Assessment of ventricular shunt patency by sonography: a new noninvasive test

This report describes a new test of patency of the ventricular component of a ventriculoperitoneal shunt that precludes injection into the shunt reservoir. Preliminary clinical experience confirms that digital compression of the ventricular shunt reservoir (Cordis standard and pediatric reservoir),...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:American journal of roentgenology (1976) 1986-08, Vol.147 (2), p.353-356
Hauptverfasser: Widder, DJ, Davis, KR, Taveras, JM
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This report describes a new test of patency of the ventricular component of a ventriculoperitoneal shunt that precludes injection into the shunt reservoir. Preliminary clinical experience confirms that digital compression of the ventricular shunt reservoir (Cordis standard and pediatric reservoir), while occluding the shunt distal to the reservoir, generates microbubbles that increase the echogenicity of the catheter, demonstrating shunt patency. In vitro, the extrusion of microbubbles from catheter side holes can be imaged during digital compression of the reservoir, but so far this has not been imaged in patients as a direct sign of shunt patency. A sonographic means of assessing shunt function offers the advantage of simultaneously imaging the ventricular catheter, determining its location, and assessing the etiology of shunt failure. Its usefulness is limited to pediatric cases in which an adequate transfontanelle cranial sonographic study can be obtained and to individuals with large craniotomy defects.
ISSN:0361-803X
1546-3141
DOI:10.2214/ajr.147.2.353