Surface oxidized zirconium total hip arthroplasty head damage due to closed reduction effects on polyethylene wear

Recent case studies of surface oxidized zirconium THA heads removed after attempted, closed reduction have shown significant surface damage that has been suggested as potentially deleterious to polyethylene wear. We obtained 4 clinically retrieved specimens, produced well-characterized surface damag...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of arthroplasty 2009-09, Vol.24 (6), p.898-902
Hauptverfasser: Jaffe, William L, Strauss, Eric J, Cardinale, M, Herrera, Lizeth, Kummer, Fred J
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 902
container_issue 6
container_start_page 898
container_title The Journal of arthroplasty
container_volume 24
creator Jaffe, William L
Strauss, Eric J
Cardinale, M
Herrera, Lizeth
Kummer, Fred J
description Recent case studies of surface oxidized zirconium THA heads removed after attempted, closed reduction have shown significant surface damage that has been suggested as potentially deleterious to polyethylene wear. We obtained 4 clinically retrieved specimens, produced well-characterized surface damage on additional heads, and tested them on a hip simulator. After 1 million cycles, the amount of polyethylene wear was related to the extent of surface damage, the most damaged clinical specimen showing more than 50 times more wear than a new head. Although all heads after failed attempted closed reduction(s) should be replaced, surface oxidized zirconium heads are of particular concern; those patients with a successful, simple closed reduction should be monitored for excessive wear.
doi_str_mv 10.1016/j.arth.2008.07.009
format Article
fullrecord <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_proquest_miscellaneous_733971794</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>34927569</sourcerecordid><originalsourceid>FETCH-LOGICAL-p156t-a813296674e1c475eaad82a46e3df8240892490a6f981511f234eebee90cd3fd3</originalsourceid><addsrcrecordid>eNp90D1PwzAQBmALgWgp_AEG5AmmBH8ltkeE-JIqMQBz5MYXkiqJg-0I0l9PEGVlund47l7pEDqnJKWE5tfb1PhYp4wQlRKZEqIP0JJmnCVKkPwQLYlSPMkE4Qt0EsKWEEqzTByjBVVKKMHkEvmX0VemBOy-GtvswOJd40vXN2OHo4umxXUz4J8e74bWhDjhGozF1nTmHbAdYWa4bF2YVz3YsYyN6zFUFZQx4DkOrp0g1lMLPeBPMP4UHVWmDXC2nyv0dn_3evuYrJ8fnm5v1slAszwmRlHOdJ5LAbQUMgNjrGJG5MBtpZggSjOhickrrWhGacW4ANgAaFJaXlm-Qle_dwfvPkYIseiaUELbmh7cGArJuZZUajHLy38lF5rJLNczvNjDcdOBLQbfdMZPxd8_-TdhkXqm</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>34927569</pqid></control><display><type>article</type><title>Surface oxidized zirconium total hip arthroplasty head damage due to closed reduction effects on polyethylene wear</title><source>MEDLINE</source><source>Elsevier ScienceDirect Journals Complete</source><creator>Jaffe, William L ; Strauss, Eric J ; Cardinale, M ; Herrera, Lizeth ; Kummer, Fred J</creator><creatorcontrib>Jaffe, William L ; Strauss, Eric J ; Cardinale, M ; Herrera, Lizeth ; Kummer, Fred J</creatorcontrib><description>Recent case studies of surface oxidized zirconium THA heads removed after attempted, closed reduction have shown significant surface damage that has been suggested as potentially deleterious to polyethylene wear. We obtained 4 clinically retrieved specimens, produced well-characterized surface damage on additional heads, and tested them on a hip simulator. After 1 million cycles, the amount of polyethylene wear was related to the extent of surface damage, the most damaged clinical specimen showing more than 50 times more wear than a new head. Although all heads after failed attempted closed reduction(s) should be replaced, surface oxidized zirconium heads are of particular concern; those patients with a successful, simple closed reduction should be monitored for excessive wear.