Reliability of Evoked Endocardial T-wave Sensing in 1,500 Pacemaker Patients

Intracardiac measurements of the evoked endocardial T‐wave were performed on 1,500 pacemaker patients. The sensed evoked T‐wave amplitude was found to be above 0.75 mV in 94.1 percent of all cases. No correlation was found between the amplitude of the spontaneous R‐wave and that of the evoked T‐wave...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 1986-11, Vol.9 (6), p.948-953
Hauptverfasser: BOUTE, W., DERRIEN, Y., WITTKAMPF, F.H.M.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 953
container_issue 6
container_start_page 948
container_title Pacing and clinical electrophysiology
container_volume 9
creator BOUTE, W.
DERRIEN, Y.
WITTKAMPF, F.H.M.
description Intracardiac measurements of the evoked endocardial T‐wave were performed on 1,500 pacemaker patients. The sensed evoked T‐wave amplitude was found to be above 0.75 mV in 94.1 percent of all cases. No correlation was found between the amplitude of the spontaneous R‐wave and that of the evoked T‐wave. Some electrode characteristics proved to be more favorable than others for evoked endocardial wave sensing, i.e., a surface area of less than 12 mm2. porous surface structure (preferably carbon), and atraumatic fixation.
doi_str_mv 10.1111/j.1540-8159.1986.tb06652.x
format Article
fullrecord <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_proquest_miscellaneous_77228053</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>77228053</sourcerecordid><originalsourceid>FETCH-LOGICAL-c2898-f7b0445374d8d3c0d3d13f26b02d0946a1191b81bb6981347d6038a33b2945e63</originalsourceid><addsrcrecordid>eNo9kE1rGzEQhkVISR2nPyGw9JBTdzP6lk4luG5aYhqTuMlRSCu5yF7vpqt1bP_7rLHxXGbgeXlhHoS-YihwP7eLAnMGucJcF1grUXQOhOCk2J6hwQmdowFgJnNFlf6MLlNaAIAAxi_QBWGUcEkHaPIUqmhdrGK3y5p5Nn5vlsFn49o3pW19tFU2yzf2PWTPoU6x_pfFOsPfOEA2tWVY2WVo-6uLoe7SFfo0t1UKX457iP7-HM9Gv_LJ4_3v0d0kL4nSKp9LB4xxKplXnpbgqcd0ToQD4kEzYTHW2CnsnNAKUya9AKospY5oxoOgQ3Rz6H1rm__rkDqziqkMVWXr0KyTkZIQBZz2wetjcO1WwZu3Nq5suzPH93v-_cA3sQq7E8Zg9p7Nwuxlmr1Ms_dsjp7N1kzvRmPNVN-QHxpi6sL21GDbpRGSSm5e_9yb0XQyfZn9eDAz-gEdpn53</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>77228053</pqid></control><display><type>article</type><title>Reliability of Evoked Endocardial T-wave Sensing in 1,500 Pacemaker Patients</title><source>MEDLINE</source><source>Access via Wiley Online Library</source><creator>BOUTE, W. ; DERRIEN, Y. ; WITTKAMPF, F.H.M.</creator><creatorcontrib>BOUTE, W. ; DERRIEN, Y. ; WITTKAMPF, F.H.M.</creatorcontrib><description>Intracardiac measurements of the evoked endocardial T‐wave were performed on 1,500 pacemaker patients. The sensed evoked T‐wave amplitude was found to be above 0.75 mV in 94.1 percent of all cases. No correlation was found between the amplitude of the spontaneous R‐wave and that of the evoked T‐wave. Some electrode characteristics proved to be more favorable than others for evoked endocardial wave sensing, i.e., a surface area of less than 12 mm2. porous surface structure (preferably carbon), and atraumatic fixation.</description><identifier>ISSN: 0147-8389</identifier><identifier>EISSN: 1540-8159</identifier><identifier>DOI: 10.1111/j.1540-8159.1986.tb06652.x</identifier><identifier>PMID: 2432573</identifier><language>eng</language><publisher>Oxford, UK: Blackwell Publishing Ltd</publisher><subject>amplitude ; Electrocardiography ; electrode characteristics ; Electrodes ; Endocardium - physiopathology ; Evoked Potentials ; evoked T-wave ; Humans ; Pacemaker, Artificial ; sensing ; spontaneous R-wave amplitude</subject><ispartof>Pacing and clinical electrophysiology, 1986-11, Vol.9 (6), p.948-953</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c2898-f7b0445374d8d3c0d3d13f26b02d0946a1191b81bb6981347d6038a33b2945e63</citedby></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://onlinelibrary.wiley.com/doi/pdf/10.1111%2Fj.1540-8159.1986.tb06652.x$$EPDF$$P50$$Gwiley$$H</linktopdf><linktohtml>$$Uhttps://onlinelibrary.wiley.com/doi/full/10.1111%2Fj.1540-8159.1986.tb06652.x$$EHTML$$P50$$Gwiley$$H</linktohtml><link.rule.ids>314,780,784,1417,27924,27925,45574,45575</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/2432573$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>BOUTE, W.</creatorcontrib><creatorcontrib>DERRIEN, Y.</creatorcontrib><creatorcontrib>WITTKAMPF, F.H.M.</creatorcontrib><title>Reliability of Evoked Endocardial T-wave Sensing in 1,500 Pacemaker Patients</title><title>Pacing and clinical electrophysiology</title><addtitle>Pacing Clin Electrophysiol</addtitle><description>Intracardiac measurements of the evoked endocardial T‐wave were performed on 1,500 pacemaker patients. The sensed evoked T‐wave amplitude was found to be above 0.75 mV in 94.1 percent of all cases. No correlation was found between the amplitude of the spontaneous R‐wave and that of the evoked T‐wave. Some electrode characteristics proved to be more favorable than others for evoked endocardial wave sensing, i.e., a surface area of less than 12 mm2. porous surface structure (preferably carbon), and atraumatic fixation.