Towards the limit of quantifying low-amplitude strains on bone and in coagulum around immediately loaded oral implants in extraction sockets
The purpose of this study was to quantify strains in coagulum around immediately loaded oral implants in extraction sockets at the ex vivo level. Bilateral maxillary premolar teeth of two fresh human cadavers were extracted and psi 4.1 x 12 mm Straumann TE implants were placed in the sockets of firs...
Gespeichert in:
Veröffentlicht in: | Medical & biological engineering & computing 2006-03, Vol.44 (1-2), p.86-94 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 94 |
---|---|
container_issue | 1-2 |
container_start_page | 86 |
container_title | Medical & biological engineering & computing |
container_volume | 44 |
creator | Cehreli, Murat Cavit Comert, Ayhan Akkocaoglu, Murat Tekdemir, Ibrahim Akca, Kivanc |
description | The purpose of this study was to quantify strains in coagulum around immediately loaded oral implants in extraction sockets at the ex vivo level. Bilateral maxillary premolar teeth of two fresh human cadavers were extracted and psi 4.1 x 12 mm Straumann TE implants were placed in the sockets of first and second premolars by utilizing mesio-distal and palatal anchorage, respectively. Installation torque value (ITV) of each implant was measured by a custom-made torque wrench and resonance frequency analyses (RFAs) were undertaken to determine intraosseous stability. Upon abutment connection, a gold coping allowing the placement of a miniature load cell to contact the underlying solid abutment was fabricated. A linear strain gauge was connected to the coping at a distance for strain measurements in coagulum around the implant neck in the extraction socket. Linear strain gauges were also bonded on the labial marginal bone of each extraction socket. Strain measurements were performed at a sample rate of 10 kHz simultaneously monitored from a computer connected to data acquisition system and under a maximum load of 100 N on each implant with or without human coagulum in the extraction socket. Low-amplitude strains were measured around immediate implants. The increase in load increased strains on labial marginal cortical bone around implants (P < 0.05). Bone strains were higher on the implant loaded, when coagulum was present in the bone defects (P < 0.05). Strains within coagulum around mesiodistally anchored implants were higher than palatally anchored implants (P < 0.05). The type of implant on anchorage and presence of coagulum has an impact mechanotransduction to buccal marginal bone around immediate implants. |
doi_str_mv | 10.1007/s11517-005-0010-4 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_754567075</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>754567075</sourcerecordid><originalsourceid>FETCH-LOGICAL-c389t-dc43f8fa7c3e954a56923212147c4c244b6df9834dfe59e314cb884ec55d18b03</originalsourceid><addsrcrecordid>eNqFkU1rFTEUhoMo9rb6A9xIcGFXU3MmySSzlGKrUHBT1yGTZGpqZnKbD9r7H_zRzeVeEFzoIgQOz_uQkxehd0AugBDxKQNwEB0hvB0gHXuBNiAYdIQx9hJtCDDSEQB5gk5zviekB96z1-gEhrEfx55v0O_b-KiTzbj8dDj4xRccZ_xQ9Vr8vPPrHQ7xsdPLNvhSrcO5JO3XjOOKp7g6rFeL_YpN1Hc11AXrFOt-tCzOel1c2DWBts7imHRo821o6rzPuKfmMsU3VY7mlyv5DXo165Dd2-N9hn5cfbm9_NrdfL_-dvn5pjNUjqWzhtFZzloY6kbONG_b0B56YMIw0zM2DXYeJWV2dnx0FJiZpGTOcG5BToSeofODd5viQ3W5qMVn40J7mos1K8EZHwQRvJEf_0kOUkhJgf4X7AmlQhJo4Ie_wPtY09rWVcMAlEsqhwbBATIp5pzcrLbJLzrtFBC1r14dqleterWvXrGWeX8U16l9_p_EsWv6DIzbqy4</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>661358386</pqid></control><display><type>article</type><title>Towards the limit of quantifying low-amplitude strains on bone and in coagulum around immediately loaded oral implants in extraction sockets</title><source>MEDLINE</source><source>SpringerLink Journals</source><source>Business Source Complete</source><creator>Cehreli, Murat Cavit ; Comert, Ayhan ; Akkocaoglu, Murat ; Tekdemir, Ibrahim ; Akca, Kivanc</creator><creatorcontrib>Cehreli, Murat Cavit ; Comert, Ayhan ; Akkocaoglu, Murat ; Tekdemir, Ibrahim ; Akca, Kivanc</creatorcontrib><description>The purpose of this study was to quantify strains in coagulum around immediately loaded oral implants in extraction sockets at the ex vivo level. Bilateral maxillary premolar teeth of two fresh human cadavers were extracted and psi 4.1 x 