Rapid agglutination testing in an ultrasonic standing wave
The time taken to perform diagnostic agglutination tests can be signifintlyreduced by applying an ultrasonic standing wave field to a droplet of reactants held in a capillary tube. Avian erythrocytes, bacteria and latex particles from commercially available test kits were agglutinated in 15 s, 5 min...
Gespeichert in:
Veröffentlicht in: | Journal of immunological methods 1993-09, Vol.165 (1), p.47-57 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The time taken to perform diagnostic agglutination tests can be signifintlyreduced by applying an ultrasonic standing wave field to a droplet of reactants held in a capillary tube. Avian erythrocytes, bacteria and latex particles from commercially available test kits were agglutinated in 15 s, 5 min, and 1 min respectively. These times compare favourably with the times of 30 min, 4 h, and 8 min required for agglutination by the methods prescribed for the respective kits. No loss in sensitivity or specificity was observed with the ultrasonic method. A multi-test procedure is also described whereby a series of five droplets loaded in a single capillary can be tested in less than 4 min by drawing the capillary along the axis of the ultrasonic field of a ring transducer. |
---|---|
ISSN: | 0022-1759 1872-7905 |
DOI: | 10.1016/0022-1759(93)90105-G |