6-Bit and 8-bit digital radiography for detecting simulated periodontal lesions
The purpose of this study was to compare the diagnostic performance of a digital radiography system that uses 6-and 8-bit displays with conventional D-speed film for the detection of simulated periodontal bone lesions. Eleven human hemimandibles were used as specimens. Simulated lesions were created...
Gespeichert in:
Veröffentlicht in: | Oral surgery, oral medicine, oral pathology oral medicine, oral pathology, 1994-04, Vol.77 (4), p.406-411 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 411 |
---|---|
container_issue | 4 |
container_start_page | 406 |
container_title | Oral surgery, oral medicine, oral pathology |
container_volume | 77 |
creator | Pass, B. Furkart, A.J. Dove, S.B. McDavid, W.D. Gregson, P.H. |
description | The purpose of this study was to compare the diagnostic performance of a digital radiography system that uses 6-and 8-bit displays with conventional D-speed film for the detection of simulated periodontal bone lesions. Eleven human hemimandibles were used as specimens. Simulated lesions were created at the buccal cortical plate in the marginal bone area with the use of a round bur 1.4 mm in diameter. Lesions were created in a defined sequence to preclude visual cues as to the depth of the lesions. Lesion size progressed in 0.5 mm increments. At each stage the mandibles were imaged with a Sens-A-Ray system (REGAM Medical Systems AB, Sundsvall, Sweden) and D-speed film. Exposure parameters for each specimen/receptor combination were standardized by either the mean optical density or mean gray value at the approximal crestal bone area. Film images and digital images displayed with 64 and 256 gray levels were presented to six observers for evaluation. Observers were ask to rate their confidence as to the presence or absence of a lesion using a 5-point confidence scale. A total of 96 lesion sites and 96 control sites were presented to the observers. Receiver operating characteristic curves were generated for each system. The area under the curve was used as the index of diagnostic accuracy. The mean receiver operating characteristic areas for 6-bit and 8-bit displays and D-speed film were 0.746 ± 0.043, 0.717 ± 0.056 and 0.742 ± 0.059, respectively. Analysis of variance was used to compare the means. No statistical difference was found between any of the three image displays (
p > 0.05). These findings are consistent with an analytic analysis of the viewing conditions and indicate that 6-bit (64 gray levels) are sufficient for the diagnosis of these simulated periodontal lesions. |
doi_str_mv | 10.1016/0030-4220(94)90205-4 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_76565775</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>0030422094902054</els_id><sourcerecordid>76565775</sourcerecordid><originalsourceid>FETCH-LOGICAL-c415t-951645c6d6300fcccc478bc4cf20819111b660bd6a6ec1c1793c17dd52e08ba73</originalsourceid><addsrcrecordid>eNp9kE1rGzEQhkVJcFy3_6CBPYTSHDYd7eprL4XWNB9g8KU9C6006yisV460DvjfV1sbH6PDSMM88yIeQr5QuKNAxXeAGkpWVfCtYbcNVMBL9oHMqZKqrKjkF2R-Rq7Ix5ReciuZqGdkpoByBXJO1qL85cfCDK5QZZtfzm_8aPoiGufDJprd86HoQiwcjmhHP2yK5Lf73ozoih1GH1wYJr7H5MOQPpHLzvQJP5_uBfl7__vP8rFcrR-elj9XpWWUj2XDqWDcCidqgM7mw6RqLbNdBYo2lNJWCGidMAIttVQ2dS7O8QpBtUbWC_L1mLuL4XWPadRbnyz2vRkw7JOWggsuJc8gO4I2hpQidnoX_dbEg6agJ496kqQnSbph-r9HzfLa9Sl_327RnZdO4vL85jQ3yZq-i2awPp0xBkyxZor5ccQwu3jzGHWyHgeLzsesU7vg3__HP3gtjaA</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>76565775</pqid></control><display><type>article</type><title>6-Bit and 8-bit digital radiography for detecting simulated periodontal lesions</title><source>MEDLINE</source><source>Alma/SFX Local Collection</source><creator>Pass, B. ; Furkart, A.J. ; Dove, S.B. ; McDavid, W.D. ; Gregson, P.H.</creator><creatorcontrib>Pass, B. ; Furkart, A.J. ; Dove, S.B. ; McDavid, W.D. ; Gregson, P.H.