12 mm Straumann TE implants were placed in the sockets of first and second premolars by utilizing mesio-distal and palatal anchorage, respectively. Installation torque value (ITV) of each implant was measured by a custom-made torque wrench and resonance frequency analyses (RFAs) were undertaken to determine intraosseous stability. Upon abutment connection, a gold coping allowing the placement of a miniature load cell to contact the underlying solid abutment was fabricated. A linear strain gauge was connected to the coping at a distance for strain measurements in coagulum around the implant neck in the extraction socket. Linear strain gauges were also bonded on the labial marginal bone of each extraction socket. Strain measurements were performed at a sample rate of 10 kHz simultaneously monitored from a computer connected to data acquisition system and under a maximum load of 100 N on each implant with or without human coagulum in the extraction socket. Low-amplitude strains were measured around immediate implants. The increase in load increased strains on labial marginal cortical bone around implants (P < 0.05). Bone strains were higher on the implant loaded, when coagulum was present in the bone defects (P < 0.05). Strains within coagulum around mesiodistally anchored implants were higher than palatally anchored implants (P < 0.05). The type of implant on anchorage and presence of coagulum has an impact mechanotransduction to buccal marginal bone around immediate implants.</description><identifier>ISSN: 0140-0118</identifier><identifier>EISSN: 1741-0444</identifier><identifier>DOI: 10.1007/s11517-005-0010-4</identifier><identifier>PMID: 16929925</identifier><language>eng</language><publisher>United States: Springer Nature B.V</publisher><subject>Blood Coagulation ; Data acquisition systems ; Dental Implantation, Endosseous ; Dental Implants, Single-Tooth ; Dental Stress Analysis ; Humans ; Models, Biological ; Strain gauges ; Stress, Mechanical ; Tooth Extraction ; Tooth Socket - physiopathology ; Transplants & implants</subject><ispartof>Medical & biological engineering & computing, 2006-03, Vol.44 (1-2), p.86-94</ispartof><rights>International Federation for Medical and Biological Engineering 2006</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c389t-dc43f8fa7c3e954a56923212147c4c244b6df9834dfe59e314cb884ec55d18b03</citedby><cites>FETCH-LOGICAL-c389t-dc43f8fa7c3e954a56923212147c4c244b6df9834dfe59e314cb884ec55d18b03</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,776,780,27901,27902</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/16929925$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Cehreli, Murat Cavit</creatorcontrib><creatorcontrib>Comert, Ayhan</creatorcontrib><creatorcontrib>Akkocaoglu, Murat</creatorcontrib><creatorcontrib>Tekdemir, Ibrahim</creatorcontrib><creatorcontrib>Akca, Kivanc</creatorcontrib><title>Towards the limit of quantifying low-amplitude strains on bone and in coagulum around immediately loaded oral implants in extraction sockets</title><title>Medical & biological engineering & computing</title><addtitle>Med Biol Eng Comput</addtitle><description>The purpose of this study was to quantify strains in coagulum around immediately loaded oral implants in extraction sockets at the ex vivo level. Bilateral maxillary premolar teeth of two fresh human cadavers were extracted and psi 4.1 x 12 mm Straumann TE implants were placed in the sockets of first and second premolars by utilizing mesio-distal and palatal anchorage, respectively. Installation torque value (ITV) of each implant was measured by a custom-made torque wrench and resonance frequency analyses (RFAs) were undertaken to determine intraosseous stability. Upon abutment connection, a gold coping allowing the placement of a miniature load cell to contact the underlying solid abutment was fabricated. A linear strain gauge was connected to the coping at a distance for strain measurements in coagulum around the implant neck in the extraction socket. Linear strain gauges were also bonded on the labial marginal bone of each extraction socket. Strain measurements were performed at a sample rate of 10 kHz simultaneously monitored from a computer connected to data acquisition system and under a maximum load of 100 N on each implant with or without human coagulum in the extraction socket. Low-amplitude strains were measured around immediate implants. The increase in load increased strains on labial marginal cortical bone around implants (P < 0.05). Bone strains were higher on the implant loaded, when coagulum was present in the bone defects (P < 0.05). Strains within coagulum around mesiodistally anchored implants were higher than palatally anchored implants (P < 0.05). The type of implant on anchorage and presence of coagulum has an impact mechanotransduction to buccal marginal bone around immediate implants.