A new experimental procedure for determination of photoelectric efficiency of a NaI(Tl) detector used for nuclear medicine liquid waste monitoring with traceability to a reference standard radionuclide calibrator
To determine photopeak efficiency for 99mTc of the NaI(Tl) detector used for liquid waste monitoring at the Nuclear Medicine Unit of IRCCS Paediatric Hospital Bambino Gesù in Rome, a specific experimental procedure, with traceability to primary standards, was developed. Working with the Italian Nati...
Gespeichert in:
Veröffentlicht in: | Applied radiation and isotopes 2010-02, Vol.68 (2), p.275-279 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 279 |
---|---|
container_issue | 2 |
container_start_page | 275 |
container_title | Applied radiation and isotopes |
container_volume | 68 |
creator | Ceccatelli, A. Campanella, F. Ciofetta, G. Marracino, F.M. Cannatà, V. |
description | To determine photopeak efficiency for
99mTc of the NaI(Tl) detector used for liquid waste monitoring at the Nuclear Medicine Unit of IRCCS Paediatric Hospital Bambino Gesù in Rome, a specific experimental procedure, with traceability to primary standards, was developed. Working with the Italian National Institute for Occupational Prevention and Safety, two different calibration source geometries were employed and the detector response dependence on geometry was investigated. The large percentage difference (almost 40%) between the two efficiency values obtained showed that geometrical effects cannot be neglected. |
doi_str_mv | 10.1016/j.apradiso.2009.10.041 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_733705873</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S0969804309006587</els_id><sourcerecordid>733705873</sourcerecordid><originalsourceid>FETCH-LOGICAL-c367t-104ca4665ef17a30b1b640037b95cd3065a0f86ca5d657147fc79f76ef7fc05b3</originalsourceid><addsrcrecordid>eNqFUU2P0zAQtRCILYW_sPINOKSMm8RJbqxWLKy0gstyjib2mHWV2l3b2dL_yQ9apy3iyGms8fvQvMfYpYCVACE_bVa4C6ht9Ks1QJeXK6jEC7YQbbMuuhbgJVtAJ7uihaq8YG9i3ABA1Xbr1-xCdJ2ooIUF-3PFHe05_d5RsFtyCUe-C16RngJx4wPXlChsrcNkvePe8N2DT55GUilYxckYqyw5dZj_kH_H2w_348cjTaXMnyLpo5Cb1EgY-JZ0Zjjio32crOZ7jIn41jub4db94nubHngKqAgHO9p04Mln5UCGQjYiHhM6jUHzOQE_61pNXOFoh4BZ5C17ZXCM9O48l-znzZf762_F3Y-vt9dXd4UqZZMKAZXCSsqajGiwhEEMsgIom6GrlS5B1gimlQprLetGVI1RTWcaSSa_oB7KJXt_0s2JPU4UU7-1UdE4oiM_xb4pywbqNo8lkyekCj7GfEm_y3FjOPQC-rnQftP_LbSfC533udBMvDxbTEMO7h_t3GAGfD4BKB_6ZCn08VhHDjnkAnrt7f88ngHEXLt7</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>733705873</pqid></control><display><type>article</type><title>A new experimental procedure for determination of photoelectric efficiency of a NaI(Tl) detector used for nuclear medicine liquid waste monitoring with traceability to a reference standard radionuclide calibrator</title><source>MEDLINE</source><source>ScienceDirect Journals (5 years ago - present)</source><creator>Ceccatelli, A. ; Campanella, F. ; Ciofetta, G. ; Marracino, F.M. ; Cannatà, V.</creator><creatorcontrib>Ceccatelli, A. ; Campanella, F. ; Ciofetta, G. ; Marracino, F.M. ; Cannatà, V.</creatorcontrib><description>To determine photopeak efficiency for
99mTc of the NaI(Tl) detector used for liquid waste monitoring at the Nuclear Medicine Unit of IRCCS Paediatric Hospital Bambino Gesù in Rome, a specific experimental procedure, with traceability to primary standards, was developed. Working with the Italian National Institute for Occupational Prevention and Safety, two different calibration source geometries were employed and the detector response dependence on geometry was investigated. The large percentage difference (almost 40%) between the two efficiency values obtained showed that geometrical effects cannot be neglected.