</creatorcontrib><description>The purpose of this study was to compare the diagnostic performance of a digital radiography system that uses 6-and 8-bit displays with conventional D-speed film for the detection of simulated periodontal bone lesions. Eleven human hemimandibles were used as specimens. Simulated lesions were created at the buccal cortical plate in the marginal bone area with the use of a round bur 1.4 mm in diameter. Lesions were created in a defined sequence to preclude visual cues as to the depth of the lesions. Lesion size progressed in 0.5 mm increments. At each stage the mandibles were imaged with a Sens-A-Ray system (REGAM Medical Systems AB, Sundsvall, Sweden) and D-speed film. Exposure parameters for each specimen/receptor combination were standardized by either the mean optical density or mean gray value at the approximal crestal bone area. Film images and digital images displayed with 64 and 256 gray levels were presented to six observers for evaluation. Observers were ask to rate their confidence as to the presence or absence of a lesion using a 5-point confidence scale. A total of 96 lesion sites and 96 control sites were presented to the observers. Receiver operating characteristic curves were generated for each system. The area under the curve was used as the index of diagnostic accuracy. The mean receiver operating characteristic areas for 6-bit and 8-bit displays and D-speed film were 0.746 ± 0.043, 0.717 ± 0.056 and 0.742 ± 0.059, respectively. Analysis of variance was used to compare the means. No statistical difference was found between any of the three image displays (
p > 0.05). These findings are consistent with an analytic analysis of the viewing conditions and indicate that 6-bit (64 gray levels) are sufficient for the diagnosis of these simulated periodontal lesions.</description><identifier>ISSN: 0030-4220</identifier><identifier>EISSN: 1878-2175</identifier><identifier>DOI: 10.1016/0030-4220(94)90205-4</identifier><identifier>PMID: 8015807</identifier><identifier>CODEN: OSOMAE</identifier><language>eng</language><publisher>Saint Louis, MO: Elsevier Inc</publisher><subject>Analysis of Variance ; Biological and medical sciences ; Contrast Sensitivity ; Dental Caries - diagnostic imaging ; Dentistry ; Humans ; Investigative techniques, diagnostic techniques (general aspects) ; Mandible - diagnostic imaging ; Medical sciences ; Otorhinolaryngology. Stomatology. Orbit ; Periodontal Diseases - diagnostic imaging ; Radiodiagnosis. Nmr imagery. Nmr spectrometry ; Radiographic Image Enhancement - methods ; Reproducibility of Results ; ROC Curve</subject><ispartof>Oral surgery, oral medicine, oral pathology, 1994-04, Vol.77 (4), p.406-411</ispartof><rights>1994</rights><rights>1994 INIST-CNRS</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c415t-951645c6d6300fcccc478bc4cf20819111b660bd6a6ec1c1793c17dd52e08ba73</citedby><cites>FETCH-LOGICAL-c415t-951645c6d6300fcccc478bc4cf20819111b660bd6a6ec1c1793c17dd52e08ba73</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27924,27925</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&idt=4048494$$DView record in Pascal Francis$$Hfree_for_read</backlink><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/8015807$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Pass, B.</creatorcontrib><creatorcontrib>Furkart, A.J.</creatorcontrib><creatorcontrib>Dove, S.B.</creatorcontrib><creatorcontrib>McDavid, W.D.</creatorcontrib><creatorcontrib>Gregson, P.H.</creatorcontrib><title>6-Bit and 8-bit digital radiography for detecting simulated periodontal lesions</title><title>Oral surgery, oral medicine, oral pathology</title><addtitle>Oral Surg Oral Med Oral Pathol</addtitle><description>The purpose of this study was to compare the diagnostic performance of a digital radiography system that uses 6-and 8-bit displays with conventional D-speed film for the detection of simulated periodontal bone lesions. Eleven human hemimandibles were used as specimens. Simulated lesions were created at the buccal cortical plate in the marginal bone area with the use of a round bur 1.4 mm in diameter. Lesions were created in a defined sequence to preclude visual cues as to the depth of the lesions. Lesion size progressed in 0.5 mm increments. At each stage the mandibles were imaged with a Sens-A-Ray system (REGAM Medical Systems AB, Sundsvall, Sweden) and D-speed