Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical considerations during total knee arthroplasty

Fifty-five unconstrained polyethylene tibial inserts were retrieved at revision total knee arthroplasty and examined for evidence of wear after a mean implantation time of 34.2 months (2.5-80 months). Twenty inserts were ultra-high molecular weight polyethylene (UHMWPE) and 35 were carbon-reinforced...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical orthopaedics and related research 1994-02, Vol.299 (299), p.31-43
Hauptverfasser: Wasielewski, R C, Galante, J O, Leighty, R M, Natarajan, R N, Rosenberg, A G
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 43
container_issue 299
container_start_page 31
container_title Clinical orthopaedics and related research
container_volume 299
creator Wasielewski, R C
Galante, J O
Leighty, R M
Natarajan, R N
Rosenberg, A G
description Fifty-five unconstrained polyethylene tibial inserts were retrieved at revision total knee arthroplasty and examined for evidence of wear after a mean implantation time of 34.2 months (2.5-80 months). Twenty inserts were ultra-high molecular weight polyethylene (UHMWPE) and 35 were carbon-reinforced polyethylene. Topographic maps of the articular and metal-backed surfaces of each component were constructed to characterize the extent and location of polyethylene degradation, identified visually by mode. In 32 of the retrieved inserts, pre- and postarthroplasty or prerevision radiographs were analyzed for component positioning, sizing, and extremity alignment. These factors then were compared with the patterns and severity of polyethylene wear on the inserts to establish correlations. Severe generalized articular wear was seen in inserts with third body wear from patellar metal-backed failure and cement debris. Severe localized delamination wear was seen in inserts with rotational-subluxation patterns of wear (p = 0.05). The external rotation subluxation wear pattern was strongly associated with knees that had lateral subluxation of the patella (p = 0.0002). Articular wear and cold flow into screw holes tended to be greater in the tightest prearthroplasty compartment (medial in the varus knee [p = 0.0157]; lateral in the valgus knees [p = 0.0226]). Fourteen of 16 knees with a preoperative varus deformities--even when corrected to a normal postarthroplasty anatomic axis--still had greater medial compartment articular wear (p = 0.001). Twelve of these knees did not have a medial release at the time of initial arthroplasty. Preoperative varus also was found to be related to the occurrence of posteromedial cold flow of polyethylene into tibial tray screw holes (p = 0.007). Increasing tibial insert posterior slope was associated with increasingly posterior articular wear track location (p = 0.03). This study indicates that unconstrained tibial component wear patterns and severity may be associated with clinical and mechanical factors under the surgeon's control, including component size and position, and knee alignment and ligament balance.
doi_str_mv 10.1097/00003086-199402000-00005
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_76383737</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>76383737</sourcerecordid><originalsourceid>FETCH-LOGICAL-c310t-e2e2444bc94f28d0cba0b357e858d1b572eb0b20504954b9353fd36a64b4cae73</originalsourceid><addsrcrecordid>eNo9kE1OwzAQhb0AlVI4ApJX7AJ2bCfxElX8SZXYgGAX2c6EGFIn2A5Sb8CxcWnpbJ7mzTcz0kMIU3JFiSyvSSpGqiKjUnKSpy7bWuIIzZPITOb07QSdhvCxBbnIZ2hWUSoJE3P08wrK41HFCN4FPDjsIXoL39Dgceg3ELtNDw5wtNqqHlsXwMeAlWtw7MD6xPcq2sGFzo44DjiC6Zw1iTXJtA343Rg3k7fuPSExzT4dAFY-dn4YexXi5gwdt6oPcL7XBXq5u31ePmSrp_vH5c0qM4ySmEEOOedcG8nbvGqI0YpoJkqoRNVQLcocNNE5EYRLwbVkgrUNK1TBNTcKSrZAl7u7ox--JgixXttgoO-Vg2EKdVmwipVsC1Y70PghBA9tPXq7Vn5TU1Jvg6__g68Pwf9ZIq1e7H9Meg3NYXGfOvsFsY-EdQ</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>76383737</pqid></control><display><type>article</type><title>Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical considerations during total knee arthroplasty</title><source>Journals@Ovid Ovid Autoload</source><source>MEDLINE</source><creator>Wasielewski, R C ; Galante, J O ; Leighty, R M ; Natarajan, R N ; Rosenberg, A