Precision and accuracy of stereotactic convergent beam irradiations from a linear accelerator
Purpose : The accuracy and the precision for radiosurgery procedures at linear accelerator facilities were investigated. Methods and Materials : The technique of convergent beam irradiation, that is a series of successive isocentric arc irradiations, is specifically considered in this paper. Accurac...
Gespeichert in:
Veröffentlicht in: | International journal of radiation oncology, biology, physics biology, physics, 1994-01, Vol.28 (2), p.481-492 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 492 |
---|---|
container_issue | 2 |
container_start_page | 481 |
container_title | International journal of radiation oncology, biology, physics |
container_volume | 28 |
creator | Hartmann, Günther H. Bauer-Kirpes, Bernhard Serago, Christopher F. Lorenz, Walter J. |
description | Purpose
: The accuracy and the precision for radiosurgery procedures at linear accelerator facilities were investigated.
Methods and Materials
: The technique of convergent beam irradiation, that is a series of successive isocentric arc irradiations, is specifically considered in this paper. Accuracy and precision depend on a sequence of methods and equipment among which localization of the target, patient alignment, and the dose delivery are the most critical steps. The purpose of the investigation was to quantitatively assess their contribution to the overall accuracy. The definitions and methods used to quantify and control accuracy are described. Measurements were carried out at a phantom to analyze the localization and positioning errors. Errors which may occur with the dose delivery technique were studied by it computer simulation.
Results
: The calculations showed that these errors are not the main contributors to the overall accuracy as long as the linac inaccuracies are in the order or less than 1 mm. The accuracy found in the localization and positioning methods was less than 1 mm.
Conclusion
: It was concluded that an overall accuracy in the order of 1 mm can be obtained also under routine conditions. The great importance of adequate quality control is emphasized. |
doi_str_mv | 10.1016/0360-3016(94)90075-2 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_76330333</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>0360301694900752</els_id><sourcerecordid>76330333</sourcerecordid><originalsourceid>FETCH-LOGICAL-c367t-802ca7edded9481c1ffd235b915371da6f071a4d487ca1e13e073e02e4e302a43</originalsourceid><addsrcrecordid>eNp9kUGLFDEQhYMo6-zoP1DIQRY9tCaddNJ9EWRxXWFBDwpeJNQk1RLp7qyVzML--007wxw3UCRQ770UXzH2Sor3UkjzQSgjGlVfbwf9bhDCdk37hG1kb4dGdd2vp2xzkjxn5zn_FUJIafUZO-tba4zpNuz3d0Ifc0wLhyVw8H5P4O95GnkuSJgK-BI992m5Q_qDS-E7hJlHIggRSjVmPlKaOfApLgi0ZuCEBCXRC_ZshCnjy-O9ZT-vPv-4vG5uvn35evnppvHK2NL0ovVgMQQMg-6ll-MYWtXtBtkpKwOYUVgJOujeepAoFQpbq0WNSrSg1ZZdHHJvKf3bYy5ujrlOMcGCaZ-dNUoJVc-W6YPQU8qZcHS3FGegeyeFW6m6FZlbkblBu_9UXVttr4_5-92M4WQ6Yqz9N8c-ZA_TSLBUqCeZ6q3ozfr7x4MMK4u7iOSyj7h4DLFuobiQ4uNzPAD-55Qo</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>76330333</pqid></control><display><type>article</type><title>Precision and accuracy of stereotactic convergent beam irradiations from a linear accelerator</title><source>MEDLINE</source><source>Elsevier ScienceDirect Journals</source><creator>Hartmann, Günther H. ; Bauer-Kirpes, Bernhard ; Serago, Christopher F. ; Lorenz, Walter J.</creator><creatorcontrib>Hartmann, Günther H. ; Bauer-Kirpes, Bernhard ; Serago, Christopher F. ; Lorenz, Walter J.</creatorcontrib><description>Purpose
: The accuracy and the precision for radiosurgery procedures at linear accelerator facilities were investigated.
Methods and Materials
: The technique of convergent beam irradiation, that is a series of successive isocentric arc irradiations, is specifically considered in this paper. Accuracy and precision depend on a sequence of methods and equipment among which localization of the target, patient alignment, and the dose delivery are the most critical steps. The purpose of the investigation was to quantitatively assess their contribution to the overall accuracy. The definitions and methods used to quantify and control accuracy are described. Measurements were carried out at a phantom to analyze the localization and positioning errors. Errors which may occur with the dose delivery technique were studied by it computer simulation.
Results
: The calculations showed that these errors are not the main contributors to the overall accuracy as long as the linac inaccuracies are in the order or less than 1 mm. The accuracy found in the localization and positioning methods was less than 1 mm.
