Combined automatic implantable cardioverter-defibrillator and pacemaker systems: Implantation techniques and follow-up

The automatic implantable cardioverter-defibrillator (AICD) effectively prevents death due to ventricular tachycardia or ventricular fibrillation. Some patients who need an AICD also require cardiac pacing to treat symptomatic bradycardia, bradycardia after defibrillation, or to provide a rate floor...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of the American College of Cardiology 1989-01, Vol.13 (1), p.121-131
Hauptverfasser: Epstein, Andrew E., Kay, G.Neal, Plumb, Vance J., Shepard, Richard B., Kirklin, James K.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 131
container_issue 1
container_start_page 121
container_title Journal of the American College of Cardiology
container_volume 13
creator Epstein, Andrew E.
Kay, G.Neal
Plumb, Vance J.
Shepard, Richard B.
Kirklin, James K.
description The automatic implantable cardioverter-defibrillator (AICD) effectively prevents death due to ventricular tachycardia or ventricular fibrillation. Some patients who need an AICD also require cardiac pacing to treat symptomatic bradycardia, bradycardia after defibrillation, or to provide a rate floor to reduce the frequency of bradycardia-related ventricular arrhythmias. Some patients also can benefit from antitachycardia pacing. A mapping technique to implant a pacemaker and AICD sensing leads is presented. For patients with a pacemaker who later need an AICD, the left ventricle is mapped with use of the AICD rate-sensing electrodes to identify a site at which the minimal pacemaker stimulus and maximal ventricular electrogram amplitudes are recorded. An external cardioverter-defibrillator that has amplifiers similar to those in the AICD is used to monitor the rate-sensing electrogram. For patients with an implanted AICD, pacemaker implantation is undertaken by mapping the right ventricle with the pacemaker lead while the AICD is in standby mode; the AICD beep monitor is then used to determine a site where pacemaker stimulus detection by the AICD does not occur. Eight patients underwent implantation of a combined AICD-pacemaker system (four ventricular antitachycardia pacemakers, three ventricular demand pacemakers and one atrial demand pacemaker). Neither inhibition of AICD arrhythmia detection nor double counting occurred. Satisfactory AICD-pacemaker function was shown in all patients postoperatively, and no pacemaker malfunction was observed. Thus, with currently available technology, a combined AICD-pacemaker system can be implanted with satisfactory function of both devices and without adverse device-device interactions.
doi_str_mv 10.1016/0735-1097(89)90559-7
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_78809533</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>0735109789905597</els_id><sourcerecordid>78809533</sourcerecordid><originalsourceid>FETCH-LOGICAL-c437t-31e683556f8bffdfc6b30605f5b34bd8756c30e3f0e9a647218261d3df37ee533</originalsourceid><addsrcrecordid>eNp9kE9P3DAQxS1URJeFb9BKOVXlELDX8b8ekKoVbZGQuMDZcuyx6jaJU9tZtN-ehF1x7GkO896beT-EPhF8TTDhN1hQVhOsxFeprhRmTNXiBK0IY7KmTIkPaPUu-YjOc_6DMeaSqDN0tlFYzYYV2m1j34YBXGWmEntTgq1CP3ZmKKbtoLImuRB3kAqk2oEPbQpdZ0pMlRlcNRoLvfkLqcr7XKDP36r7o7uEOFQF7O8h_Jsgv8l97Lr4Uk_jBTr1pstweZxr9Pzj7mn7q354_Hm__f5Q24aKUlMCXFLGuJet985b3lLMMfOspU3rpGDcUgzUY1CGN2JD5IYTR52nAoBRukZfDrljissXRfchW5gbDBCnrIWUMwi6CJuD0KaYcwKvxxR6k_aaYL3g1gtLvbDUUuk33FrMts_H_Kntwb2bjnzn_e1hD3PJXYCksw0wWHAhgS3axfD_A6-QgJH5</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>78809533</pqid></control><display><type>article</type><title>Combined automatic implantable cardioverter-defibrillator and pacemaker systems: Implantation techniques and follow-up</title><source>MEDLINE</source><source>Access via ScienceDirect (Elsevier)</source><source>EZB-FREE-00999 freely available EZB journals</source><creator>Epstein, Andrew E. ; Kay, G.Neal ; Plumb, Vance J. ; Shepard, Richard B. ; Kirklin, James K.</creator><creatorcontrib>Epstein, Andrew E. ; Kay, G.Neal ; Plumb, Vance J. ; Shepard, Richard B. ; Kirklin, James K.