatorcontrib>BOUTE, W.</creatorcontrib><creatorcontrib>BEGEMANN, M.J.S.</creatorcontrib><creatorcontrib>COBBOLD, J.P.</creatorcontrib><creatorcontrib>PERRINS, E.J.</creatorcontrib><title>Improved Pattern of Rate Responsiveness with Dynamic Slope Setting for the QT Sensing Pacemaker</title><title>Pacing and clinical electrophysiology</title><addtitle>Pacing Clin Electrophysiol</addtitle><description>We have recently described the electrophysiological basis of a new algorithm for the QT (TX) sensing rate responsive pacemaker. By using the new software program running on the standard programmer it has been possible to simulate the new algorithm in ten patients with complete heart block (seven patients had implanted TX units and three were paced with an external TX pacemaker) during routine exercise testing. In this way a single‐blind, intra‐patient comparison of the pattern of pacing rote change using both the existing and new algorithms was possible. In nine out of the ten cases the time taken to increase the pacing rate from 70 to 80 bpm was reduced significantly when the new algorithm was used (P = 0.037). Additionally, the correlation between the atrial and ventricular rates in those patients with normal sinus node function (seven patients) was determined. In all cases we have observed a significantly improved correlation between the atrial and ventricular paced rates during exercise with the new algorithm (P &lt; 0.001).</description><subject>Adult</subject><subject>Aged</subject><subject>algorithm</subject><subject>Computer Simulation</subject><subject>dynamic slope</subject><subject>Electrocardiography</subject><subject>Exercise Test</subject><subject>Female</subject><subject>Heart Block - therapy</subject><subject>Heart Rate</subject><subject>Humans</subject><subject>Male</subject><subject>Middle Aged</subject><subject>Models, Cardiovascular</subject><subject>Pacemaker, Artificial</subject><subject>QT sensing pacemaker</subject><issn>0147-8389</issn><issn>1540-8159</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>1989</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNqVkE1PGzEQhq2qFU1pf0Ili0Nvu7Xjby4IBUiRUBsgFdws72YWNuxHsB1I_n29SpR75zLSvPO-o3kQOqEkp6l-LnMqOMk0FSanRps8FmQsJc83H9DoIH1EI0K5yjTT5jP6EsKSECIJF0foaMylppyOkL1uV75_gwWeuRjBd7iv8J2LgO8grPou1G_QQQj4vY7P-GLbubYu8X3TrwDfQ4x194Sr3uP4DPh2nkbJkUYzV0LrXsB_RZ8q1wT4tu_H6O_V5XzyK7v5M72enN9kJSeaZqWWoKoCxpqUBYAkhdDOGU6EMgacWChFCKskcEYNI4Wh1BXpZSFgYYQ07Bj92OWmb17XEKJt61BC07gO-nWwShsmtdBp8XS3WPo-BA-VXfm6dX5rKbEDXbu0A0I7ILQDXbunazfJ_H1_ZV20sDhY9ziTfrbT3-sGtv-RbGfnk0tGh4Rsl1CHCJtDgvMvViqmhH34PbXm4lbNH6czK9k_itKZhg</recordid><startdate>198902</startdate><enddate>198902</enddate><creator>BAIG, M.W.</creator><creator>WILSON, J.</creator><creator>BOUTE, W.</creator><creator>BEGEMANN, M.J.S.</creator><creator>COBBOLD, J.P.</creator><creator>PERRINS, E.J.</creator><general>Blackwell Publishing Ltd</general><scope>BSCLL</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>198902</creationdate><title>Improved Pattern of Rate Responsiveness with Dynamic Slope Setting for the QT Sensing Pacemaker</title><author>BAIG, M.W. ; WILSON, J. ; BOUTE, W. ; BEGEMANN, M.J.S. ; COBBOLD, J.P. ; PERRINS, E.J.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c4081-c86e7fbe280cbee60b58aa9405799ea5d77003f6e431930b911ab26655ed95693</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>1989</creationdate><topic>Adult</topic><topic>Aged</topic><topic>algorithm</topic><topic>Computer Simulation</topic><topic>dynamic slope</topic><topic>Electrocardiography</topic><topic>Exercise Test</topic><topic>Female</topic><topic>Heart Block - therapy</topic><topic>Heart Rate</topic><topic>Humans</topic><topic>Male</topic><topic>Middle Aged</topic><topic>Models, Cardiovascular</topic><topic>Pacemaker, Artificial</topic><topic>QT sensing pacemaker</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>BAIG, M.W.