Report on the ECR2003 (European Congress of Radiology): comparison of apparent diffusion coefficient (ADC) between different MRI scanners
To analyze reproducibility of ADC measurements on different MRI scanners. Diffusion-weighted imaging (DWI) of a home-made phantom (gelatine of different concentration was enclosed in case of circle pillar shape) was performed with the following protocol; spin-echo type echo planar imaging; TR=7000 m...
Gespeichert in:
Veröffentlicht in: | Nippon Hōshasen Gijutsu Gakkai zasshi 2003-07, Vol.59 (7), p.825-826 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | jpn |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 826 |
---|---|
container_issue | 7 |
container_start_page | 825 |
container_title | Nippon Hōshasen Gijutsu Gakkai zasshi |
container_volume | 59 |
creator | Koizumi, Koji Masuda, K Komizu, M Ikemoto, Y Yoshimura, M Iguchi, H Hiraga, A Kobayashi, A |
description | To analyze reproducibility of ADC measurements on different MRI scanners.
Diffusion-weighted imaging (DWI) of a home-made phantom (gelatine of different concentration was enclosed in case of circle pillar shape) was performed with the following protocol; spin-echo type echo planar imaging; TR=7000 msec, TE=minimal, matrix=128 x 128, FOV=220 mm, slice thickness=5mm, NEX=1, b-factor=0, 500, and 1000 sec/mm(2) on four different 1.5T MR scanners [GE Signa Horizon (two of them), GE Signa Horizon Lx, and Siemens Magnetom Symphony]. Image analysis were performed on a workstation (GE Medical Systems) or on an operators console of Symphony (for data of Symphony). We used two-point method. Regions of interest (ROIs) were set on every part of ADC map of home-made phantom and ADCs were measured. A linear approximation graphs were also made with every MRI scanner and b-factor.
A linear approximation was performed, formulas were y=-0.08x+2.32, r(2)=0.84(Lx), y=-0.09x+2.58, r(2)=0.89(Horizon-1) y=-0.11x+2.67, r(2)=0.95(Horizon-2), y=-0.08x+2.42, r(2)=0.97(Symphony) at b-factor of 500sec/mm(2). y=-0.08x+2.33, r(2)=0.91(Lx), y=-0.09x+2.43, r(2)=0.93(Horizon-1), y=-0.08x +2.48, r(2)=0.92(Horizon-2), y=-0.07x+2.39, r(2)=0.96(Symphony) at b-factor of 1000 sec/mm(2).
ADC values measured at two MRI scanners (same model and older than the other scanners) are higher than other scanners on b-factor=500 sec/mm(2), but at b-factor=1000 sec/mm(2), a good relationship between every scanners was obtained. This result seems to depend on generation of the MRI scanners; newly scanners have good relations b-factor 500 and 1000 sec/mm(2). Pulse sequence design may be improved and optimized in newly scanners. When ADCs were used as quantitative value, a proper b-factor must be used. |
doi_str_mv | 10.6009/jjrt.KJ00003174210 |
format | Article |
fullrecord | <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_proquest_miscellaneous_73570535</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>73570535</sourcerecordid><originalsourceid>FETCH-LOGICAL-p122t-dabc3759f15c3919c13fc09960be5303cfe687425f4b58e416f57893ba712943</originalsourceid><addsrcrecordid>eNo1kM9OAjEQxnvQCEFewIPpycABbLfbXeqNrPgXY0K4k26ZYsluu7a7MTyCb21FnMtkvu83XzKD0BUl04wQcbvf-3b6-kJiMZqnCSVnqE9YJiYpI7yHhiGYMpoiSiS9QD2aCJanJOmj7xU0zrfYWdx-AF4UqySG4NGi864BaXHh7M5DCNhpvJJb4yq3O4zvsHJ1I70JcTE6sokD2BZvjdZdMFFVDrQ2yvyqo_l9McYltF8A9sjAkX5bPeOgpLXgwyU617IKMDz1AVo_LNbF02T5_vhczJeThiZJO9nKUrGcC025YoIKRZlWRIiMlMAZYUpDNos_4Dot-QxSmmmezwQrZR6vTtkA3fzFNt59dhDaTW2CgqqSFlwXNjnjOeGMR_D6BHZlDdtN400t_WHz_zv2Az9HcIA</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>73570535</pqid></control><display><type>article</type><title>Report on the ECR2003 (European Congress of Radiology): comparison of apparent diffusion coefficient (ADC) between different MRI scanners</title><source>MEDLINE</source><source>Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals</source><creator>Koizumi, Koji ; Masuda, K ; Komizu, M ; Ikemoto, Y ; Yoshimura, M ; Iguchi, H ; Hiraga, A ; Kobayashi, A</creator><creatorcontrib>Koizumi, Koji ; Masuda, K ; Komizu, M ; Ikemoto, Y ; Yoshimura, M ; Iguchi, H ; Hiraga, A ; Kobayashi, A</creatorcontrib><description>To analyze reproducibility of ADC measurements on different MRI scanners.