G</creator><creatorcontrib>Wasielewski, R C ; Galante, J O ; Leighty, R M ; Natarajan, R N ; Rosenberg, A G</creatorcontrib><description>Fifty-five unconstrained polyethylene tibial inserts were retrieved at revision total knee arthroplasty and examined for evidence of wear after a mean implantation time of 34.2 months (2.5-80 months). Twenty inserts were ultra-high molecular weight polyethylene (UHMWPE) and 35 were carbon-reinforced polyethylene. Topographic maps of the articular and metal-backed surfaces of each component were constructed to characterize the extent and location of polyethylene degradation, identified visually by mode. In 32 of the retrieved inserts, pre- and postarthroplasty or prerevision radiographs were analyzed for component positioning, sizing, and extremity alignment. These factors then were compared with the patterns and severity of polyethylene wear on the inserts to establish correlations. Severe generalized articular wear was seen in inserts with third body wear from patellar metal-backed failure and cement debris. Severe localized delamination wear was seen in inserts with rotational-subluxation patterns of wear (p = 0.05). The external rotation subluxation wear pattern was strongly associated with knees that had lateral subluxation of the patella (p = 0.0002). Articular wear and cold flow into screw holes tended to be greater in the tightest prearthroplasty compartment (medial in the varus knee [p = 0.0157]; lateral in the valgus knees [p = 0.0226]). Fourteen of 16 knees with a preoperative varus deformities--even when corrected to a normal postarthroplasty anatomic axis--still had greater medial compartment articular wear (p = 0.001). Twelve of these knees did not have a medial release at the time of initial arthroplasty. Preoperative varus also was found to be related to the occurrence of posteromedial cold flow of polyethylene into tibial tray screw holes (p = 0.007). Increasing tibial insert posterior slope was associated with increasingly posterior articular wear track location (p = 0.03). This study indicates that unconstrained tibial component wear patterns and severity may be associated with clinical and mechanical factors under the surgeon's control, including component size and position, and knee alignment and ligament balance.</description><identifier>ISSN: 0009-921X</identifier><identifier>DOI: 10.1097/00003086-199402000-00005</identifier><identifier>PMID: 8119035</identifier><language>eng</language><publisher>United States</publisher><subject>Adult ; Aged ; Biomechanical Phenomena ; Humans ; Knee Joint - diagnostic imaging ; Knee Prosthesis - statistics &amp; numerical data ; Logistic Models ; Middle Aged ; Polyethylenes ; Prosthesis Design - statistics &amp; numerical data ; Radiography ; Reoperation ; Surface Properties ; Tibia</subject><ispartof>Clinical orthopaedics and related research, 1994-02, Vol.299 (299), p.31-43</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c310t-e2e2444bc94f28d0cba0b357e858d1b572eb0b20504954b9353fd36a64b4cae73</citedby></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,776,780,27901,27902</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/8119035$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Wasielewski, R C</creatorcontrib><creatorcontrib>Galante, J O</creatorcontrib><creatorcontrib>Leighty, R M</creatorcontrib><creatorcontrib>Natarajan, R N</creatorcontrib><creatorcontrib>Rosenberg, A G</creatorcontrib><title>Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical considerations during total knee arthroplasty</title><title>Clinical orthopaedics and related research</title><addtitle>Clin Orthop Relat Res</addtitle><description>Fifty-five unconstrained polyethylene tibial inserts were retrieved at revision total knee arthroplasty and examined for evidence of wear after a mean implantation time of 34.2 months (2.5-80 months). Twenty inserts were ultra-high molecular weight polyethylene (UHMWPE) and 35 were carbon-reinforced polyethylene. Topographic maps of the articular and metal-backed surfaces of each component were constructed to characterize the extent and location of polyethylene degradation, identified visually by mode. In 32 