Improved Pattern of Rate Responsiveness with Dynamic Slope Setting for the QT Sensing Pacemaker
We have recently described the electrophysiological basis of a new algorithm for the QT (TX) sensing rate responsive pacemaker. By using the new software program running on the standard programmer it has been possible to simulate the new algorithm in ten patients with complete heart block (seven pat...
Gespeichert in:
Veröffentlicht in: | Pacing and clinical electrophysiology 1989-02, Vol.12 (2), p.311-320 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 320 |
---|---|
container_issue | 2 |
container_start_page | 311 |
container_title | Pacing and clinical electrophysiology |
container_volume | 12 |
creator | BAIG, M.W. WILSON, J. BOUTE, W. BEGEMANN, M.J.S. COBBOLD, J.P. PERRINS, E.J. |
description | We have recently described the electrophysiological basis of a new algorithm for the QT (TX) sensing rate responsive pacemaker. By using the new software program running on the standard programmer it has been possible to simulate the new algorithm in ten patients with complete heart block (seven patients had implanted TX units and three were paced with an external TX pacemaker) during routine exercise testing. In this way a single‐blind, intra‐patient comparison of the pattern of pacing rote change using both the existing and new algorithms was possible. In nine out of the ten cases the time taken to increase the pacing rate from 70 to 80 bpm was reduced significantly when the new algorithm was used (P = 0.037). Additionally, the correlation between the atrial and ventricular rates in those patients with normal sinus node function (seven patients) was determined. In all cases we have observed a significantly improved correlation between the atrial and ventricular paced rates during exercise with the new algorithm (P < 0.001). |
doi_str_mv | 10.1111/j.1540-8159.1989.tb02664.x |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_78936858</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>78936858</sourcerecordid><originalsourceid>FETCH-LOGICAL-c4081-c86e7fbe280cbee60b58aa9405799ea5d77003f6e431930b911ab26655ed95693</originalsourceid><addsrcrecordid>eNqVkE1PGzEQhq2qFU1pf0Ili0Nvu7Xjby4IBUiRUBsgFdws72YWNuxHsB1I_n29SpR75zLSvPO-o3kQOqEkp6l-LnMqOMk0FSanRps8FmQsJc83H9DoIH1EI0K5yjTT5jP6EsKSECIJF0foaMylppyOkL1uV75_gwWeuRjBd7iv8J2LgO8grPou1G_QQQj4vY7P-GLbubYu8X3TrwDfQ4x194Sr3uP4DPh2nkbJkUYzV0LrXsB_RZ8q1wT4tu_H6O_V5XzyK7v5M72enN9kJSeaZqWWoKoCxpqUBYAkhdDOGU6EMgacWChFCKskcEYNI4Wh1BXpZSFgYYQ07Bj92OWmb17XEKJt61BC07gO-nWwShsmtdBp8XS3WPo-BA-VXfm6dX5rKbEDXbu0A0I7ILQDXbunazfJ_H1_ZV20sDhY9ziTfrbT3-sGtv-RbGfnk0tGh4Rsl1CHCJtDgvMvViqmhH34PbXm4lbNH6czK9k_itKZhg</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>78936858</pqid></control><display><type>article</type><title>Improved Pattern of Rate Responsiveness with Dynamic Slope Setting for the QT Sensing Pacemaker</title><source>MEDLINE</source><source>Wiley Online Library Journals Frontfile Complete</source><creator>BAIG, M.W. ; WILSON, J. ; BOUTE, W. ; BEGEMANN, M.J.S. ; COBBOLD, J.P. ; PERRINS, E.J.</creator><creatorcontrib>BAIG, M.W. ; WILSON, J. ; BOUTE, W. ; BEGEMANN, M.J.S. ; COBBOLD, J.P. ; PERRINS, E.J.</creatorcontrib><description>We have recently described the electrophysiological basis of a new algorithm for the QT (TX) sensing rate responsive pacemaker. By using the new software program running on the standard programmer it has been possible to simulate the new algorithm in ten patients with complete heart block (seven patients had implanted TX units and three were paced with an external TX pacemaker) during routine exercise testing. In this way a single‐blind, intra‐patient comparison of the pattern of pacing rote change using both the existing and new algorithms was possible. In nine out of the ten cases the time taken to increase the pacing rate from 70 to 80 bpm was reduced significantly when the new algorithm was used (P = 0.037). Additionally, the correlation between the atrial and ventricular rates in those patients with normal sinus node function (seven patients) was determined. In all cases we have observed a significantly improved correlation between the atrial and ventricular paced rates during exercise with the new algorithm (P < 0.001).