Diffusion-weighted imaging (DWI) of a home-made phantom (gelatine of different concentration was enclosed in case of circle pillar shape) was performed with the following protocol; spin-echo type echo planar imaging; TR=7000 msec, TE=minimal, matrix=128 x 128, FOV=220 mm, slice thickness=5mm, NEX=1, b-factor=0, 500, and 1000 sec/mm(2) on four different 1.5T MR scanners [GE Signa Horizon (two of them), GE Signa Horizon Lx, and Siemens Magnetom Symphony]. Image analysis were performed on a workstation (GE Medical Systems) or on an operators console of Symphony (for data of Symphony). We used two-point method. Regions of interest (ROIs) were set on every part of ADC map of home-made phantom and ADCs were measured. A linear approximation graphs were also made with every MRI scanner and b-factor.
A linear approximation was performed, formulas were y=-0.08x+2.32, r(2)=0.84(Lx), y=-0.09x+2.58, r(2)=0.89(Horizon-1) y=-0.11x+2.67, r(2)=0.95(Horizon-2), y=-0.08x+2.42, r(2)=0.97(Symphony) at b-factor of 500sec/mm(2). y=-0.08x+2.33, r(2)=0.91(Lx), y=-0.09x+2.43, r(2)=0.93(Horizon-1), y=-0.08x +2.48, r(2)=0.92(Horizon-2), y=-0.07x+2.39, r(2)=0.96(Symphony) at b-factor of 1000 sec/mm(2).
ADC values measured at two MRI scanners (same model and older than the other scanners) are higher than other scanners on b-factor=500 sec/mm(2), but at b-factor=1000 sec/mm(2), a good relationship between every scanners was obtained. This result seems to depend on generation of the MRI scanners; newly scanners have good relations b-factor 500 and 1000 sec/mm(2). Pulse sequence design may be improved and optimized in newly scanners. When ADCs were used as quantitative value, a proper b-factor must be used.</description><identifier>ISSN: 0369-4305</identifier><identifier>DOI: 10.6009/jjrt.KJ00003174210</identifier><identifier>PMID: 12937402</identifier><language>jpn</language><publisher>Japan</publisher><subject>Diffusion Magnetic Resonance Imaging - instrumentation ; Humans ; Phantoms, Imaging</subject><ispartof>Nippon Hōshasen Gijutsu Gakkai zasshi, 2003-07, Vol.59 (7), p.825-826</ispartof><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>309,310,314,776,780,785,786,23909,23910,25118,27901,27902</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/12937402$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Koizumi, Koji</creatorcontrib><creatorcontrib>Masuda, K</creatorcontrib><creatorcontrib>Komizu, M</creatorcontrib><creatorcontrib>Ikemoto, Y</creatorcontrib><creatorcontrib>Yoshimura, M</creatorcontrib><creatorcontrib>Iguchi, H</creatorcontrib><creatorcontrib>Hiraga, A</creatorcontrib><creatorcontrib>Kobayashi, A</creatorcontrib><title>Report on the ECR2003 (European Congress of Radiology): comparison of apparent diffusion coefficient (ADC) between different MRI scanners</title><title>Nippon Hōshasen Gijutsu Gakkai zasshi</title><addtitle>Nihon Hoshasen Gijutsu Gakkai Zasshi</addtitle><description>To analyze reproducibility of ADC measurements on different MRI scanners.
