Reliability of Evoked Endocardial T-wave Sensing in 1,500 Pacemaker Patients
Intracardiac measurements of the evoked endocardial T‐wave were performed on 1,500 pacemaker patients. The sensed evoked T‐wave amplitude was found to be above 0.75 mV in 94.1 percent of all cases. No correlation was found between the amplitude of the spontaneous R‐wave and that of the evoked T‐wave...
Gespeichert in:
Veröffentlicht in: | Pacing and clinical electrophysiology 1986-11, Vol.9 (6), p.948-953 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 953 |
---|---|
container_issue | 6 |
container_start_page | 948 |
container_title | Pacing and clinical electrophysiology |
container_volume | 9 |
creator | BOUTE, W. DERRIEN, Y. WITTKAMPF, F.H.M. |
description | Intracardiac measurements of the evoked endocardial T‐wave were performed on 1,500 pacemaker patients. The sensed evoked T‐wave amplitude was found to be above 0.75 mV in 94.1 percent of all cases. No correlation was found between the amplitude of the spontaneous R‐wave and that of the evoked T‐wave. Some electrode characteristics proved to be more favorable than others for evoked endocardial wave sensing, i.e., a surface area of less than 12 mm2. porous surface structure (preferably carbon), and atraumatic fixation. |
doi_str_mv | 10.1111/j.1540-8159.1986.tb06652.x |
format | Article |
fullrecord | <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_proquest_miscellaneous_77228053</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>77228053</sourcerecordid><originalsourceid>FETCH-LOGICAL-c2898-f7b0445374d8d3c0d3d13f26b02d0946a1191b81bb6981347d6038a33b2945e63</originalsourceid><addsrcrecordid>eNo9kE1rGzEQhkVISR2nPyGw9JBTdzP6lk4luG5aYhqTuMlRSCu5yF7vpqt1bP_7rLHxXGbgeXlhHoS-YihwP7eLAnMGucJcF1grUXQOhOCk2J6hwQmdowFgJnNFlf6MLlNaAIAAxi_QBWGUcEkHaPIUqmhdrGK3y5p5Nn5vlsFn49o3pW19tFU2yzf2PWTPoU6x_pfFOsPfOEA2tWVY2WVo-6uLoe7SFfo0t1UKX457iP7-HM9Gv_LJ4_3v0d0kL4nSKp9LB4xxKplXnpbgqcd0ToQD4kEzYTHW2CnsnNAKUya9AKospY5oxoOgQ3Rz6H1rm__rkDqziqkMVWXr0KyTkZIQBZz2wetjcO1WwZu3Nq5suzPH93v-_cA3sQq7E8Zg9p7Nwuxlmr1Ms_dsjp7N1kzvRmPNVN-QHxpi6sL21GDbpRGSSm5e_9yb0XQyfZn9eDAz-gEdpn53</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>77228053</pqid></control><display><type>article</type><title>Reliability of Evoked Endocardial T-wave Sensing in 1,500 Pacemaker Patients</title><source>MEDLINE</source><source>Access via Wiley Online Library</source><creator>BOUTE, W. ; DERRIEN, Y. ; WITTKAMPF, F.H.M.</creator><creatorcontrib>BOUTE, W. ; DERRIEN, Y. ; WITTKAMPF, F.H.M.</creatorcontrib><description>Intracardiac measurements of the evoked endocardial T‐wave were performed on 1,500 pacemaker patients. The sensed evoked T‐wave amplitude was found to be above 0.75 mV in 94.1 percent of all cases. No correlation was found between the amplitude of the spontaneous R‐wave and that of the evoked T‐wave. Some electrode characteristics proved to be more favorable than others for evoked endocardial wave sensing, i.e., a surface area of less than 12 mm2. porous surface structure (preferably carbon), and atraumatic fixation.