</description><subject>amplitude</subject><subject>Electrocardiography</subject><subject>electrode characteristics</subject><subject>Electrodes</subject><subject>Endocardium - physiopathology</subject><subject>Evoked Potentials</subject><subject>evoked T-wave</subject><subject>Humans</subject><subject>Pacemaker, Artificial</subject><subject>sensing</subject><subject>spontaneous R-wave amplitude</subject><issn>0147-8389</issn><issn>1540-8159</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>1986</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNo9kE1rGzEQhkVISR2nPyGw9JBTdzP6lk4luG5aYhqTuMlRSCu5yF7vpqt1bP_7rLHxXGbgeXlhHoS-YihwP7eLAnMGucJcF1grUXQOhOCk2J6hwQmdowFgJnNFlf6MLlNaAIAAxi_QBWGUcEkHaPIUqmhdrGK3y5p5Nn5vlsFn49o3pW19tFU2yzf2PWTPoU6x_pfFOsPfOEA2tWVY2WVo-6uLoe7SFfo0t1UKX457iP7-HM9Gv_LJ4_3v0d0kL4nSKp9LB4xxKplXnpbgqcd0ToQD4kEzYTHW2CnsnNAKUya9AKospY5oxoOgQ3Rz6H1rm__rkDqziqkMVWXr0KyTkZIQBZz2wetjcO1WwZu3Nq5suzPH93v-_cA3sQq7E8Zg9p7Nwuxlmr1Ms_dsjp7N1kzvRmPNVN-QHxpi6sL21GDbpRGSSm5e_9yb0XQyfZn9eDAz-gEdpn53</recordid><startdate>198611</startdate><enddate>198611</enddate><creator>BOUTE, W.</creator><creator>DERRIEN, Y.</creator><creator>WITTKAMPF, F.H.M.</creator><general>Blackwell Publishing Ltd</general><scope>BSCLL</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>7X8</scope></search><sort><creationdate>198611</creationdate><title>Reliability of Evoked Endocardial T-wave Sensing in 1,500 Pacemaker Patients</title><author>BOUTE, W. ; DERRIEN, Y. ; WITTKAMPF, F.H.M.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c2898-f7b0445374d8d3c0d3d13f26b02d0946a1191b81bb6981347d6038a33b2945e63</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>1986</creationdate><topic>amplitude</topic><topic>Electrocardiography</topic><topic>electrode characteristics</topic><topic>Electrodes</topic><topic>Endocardium - physiopathology</topic><topic>Evoked Potentials</topic><topic>evoked T-wave</topic><topic>Humans</topic><topic>Pacemaker, Artificial</topic><topic>sensing</topic><topic>spontaneous R-wave amplitude</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>BOUTE, W.</creatorcontrib><creatorcontrib>DERRIEN, Y.</creatorcontrib><creatorcontrib>WITTKAMPF, F.H.M.</creatorcontrib><collection>Istex</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>MEDLINE - Academic</collection><jtitle>Pacing and clinical electrophysiology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>BOUTE, W.</au><au>DERRIEN, Y.</au><au>WITTKAMPF, F.H.M.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Reliability of Evoked Endocardial T-wave Sensing in 1,500 Pacemaker Patients</atitle><jtitle>Pacing and clinical electrophysiology</jtitle><addtitle>Pacing Clin Electrophysiol</addtitle><date>1986-11</date><risdate>1986</risdate><volume>9</volume><issue>6</issue><spage>948</spage><epage>953</epage><pages>948-953</pages><issn>0147-8389</issn><eissn>1540-8159</eissn><abstract>Intracardiac measurements of the evoked endocardial T‐wave were performed on 1,500 pacemaker patients. The sensed evoked T‐wave amplitude was found to be above 0.75 mV in 94.1 percent of all cases. No correlation was found between the amplitude of the spontaneous R‐wave and that of the evoked T‐wave. Some electrode characteristics proved to be more favorable than others for evoked endocardial wave sensing, i.e., a surface area of less than 12 mm2. porous surface structure (preferably carbon), and atraumatic fixation.</abstract><cop>Oxford, UK</cop><pub>Blackwell Publishing Ltd</pub><pmid>2432573</pmid><doi>10.1111/j.1540-8159.1986.tb06652.x</doi><tpages>6</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0147-8389
ispartof Pacing and clinical electrophysiology, 1986-11, Vol.9 (6), p.948-953
issn 0147-8389
1540-8159
language eng
recordid cdi_proquest_miscellaneous_77228053
source MEDLINE; Access via Wiley Online Library
subjects amplitude
Electrocardiography
electrode characteristics
Electrodes
Endocardium - physiopathology
Evoked Potentials
evoked T-wave
Humans
Pacemaker, Artificial
sensing
spontaneous R-wave amplitude
title Reliability of Evoked Endocardial T-wave Sensing in 1,500 Pacemaker Patients
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-26T02%3A50%3A59IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Reliability%20of%20Evoked%20Endocardial%20T-wave%20Sensing%20in%201,500%20Pacemaker%20Patients&rft.jtitle=Pacing%20and%20clinical%20electrophysiology&rft.au=BOUTE,%20W.&rft.date=1986-11&rft.volume=9&rft.issue=6&rft.spage=948&rft.epage=953&rft.pages=948-953&rft.issn=0147-8389&rft.eissn=1540-8159&rft_id=info:doi/10.1111/j.1540-8159.1986.tb06652.x&rft_dat=%3Cproquest_pubme%3E77228053%3C/proquest_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=77228053&rft_id=info:pmid/2432573&rfr_iscdi=true