Conclusion
: It was concluded that an overall accuracy in the order of 1 mm can be obtained also under routine conditions. The great importance of adequate quality control is emphasized.</description><identifier>ISSN: 0360-3016</identifier><identifier>EISSN: 1879-355X</identifier><identifier>DOI: 10.1016/0360-3016(94)90075-2</identifier><identifier>PMID: 8276665</identifier><identifier>CODEN: IOBPD3</identifier><language>eng</language><publisher>New York, NY: Elsevier Inc</publisher><subject>Accuracy ; Biological and medical sciences ; Diseases of the nervous system ; Humans ; Linear accelerator ; Medical sciences ; Particle Accelerators ; Precision ; Radiosurgery - methods ; Radiotherapy. Instrumental treatment. Physiotherapy. Reeducation. Rehabilitation, orthophony, crenotherapy. Diet therapy and various other treatments (general aspects) ; Stereotactic radiosurgery</subject><ispartof>International journal of radiation oncology, biology, physics, 1994-01, Vol.28 (2), p.481-492</ispartof><rights>1993</rights><rights>1994 INIST-CNRS</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c367t-802ca7edded9481c1ffd235b915371da6f071a4d487ca1e13e073e02e4e302a43</citedby><cites>FETCH-LOGICAL-c367t-802ca7edded9481c1ffd235b915371da6f071a4d487ca1e13e073e02e4e302a43</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://www.sciencedirect.com/science/article/pii/0360301694900752$$EHTML$$P50$$Gelsevier$$H</linktohtml><link.rule.ids>314,776,780,3537,27901,27902,65306</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&idt=3870863$$DView record in Pascal Francis$$Hfree_for_read</backlink><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/8276665$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Hartmann, Günther H.</creatorcontrib><creatorcontrib>Bauer-Kirpes, Bernhard</creatorcontrib><creatorcontrib>Serago, Christopher F.</creatorcontrib><creatorcontrib>Lorenz, Walter J.</creatorcontrib><title>Precision and accuracy of stereotactic convergent beam irradiations from a linear accelerator</title><title>International journal of radiation oncology, biology, physics</title><addtitle>Int J Radiat Oncol Biol Phys</addtitle><description>Purpose
: The accuracy and the precision for radiosurgery procedures at linear accelerator facilities were investigated.
Methods and Materials
: The technique of convergent beam irradiation, that is a series of successive isocentric arc irradiations, is specifically considered in this paper. Accuracy and precision depend on a sequence of methods and equipment among which localization of the target, patient alignment, and the dose delivery are the most critical steps. The purpose of the investigation was to quantitatively assess their contribution to the overall accuracy. The definitions and methods used to quantify and control accuracy are described. Measurements were carried out at a phantom to analyze the localization and positioning errors. Errors which may occur with the dose delivery technique were studied by it computer simulation.
Results
: The calculations showed that these errors are not the main contributors to the overall accuracy as long as the linac inaccuracies are in the order or less than 1 mm. The accuracy found in the localization and positioning methods was less than 1 mm.
Conclusion
: It was concluded that an overall accuracy in the order of 1 mm can be obtained also under routine conditions. The great importance of adequate quality control is emphasized.</description><subject>Accuracy</subject><subject>Biological and medical sciences</subject><subject>Diseases of the nervous system</subject><subject>Humans</subject><subject>Linear accelerator</subject><subject>Medical sciences</subject><subject>Particle Accelerators</subject><subject>Precision</subject><subject>Radiosurgery - methods</subject><subject>Radiotherapy. Instrumental treatment. Physiotherapy. Reeducation. Rehabilitation, orthophony, crenotherapy. Diet therapy and various other treatments (general aspects)</subject><subject>Stereotactic radiosurgery</subject><issn>0360-3016</issn><issn>1879-355X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>1994</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp9kUGLFDEQhYMo6-zoP1DIQRY9tCaddNJ9EWRxXWFBDwpeJNQk1RLp7qyVzML--007wxw3UCRQ770UXzH2Sor3UkjzQSgjGlVfbwf9bhDCdk37hG1kb4dGdd2vp2xzkjxn5zn_FUJIafUZO-tba4zpNuz3d0Ifc0wLhyVw8H5P4O95GnkuSJgK-BI992m5Q_qDS-E7hJlHIggRSjVmPlKaOfApLgi0ZuCEBCXRC_ZshCnjy-O9ZT-vPv-4vG5uvn35evnppvHK2NL0ovVgMQQMg-6ll-MYWtXtBtkpKwOYUVgJOujeepAoFQpbq0WNSrSg1ZZdHHJvKf3bYy5ujrlOMcGCaZ-dNUoJVc-W6YPQU8qZcHS3FGegeyeFW6m6FZlbkblBu_9UXVttr4_5-92M4WQ6Yqz9N8c-ZA_TSLBUqCeZ6q3ozfr7x4MMK4u7iOSyj7h4DLFuobiQ4uNzPAD-55Qo</recordid><startdate>19940115</startdate><enddate>19940115</enddate><creator>Hartmann, Günther H.</creator><creator>Bauer-Kirpes, Bernhard</creator><creator>Serago, Christopher F.</creator><creator>Lorenz, Walter J.