</creatorcontrib><creatorcontrib>WILSON, J.</creatorcontrib><creatorcontrib>BOUTE, W.</creatorcontrib><creatorcontrib>BEGEMANN, M.J.S.</creatorcontrib><creatorcontrib>COBBOLD, J.P.</creatorcontrib><creatorcontrib>PERRINS, E.J.</creatorcontrib><collection>Istex</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Pacing and clinical electrophysiology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>BAIG, M.W.</au><au>WILSON, J.</au><au>BOUTE, W.</au><au>BEGEMANN, M.J.S.</au><au>COBBOLD, J.P.</au><au>PERRINS, E.J.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Improved Pattern of Rate Responsiveness with Dynamic Slope Setting for the QT Sensing Pacemaker</atitle><jtitle>Pacing and clinical electrophysiology</jtitle><addtitle>Pacing Clin Electrophysiol</addtitle><date>1989-02</date><risdate>1989</risdate><volume>12</volume><issue>2</issue><spage>311</spage><epage>320</epage><pages>311-320</pages><issn>0147-8389</issn><eissn>1540-8159</eissn><abstract>We have recently described the electrophysiological basis of a new algorithm for the QT (TX) sensing rate responsive pacemaker. By using the new software program running on the standard programmer it has been possible to simulate the new algorithm in ten patients with complete heart block (seven patients had implanted TX units and three were paced with an external TX pacemaker) during routine exercise testing. In this way a single‐blind, intra‐patient comparison of the pattern of pacing rote change using both the existing and new algorithms was possible. In nine out of the ten cases the time taken to increase the pacing rate from 70 to 80 bpm was reduced significantly when the new algorithm was used (P = 0.037). Additionally, the correlation between the atrial and ventricular rates in those patients with normal sinus node function (seven patients) was determined. In all cases we have observed a significantly improved correlation between the atrial and ventricular paced rates during exercise with the new algorithm (P &lt; 0.001).</abstract><cop>Oxford, UK</cop><pub>Blackwell Publishing Ltd</pub><pmid>2468141</pmid><doi>10.1111/j.1540-8159.1989.tb02664.x</doi><tpages>10</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0147-8389
ispartof Pacing and clinical electrophysiology, 1989-02, Vol.12 (2), p.311-320
issn 0147-8389
1540-8159
language eng
recordid cdi_proquest_miscellaneous_78936858
source MEDLINE; Wiley Online Library Journals Frontfile Complete
subjects Adult
Aged
algorithm
Computer Simulation
dynamic slope
Electrocardiography
Exercise Test
Female
Heart Block - therapy
Heart Rate
Humans
Male
Middle Aged
Models, Cardiovascular
Pacemaker, Artificial
QT sensing pacemaker
title Improved Pattern of Rate Responsiveness with Dynamic Slope Setting for the QT Sensing Pacemaker
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-26T06%3A23%3A45IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Improved%20Pattern%20of%20Rate%20Responsiveness%20with%20Dynamic%20Slope%20Setting%20for%20the%20QT%20Sensing%20Pacemaker&rft.jtitle=Pacing%20and%20clinical%20electrophysiology&rft.au=BAIG,%20M.W.&rft.date=1989-02&rft.volume=12&rft.issue=2&rft.spage=311&rft.epage=320&rft.pages=311-320&rft.issn=0147-8389&rft.eissn=1540-8159&rft_id=info:doi/10.1111/j.1540-8159.1989.tb02664.x&rft_dat=%3Cproquest_cross%3E78936858%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=78936858&rft_id=info:pmid/2468141&rfr_iscdi=true