film. Exposure parameters for each specimen/receptor combination were standardized by either the mean optical density or mean gray value at the approximal crestal bone area. Film images and digital images displayed with 64 and 256 gray levels were presented to six observers for evaluation. Observers were ask to rate their confidence as to the presence or absence of a lesion using a 5-point confidence scale. A total of 96 lesion sites and 96 control sites were presented to the observers. Receiver operating characteristic curves were generated for each system. The area under the curve was used as the index of diagnostic accuracy. The mean receiver operating characteristic areas for 6-bit and 8-bit displays and D-speed film were 0.746 ± 0.043, 0.717 ± 0.056 and 0.742 ± 0.059, respectively. Analysis of variance was used to compare the means. No statistical difference was found between any of the three image displays (
p > 0.05). These findings are consistent with an analytic analysis of the viewing conditions and indicate that 6-bit (64 gray levels) are sufficient for the diagnosis of these simulated periodontal lesions.</description><subject>Analysis of Variance</subject><subject>Biological and medical sciences</subject><subject>Contrast Sensitivity</subject><subject>Dental Caries - diagnostic imaging</subject><subject>Dentistry</subject><subject>Humans</subject><subject>Investigative techniques, diagnostic techniques (general aspects)</subject><subject>Mandible - diagnostic imaging</subject><subject>Medical sciences</subject><subject>Otorhinolaryngology. Stomatology. Orbit</subject><subject>Periodontal Diseases - diagnostic imaging</subject><subject>Radiodiagnosis. Nmr imagery. Nmr spectrometry</subject><subject>Radiographic Image Enhancement - methods</subject><subject>Reproducibility of Results</subject><subject>ROC Curve</subject><issn>0030-4220</issn><issn>1878-2175</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>1994</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp9kE1rGzEQhkVJcFy3_6CBPYTSHDYd7eprL4XWNB9g8KU9C6006yisV460DvjfV1sbH6PDSMM88yIeQr5QuKNAxXeAGkpWVfCtYbcNVMBL9oHMqZKqrKjkF2R-Rq7Ix5ReciuZqGdkpoByBXJO1qL85cfCDK5QZZtfzm_8aPoiGufDJprd86HoQiwcjmhHP2yK5Lf73ozoih1GH1wYJr7H5MOQPpHLzvQJP5_uBfl7__vP8rFcrR-elj9XpWWUj2XDqWDcCidqgM7mw6RqLbNdBYo2lNJWCGidMAIttVQ2dS7O8QpBtUbWC_L1mLuL4XWPadRbnyz2vRkw7JOWggsuJc8gO4I2hpQidnoX_dbEg6agJ496kqQnSbph-r9HzfLa9Sl_327RnZdO4vL85jQ3yZq-i2awPp0xBkyxZor5ccQwu3jzGHWyHgeLzsesU7vg3__HP3gtjaA</recordid><startdate>19940401</startdate><enddate>19940401</enddate><creator>Pass, B.</creator><creator>Furkart, A.J.</creator><creator>Dove, S.B.</creator><creator>McDavid, W.D.</creator><creator>Gregson, P.H.</creator><general>Elsevier Inc</general><general>Mosby</general><scope>IQODW</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>19940401</creationdate><title>6-Bit and 8-bit digital radiography for detecting simulated periodontal lesions</title><author>Pass, B. ; Furkart, A.J. ; Dove, S.B. ; McDavid, W.D. ; Gregson, P.H.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c415t-951645c6d6300fcccc478bc4cf20819111b660bd6a6ec1c1793c17dd52e08ba73</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>1994</creationdate><topic>Analysis of Variance</topic><topic>Biological and medical sciences</topic><topic>Contrast Sensitivity</topic><topic>Dental Caries - diagnostic imaging</topic><topic>Dentistry</topic><topic>Humans</topic><topic>Investigative techniques, diagnostic techniques (general aspects)</topic><topic>Mandible - diagnostic imaging</topic><topic>Medical sciences</topic><topic>Otorhinolaryngology. Stomatology. Orbit</topic><topic>Periodontal Diseases - diagnostic imaging</topic><topic>Radiodiagnosis. Nmr imagery. Nmr spectrometry</topic><topic>Radiographic Image Enhancement - methods</topic><topic>Reproducibility of Results</topic><topic>ROC Curve</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Pass, B.</creatorcontrib><creatorcontrib>Furkart, A.J.</creatorcontrib><creatorcontrib>Dove, S.B.</creatorcontrib><creatorcontrib>McDavid, W.D.</creatorcontrib><creatorcontrib>Gregson, P.H.