Does the Use of Hydrochloric Acid Damage Silicone Rubber Central Venous Catheters?
Background: Hydrochloric acid (HCl) is commonly used to clear obstructions in central venous catheters (CVC). Methods: To determine if HCl adversely affects the integrity of a CVC we infused 0.1 N HCl daily into CVC in vitro. At weekly intervals sections of the CVC were removed and examined using sc...
Gespeichert in:
Veröffentlicht in: | JPEN. Journal of parenteral and enteral nutrition 1995-09, Vol.19 (5), p.407-409 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background: Hydrochloric acid (HCl) is commonly used to clear obstructions in central venous catheters (CVC). Methods: To determine if HCl adversely affects the integrity of a CVC we infused 0.1 N HCl daily into CVC in vitro. At weekly intervals sections of the CVC were removed and examined using scanning electron microscopy. Results: Over 8 weeks, no damage to the CVC was visible. Conclusions: HCl administration poses no threat to the structural integrity of a CVC. (Journal of Parenteral and Enteral Nutrition
19:407-409, 1995) |
---|---|
ISSN: | 0148-6071 1941-2444 |
DOI: | 10.1177/0148607195019005407 |