</creatorcontrib><description>The automatic implantable cardioverter-defibrillator (AICD) effectively prevents death due to ventricular tachycardia or ventricular fibrillation. Some patients who need an AICD also require cardiac pacing to treat symptomatic bradycardia, bradycardia after defibrillation, or to provide a rate floor to reduce the frequency of bradycardia-related ventricular arrhythmias. Some patients also can benefit from antitachycardia pacing. A mapping technique to implant a pacemaker and AICD sensing leads is presented. For patients with a pacemaker who later need an AICD, the left ventricle is mapped with use of the AICD rate-sensing electrodes to identify a site at which the minimal pacemaker stimulus and maximal ventricular electrogram amplitudes are recorded. An external cardioverter-defibrillator that has amplifiers similar to those in the AICD is used to monitor the rate-sensing electrogram. For patients with an implanted AICD, pacemaker implantation is undertaken by mapping the right ventricle with the pacemaker lead while the AICD is in standby mode; the AICD beep monitor is then used to determine a site where pacemaker stimulus detection by the AICD does not occur. Eight patients underwent implantation of a combined AICD-pacemaker system (four ventricular antitachycardia pacemakers, three ventricular demand pacemakers and one atrial demand pacemaker). Neither inhibition of AICD arrhythmia detection nor double counting occurred. Satisfactory AICD-pacemaker function was shown in all patients postoperatively, and no pacemaker malfunction was observed. Thus, with currently available technology, a combined AICD-pacemaker system can be implanted with satisfactory function of both devices and without adverse device-device interactions.</description><identifier>ISSN: 0735-1097</identifier><identifier>EISSN: 1558-3597</identifier><identifier>DOI: 10.1016/0735-1097(89)90559-7</identifier><identifier>PMID: 2909559</identifier><language>eng</language><publisher>United States: Elsevier Inc</publisher><subject>Aged ; Arrhythmias, Cardiac - physiopathology ; Arrhythmias, Cardiac - therapy ; Cardiac Pacing, Artificial - methods ; Electric Countershock - instrumentation ; Electric Countershock - methods ; Electrophysiology ; Female ; Follow-Up Studies ; Humans ; Male ; Middle Aged ; Pacemaker, Artificial ; Prostheses and Implants ; Tachycardia, Supraventricular - physiopathology ; Tachycardia, Supraventricular - therapy</subject><ispartof>Journal of the American College of Cardiology, 1989-01, Vol.13 (1), p.121-131</ispartof><rights>1989</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c437t-31e683556f8bffdfc6b30605f5b34bd8756c30e3f0e9a647218261d3df37ee533</citedby><cites>FETCH-LOGICAL-c437t-31e683556f8bffdfc6b30605f5b34bd8756c30e3f0e9a647218261d3df37ee533</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://dx.doi.org/10.1016/0735-1097(89)90559-7$$EHTML$$P50$$Gelsevier$$Hfree_for_read</linktohtml><link.rule.ids>315,782,786,3554,27933,27934,46004</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/2909559$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Epstein, Andrew E.</creatorcontrib><creatorcontrib>Kay, G.Neal</creatorcontrib><creatorcontrib>Plumb, Vance J.</creatorcontrib><creatorcontrib>Shepard, Richard B.</creatorcontrib><creatorcontrib>Kirklin, James K.</creatorcontrib><title>Combined automatic implantable cardioverter-defibrillator and pacemaker systems: Implantation techniques and follow-up</title><title>Journal of the American College of Cardiology</title><addtitle>J Am Coll Cardiol</addtitle><description>The automatic implantable cardioverter-defibrillator (AICD) effectively prevents death due to ventricular tachycardia or ventricular fibrillation. Some patients who need an AICD also require cardiac pacing to treat symptomatic bradycardia, bradycardia after defibrillation, or to provide a rate floor to reduce the frequency of bradycardia-related ventricular arrhythmias. Some patients also can benefit from antitachycardia pacing. A mapping technique to implant a pacemaker and AICD sensing leads is presented. For patients with a pacemaker who later need an AICD, the left ventricle is mapped with use of the AICD rate-sensing electrodes to identify a