</description><subject>Blood Coagulation</subject><subject>Data acquisition systems</subject><subject>Dental Implantation, Endosseous</subject><subject>Dental Implants, Single-Tooth</subject><subject>Dental Stress Analysis</subject><subject>Humans</subject><subject>Models, Biological</subject><subject>Strain gauges</subject><subject>Stress, Mechanical</subject><subject>Tooth Extraction</subject><subject>Tooth Socket - physiopathology</subject><subject>Transplants & implants</subject><issn>0140-0118</issn><issn>1741-0444</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2006</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><sourceid>BENPR</sourceid><recordid>eNqFkU1rFTEUhoMo9rb6A9xIcGFXU3MmySSzlGKrUHBT1yGTZGpqZnKbD9r7H_zRzeVeEFzoIgQOz_uQkxehd0AugBDxKQNwEB0hvB0gHXuBNiAYdIQx9hJtCDDSEQB5gk5zviekB96z1-gEhrEfx55v0O_b-KiTzbj8dDj4xRccZ_xQ9Vr8vPPrHQ7xsdPLNvhSrcO5JO3XjOOKp7g6rFeL_YpN1Hc11AXrFOt-tCzOel1c2DWBts7imHRo821o6rzPuKfmMsU3VY7mlyv5DXo165Dd2-N9hn5cfbm9_NrdfL_-dvn5pjNUjqWzhtFZzloY6kbONG_b0B56YMIw0zM2DXYeJWV2dnx0FJiZpGTOcG5BToSeofODd5viQ3W5qMVn40J7mos1K8EZHwQRvJEf_0kOUkhJgf4X7AmlQhJo4Ie_wPtY09rWVcMAlEsqhwbBATIp5pzcrLbJLzrtFBC1r14dqleterWvXrGWeX8U16l9_p_EsWv6DIzbqy4</recordid><startdate>20060301</startdate><enddate>20060301</enddate><creator>Cehreli, Murat Cavit</creator><creator>Comert, Ayhan</creator><creator>Akkocaoglu, Murat</creator><creator>Tekdemir, Ibrahim</creator><creator>Akca, Kivanc</creator><general>Springer Nature B.V</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>3V.</scope><scope>7RV</scope><scope>7SC</scope><scope>7TB</scope><scope>7TS</scope><scope>7WY</scope><scope>7WZ</scope><scope>7X7</scope><scope>7XB</scope><scope>87Z</scope><scope>88A</scope><scope>88E</scope><scope>88I</scope><scope>8AL</scope><scope>8AO</scope><scope>8FD</scope><scope>8FE</scope><scope>8FG</scope><scope>8FH</scope><scope>8FI</scope><scope>8FJ</scope><scope>8FK</scope><scope>8FL</scope><scope>ABUWG</scope><scope>AFKRA</scope><scope>ARAPS</scope><scope>AZQEC</scope><scope>BBNVY</scope><scope>BENPR</scope><scope>BEZIV</scope><scope>BGLVJ</scope><scope>BHPHI</scope><scope>CCPQU</scope><scope>DWQXO</scope><scope>FR3</scope><scope>FRNLG</scope><scope>FYUFA</scope><scope>F~G</scope><scope>GHDGH</scope><scope>GNUQQ</scope><scope>HCIFZ</scope><scope>JQ2</scope><scope>K60</scope><scope>K6~</scope><scope>K7-</scope><scope>K9.</scope><scope>KB0</scope><scope>L.-</scope><scope>L7M</scope><scope>LK8</scope><scope>L~C</scope><scope>L~D</scope><scope>M0C</scope><scope>M0N</scope><scope>M0S</scope><scope>M1P</scope><scope>M2P</scope><scope>M7P</scope><scope>M7Z</scope><scope>NAPCQ</scope><scope>P5Z</scope><scope>P62</scope><scope>P64</scope><scope>PQBIZ</scope><scope>PQBZA</scope><scope>PQEST</scope><scope>PQQKQ</scope><scope>PQUKI</scope><scope>PRINS</scope><scope>Q9U</scope><scope>7QO</scope><scope>7X8</scope><scope>7QP</scope></search><sort><creationdate>20060301</creationdate><title>Towards the limit of quantifying low-amplitude strains on bone and in coagulum around immediately loaded oral implants in extraction sockets</title><author>Cehreli, Murat Cavit ; Comert, Ayhan ; Akkocaoglu, Murat ; Tekdemir, Ibrahim ; Akca, Kivanc</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c389t-dc43f8fa7c3e954a56923212147c4c244b6df9834dfe59e314cb884ec55d18b03</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2006</creationdate><topic>Blood Coagulation</topic><topic>Data acquisition systems</topic><topic>Dental Implantation, Endosseous</topic><topic>Dental Implants, Single-Tooth</topic><topic>Dental Stress Analysis</topic><topic>Humans</topic><topic>Models, Biological</topic><topic>Strain gauges</topic><topic>Stress, Mechanical</topic><topic>Tooth Extraction</topic><topic>Tooth