Air Entrapment as a Cause of Transient Cardiac Pacemaker Malfunction
Air within a pacemaker pocket may cause transient malfunction of a replacement unipolar cardiac pacemaker. This was noted in a patient in whom entrapped air prevented tissue contact of the anode, resulting in complete cessation of myocardlal stimulation and absence of an electrocardio‐graphic pacema...
Gespeichert in:
Veröffentlicht in: | Pacing and clinical electrophysiology 1979-11, Vol.2 (6), p.641-644 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Air within a pacemaker pocket may cause transient malfunction of a replacement unipolar cardiac pacemaker. This was noted in a patient in whom entrapped air prevented tissue contact of the anode, resulting in complete cessation of myocardlal stimulation and absence of an electrocardio‐graphic pacemaker artifact, Methods for avoiding this complication are proposed. |
---|---|
ISSN: | 0147-8389 1540-8159 |
DOI: | 10.1111/j.1540-8159.1979.tb04284.x |