Three-dimensional simulation of near-infrared diffusion in tissue: boundary condition and geometry analysis for finite-element image reconstruction

Imaging of tissue with near-infrared spectral tomography is emerging as a practicable method to map hemoglobin concentrations within tissue. However, the accurate recovery of images by using modeling methods requires a good match between experiments and the model prediction of light transport in tis...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Applied Optics 2001-02, Vol.40 (4), p.588-600
Hauptverfasser: Pogue, B W, Geimer, S, McBride, T O, Jiang, S, Osterberg, U L, Paulsen, K D
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Imaging of tissue with near-infrared spectral tomography is emerging as a practicable method to map hemoglobin concentrations within tissue. However, the accurate recovery of images by using modeling methods requires a good match between experiments and the model prediction of light transport in tissue. We illustrate the potential for a match between (i) three-dimensional (3-D) frequency-domain diffusion theory, (ii) two-dimensional diffusion theory, (iii) Monte Carlo simulations, and (iv) experimental data from tissue-simulating phantoms. Robin-type boundary conditions are imposed in the 3-D model, which can be implemented with a scalar coupling coefficient relating the flux through the surface to the diffuse fluence rate at the same location. A comparison of 3-D mesh geometries for breast imaging indicates that relative measurements are sufficiently similar when calculated on either cylindrical or female breast shapes, suggesting that accurate reconstruction may be achieved with the simpler cylindrical mesh. Simulation studies directly assess the effects from objects extending out of the image plane, with results suggesting that spherically shaped objects reconstruct at lower contrast when their diameters are less than 15-20 mm. The algorithm presented here illustrates that a 3-D forward diffusion model can be used with circular tomographic measurements to reconstruct two-dimensional images of the interior absorption coefficient.
ISSN:1559-128X
0003-6935
1539-4522
DOI:10.1364/ao.40.000588