Flow Resistance of Expiratory Positive-Pressure Systems
We measured the flow-resistance of five commercially available 10 cm H2O expiratory positive-pressure (EPP) valves (n = five per valve type) at bias flows of between 0 and 2,000 ml/s. We found that individual valves of each type and manufacturer functioned similarly. Different valve types, however,...
Gespeichert in:
Veröffentlicht in: | Chest 1988-10, Vol.94 (4), p.788-791 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We measured the flow-resistance of five commercially available 10 cm H2O expiratory positive-pressure (EPP) valves (n = five per valve type) at bias flows of between 0 and 2,000 ml/s. We found that individual valves of each type and manufacturer functioned similarly. Different valve types, however, functioned differently: with one type, system pressure was higher than rated (p |
---|---|
ISSN: | 0012-3692 1931-3543 |