</description><identifier>ISSN: 0147-8389</identifier><identifier>EISSN: 1540-8159</identifier><identifier>DOI: 10.1111/j.1540-8159.1986.tb06652.x</identifier><identifier>PMID: 2432573</identifier><language>eng</language><publisher>Oxford, UK: Blackwell Publishing Ltd</publisher><subject>amplitude ; Electrocardiography ; electrode characteristics ; Electrodes ; Endocardium - physiopathology ; Evoked Potentials ; evoked T-wave ; Humans ; Pacemaker, Artificial ; sensing ; spontaneous R-wave amplitude</subject><ispartof>Pacing and clinical electrophysiology, 1986-11, Vol.9 (6), p.948-953</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c2898-f7b0445374d8d3c0d3d13f26b02d0946a1191b81bb6981347d6038a33b2945e63</citedby></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://onlinelibrary.wiley.com/doi/pdf/10.1111%2Fj.1540-8159.1986.tb06652.x$$EPDF$$P50$$Gwiley$$H</linktopdf><linktohtml>$$Uhttps://onlinelibrary.wiley.com/doi/full/10.1111%2Fj.1540-8159.1986.tb06652.x$$EHTML$$P50$$Gwiley$$H</linktohtml><link.rule.ids>314,780,784,1417,27924,27925,45574,45575</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/2432573$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>BOUTE, W.</creatorcontrib><creatorcontrib>DERRIEN, Y.</creatorcontrib><creatorcontrib>WITTKAMPF, F.H.M.</creatorcontrib><title>Reliability of Evoked Endocardial T-wave Sensing in 1,500 Pacemaker Patients</title><title>Pacing and clinical electrophysiology</title><addtitle>Pacing Clin Electrophysiol</addtitle><description>Intracardiac measurements of the evoked endocardial T‐wave were performed on 1,500 pacemaker patients. The sensed evoked T‐wave amplitude was found to be above 0.75 mV in 94.1 percent of all cases. No correlation was found between the amplitude of the spontaneous R‐wave and that of the evoked T‐wave. Some electrode characteristics proved to be more favorable than others for evoked endocardial wave sensing, i.e., a surface area of less than 12 mm2. porous surface structure (preferably carbon), and atraumatic fixation.</description><subject>amplitude</subject><subject>Electrocardiography</subject><subject>electrode characteristics</subject><subject>Electrodes</subject><subject>Endocardium - physiopathology</subject><subject>Evoked Potentials</subject><subject>evoked T-wave</subject><subject>Humans</subject><subject>Pacemaker, Artificial</subject><subject>sensing</subject><subject>spontaneous R-wave amplitude</subject><issn>0147-8389</issn><issn>1540-8159</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>1986</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNo9kE1rGzEQhkVISR2nPyGw9JBTdzP6lk4luG5aYhqTuMlRSCu5yF7vpqt1bP_7rLHxXGbgeXlhHoS-YihwP7eLAnMGucJcF1grUXQOhOCk2J6hwQmdowFgJnNFlf6MLlNaAIAAxi_QBWGUcEkHaPIUqmhdrGK3y5p5Nn5vlsFn49o3pW19tFU2yzf2PWTPoU6x_pfFOsPfOEA2tWVY2WVo-6uLoe7SFfo0t1UKX457iP7-HM9Gv_LJ4_3v0d0kL4nSKp9LB4xxKplXnpbgqcd0ToQD4kEzYTHW2CnsnNAKUya9AKospY5oxoOgQ3Rz6H1rm__rkDqziqkMVWXr0KyTkZIQBZz2wetjcO1WwZu3Nq5suzPH93v-_cA3sQq7E8Zg9p7Nwuxlmr1Ms_dsjp7N1kzvRmPNVN-QHxpi6sL21GDbpRGSSm5e_9yb0XQyfZn9eDAz-gEdpn53</recordid><startdate>198611</startdate><enddate>198611</enddate><creator>BOUTE, W.</creator><creator>DERRIEN, Y.</creator><creator>WITTKAMPF, F.H.M.</creator><general>Blackwell Publishing Ltd</general><scope>BSCLL</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>7X8</scope></search><sort><creationdate>198611</creationdate><title>Reliability of Evoked Endocardial T-wave Sensing in 1,500 Pacemaker Patients</title><author>BOUTE, W. ; DERRIEN, Y. ; WITTKAMPF, F.H.M.