of the retrieved inserts, pre- and postarthroplasty or prerevision radiographs were analyzed for component positioning, sizing, and extremity alignment. These factors then were compared with the patterns and severity of polyethylene wear on the inserts to establish correlations. Severe generalized articular wear was seen in inserts with third body wear from patellar metal-backed failure and cement debris. Severe localized delamination wear was seen in inserts with rotational-subluxation patterns of wear (p = 0.05). The external rotation subluxation wear pattern was strongly associated with knees that had lateral subluxation of the patella (p = 0.0002). Articular wear and cold flow into screw holes tended to be greater in the tightest prearthroplasty compartment (medial in the varus knee [p = 0.0157]; lateral in the valgus knees [p = 0.0226]). Fourteen of 16 knees with a preoperative varus deformities--even when corrected to a normal postarthroplasty anatomic axis--still had greater medial compartment articular wear (p = 0.001). Twelve of these knees did not have a medial release at the time of initial arthroplasty. Preoperative varus also was found to be related to the occurrence of posteromedial cold flow of polyethylene into tibial tray screw holes (p = 0.007). Increasing tibial insert posterior slope was associated with increasingly posterior articular wear track location (p = 0.03). This study indicates that unconstrained tibial component wear patterns and severity may be associated with clinical and mechanical factors under the surgeon's control, including component size and position, and knee alignment and ligament balance.</description><subject>Adult</subject><subject>Aged</subject><subject>Biomechanical Phenomena</subject><subject>Humans</subject><subject>Knee Joint - diagnostic imaging</subject><subject>Knee Prosthesis - statistics &amp; numerical data</subject><subject>Logistic Models</subject><subject>Middle Aged</subject><subject>Polyethylenes</subject><subject>Prosthesis Design - statistics &amp; numerical data</subject><subject>Radiography</subject><subject>Reoperation</subject><subject>Surface Properties</subject><subject>Tibia</subject><issn>0009-921X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>1994</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNo9kE1OwzAQhb0AlVI4ApJX7AJ2bCfxElX8SZXYgGAX2c6EGFIn2A5Sb8CxcWnpbJ7mzTcz0kMIU3JFiSyvSSpGqiKjUnKSpy7bWuIIzZPITOb07QSdhvCxBbnIZ2hWUSoJE3P08wrK41HFCN4FPDjsIXoL39Dgceg3ELtNDw5wtNqqHlsXwMeAlWtw7MD6xPcq2sGFzo44DjiC6Zw1iTXJtA343Rg3k7fuPSExzT4dAFY-dn4YexXi5gwdt6oPcL7XBXq5u31ePmSrp_vH5c0qM4ySmEEOOedcG8nbvGqI0YpoJkqoRNVQLcocNNE5EYRLwbVkgrUNK1TBNTcKSrZAl7u7ox--JgixXttgoO-Vg2EKdVmwipVsC1Y70PghBA9tPXq7Vn5TU1Jvg6__g68Pwf9ZIq1e7H9Meg3NYXGfOvsFsY-EdQ</recordid><startdate>19940201</startdate><enddate>19940201</enddate><creator>Wasielewski, R C</creator><creator>Galante, J O</creator><creator>Leighty, R M</creator><creator>Natarajan, R N</creator><creator>Rosenberg, A G</creator><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>19940201</creationdate><title>Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical considerations during total knee arthroplasty</title><author>Wasielewski, R C ; Galante, J O ; Leighty, R M ; Natarajan, R N ; Rosenberg, A G</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c310t-e2e2444bc94f28d0cba0b357e858d1b572eb0b20504954b9353fd36a64b4cae73</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>1994</creationdate><topic>Adult</topic><topic>Aged</topic><topic>Biomechanical Phenomena</topic><topic>Humans</topic><topic>Knee Joint - diagnostic imaging</topic><topic>Knee Prosthesis - statistics &amp; numerical data</topic><topic>Logistic Models</topic><topic>Middle Aged</topic><topic>Polyethylenes</topic><topic>Prosthesis Design - statistics &amp; numerical data</topic><topic>Radiography</topic><topic>Reoperation</topic><topic>Surface Properties</topic><topic>Tibia</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Wasielewski, R C</creatorcontrib><creatorcontrib>Galante, J O</creatorcontrib><creatorcontrib>Leighty, R M</creatorcontrib><creatorcontrib>Natarajan, R N</creatorcontrib><creatorcontrib>Rosenberg, A