site at which the minimal pacemaker stimulus and maximal ventricular electrogram amplitudes are recorded. An external cardioverter-defibrillator that has amplifiers similar to those in the AICD is used to monitor the rate-sensing electrogram. For patients with an implanted AICD, pacemaker implantation is undertaken by mapping the right ventricle with the pacemaker lead while the AICD is in standby mode; the AICD beep monitor is then used to determine a site where pacemaker stimulus detection by the AICD does not occur. Eight patients underwent implantation of a combined AICD-pacemaker system (four ventricular antitachycardia pacemakers, three ventricular demand pacemakers and one atrial demand pacemaker). Neither inhibition of AICD arrhythmia detection nor double counting occurred. Satisfactory AICD-pacemaker function was shown in all patients postoperatively, and no pacemaker malfunction was observed. Thus, with currently available technology, a combined AICD-pacemaker system can be implanted with satisfactory function of both devices and without adverse device-device interactions.</description><subject>Aged</subject><subject>Arrhythmias, Cardiac - physiopathology</subject><subject>Arrhythmias, Cardiac - therapy</subject><subject>Cardiac Pacing, Artificial - methods</subject><subject>Electric Countershock - instrumentation</subject><subject>Electric Countershock - methods</subject><subject>Electrophysiology</subject><subject>Female</subject><subject>Follow-Up Studies</subject><subject>Humans</subject><subject>Male</subject><subject>Middle Aged</subject><subject>Pacemaker, Artificial</subject><subject>Prostheses and Implants</subject><subject>Tachycardia, Supraventricular - physiopathology</subject><subject>Tachycardia, Supraventricular - therapy</subject><issn>0735-1097</issn><issn>1558-3597</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>1989</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp9kE9P3DAQxS1URJeFb9BKOVXlELDX8b8ekKoVbZGQuMDZcuyx6jaJU9tZtN-ehF1x7GkO896beT-EPhF8TTDhN1hQVhOsxFeprhRmTNXiBK0IY7KmTIkPaPUu-YjOc_6DMeaSqDN0tlFYzYYV2m1j34YBXGWmEntTgq1CP3ZmKKbtoLImuRB3kAqk2oEPbQpdZ0pMlRlcNRoLvfkLqcr7XKDP36r7o7uEOFQF7O8h_Jsgv8l97Lr4Uk_jBTr1pstweZxr9Pzj7mn7q354_Hm__f5Q24aKUlMCXFLGuJet985b3lLMMfOspU3rpGDcUgzUY1CGN2JD5IYTR52nAoBRukZfDrljissXRfchW5gbDBCnrIWUMwi6CJuD0KaYcwKvxxR6k_aaYL3g1gtLvbDUUuk33FrMts_H_Kntwb2bjnzn_e1hD3PJXYCksw0wWHAhgS3axfD_A6-QgJH5</recordid><startdate>198901</startdate><enddate>198901</enddate><creator>Epstein, Andrew E.</creator><creator>Kay, G.Neal</creator><creator>Plumb, Vance J.</creator><creator>Shepard, Richard B.</creator><creator>Kirklin, James K.</creator><general>Elsevier Inc</general><scope>6I.</scope><scope>AAFTH</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>198901</creationdate><title>Combined automatic implantable cardioverter-defibrillator and pacemaker systems: Implantation techniques and follow-up</title><author>Epstein, Andrew E. ; Kay, G.Neal ; Plumb, Vance J. ; Shepard, Richard B. ; Kirklin, James K.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c437t-31e683556f8bffdfc6b30605f5b34bd8756c30e3f0e9a647218261d3df37ee533</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>1989</creationdate><topic>Aged</topic><topic>Arrhythmias, Cardiac - physiopathology</topic><topic>Arrhythmias, Cardiac - therapy</topic><topic>Cardiac Pacing, Artificial - methods</topic><topic>Electric Countershock - instrumentation</topic><topic>Electric Countershock - methods</topic><topic>Electrophysiology</topic><topic>Female</topic><topic>Follow-Up Studies</topic><topic>Humans</topic><topic>Male</topic><topic>Middle Aged</topic><topic>Pacemaker, Artificial</topic><topic>Prostheses and Implants</topic><topic>Tachycardia, Supraventricular - physiopathology</topic><topic>Tachycardia, Supraventricular - therapy</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Epstein, Andrew E.</creatorcontrib><creatorcontrib>Kay, G.Neal</creatorcontrib><creatorcontrib>Plumb, Vance J.</creatorcontrib><creatorcontrib>Shepard, Richard B.</creatorcontrib><creatorcontrib>Kirklin, James K.