Socket - physiopathology</topic><topic>Transplants & implants</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Cehreli, Murat Cavit</creatorcontrib><creatorcontrib>Comert, Ayhan</creatorcontrib><creatorcontrib>Akkocaoglu, Murat</creatorcontrib><creatorcontrib>Tekdemir, Ibrahim</creatorcontrib><creatorcontrib>Akca, Kivanc</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>ProQuest Central (Corporate)</collection><collection>Nursing & Allied Health Database</collection><collection>Computer and Information Systems Abstracts</collection><collection>Mechanical & Transportation Engineering Abstracts</collection><collection>Physical Education Index</collection><collection>ABI/INFORM Collection</collection><collection>ABI/INFORM Global (PDF only)</collection><collection>Health & Medical Collection</collection><collection>ProQuest Central (purchase pre-March 2016)</collection><collection>ABI/INFORM Global (Alumni Edition)</collection><collection>Biology Database (Alumni Edition)</collection><collection>Medical Database (Alumni Edition)</collection><collection>Science Database (Alumni Edition)</collection><collection>Computing Database (Alumni Edition)</collection><collection>ProQuest Pharma Collection</collection><collection>Technology Research Database</collection><collection>ProQuest SciTech Collection</collection><collection>ProQuest Technology Collection</collection><collection>ProQuest Natural Science Collection</collection><collection>Hospital Premium Collection</collection><collection>Hospital Premium Collection (Alumni Edition)</collection><collection>ProQuest Central (Alumni) (purchase pre-March 2016)</collection><collection>ABI/INFORM Collection (Alumni Edition)</collection><collection>ProQuest Central (Alumni Edition)</collection><collection>ProQuest Central UK/Ireland</collection><collection>Advanced Technologies & Aerospace Collection</collection><collection>ProQuest Central Essentials</collection><collection>Biological Science Collection</collection><collection>ProQuest Central</collection><collection>Business Premium Collection</collection><collection>Technology Collection</collection><collection>Natural Science Collection</collection><collection>ProQuest One Community College</collection><collection>ProQuest Central Korea</collection><collection>Engineering Research Database</collection><collection>Business Premium Collection (Alumni)</collection><collection>Health Research Premium Collection</collection><collection>ABI/INFORM Global (Corporate)</collection><collection>Health Research Premium Collection (Alumni)</collection><collection>ProQuest Central Student</collection><collection>SciTech Premium Collection</collection><collection>ProQuest Computer Science Collection</collection><collection>ProQuest Business Collection (Alumni Edition)</collection><collection>ProQuest Business Collection</collection><collection>Computer Science Database</collection><collection>ProQuest Health & Medical Complete (Alumni)</collection><collection>Nursing & Allied Health Database (Alumni Edition)</collection><collection>ABI/INFORM Professional Advanced</collection><collection>Advanced Technologies Database with Aerospace</collection><collection>ProQuest Biological Science Collection</collection><collection>Computer and Information Systems Abstracts Academic</collection><collection>Computer and Information Systems Abstracts Professional</collection><collection>ABI/INFORM Global</collection><collection>Computing Database</collection><collection>Health & Medical Collection (Alumni Edition)</collection><collection>Medical Database</collection><collection>Science Database</collection><collection>Biological Science Database</collection><collection>Biochemistry Abstracts 1</collection><collection>Nursing & Allied Health Premium</collection><collection>Advanced Technologies & Aerospace Database</collection><collection>ProQuest Advanced Technologies & Aerospace Collection</collection><collection>Biotechnology and BioEngineering Abstracts</collection><collection>ProQuest One Business</collection><collection>ProQuest One Business (Alumni)</collection><collection>ProQuest One Academic Eastern Edition (DO NOT USE)</collection><collection>ProQuest One Academic</collection><collection>ProQuest One Academic UKI