The heat balance of small babies nursed in incubators and under radiant warmers

The heat balance of 12 healthy preterm babies (mean birth weight 1.58 kg, gestation 32 weeks, age 7 days) was studied first in an incubator and then under a radiant warmer during normal nursing. Heat production and heat loss by radiation, convection and evaporation were measured in presumed thermone...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Early human development 1982-04, Vol.6 (2), p.131-143
Hauptverfasser: Wheldon, A.E., Rutter, N.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 143
container_issue 2
container_start_page 131
container_title Early human development
container_volume 6
creator Wheldon, A.E.
Rutter, N.
description The heat balance of 12 healthy preterm babies (mean birth weight 1.58 kg, gestation 32 weeks, age 7 days) was studied first in an incubator and then under a radiant warmer during normal nursing. Heat production and heat loss by radiation, convection and evaporation were measured in presumed thermoneutral conditions. Although rectal and mean skin temperatures were normal and the same in both environments, there were important differences. Radiation was the major source of heat loss in the incubator and convective losses were low. Under the radiant warmer convection was the major source of heat loss and heat was gained by radiation. A small rise in metabolic heat production occurred under the radiant warmer. Respiratory water loss was low in both environments. Skin water loss was significantly higher under the radiant warmer. The most immature baby (gestation 28 weeks) could not be kept warm in the incubator despite high air temperature, because the evaporative heat loss from her skin was very high. Her body temperatures were normal under the radiant warmer. It is concluded that both devices provide acceptable thermal environments for most preterm babies but that incubators without humidification may be inadequate for immature babies with a high skin water loss.
doi_str_mv 10.1016/0378-3782(82)90100-1
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_74113817</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>0378378282901001</els_id><sourcerecordid>74113817</sourcerecordid><originalsourceid>FETCH-LOGICAL-c357t-8bbfa085b64b30751b1cfefb7b990bb3218d6feac141671546bca2e9398219353</originalsourceid><addsrcrecordid>eNp9kE1LAzEQhoMotVb_gUJOoofVzGZ3k70IUvyCQi_1HJLsLI3sR012Ff-9qS09CjMMzLzzDvMQcgnsDhgU94wLmcRMb2R6WzJgLIEjMgUp0qRIeXpMpgfJKTkL4YMxlsuSTchEsDKTOUzJcrVGukY9UKMb3VmkfU1Dq5smNozDQLvRB6yo62LY0eih94HqrqJjV6GnXldOdwP91r5FH87JSa2bgBf7OiPvz0-r-WuyWL68zR8XieW5GBJpTK2ZzE2RGc5EDgZsjbURpiyZMTwFWRU1agsZFALyrDBWp1jyUqZQ8pzPyPXOd-P7zxHDoFoXLDbxB-zHoEQGwCWIKMx2Quv7EDzWauNdq_2PAqa2HNUWktpCUjH-OCqIa1d7_9G0WB2W9uDi_GE3x_jkl0OvgnUY-VXOox1U1bv_D_wCBqWBQQ</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>74113817</pqid></control><display><type>article</type><title>The heat balance of small babies nursed in incubators and under radiant warmers</title><source>MEDLINE</source><source>Access via ScienceDirect (Elsevier)</source><creator>Wheldon, A.E. ; Rutter, N.</creator><creatorcontrib>Wheldon, A.E. ; Rutter, N.</creatorcontrib><description>The heat balance of 12 healthy preterm babies (mean birth weight 1.58 kg, gestation 32 weeks, age 7 days) was studied first in an incubator and then under a radiant warmer during normal nursing. Heat production and heat loss by radiation, convection and evaporation were measured in presumed thermoneutral conditions. Although rectal and mean skin temperatures were normal and the same in both environments, there were important differences. Radiation was the major source of heat loss in the incubator and convective losses were low. Under the radiant warmer convection was the major source of heat loss and heat was gained by radiation. A small rise in metabolic heat production occurred under the radiant warmer. Respiratory water loss was low in both environments. Skin water loss was significantly higher under the radiant warmer. The most immature baby (gestation 28 weeks) could not be kept warm in the incubator despite high air temperature, because the evaporative heat loss from her skin was very high. Her body temperatures were normal under the radiant warmer. It is concluded that both devices provide acceptable thermal environments for most preterm babies but that incubators without humidification may be inadequate for immature babies with a high skin water loss.