</creator><general>Elsevier Inc</general><general>Elsevier</general><scope>IQODW</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>19940115</creationdate><title>Precision and accuracy of stereotactic convergent beam irradiations from a linear accelerator</title><author>Hartmann, Günther H. ; Bauer-Kirpes, Bernhard ; Serago, Christopher F. ; Lorenz, Walter J.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c367t-802ca7edded9481c1ffd235b915371da6f071a4d487ca1e13e073e02e4e302a43</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>1994</creationdate><topic>Accuracy</topic><topic>Biological and medical sciences</topic><topic>Diseases of the nervous system</topic><topic>Humans</topic><topic>Linear accelerator</topic><topic>Medical sciences</topic><topic>Particle Accelerators</topic><topic>Precision</topic><topic>Radiosurgery - methods</topic><topic>Radiotherapy. Instrumental treatment. Physiotherapy. Reeducation. Rehabilitation, orthophony, crenotherapy. Diet therapy and various other treatments (general aspects)</topic><topic>Stereotactic radiosurgery</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Hartmann, Günther H.</creatorcontrib><creatorcontrib>Bauer-Kirpes, Bernhard</creatorcontrib><creatorcontrib>Serago, Christopher F.</creatorcontrib><creatorcontrib>Lorenz, Walter J.</creatorcontrib><collection>Pascal-Francis</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>International journal of radiation oncology, biology, physics</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Hartmann, Günther H.</au><au>Bauer-Kirpes, Bernhard</au><au>Serago, Christopher F.</au><au>Lorenz, Walter J.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Precision and accuracy of stereotactic convergent beam irradiations from a linear accelerator</atitle><jtitle>International journal of radiation oncology, biology, physics</jtitle><addtitle>Int J Radiat Oncol Biol Phys</addtitle><date>1994-01-15</date><risdate>1994</risdate><volume>28</volume><issue>2</issue><spage>481</spage><epage>492</epage><pages>481-492</pages><issn>0360-3016</issn><eissn>1879-355X</eissn><coden>IOBPD3</coden><abstract>Purpose
: The accuracy and the precision for radiosurgery procedures at linear accelerator facilities were investigated.
Methods and Materials
: The technique of convergent beam irradiation, that is a series of successive isocentric arc irradiations, is specifically considered in this paper. Accuracy and precision depend on a sequence of methods and equipment among which localization of the target, patient alignment, and the dose delivery are the most critical steps. The purpose of the investigation was to quantitatively assess their contribution to the overall accuracy. The definitions and methods used to quantify and control accuracy are described. Measurements were carried out at a phantom to analyze the localization and positioning errors. Errors which may occur with the dose delivery technique were studied by it computer simulation.
Results
: The calculations showed that these errors are not the main contributors to the overall accuracy as long as the linac inaccuracies are in the order or less than 1 mm. The accuracy found in the localization and positioning methods was less than 1 mm.
Conclusion
: It was concluded that an overall accuracy in the order of 1 mm can be obtained also under routine conditions. The great importance of adequate quality control is emphasized.</abstract><cop>New York, NY</cop><pub>Elsevier Inc</pub><pmid>8276665</pmid><doi>10.1016/0360-3016(94)90075-2</doi><tpages>12</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0360-3016 |
ispartof | International journal of radiation oncology, biology, physics, 1994-01, Vol.28 (2), p.481-492 |
issn | 0360-3016 1879-355X |
language | eng |
recordid | cdi_proquest_miscellaneous_76330333 |
source | MEDLINE; Elsevier ScienceDirect Journals |
subjects | Accuracy Biological and medical sciences Diseases of the nervous system Humans Linear accelerator Medical sciences Particle Accelerators Precision Radiosurgery - methods Radiotherapy. Instrumental treatment. Physiotherapy. Reeducation. Rehabilitation, orthophony, crenotherapy. Diet therapy and various other treatments (general aspects) Stereotactic radiosurgery |
title | Precision and accuracy of stereotactic convergent beam irradiations from a linear accelerator |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-09T22%3A20%3A40IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Precision%20and%20accuracy%20of%20stereotactic%20convergent%20beam%20irradiations%20from%20a%20linear%20accelerator&rft.jtitle=International%20journal%20of%20radiation%20oncology,%20biology,%20physics&rft.au=Hartmann,%20G%C3%BCnther%20H.&rft.date=1994-01-15&rft.volume=28&rft.issue=2&rft.spage=481&rft.epage=492&rft.pages=481-492&rft.issn=0360-3016&rft.eissn=1879-355X&rft.coden=IOBPD3&rft_id=info:doi/10.1016/0360-3016(94)90075-2&rft_dat=%3Cproquest_cross%3E76330333%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=76330333&rft_id=info:pmid/8276665&rft_els_id=0360301694900752&rfr_iscdi=true |