Diffusion-weighted imaging (DWI) of a home-made phantom (gelatine of different concentration was enclosed in case of circle pillar shape) was performed with the following protocol; spin-echo type echo planar imaging; TR=7000 msec, TE=minimal, matrix=128 x 128, FOV=220 mm, slice thickness=5mm, NEX=1, b-factor=0, 500, and 1000 sec/mm(2) on four different 1.5T MR scanners [GE Signa Horizon (two of them), GE Signa Horizon Lx, and Siemens Magnetom Symphony]. Image analysis were performed on a workstation (GE Medical Systems) or on an operators console of Symphony (for data of Symphony). We used two-point method. Regions of interest (ROIs) were set on every part of ADC map of home-made phantom and ADCs were measured. A linear approximation graphs were also made with every MRI scanner and b-factor.
A linear approximation was performed, formulas were y=-0.08x+2.32, r(2)=0.84(Lx), y=-0.09x+2.58, r(2)=0.89(Horizon-1) y=-0.11x+2.67, r(2)=0.95(Horizon-2), y=-0.08x+2.42, r(2)=0.97(Symphony) at b-factor of 500sec/mm(2). y=-0.08x+2.33, r(2)=0.91(Lx), y=-0.09x+2.43, r(2)=0.93(Horizon-1), y=-0.08x +2.48, r(2)=0.92(Horizon-2), y=-0.07x+2.39, r(2)=0.96(Symphony) at b-factor of 1000 sec/mm(2).
ADC values measured at two MRI scanners (same model and older than the other scanners) are higher than other scanners on b-factor=500 sec/mm(2), but at b-factor=1000 sec/mm(2), a good relationship between every scanners was obtained. This result seems to depend on generation of the MRI scanners; newly scanners have good relations b-factor 500 and 1000 sec/mm(2). Pulse sequence design may be improved and optimized in newly scanners. When ADCs were used as quantitative value, a proper b-factor must be used.</description><subject>Diffusion Magnetic Resonance Imaging - instrumentation</subject><subject>Humans</subject><subject>Phantoms, Imaging</subject><issn>0369-4305</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2003</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNo1kM9OAjEQxnvQCEFewIPpycABbLfbXeqNrPgXY0K4k26ZYsluu7a7MTyCb21FnMtkvu83XzKD0BUl04wQcbvf-3b6-kJiMZqnCSVnqE9YJiYpI7yHhiGYMpoiSiS9QD2aCJanJOmj7xU0zrfYWdx-AF4UqySG4NGi864BaXHh7M5DCNhpvJJb4yq3O4zvsHJ1I70JcTE6sokD2BZvjdZdMFFVDrQ2yvyqo_l9McYltF8A9sjAkX5bPeOgpLXgwyU617IKMDz1AVo_LNbF02T5_vhczJeThiZJO9nKUrGcC025YoIKRZlWRIiMlMAZYUpDNos_4Dot-QxSmmmezwQrZR6vTtkA3fzFNt59dhDaTW2CgqqSFlwXNjnjOeGMR_D6BHZlDdtN400t_WHz_zv2Az9HcIA</recordid><startdate>200307</startdate><enddate>200307</enddate><creator>Koizumi, Koji</creator><creator>Masuda, K</creator><creator>Komizu, M</creator><creator>Ikemoto, Y</creator><creator>Yoshimura, M</creator><creator>Iguchi, H</creator><creator>Hiraga, A</creator><creator>Kobayashi, A</creator><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>7X8</scope></search><sort><creationdate>200307</creationdate><title>Report on the ECR2003 (European Congress of Radiology): comparison of apparent diffusion coefficient (ADC) between different MRI scanners</title><author>Koizumi, Koji ; Masuda, K ; Komizu, M ; Ikemoto, Y ; Yoshimura, M ; Iguchi, H ; Hiraga, A ; Kobayashi, A</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-p122t-dabc3759f15c3919c13fc09960be5303cfe687425f4b58e416f57893ba712943</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>jpn</language><creationdate>2003</creationdate><topic>Diffusion Magnetic Resonance Imaging - instrumentation</topic><topic>Humans</topic><topic>Phantoms, Imaging</topic><toplevel>online_resources</toplevel><creatorcontrib>Koizumi, Koji</creatorcontrib><creatorcontrib>Masuda, K</creatorcontrib><creatorcontrib>Komizu, M</creatorcontrib><creatorcontrib>Ikemoto, Y</creatorcontrib><creatorcontrib>Yoshimura, M</creatorcontrib><creatorcontrib>Iguchi, H</creatorcontrib><creatorcontrib>Hiraga, A</creatorcontrib><creatorcontrib>Kobayashi, A</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>MEDLINE - Academic</collection><jtitle>Nippon Hōshasen Gijutsu Gakkai zasshi</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Koizumi, Koji</au><au>Masuda, K</au><au>Komizu, M</au><au>Ikemoto, Y</au><au>Yoshimura, M</au><au>Iguchi, H</au><au>Hiraga, A</au><au>Kobayashi, A</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Report on the ECR2003 (European Congress of Radiology): comparison of apparent diffusion coefficient (ADC) between different MRI scanners</atitle><jtitle>Nippon Hōshasen Gijutsu Gakkai zasshi</jtitle><addtitle>Nihon Hoshasen Gijutsu Gakkai Zasshi</addtitle><date>2003-07</date><risdate>2003</risdate><volume>59</volume><issue>7</issue><spage>825</spage><epage>826</epage><pages>825-826</pages><issn>0369-4305</issn><abstract>To analyze reproducibility of ADC measurements on different MRI scanners.