</description><identifier>ISSN: 0378-3782</identifier><identifier>EISSN: 1872-6232</identifier><identifier>DOI: 10.1016/0378-3782(82)90100-1</identifier><identifier>PMID: 7094851</identifier><language>eng</language><publisher>Ireland: Elsevier Ireland Ltd</publisher><subject>Body Temperature ; Body Temperature Regulation ; Body Water - metabolism ; Female ; heat balance ; Heating ; Humans ; Humidity ; incubator ; Incubators, Infant ; Infant, Newborn ; Infant, Premature ; Male ; premature baby ; radiant warmer ; Skin Temperature ; skin water loss ; thermal environment ; Water Loss, Insensible</subject><ispartof>Early human development, 1982-04, Vol.6 (2), p.131-143</ispartof><rights>1982</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c357t-8bbfa085b64b30751b1cfefb7b990bb3218d6feac141671546bca2e9398219353</citedby><cites>FETCH-LOGICAL-c357t-8bbfa085b64b30751b1cfefb7b990bb3218d6feac141671546bca2e9398219353</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://dx.doi.org/10.1016/0378-3782(82)90100-1$$EHTML$$P50$$Gelsevier$$H</linktohtml><link.rule.ids>314,780,784,3550,27924,27925,45995</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/7094851$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Wheldon, A.E.</creatorcontrib><creatorcontrib>Rutter, N.</creatorcontrib><title>The heat balance of small babies nursed in incubators and under radiant warmers</title><title>Early human development</title><addtitle>Early Hum Dev</addtitle><description>The heat balance of 12 healthy preterm babies (mean birth weight 1.58 kg, gestation 32 weeks, age 7 days) was studied first in an incubator and then under a radiant warmer during normal nursing. Heat production and heat loss by radiation, convection and evaporation were measured in presumed thermoneutral conditions. Although rectal and mean skin temperatures were normal and the same in both environments, there were important differences. Radiation was the major source of heat loss in the incubator and convective losses were low. Under the radiant warmer convection was the major source of heat loss and heat was gained by radiation. A small rise in metabolic heat production occurred under the radiant warmer. Respiratory water loss was low in both environments. Skin water loss was significantly higher under the radiant warmer. The most immature baby (gestation 28 weeks) could not be kept warm in the incubator despite high air temperature, because the evaporative heat loss from her skin was very high. Her body temperatures were normal under the radiant warmer. It is concluded that both devices provide acceptable thermal environments for most preterm babies but that incubators without humidification may be inadequate for immature babies with a high skin water loss.</description><subject>Body Temperature</subject><subject>Body Temperature Regulation</subject><subject>Body Water - metabolism</subject><subject>Female</subject><subject>heat balance</subject><subject>Heating</subject><subject>Humans</subject><subject>Humidity</subject><subject>incubator</subject><subject>Incubators, Infant</subject><subject>Infant, Newborn</subject><subject>Infant, Premature</subject><subject>Male</subject><subject>premature baby</subject><subject>radiant warmer</subject><subject>Skin Temperature</subject><subject>skin water loss</subject><subject>thermal environment</subject><subject>Water Loss, Insensible</subject><issn>0378-3782</issn><issn>1872-6232</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>1982</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp9kE1LAzEQhoMotVb_gUJOoofVzGZ3k70IUvyCQi_1HJLsLI3sR012Ff-9qS09CjMMzLzzDvMQcgnsDhgU94wLmcRMb2R6WzJgLIEjMgUp0qRIeXpMpgfJKTkL4YMxlsuSTchEsDKTOUzJcrVGukY9UKMb3VmkfU1Dq5smNozDQLvRB6yo62LY0eih94HqrqJjV6GnXldOdwP91r5FH87JSa2bgBf7OiPvz0-r-WuyWL68zR8XieW5GBJpTK2ZzE2RGc5EDgZsjbURpiyZMTwFWRU1agsZFALyrDBWp1jyUqZQ8pzPyPXOd-P7zxHDoFoXLDbxB-zHoEQGwCWIKMx2Quv7EDzWauNdq_2PAqa2HNUWktpCUjH-OCqIa1d7_9G0WB2W9uDi_GE3x_jkl0OvgnUY-VXOox1U1bv_D_wCBqWBQQ</recordid><startdate>198204</startdate><enddate>198204</enddate><creator>Wheldon, A.E.</creator><creator>Rutter, N.