Air Entrapment as a Cause of Transient Cardiac Pacemaker Malfunction

Air within a pacemaker pocket may cause transient malfunction of a replacement unipolar cardiac pacemaker. This was noted in a patient in whom entrapped air prevented tissue contact of the anode, resulting in complete cessation of myocardlal stimulation and absence of an electrocardio‐graphic pacema...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 1979-11, Vol.2 (6), p.641-644
Hauptverfasser: KREIS, DAVID J., LICALZI, LUKE, SHAW, RICHARD K.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Air within a pacemaker pocket may cause transient malfunction of a replacement unipolar cardiac pacemaker. This was noted in a patient in whom entrapped air prevented tissue contact of the anode, resulting in complete cessation of myocardlal stimulation and absence of an electrocardio‐graphic pacemaker artifact, Methods for avoiding this complication are proposed.
ISSN:0147-8389
1540-8159
DOI:10.1111/j.1540-8159.1979.tb04284.x