Edition</collection><collection>ProQuest Central China</collection><collection>ProQuest Central Basic</collection><collection>Biotechnology Research Abstracts</collection><collection>MEDLINE - Academic</collection><collection>Calcium & Calcified Tissue Abstracts</collection><jtitle>Medical & biological engineering & computing</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Cehreli, Murat Cavit</au><au>Comert, Ayhan</au><au>Akkocaoglu, Murat</au><au>Tekdemir, Ibrahim</au><au>Akca, Kivanc</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Towards the limit of quantifying low-amplitude strains on bone and in coagulum around immediately loaded oral implants in extraction sockets</atitle><jtitle>Medical & biological engineering & computing</jtitle><addtitle>Med Biol Eng Comput</addtitle><date>2006-03-01</date><risdate>2006</risdate><volume>44</volume><issue>1-2</issue><spage>86</spage><epage>94</epage><pages>86-94</pages><issn>0140-0118</issn><eissn>1741-0444</eissn><abstract>The purpose of this study was to quantify strains in coagulum around immediately loaded oral implants in extraction sockets at the ex vivo level. Bilateral maxillary premolar teeth of two fresh human cadavers were extracted and psi 4.1 x 12 mm Straumann TE implants were placed in the sockets of first and second premolars by utilizing mesio-distal and palatal anchorage, respectively. Installation torque value (ITV) of each implant was measured by a custom-made torque wrench and resonance frequency analyses (RFAs) were undertaken to determine intraosseous stability. Upon abutment connection, a gold coping allowing the placement of a miniature load cell to contact the underlying solid abutment was fabricated. A linear strain gauge was connected to the coping at a distance for strain measurements in coagulum around the implant neck in the extraction socket. Linear strain gauges were also bonded on the labial marginal bone of each extraction socket. Strain measurements were performed at a sample rate of 10 kHz simultaneously monitored from a computer connected to data acquisition system and under a maximum load of 100 N on each implant with or without human coagulum in the extraction socket. Low-amplitude strains were measured around immediate implants. The increase in load increased strains on labial marginal cortical bone around implants (P < 0.05). Bone strains were higher on the implant loaded, when coagulum was present in the bone defects (P < 0.05). Strains within coagulum around mesiodistally anchored implants were higher than palatally anchored implants (P < 0.05). The type of implant on anchorage and presence of coagulum has an impact mechanotransduction to buccal marginal bone around immediate implants.</abstract><cop>United States</cop><pub>Springer Nature B.V</pub><pmid>16929925</pmid><doi>10.1007/s11517-005-0010-4</doi><tpages>9</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0140-0118 |
ispartof | Medical & biological engineering & computing, 2006-03, Vol.44 (1-2), p.86-94 |
issn | 0140-0118 1741-0444 |
language | eng |
recordid | cdi_proquest_miscellaneous_754567075 |
source | MEDLINE; SpringerLink Journals; Business Source Complete |
subjects | Blood Coagulation Data acquisition systems Dental Implantation, Endosseous Dental Implants, Single-Tooth Dental Stress Analysis Humans Models, Biological Strain gauges Stress, Mechanical Tooth Extraction Tooth Socket - physiopathology Transplants & implants |
title | Towards the limit of quantifying low-amplitude strains on bone and in coagulum around immediately loaded oral implants in extraction sockets |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-02T07%3A27%3A10IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Towards%20the%20limit%20of%20quantifying%20low-amplitude%20strains%20on%20bone%20and%20in%20coagulum%20around%20immediately%20loaded%20oral%20implants%20in%20extraction%20sockets&rft.jtitle=Medical%20&%20biological%20engineering%20&%20computing&rft.au=Cehreli,%20Murat%20Cavit&rft.date=2006-03-01&rft.volume=44&rft.issue=1-2&rft.spage=86&rft.epage=94&rft.pages=86-94&rft.issn=0140-0118&rft.eissn=1741-0444&rft_id=info:doi/10.1007/s11517-005-0010-4&rft_dat=%3Cproquest_cross%3E754567075%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=661358386&rft_id=info:pmid/16929925&rfr_iscdi=true |