</creator><general>Elsevier Ireland Ltd</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>198204</creationdate><title>The heat balance of small babies nursed in incubators and under radiant warmers</title><author>Wheldon, A.E. ; Rutter, N.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c357t-8bbfa085b64b30751b1cfefb7b990bb3218d6feac141671546bca2e9398219353</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>1982</creationdate><topic>Body Temperature</topic><topic>Body Temperature Regulation</topic><topic>Body Water - metabolism</topic><topic>Female</topic><topic>heat balance</topic><topic>Heating</topic><topic>Humans</topic><topic>Humidity</topic><topic>incubator</topic><topic>Incubators, Infant</topic><topic>Infant, Newborn</topic><topic>Infant, Premature</topic><topic>Male</topic><topic>premature baby</topic><topic>radiant warmer</topic><topic>Skin Temperature</topic><topic>skin water loss</topic><topic>thermal environment</topic><topic>Water Loss, Insensible</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Wheldon, A.E.</creatorcontrib><creatorcontrib>Rutter, N.</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Early human development</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Wheldon, A.E.</au><au>Rutter, N.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>The heat balance of small babies nursed in incubators and under radiant warmers</atitle><jtitle>Early human development</jtitle><addtitle>Early Hum Dev</addtitle><date>1982-04</date><risdate>1982</risdate><volume>6</volume><issue>2</issue><spage>131</spage><epage>143</epage><pages>131-143</pages><issn>0378-3782</issn><eissn>1872-6232</eissn><abstract>The heat balance of 12 healthy preterm babies (mean birth weight 1.58 kg, gestation 32 weeks, age 7 days) was studied first in an incubator and then under a radiant warmer during normal nursing. Heat production and heat loss by radiation, convection and evaporation were measured in presumed thermoneutral conditions. Although rectal and mean skin temperatures were normal and the same in both environments, there were important differences. Radiation was the major source of heat loss in the incubator and convective losses were low. Under the radiant warmer convection was the major source of heat loss and heat was gained by radiation. A small rise in metabolic heat production occurred under the radiant warmer. Respiratory water loss was low in both environments. Skin water loss was significantly higher under the radiant warmer. The most immature baby (gestation 28 weeks) could not be kept warm in the incubator despite high air temperature, because the evaporative heat loss from her skin was very high. Her body temperatures were normal under the radiant warmer. It is concluded that both devices provide acceptable thermal environments for most preterm babies but that incubators without humidification may be inadequate for immature babies with a high skin water loss.</abstract><cop>Ireland</cop><pub>Elsevier Ireland Ltd</pub><pmid>7094851</pmid><doi>10.1016/0378-3782(82)90100-1</doi><tpages>13</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0378-3782
ispartof Early human development, 1982-04, Vol.6 (2), p.131-143
issn 0378-3782
1872-6232
language eng
recordid cdi_proquest_miscellaneous_74113817
source MEDLINE; Access via ScienceDirect (Elsevier)
subjects Body Temperature
Body Temperature Regulation
Body Water - metabolism
Female
heat balance
Heating
Humans
Humidity
incubator
Incubators, Infant
Infant, Newborn
Infant, Premature
Male
premature baby
radiant warmer
Skin Temperature
skin water loss
thermal environment
Water Loss, Insensible
title The heat balance of small babies nursed in incubators and under radiant warmers
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-29T18%3A02%3A52IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=The%20heat%20balance%20of%20small%20babies%20nursed%20in%20incubators%20and%20under%20radiant%20warmers&rft.jtitle=Early%20human%20development&rft.au=Wheldon,%20A.E.&rft.date=1982-04&rft.volume=6&rft.issue=2&rft.spage=131&rft.epage=143&rft.pages=131-143&rft.issn=0378-3782&rft.eissn=1872-6232&rft_id=info:doi/10.1016/0378-3782(82)90100-1&rft_dat=%3Cproquest_cross%3E74113817%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=74113817&rft_id=info:pmid/7094851&rft_els_id=0378378282901001&rfr_iscdi=true