</description><identifier>ISSN: 0969-8043</identifier><identifier>EISSN: 1872-9800</identifier><identifier>DOI: 10.1016/j.apradiso.2009.10.041</identifier><identifier>PMID: 19914080</identifier><language>eng</language><publisher>England: Elsevier Ltd</publisher><subject>Calibration ; Efficiency calibration ; Equipment Design ; Equipment Failure Analysis ; Experimental procedure accuracy ; Geometrical effect ; Liquid waste monitoring ; Measurement geometry ; Medical Waste - analysis ; Nuclear medicine ; Radiation Dosage ; Radiation Monitoring - instrumentation ; Radiation Monitoring - standards ; Radioactive Waste - analysis ; Radiopharmaceuticals - analysis ; Reference Standards ; Reproducibility of Results ; Rome ; Sensitivity and Specificity ; Sodium Iodide - radiation effects ; Solutions ; Traceability</subject><ispartof>Applied radiation and isotopes, 2010-02, Vol.68 (2), p.275-279</ispartof><rights>2009 Elsevier Ltd</rights><rights>Copyright (c) 2009 Elsevier Ltd. All rights reserved.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c367t-104ca4665ef17a30b1b640037b95cd3065a0f86ca5d657147fc79f76ef7fc05b3</citedby><cites>FETCH-LOGICAL-c367t-104ca4665ef17a30b1b640037b95cd3065a0f86ca5d657147fc79f76ef7fc05b3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://dx.doi.org/10.1016/j.apradiso.2009.10.041$$EHTML$$P50$$Gelsevier$$H</linktohtml><link.rule.ids>314,780,784,3550,27924,27925,45995</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/19914080$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Ceccatelli, A.</creatorcontrib><creatorcontrib>Campanella, F.</creatorcontrib><creatorcontrib>Ciofetta, G.</creatorcontrib><creatorcontrib>Marracino, F.M.</creatorcontrib><creatorcontrib>Cannatà, V.</creatorcontrib><title>A new experimental procedure for determination of photoelectric efficiency of a NaI(Tl) detector used for nuclear medicine liquid waste monitoring with traceability to a reference standard radionuclide calibrator</title><title>Applied radiation and isotopes</title><addtitle>Appl Radiat Isot</addtitle><description>To determine photopeak efficiency for
99mTc of the NaI(Tl) detector used for liquid waste monitoring at the Nuclear Medicine Unit of IRCCS Paediatric Hospital Bambino Gesù in Rome, a specific experimental procedure, with traceability to primary standards, was developed. Working with the Italian National Institute for Occupational Prevention and Safety, two different calibration source geometries were employed and the detector response dependence on geometry was investigated. The large percentage difference (almost 40%) between the two efficiency values obtained showed that geometrical effects cannot be neglected.</description><subject>Calibration</subject><subject>Efficiency calibration</subject><subject>Equipment Design</subject><subject>Equipment Failure Analysis</subject><subject>Experimental procedure accuracy</subject><subject>Geometrical effect</subject><subject>Liquid waste monitoring</subject><subject>Measurement geometry</subject><subject>Medical Waste - analysis</subject><subject>Nuclear medicine</subject><subject>Radiation Dosage</subject><subject>Radiation Monitoring - instrumentation</subject><subject>Radiation Monitoring - standards</subject><subject>Radioactive Waste - analysis</subject><subject>Radiopharmaceuticals - analysis</subject><subject>Reference Standards</subject><subject>Reproducibility of Results</subject><subject>Rome</subject><subject>Sensitivity and Specificity</subject><subject>Sodium Iodide - radiation effects</subject><subject>Solutions</subject><subject>Traceability</subject><issn>0969-8043</issn><issn>1872-9800</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2010</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNqFUU2P0zAQtRCILYW_sPINOKSMm8RJbqxWLKy0gstyjib2mHWV2l3b2dL_yQ9apy3iyGms8fvQvMfYpYCVACE_bVa4C6ht9Ks1QJeXK6jEC7YQbbMuuhbgJVtAJ7uihaq8YG9i3ABA1Xbr1-xCdJ2ooIUF-3PFHe05_d5RsFtyCUe-C16RngJx4wPXlChsrcNkvePe8N2DT55GUilYxckYqyw5dZj_kH_H2w_348cjTaXMnyLpo5Cb1EgY-JZ0Zjjio32crOZ7jIn41jub4db94nubHngKqAgHO9p04Mln5UCGQjYiHhM6jUHzOQE_61pNXOFoh4BZ5C17ZXCM9O48l-znzZf762_F3Y-vt9dXd4UqZZMKAZXCSsqajGiwhEEMsgIom6GrlS5B1gimlQprLetGVI1RTWcaSSa_oB7KJXt_0s2JPU4UU7-1UdE4oiM_xb4pywbqNo8lkyekCj7GfEm_y3FjOPQC-rnQftP_LbSfC533udBMvDxbTEMO7h_t3GAGfD4BKB_6ZCn08VhHDjnkAnrt7f88ngHEXLt7</recordid><startdate>20100201</startdate><enddate>20100201</enddate><creator>Ceccatelli, A.