</creatorcontrib><collection>ScienceDirect Open Access Titles</collection><collection>Elsevier:ScienceDirect:Open Access</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Journal of the American College of Cardiology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Epstein, Andrew E.</au><au>Kay, G.Neal</au><au>Plumb, Vance J.</au><au>Shepard, Richard B.</au><au>Kirklin, James K.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Combined automatic implantable cardioverter-defibrillator and pacemaker systems: Implantation techniques and follow-up</atitle><jtitle>Journal of the American College of Cardiology</jtitle><addtitle>J Am Coll Cardiol</addtitle><date>1989-01</date><risdate>1989</risdate><volume>13</volume><issue>1</issue><spage>121</spage><epage>131</epage><pages>121-131</pages><issn>0735-1097</issn><eissn>1558-3597</eissn><abstract>The automatic implantable cardioverter-defibrillator (AICD) effectively prevents death due to ventricular tachycardia or ventricular fibrillation. Some patients who need an AICD also require cardiac pacing to treat symptomatic bradycardia, bradycardia after defibrillation, or to provide a rate floor to reduce the frequency of bradycardia-related ventricular arrhythmias. Some patients also can benefit from antitachycardia pacing. A mapping technique to implant a pacemaker and AICD sensing leads is presented. For patients with a pacemaker who later need an AICD, the left ventricle is mapped with use of the AICD rate-sensing electrodes to identify a site at which the minimal pacemaker stimulus and maximal ventricular electrogram amplitudes are recorded. An external cardioverter-defibrillator that has amplifiers similar to those in the AICD is used to monitor the rate-sensing electrogram. For patients with an implanted AICD, pacemaker implantation is undertaken by mapping the right ventricle with the pacemaker lead while the AICD is in standby mode; the AICD beep monitor is then used to determine a site where pacemaker stimulus detection by the AICD does not occur. Eight patients underwent implantation of a combined AICD-pacemaker system (four ventricular antitachycardia pacemakers, three ventricular demand pacemakers and one atrial demand pacemaker). Neither inhibition of AICD arrhythmia detection nor double counting occurred. Satisfactory AICD-pacemaker function was shown in all patients postoperatively, and no pacemaker malfunction was observed. Thus, with currently available technology, a combined AICD-pacemaker system can be implanted with satisfactory function of both devices and without adverse device-device interactions.</abstract><cop>United States</cop><pub>Elsevier Inc</pub><pmid>2909559</pmid><doi>10.1016/0735-1097(89)90559-7</doi><tpages>11</tpages><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 0735-1097
ispartof Journal of the American College of Cardiology, 1989-01, Vol.13 (1), p.121-131
issn 0735-1097
1558-3597
language eng
recordid cdi_proquest_miscellaneous_78809533
source MEDLINE; Access via ScienceDirect (Elsevier); EZB-FREE-00999 freely available EZB journals
subjects Aged
Arrhythmias, Cardiac - physiopathology
Arrhythmias, Cardiac - therapy
Cardiac Pacing, Artificial - methods
Electric Countershock - instrumentation
Electric Countershock - methods
Electrophysiology
Female
Follow-Up Studies
Humans
Male
Middle Aged
Pacemaker, Artificial
Prostheses and Implants
Tachycardia, Supraventricular - physiopathology
Tachycardia, Supraventricular - therapy
title Combined automatic implantable cardioverter-defibrillator and pacemaker systems: Implantation techniques and follow-up
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-03T08%3A33%3A38IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Combined%20automatic%20implantable%20cardioverter-defibrillator%20and%20pacemaker%20systems:%20Implantation%20techniques%20and%20follow-up&rft.jtitle=Journal%20of%20the%20American%20College%20of%20Cardiology&rft.au=Epstein,%20Andrew%20E.&rft.date=1989-01&rft.volume=13&rft.issue=1&rft.spage=121&rft.epage=131&rft.pages=121-131&rft.issn=0735-1097&rft.eissn=1558-3597&rft_id=info:doi/10.1016/0735-1097(89)90559-7&rft_dat=%3Cproquest_cross%3E78809533%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=78809533&rft_id=info:pmid/2909559&rft_els_id=0735109789905597&rfr_iscdi=true