</creatorcontrib><collection>Pascal-Francis</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Oral surgery, oral medicine, oral pathology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Pass, B.</au><au>Furkart, A.J.</au><au>Dove, S.B.</au><au>McDavid, W.D.</au><au>Gregson, P.H.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>6-Bit and 8-bit digital radiography for detecting simulated periodontal lesions</atitle><jtitle>Oral surgery, oral medicine, oral pathology</jtitle><addtitle>Oral Surg Oral Med Oral Pathol</addtitle><date>1994-04-01</date><risdate>1994</risdate><volume>77</volume><issue>4</issue><spage>406</spage><epage>411</epage><pages>406-411</pages><issn>0030-4220</issn><eissn>1878-2175</eissn><coden>OSOMAE</coden><abstract>The purpose of this study was to compare the diagnostic performance of a digital radiography system that uses 6-and 8-bit displays with conventional D-speed film for the detection of simulated periodontal bone lesions. Eleven human hemimandibles were used as specimens. Simulated lesions were created at the buccal cortical plate in the marginal bone area with the use of a round bur 1.4 mm in diameter. Lesions were created in a defined sequence to preclude visual cues as to the depth of the lesions. Lesion size progressed in 0.5 mm increments. At each stage the mandibles were imaged with a Sens-A-Ray system (REGAM Medical Systems AB, Sundsvall, Sweden) and D-speed film. Exposure parameters for each specimen/receptor combination were standardized by either the mean optical density or mean gray value at the approximal crestal bone area. Film images and digital images displayed with 64 and 256 gray levels were presented to six observers for evaluation. Observers were ask to rate their confidence as to the presence or absence of a lesion using a 5-point confidence scale. A total of 96 lesion sites and 96 control sites were presented to the observers. Receiver operating characteristic curves were generated for each system. The area under the curve was used as the index of diagnostic accuracy. The mean receiver operating characteristic areas for 6-bit and 8-bit displays and D-speed film were 0.746 ± 0.043, 0.717 ± 0.056 and 0.742 ± 0.059, respectively. Analysis of variance was used to compare the means. No statistical difference was found between any of the three image displays (
p > 0.05). These findings are consistent with an analytic analysis of the viewing conditions and indicate that 6-bit (64 gray levels) are sufficient for the diagnosis of these simulated periodontal lesions.</abstract><cop>Saint Louis, MO</cop><pub>Elsevier Inc</pub><pmid>8015807</pmid><doi>10.1016/0030-4220(94)90205-4</doi><tpages>6</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0030-4220 |
ispartof | Oral surgery, oral medicine, oral pathology, 1994-04, Vol.77 (4), p.406-411 |
issn | 0030-4220 1878-2175 |
language | eng |
recordid | cdi_proquest_miscellaneous_76565775 |
source | MEDLINE; Alma/SFX Local Collection |
subjects | Analysis of Variance Biological and medical sciences Contrast Sensitivity Dental Caries - diagnostic imaging Dentistry Humans Investigative techniques, diagnostic techniques (general aspects) Mandible - diagnostic imaging Medical sciences Otorhinolaryngology. Stomatology. Orbit Periodontal Diseases - diagnostic imaging Radiodiagnosis. Nmr imagery. Nmr spectrometry Radiographic Image Enhancement - methods Reproducibility of Results ROC Curve |
title | 6-Bit and 8-bit digital radiography for detecting simulated periodontal lesions |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-19T10%3A55%3A57IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=6-Bit%20and%208-bit%20digital%20radiography%20for%20detecting%20simulated%20periodontal%20lesions&rft.jtitle=Oral%20surgery,%20oral%20medicine,%20oral%20pathology&rft.au=Pass,%20B.&rft.date=1994-04-01&rft.volume=77&rft.issue=4&rft.spage=406&rft.epage=411&rft.pages=406-411&rft.issn=0030-4220&rft.eissn=1878-2175&rft.coden=OSOMAE&rft_id=info:doi/10.1016/0030-4220(94)90205-4&rft_dat=%3Cproquest_cross%3E76565775%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=76565775&rft_id=info:pmid/8015807&rft_els_id=0030422094902054&rfr_iscdi=true |