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c2898-f7b0445374d8d3c0d3d13f26b02d0946a1191b81bb6981347d6038a33b2945e63</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>1986</creationdate><topic>amplitude</topic><topic>Electrocardiography</topic><topic>electrode characteristics</topic><topic>Electrodes</topic><topic>Endocardium - physiopathology</topic><topic>Evoked Potentials</topic><topic>evoked T-wave</topic><topic>Humans</topic><topic>Pacemaker, Artificial</topic><topic>sensing</topic><topic>spontaneous R-wave amplitude</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>BOUTE, W.</creatorcontrib><creatorcontrib>DERRIEN, Y.</creatorcontrib><creatorcontrib>WITTKAMPF, F.H.M.</creatorcontrib><collection>Istex</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>MEDLINE - Academic</collection><jtitle>Pacing and clinical electrophysiology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>BOUTE, W.</au><au>DERRIEN, Y.</au><au>WITTKAMPF, F.H.M.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Reliability of Evoked Endocardial T-wave Sensing in 1,500 Pacemaker Patients</atitle><jtitle>Pacing and clinical electrophysiology</jtitle><addtitle>Pacing Clin Electrophysiol</addtitle><date>1986-11</date><risdate>1986</risdate><volume>9</volume><issue>6</issue><spage>948</spage><epage>953</epage><pages>948-953</pages><issn>0147-8389</issn><eissn>1540-8159</eissn><abstract>Intracardiac measurements of the evoked endocardial T‐wave were performed on 1,500 pacemaker patients. The sensed evoked T‐wave amplitude was found to be above 0.75 mV in 94.1 percent of all cases. No correlation was found between the amplitude of the spontaneous R‐wave and that of the evoked T‐wave. Some electrode characteristics proved to be more favorable than others for evoked endocardial wave sensing, i.e., a surface area of less than 12 mm2. porous surface structure (preferably carbon), and atraumatic fixation.</abstract><cop>Oxford, UK</cop><pub>Blackwell Publishing Ltd</pub><pmid>2432573</pmid><doi>10.1111/j.1540-8159.1986.tb06652.x</doi><tpages>6</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0147-8389 |
ispartof | Pacing and clinical electrophysiology, 1986-11, Vol.9 (6), p.948-953 |
issn | 0147-8389 1540-8159 |
language | eng |
recordid | cdi_proquest_miscellaneous_77228053 |
source | MEDLINE; Access via Wiley Online Library |
subjects | amplitude Electrocardiography electrode characteristics Electrodes Endocardium - physiopathology Evoked Potentials evoked T-wave Humans Pacemaker, Artificial sensing spontaneous R-wave amplitude |
title | Reliability of Evoked Endocardial T-wave Sensing in 1,500 Pacemaker Patients |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-26T02%3A50%3A59IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Reliability%20of%20Evoked%20Endocardial%20T-wave%20Sensing%20in%201,500%20Pacemaker%20Patients&rft.jtitle=Pacing%20and%20clinical%20electrophysiology&rft.au=BOUTE,%20W.&rft.date=1986-11&rft.volume=9&rft.issue=6&rft.spage=948&rft.epage=953&rft.pages=948-953&rft.issn=0147-8389&rft.eissn=1540-8159&rft_id=info:doi/10.1111/j.1540-8159.1986.tb06652.x&rft_dat=%3Cproquest_pubme%3E77228053%3C/proquest_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=77228053&rft_id=info:pmid/2432573&rfr_iscdi=true |