AN AUTOMATED INTERFEROMETER FOR THE ANALYSIS OF ANAESTHETIC GAS MIXTURES

A microprocessor-controlled interferometer is described. The eyepiece of a conventional Jamin type interferometer has been replaced by an array of photocells which records the intensity across the interference pattern. Mathematical correlation procedures are used to locate the principal interference...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:British journal of anaesthesia : BJA 1988-10, Vol.61 (4), p.484-491
Hauptverfasser: SUGG, B.R., PALAYIWA, E, DAVIES, W.L., JACKSON, R, MCGRAGHAN, T, SHADBOLT, P, WELLER, S.J., HAHN, C.E.W.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A microprocessor-controlled interferometer is described. The eyepiece of a conventional Jamin type interferometer has been replaced by an array of photocells which records the intensity across the interference pattern. Mathematical correlation procedures are used to locate the principal interference pattern maximum and, by sequential analysis of a fresh gas mixture followed by fresh gas plus vapour, it is possible to determine both oxygen and vapour concentrations. The instrument was used to analyse mixtures of oxygen and nitrous oxide and also oxygen, nitrous oxide plus halothane. It was found that the oxygen concentration could be determined to an accuracy of ±1% v/v and the vapour concentration to ±0.1% v/v. The instrument is suitable for monitoring concentrations delivered by an anaesthetic machine and may be included in a microprocessor-controlled anaesthetic machine.
ISSN:0007-0912
1471-6771
DOI:10.1093/bja/61.4.484