Diffusion-weighted imaging (DWI) of a home-made phantom (gelatine of different concentration was enclosed in case of circle pillar shape) was performed with the following protocol; spin-echo type echo planar imaging; TR=7000 msec, TE=minimal, matrix=128 x 128, FOV=220 mm, slice thickness=5mm, NEX=1, b-factor=0, 500, and 1000 sec/mm(2) on four different 1.5T MR scanners [GE Signa Horizon (two of them), GE Signa Horizon Lx, and Siemens Magnetom Symphony]. Image analysis were performed on a workstation (GE Medical Systems) or on an operators console of Symphony (for data of Symphony). We used two-point method. Regions of interest (ROIs) were set on every part of ADC map of home-made phantom and ADCs were measured. A linear approximation graphs were also made with every MRI scanner and b-factor.
A linear approximation was performed, formulas were y=-0.08x+2.32, r(2)=0.84(Lx), y=-0.09x+2.58, r(2)=0.89(Horizon-1) y=-0.11x+2.67, r(2)=0.95(Horizon-2), y=-0.08x+2.42, r(2)=0.97(Symphony) at b-factor of 500sec/mm(2). y=-0.08x+2.33, r(2)=0.91(Lx), y=-0.09x+2.43, r(2)=0.93(Horizon-1), y=-0.08x +2.48, r(2)=0.92(Horizon-2), y=-0.07x+2.39, r(2)=0.96(Symphony) at b-factor of 1000 sec/mm(2).
ADC values measured at two MRI scanners (same model and older than the other scanners) are higher than other scanners on b-factor=500 sec/mm(2), but at b-factor=1000 sec/mm(2), a good relationship between every scanners was obtained. This result seems to depend on generation of the MRI scanners; newly scanners have good relations b-factor 500 and 1000 sec/mm(2). Pulse sequence design may be improved and optimized in newly scanners. When ADCs were used as quantitative value, a proper b-factor must be used.</abstract><cop>Japan</cop><pmid>12937402</pmid><doi>10.6009/jjrt.KJ00003174210</doi><tpages>2</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0369-4305 |
ispartof | Nippon Hōshasen Gijutsu Gakkai zasshi, 2003-07, Vol.59 (7), p.825-826 |
issn | 0369-4305 |
language | jpn |
recordid | cdi_proquest_miscellaneous_73570535 |
source | MEDLINE; Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals |
subjects | Diffusion Magnetic Resonance Imaging - instrumentation Humans Phantoms, Imaging |
title | Report on the ECR2003 (European Congress of Radiology): comparison of apparent diffusion coefficient (ADC) between different MRI scanners |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-31T13%3A08%3A43IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Report%20on%20the%20ECR2003%20(European%20Congress%20of%20Radiology):%20comparison%20of%20apparent%20diffusion%20coefficient%20(ADC)%20between%20different%20MRI%20scanners&rft.jtitle=Nippon%20Ho%CC%84shasen%20Gijutsu%20Gakkai%20zasshi&rft.au=Koizumi,%20Koji&rft.date=2003-07&rft.volume=59&rft.issue=7&rft.spage=825&rft.epage=826&rft.pages=825-826&rft.issn=0369-4305&rft_id=info:doi/10.6009/jjrt.KJ00003174210&rft_dat=%3Cproquest_pubme%3E73570535%3C/proquest_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=73570535&rft_id=info:pmid/12937402&rfr_iscdi=true |