</description><identifier>ISSN: 0883-5403</identifier><identifier>EISSN: 1532-8406</identifier><identifier>DOI: 10.1016/j.arth.2008.07.009</identifier><identifier>PMID: 18848427</identifier><language>eng</language><publisher>United States</publisher><subject>Arthroplasty, Replacement, Hip - instrumentation ; Biomechanical Phenomena ; Device Removal ; Equipment Failure Analysis - methods ; Hip Prosthesis - adverse effects ; Humans ; Materials Testing ; Polyethylene ; Prosthesis Design ; Prosthesis Failure ; Retrospective Studies ; Zirconium</subject><ispartof>The Journal of arthroplasty, 2009-09, Vol.24 (6), p.898-902</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27924,27925</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/18848427$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Jaffe, William L</creatorcontrib><creatorcontrib>Strauss, Eric J</creatorcontrib><creatorcontrib>Cardinale, M</creatorcontrib><creatorcontrib>Herrera, Lizeth</creatorcontrib><creatorcontrib>Kummer, Fred J</creatorcontrib><title>Surface oxidized zirconium total hip arthroplasty head damage due to closed reduction effects on polyethylene wear</title><title>The Journal of arthroplasty</title><addtitle>J Arthroplasty</addtitle><description>Recent case studies of surface oxidized zirconium THA heads removed after attempted, closed reduction have shown significant surface damage that has been suggested as potentially deleterious to polyethylene wear. We obtained 4 clinically retrieved specimens, produced well-characterized surface damage on additional heads, and tested them on a hip simulator. After 1 million cycles, the amount of polyethylene wear was related to the extent of surface damage, the most damaged clinical specimen showing more than 50 times more wear than a new head. Although all heads after failed attempted closed reduction(s) should be replaced, surface oxidized zirconium heads are of particular concern; those patients with a successful, simple closed reduction should be monitored for excessive wear.</description><subject>Arthroplasty, Replacement, Hip - instrumentation</subject><subject>Biomechanical Phenomena</subject><subject>Device Removal</subject><subject>Equipment Failure Analysis - methods</subject><subject>Hip Prosthesis - adverse effects</subject><subject>Humans</subject><subject>Materials Testing</subject><subject>Polyethylene</subject><subject>Prosthesis Design</subject><subject>Prosthesis Failure</subject><subject>Retrospective Studies</subject><subject>Zirconium</subject><issn>0883-5403</issn><issn>1532-8406</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2009</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp90D1PwzAQBmALgWgp_AEG5AmmBH8ltkeE-JIqMQBz5MYXkiqJg-0I0l9PEGVlund47l7pEDqnJKWE5tfb1PhYp4wQlRKZEqIP0JJmnCVKkPwQLYlSPMkE4Qt0EsKWEEqzTByjBVVKKMHkEvmX0VemBOy-GtvswOJd40vXN2OHo4umxXUz4J8e74bWhDjhGozF1nTmHbAdYWa4bF2YVz3YsYyN6zFUFZQx4DkOrp0g1lMLPeBPMP4UHVWmDXC2nyv0dn_3evuYrJ8fnm5v1slAszwmRlHOdJ5LAbQUMgNjrGJG5MBtpZggSjOhickrrWhGacW4ANgAaFJaXlm-Qle_dwfvPkYIseiaUELbmh7cGArJuZZUajHLy38lF5rJLNczvNjDcdOBLQbfdMZPxd8_-TdhkXqm</recordid><startdate>200909</startdate><enddate>200909</enddate><creator>Jaffe, William L</creator><creator>Strauss, Eric J</creator><creator>Cardinale, M</creator><creator>Herrera, Lizeth</creator><creator>Kummer, Fred J</creator><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>7QQ</scope><scope>8FD</scope><scope>JG9</scope><scope>7X8</scope></search><sort><creationdate>200909</creationdate><title>Surface oxidized zirconium total hip arthroplasty head damage due to closed reduction effects on polyethylene wear</title><author>Jaffe, William L ; Strauss, Eric J ; Cardinale, M ; Herrera, Lizeth ; Kummer, Fred J</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-p156t-a813296674e1c475eaad82a46e3df8240892490a6f981511f234eebee90cd3fd3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2009</creationdate><topic>Arthroplasty, Replacement, Hip - instrumentation</topic><topic>Biomechanical Phenomena</topic><topic>Device Removal</topic><topic>Equipment Failure Analysis - methods</topic><topic>Hip Prosthesis - adverse effects</topic><topic>Humans</topic><topic>Materials Testing</topic><topic>Polyethylene</topic><topic>Prosthesis Design</topic><topic>Prosthesis Failure</topic><topic>Retrospective Studies</topic><topic>Zirconium</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Jaffe, William L</creatorcontrib><creatorcontrib>Strauss, Eric J</creatorcontrib><creatorcontrib>Cardinale, M</creatorcontrib><creatorcontrib>Herrera, Lizeth</creatorcontrib><creatorcontrib>Kummer, Fred J</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>Ceramic Abstracts</collection><collection>Technology Research Database</collection><collection>Materials Research Database</collection><collection>MEDLINE - Academic</collection><jtitle>The Journal of arthroplasty</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Jaffe, William L</au><au>Strauss, Eric J</au><au>Cardinale, M</au><au>Herrera, Lizeth</au><au>Kummer, Fred J</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Surface oxidized zirconium total hip arthroplasty head damage due to closed reduction effects on polyethylene wear</atitle><jtitle>The Journal of arthroplasty</jtitle><addtitle>J Arthroplasty</addtitle><date>2009-09</date><risdate>2009</risdate><volume>24</volume><issue>6</issue><spage>898</spage><epage>902</epage><pages>898-902</pages><issn>0883-5403</issn><eissn>1532-8406</eissn><abstract>Recent case studies of surface oxidized zirconium THA heads removed after attempted, closed reduction have shown significant surface damage that has been suggested as potentially deleterious to polyethylene wear. We obtained 4 clinically retrieved specimens, produced well-characterized surface damage on additional heads, and tested them on a hip simulator. After 1 million cycles, the amount of polyethylene wear was related to the extent of surface damage, the most damaged clinical specimen showing more than 50 times more wear than a new head. Although all heads after failed attempted closed reduction(s) should be replaced, surface oxidized zirconium heads are of particular concern; those patients with a successful, simple closed reduction should be monitored for excessive wear.</abstract><cop>United States</cop><pmid>18848427</pmid><doi>10.1016/j.arth.2008.07.009</doi><tpages>5</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0883-5403
ispartof The Journal of arthroplasty, 2009-09, Vol.24 (6), p.898-902
issn 0883-5403
1532-8406
language eng
recordid cdi_proquest_miscellaneous_733971794
source MEDLINE; Elsevier ScienceDirect Journals Complete
subjects Arthroplasty, Replacement, Hip - instrumentation
Biomechanical Phenomena
Device Removal
Equipment Failure Analysis - methods
Hip Prosthesis - adverse effects
Humans
Materials Testing
Polyethylene
Prosthesis Design
Prosthesis Failure
Retrospective Studies
Zirconium
title Surface oxidized zirconium total hip arthroplasty head damage due to closed reduction effects on polyethylene wear
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-24T01%3A13%3A31IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Surface%20oxidized%20zirconium%20total%20hip%20arthroplasty%20head%20damage%20due%20to%20closed%20reduction%20effects%20on%20polyethylene%20wear&rft.jtitle=The%20Journal%20of%20arthroplasty&rft.au=Jaffe,%20William%20L&rft.date=2009-09&rft.volume=24&rft.issue=6&rft.spage=898&rft.epage=902&rft.pages=898-902&rft.issn=0883-5403&rft.eissn=1532-8406&rft_id=info:doi/10.1016/j.arth.2008.07.009&rft_dat=%3Cproquest_pubme%3E34927569%3C/proquest_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=34927569&rft_id=info:pmid/18848427&rfr_iscdi=true