G</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Clinical orthopaedics and related research</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Wasielewski, R C</au><au>Galante, J O</au><au>Leighty, R M</au><au>Natarajan, R N</au><au>Rosenberg, A G</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical considerations during total knee arthroplasty</atitle><jtitle>Clinical orthopaedics and related research</jtitle><addtitle>Clin Orthop Relat Res</addtitle><date>1994-02-01</date><risdate>1994</risdate><volume>299</volume><issue>299</issue><spage>31</spage><epage>43</epage><pages>31-43</pages><issn>0009-921X</issn><abstract>Fifty-five unconstrained polyethylene tibial inserts were retrieved at revision total knee arthroplasty and examined for evidence of wear after a mean implantation time of 34.2 months (2.5-80 months). Twenty inserts were ultra-high molecular weight polyethylene (UHMWPE) and 35 were carbon-reinforced polyethylene. Topographic maps of the articular and metal-backed surfaces of each component were constructed to characterize the extent and location of polyethylene degradation, identified visually by mode. In 32 of the retrieved inserts, pre- and postarthroplasty or prerevision radiographs were analyzed for component positioning, sizing, and extremity alignment. These factors then were compared with the patterns and severity of polyethylene wear on the inserts to establish correlations. Severe generalized articular wear was seen in inserts with third body wear from patellar metal-backed failure and cement debris. Severe localized delamination wear was seen in inserts with rotational-subluxation patterns of wear (p = 0.05). The external rotation subluxation wear pattern was strongly associated with knees that had lateral subluxation of the patella (p = 0.0002). Articular wear and cold flow into screw holes tended to be greater in the tightest prearthroplasty compartment (medial in the varus knee [p = 0.0157]; lateral in the valgus knees [p = 0.0226]). Fourteen of 16 knees with a preoperative varus deformities--even when corrected to a normal postarthroplasty anatomic axis--still had greater medial compartment articular wear (p = 0.001). Twelve of these knees did not have a medial release at the time of initial arthroplasty. Preoperative varus also was found to be related to the occurrence of posteromedial cold flow of polyethylene into tibial tray screw holes (p = 0.007). Increasing tibial insert posterior slope was associated with increasingly posterior articular wear track location (p = 0.03). This study indicates that unconstrained tibial component wear patterns and severity may be associated with clinical and mechanical factors under the surgeon's control, including component size and position, and knee alignment and ligament balance.</abstract><cop>United States</cop><pmid>8119035</pmid><doi>10.1097/00003086-199402000-00005</doi><tpages>13</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0009-921X
ispartof Clinical orthopaedics and related research, 1994-02, Vol.299 (299), p.31-43
issn 0009-921X
language eng
recordid cdi_proquest_miscellaneous_76383737
source Journals@Ovid Ovid Autoload; MEDLINE
subjects Adult
Aged
Biomechanical Phenomena
Humans
Knee Joint - diagnostic imaging
Knee Prosthesis - statistics & numerical data
Logistic Models
Middle Aged
Polyethylenes
Prosthesis Design - statistics & numerical data
Radiography
Reoperation
Surface Properties
Tibia
title Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical considerations during total knee arthroplasty
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-18T15%3A33%3A02IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Wear%20patterns%20on%20retrieved%20polyethylene%20tibial%20inserts%20and%20their%20relationship%20to%20technical%20considerations%20during%20total%20knee%20arthroplasty&rft.jtitle=Clinical%20orthopaedics%20and%20related%20research&rft.au=Wasielewski,%20R%20C&rft.date=1994-02-01&rft.volume=299&rft.issue=299&rft.spage=31&rft.epage=43&rft.pages=31-43&rft.issn=0009-921X&rft_id=info:doi/10.1097/00003086-199402000-00005&rft_dat=%3Cproquest_cross%3E76383737%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=76383737&rft_id=info:pmid/8119035&rfr_iscdi=true