Evaluation of a simple device for regulating intravenous infusions

Dial-A-Flo, an infusion control device, is assessed for accuracy, in vitro and in vivo. The unit did not permit accurate control and particularly so at lower flow rates. The normal spontaneous variations in venous pressure that occur around an infusion cannula tip, further increase the error of grav...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Intensive care medicine 1979-03, Vol.5 (1), p.41-43
Hauptverfasser: Rithalia, S V, Rozkovec, A
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Dial-A-Flo, an infusion control device, is assessed for accuracy, in vitro and in vivo. The unit did not permit accurate control and particularly so at lower flow rates. The normal spontaneous variations in venous pressure that occur around an infusion cannula tip, further increase the error of gravity-fed infusions. Dial-A-Flo has no advantage over a standard administration set alone, and should not be used for controlled administration of cardiac drugs.
ISSN:0342-4642
1432-1238
DOI:10.1007/BF01739002