</description><identifier>ISSN: 0147-8389</identifier><identifier>EISSN: 1540-8159</identifier><identifier>DOI: 10.1111/j.1540-8159.1989.tb02664.x</identifier><identifier>PMID: 2468141</identifier><language>eng</language><publisher>Oxford, UK: Blackwell Publishing Ltd</publisher><subject>Adult ; Aged ; algorithm ; Computer Simulation ; dynamic slope ; Electrocardiography ; Exercise Test ; Female ; Heart Block - therapy ; Heart Rate ; Humans ; Male ; Middle Aged ; Models, Cardiovascular ; Pacemaker, Artificial ; QT sensing pacemaker</subject><ispartof>Pacing and clinical electrophysiology, 1989-02, Vol.12 (2), p.311-320</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c4081-c86e7fbe280cbee60b58aa9405799ea5d77003f6e431930b911ab26655ed95693</citedby><cites>FETCH-LOGICAL-c4081-c86e7fbe280cbee60b58aa9405799ea5d77003f6e431930b911ab26655ed95693</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://onlinelibrary.wiley.com/doi/pdf/10.1111%2Fj.1540-8159.1989.tb02664.x$$EPDF$$P50$$Gwiley$$H</linktopdf><linktohtml>$$Uhttps://onlinelibrary.wiley.com/doi/full/10.1111%2Fj.1540-8159.1989.tb02664.x$$EHTML$$P50$$Gwiley$$H</linktohtml><link.rule.ids>314,776,780,1411,27903,27904,45553,45554</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/2468141$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>BAIG, M.W.</creatorcontrib><creatorcontrib>WILSON, J.</creatorcontrib><creatorcontrib>BOUTE, W.</creatorcontrib><creatorcontrib>BEGEMANN, M.J.S.</creatorcontrib><creatorcontrib>COBBOLD, J.P.</creatorcontrib><creatorcontrib>PERRINS, E.J.</creatorcontrib><title>Improved Pattern of Rate Responsiveness with Dynamic Slope Setting for the QT Sensing Pacemaker</title><title>Pacing and clinical electrophysiology</title><addtitle>Pacing Clin Electrophysiol</addtitle><description>We have recently described the electrophysiological basis of a new algorithm for the QT (TX) sensing rate responsive pacemaker. By using the new software program running on the standard programmer it has been possible to simulate the new algorithm in ten patients with complete heart block (seven patients had implanted TX units and three were paced with an external TX pacemaker) during routine exercise testing. In this way a single‐blind, intra‐patient comparison of the pattern of pacing rote change using both the existing and new algorithms was possible. In nine out of the ten cases the time taken to increase the pacing rate from 70 to 80 bpm was reduced significantly when the new algorithm was used (P = 0.037). Additionally, the correlation between the atrial and ventricular rates in those patients with normal sinus node function (seven patients) was determined. In all cases we have observed a significantly improved correlation between the atrial and ventricular paced rates during exercise with the new algorithm (P < 0.001).</description><subject>Adult</subject><subject>Aged</subject><subject>algorithm</subject><subject>Computer Simulation</subject><subject>dynamic slope</subject><subject>Electrocardiography</subject><subject>Exercise Test</subject><subject>Female</subject><subject>Heart Block - therapy</subject><subject>Heart Rate</subject><subject>Humans</subject><subject>Male</subject><subject>Middle Aged</subject><subject>Models, Cardiovascular</subject><subject>Pacemaker, Artificial</subject><subject>QT sensing pacemaker</subject><issn>0147-8389</issn><issn>1540-8159</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>1989</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNqVkE1PGzEQhq2qFU1pf0Ili0Nvu7Xjby4IBUiRUBsgFdws72YWNuxHsB1I_n29SpR75zLSvPO-o3kQOqEkp6l-LnMqOMk0FSanRps8FmQsJc83H9DoIH1EI0K5yjTT5jP6EsKSECIJF0foaMylppyOkL1uV75_gwWeuRjBd7iv8J2LgO8grPou1G_QQQj4vY7P-GLbubYu8X3TrwDfQ4x194Sr3uP4DPh2nkbJkUYzV0LrXsB_RZ8q1wT4tu_H6O_V5XzyK7v5M72enN9kJSeaZqWWoKoCxpqUBYAkhdDOGU6EMgacWChFCKskcEYNI4Wh1BXpZSFgYYQ07Bj92OWmb17XEKJt61BC07gO-nWwShsmtdBp8XS3WPo-BA-VXfm6dX5rKbEDXbu0A0I7ILQDXbunazfJ_H1_ZV20sDhY9ziTfrbT3-sGtv-RbGfnk0tGh4Rsl1CHCJtDgvMvViqmhH34PbXm4lbNH6czK9k_itKZhg</recordid><startdate>198902</startdate><enddate>198902</enddate><creator>BAIG, M.W.</creator><creator>WILSON, J.</creator><creator>BOUTE, W.</creator><creator>BEGEMANN, M.J.S.</creator><creator>COBBOLD, J.P.</creator><creator>PERRINS, E.J.