Evaluation of two automated methods for blood lactate levels
We evaluated two automated methods for measuring blood levels of lactates with Integra 800 (Roche Diagnostics) and ABL 725 (Radiometer). Our evaluation had shown a within run imprecision of 1% (Intégra 800) and 2% (ABL 725) and a between-assay imprecision of 2% (Intégra 800) and 5% (ABL 725). The tw...
Gespeichert in:
Veröffentlicht in: | Annales de biologie clinique (Paris) 2009-09, Vol.67 (5), p.573-576 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | fre |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We evaluated two automated methods for measuring blood levels of lactates with Integra 800 (Roche Diagnostics) and ABL 725 (Radiometer). Our evaluation had shown a within run imprecision of 1% (Intégra 800) and 2% (ABL 725) and a between-assay imprecision of 2% (Intégra 800) and 5% (ABL 725). The two methodologies appeared very well associated. The selective electrode remains expensive but it is very interesting because it saves a blood sample. |
---|---|