Visualising velocity profiles in great arteries—A preliminary report
Using a constant temperature anemometer and an L-shaped 1 mm conical hot-film anemometer probe, blood velocities were registered in a patient after implantation of a Starr-Edwards ball valve in the aortic position. Velocities were registered at 41 predetermined points in the ascending aorta, equally...
Gespeichert in:
Veröffentlicht in: | Journal of biomechanics 1982, Vol.15 (1), p.61-63 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Using a constant temperature anemometer and an L-shaped 1 mm conical hot-film anemometer probe, blood velocities were registered in a patient after implantation of a Starr-Edwards ball valve in the aortic position. Velocities were registered at 41 predetermined points in the ascending aorta, equally distributed over the cross-sectional area. Data were recorded on tape and later analysed to make three-dimensional plots of the velocity profile at 200 time intervals during one “mean” heart cycle, and a film of the velocity profile in this cycle was made. The film gave the impression of great velocity fluctuations at different places in the cross-sectional area. |
---|---|
ISSN: | 0021-9290 1873-2380 |
DOI: | 10.1016/0021-9290(82)90035-5 |