</creator><general>Blackwell Publishing Ltd</general><scope>BSCLL</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>198902</creationdate><title>Improved Pattern of Rate Responsiveness with Dynamic Slope Setting for the QT Sensing Pacemaker</title><author>BAIG, M.W. ; WILSON, J. ; BOUTE, W. ; BEGEMANN, M.J.S. ; COBBOLD, J.P. ; PERRINS, E.J.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c4081-c86e7fbe280cbee60b58aa9405799ea5d77003f6e431930b911ab26655ed95693</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>1989</creationdate><topic>Adult</topic><topic>Aged</topic><topic>algorithm</topic><topic>Computer Simulation</topic><topic>dynamic slope</topic><topic>Electrocardiography</topic><topic>Exercise Test</topic><topic>Female</topic><topic>Heart Block - therapy</topic><topic>Heart Rate</topic><topic>Humans</topic><topic>Male</topic><topic>Middle Aged</topic><topic>Models, Cardiovascular</topic><topic>Pacemaker, Artificial</topic><topic>QT sensing pacemaker</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>BAIG, M.W.</creatorcontrib><creatorcontrib>WILSON, J.</creatorcontrib><creatorcontrib>BOUTE, W.</creatorcontrib><creatorcontrib>BEGEMANN, M.J.S.</creatorcontrib><creatorcontrib>COBBOLD, J.P.</creatorcontrib><creatorcontrib>PERRINS, E.J.</creatorcontrib><collection>Istex</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Pacing and clinical electrophysiology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>BAIG, M.W.</au><au>WILSON, J.</au><au>BOUTE, W.</au><au>BEGEMANN, M.J.S.</au><au>COBBOLD, J.P.</au><au>PERRINS, E.J.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Improved Pattern of Rate Responsiveness with Dynamic Slope Setting for the QT Sensing Pacemaker</atitle><jtitle>Pacing and clinical electrophysiology</jtitle><addtitle>Pacing Clin Electrophysiol</addtitle><date>1989-02</date><risdate>1989</risdate><volume>12</volume><issue>2</issue><spage>311</spage><epage>320</epage><pages>311-320</pages><issn>0147-8389</issn><eissn>1540-8159</eissn><abstract>We have recently described the electrophysiological basis of a new algorithm for the QT (TX) sensing rate responsive pacemaker. By using the new software program running on the standard programmer it has been possible to simulate the new algorithm in ten patients with complete heart block (seven patients had implanted TX units and three were paced with an external TX pacemaker) during routine exercise testing. In this way a single‐blind, intra‐patient comparison of the pattern of pacing rote change using both the existing and new algorithms was possible. In nine out of the ten cases the time taken to increase the pacing rate from 70 to 80 bpm was reduced significantly when the new algorithm was used (P = 0.037). Additionally, the correlation between the atrial and ventricular rates in those patients with normal sinus node function (seven patients) was determined. In all cases we have observed a significantly improved correlation between the atrial and ventricular paced rates during exercise with the new algorithm (P < 0.001).</abstract><cop>Oxford, UK</cop><pub>Blackwell Publishing Ltd</pub><pmid>2468141</pmid><doi>10.1111/j.1540-8159.1989.tb02664.x</doi><tpages>10</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0147-8389 |
ispartof | Pacing and clinical electrophysiology, 1989-02, Vol.12 (2), p.311-320 |
issn | 0147-8389 1540-8159 |
language | eng |
recordid | cdi_proquest_miscellaneous_78936858 |
source | MEDLINE; Wiley Online Library Journals Frontfile Complete |
subjects | Adult Aged algorithm Computer Simulation dynamic slope Electrocardiography Exercise Test Female Heart Block - therapy Heart Rate Humans Male Middle Aged Models, Cardiovascular Pacemaker, Artificial QT sensing pacemaker |
title | Improved Pattern of Rate Responsiveness with Dynamic Slope Setting for the QT Sensing Pacemaker |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-26T06%3A23%3A45IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Improved%20Pattern%20of%20Rate%20Responsiveness%20with%20Dynamic%20Slope%20Setting%20for%20the%20QT%20Sensing%20Pacemaker&rft.jtitle=Pacing%20and%20clinical%20electrophysiology&rft.au=BAIG,%20M.W.&rft.date=1989-02&rft.volume=12&rft.issue=2&rft.spage=311&rft.epage=320&rft.pages=311-320&rft.issn=0147-8389&rft.eissn=1540-8159&rft_id=info:doi/10.1111/j.1540-8159.1989.tb02664.x&rft_dat=%3Cproquest_cross%3E78936858%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=78936858&rft_id=info:pmid/2468141&rfr_iscdi=true |