Large discrepancy between prostate-specific antigen results from different assays during longitudinal follow-up of a prostate cancer patient
A case is presented of a patient with stage D prostatic carcinoma, from whom a serum sample proved to be an outlier in a correlation study performed with a 2nd-generation prostate-specific antigen (PSA) assay on the Immulite system (6.4 micrograms/L) and IMx (101 micrograms/L). Clearly, the PSA resu...
Gespeichert in:
Veröffentlicht in: | Clinical chemistry (Baltimore, Md.) Md.), 1996-04, Vol.42 (4), p.637-641 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A case is presented of a patient with stage D prostatic carcinoma, from whom a serum sample proved to be an outlier in a correlation study performed with a 2nd-generation prostate-specific antigen (PSA) assay on the Immulite system (6.4 micrograms/L) and IMx (101 micrograms/L). Clearly, the PSA result reported by Immulite was falsely low. For nine longitudinal samples, Immulite results were approximately 20-fold lower than the IMx value (range of IMx results 5-275 micrograms/L). A selection of the samples was analyzed with the ACS:180, ES-600, and IMx (all > 180 micrograms/L); Immulite, DPC Coat-A-Count IRMA, Immuno 1, AIA-pack, and Tandem-R (all |
---|---|
ISSN: | 0009-9147 1530-8561 |
DOI: | 10.1093/clinchem/42.4.637 |