Improving the stability of cryogenic current comparator setups
We have realized an improved resistance calibration setup based on a cryogenic current comparator (CCC). The comparator, with 18 windings and 4647 turns in total, is well-suited for all the necessary comparisons with a 100 Omega standard resistor with the quantum Hall effect and inside the range of...
Gespeichert in:
Veröffentlicht in: | Superconductor science & technology 2009-11, Vol.22 (11), p.114004-114004 (8) |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We have realized an improved resistance calibration setup based on a cryogenic current comparator (CCC). The comparator, with 18 windings and 4647 turns in total, is well-suited for all the necessary comparisons with a 100 Omega standard resistor with the quantum Hall effect and inside the range of standard resistance values from 1 Omega to 1 MOmega. The new state-of-the-art setup is equipped with a low-noise dc SQUID (superconducting quantum interference device) and a digital double current source. Proper damping of the CCC resonance, careful electronics design ('box-in-box') and application of internal wideband feedback to the SQUID sensor improve the dynamic stability considerably and lower the overall measurement time as compared against our previous setup. |
---|---|
ISSN: | 0953-2048 1361-6668 |
DOI: | 10.1088/0953-2048/22/11/114004 |