Stabilization of the short sagittal split osteotomy: in vitro testing of different plate and screw configurations
Using a 3-dimensional in vitro model, the stability of different types of osteosynthesis for the short sagittal osteotomy was tested. The following four groups of plate and screw configurations were evaluated: • Group I: Fixation with miniplates using monocortical screws only • Group II: Fixation wi...
Gespeichert in:
Veröffentlicht in: | Journal of cranio-maxillo-facial surgery 1995-10, Vol.23 (5), p.321-324 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Using a 3-dimensional in vitro model, the stability of different types of osteosynthesis for the short sagittal osteotomy was tested.
The following four groups of plate and screw configurations were evaluated:
•
Group I: Fixation with miniplates using monocortical screws only
•
Group II: Fixation with miniplates, but with two of the screws engaging both fragments
•
Group III: Same as group II with an additional position screw
•
Group IV: Fixation with 3 position screws.
The stability obtained with miniplate fixation using monocortical screws only (group I) was by a factor of 2.9 less than position screw fixation (group IV), which is considered to be an approved standard. In order to increase the stability of miniplate fixation, the screws in the area of overlapping bone should engage both fragments. |
---|---|