Reliability of Evoked Endocardial T-wave Sensing in 1,500 Pacemaker Patients

Intracardiac measurements of the evoked endocardial T‐wave were performed on 1,500 pacemaker patients. The sensed evoked T‐wave amplitude was found to be above 0.75 mV in 94.1 percent of all cases. No correlation was found between the amplitude of the spontaneous R‐wave and that of the evoked T‐wave...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 1986-11, Vol.9 (6), p.948-953
Hauptverfasser: BOUTE, W., DERRIEN, Y., WITTKAMPF, F.H.M.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Intracardiac measurements of the evoked endocardial T‐wave were performed on 1,500 pacemaker patients. The sensed evoked T‐wave amplitude was found to be above 0.75 mV in 94.1 percent of all cases. No correlation was found between the amplitude of the spontaneous R‐wave and that of the evoked T‐wave. Some electrode characteristics proved to be more favorable than others for evoked endocardial wave sensing, i.e., a surface area of less than 12 mm2. porous surface structure (preferably carbon), and atraumatic fixation.