</creator><creator>Campanella, F.</creator><creator>Ciofetta, G.</creator><creator>Marracino, F.M.</creator><creator>Cannatà, V.</creator><general>Elsevier Ltd</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>20100201</creationdate><title>A new experimental procedure for determination of photoelectric efficiency of a NaI(Tl) detector used for nuclear medicine liquid waste monitoring with traceability to a reference standard radionuclide calibrator</title><author>Ceccatelli, A. ; Campanella, F. ; Ciofetta, G. ; Marracino, F.M. ; Cannatà, V.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c367t-104ca4665ef17a30b1b640037b95cd3065a0f86ca5d657147fc79f76ef7fc05b3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2010</creationdate><topic>Calibration</topic><topic>Efficiency calibration</topic><topic>Equipment Design</topic><topic>Equipment Failure Analysis</topic><topic>Experimental procedure accuracy</topic><topic>Geometrical effect</topic><topic>Liquid waste monitoring</topic><topic>Measurement geometry</topic><topic>Medical Waste - analysis</topic><topic>Nuclear medicine</topic><topic>Radiation Dosage</topic><topic>Radiation Monitoring - instrumentation</topic><topic>Radiation Monitoring - standards</topic><topic>Radioactive Waste - analysis</topic><topic>Radiopharmaceuticals - analysis</topic><topic>Reference Standards</topic><topic>Reproducibility of Results</topic><topic>Rome</topic><topic>Sensitivity and Specificity</topic><topic>Sodium Iodide - radiation effects</topic><topic>Solutions</topic><topic>Traceability</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Ceccatelli, A.</creatorcontrib><creatorcontrib>Campanella, F.</creatorcontrib><creatorcontrib>Ciofetta, G.</creatorcontrib><creatorcontrib>Marracino, F.M.</creatorcontrib><creatorcontrib>Cannatà, V.</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Applied radiation and isotopes</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Ceccatelli, A.</au><au>Campanella, F.</au><au>Ciofetta, G.</au><au>Marracino, F.M.</au><au>Cannatà, V.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>A new experimental procedure for determination of photoelectric efficiency of a NaI(Tl) detector used for nuclear medicine liquid waste monitoring with traceability to a reference standard radionuclide calibrator</atitle><jtitle>Applied radiation and isotopes</jtitle><addtitle>Appl Radiat Isot</addtitle><date>2010-02-01</date><risdate>2010</risdate><volume>68</volume><issue>2</issue><spage>275</spage><epage>279</epage><pages>275-279</pages><issn>0969-8043</issn><eissn>1872-9800</eissn><abstract>To determine photopeak efficiency for
99mTc of the NaI(Tl) detector used for liquid waste monitoring at the Nuclear Medicine Unit of IRCCS Paediatric Hospital Bambino Gesù in Rome, a specific experimental procedure, with traceability to primary standards, was developed. Working with the Italian National Institute for Occupational Prevention and Safety, two different calibration source geometries were employed and the detector response dependence on geometry was investigated. The large percentage difference (almost 40%) between the two efficiency values obtained showed that geometrical effects cannot be neglected.</abstract><cop>England</cop><pub>Elsevier Ltd</pub><pmid>19914080</pmid><doi>10.1016/j.apradiso.2009.10.041</doi><tpages>5</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0969-8043 |
ispartof | Applied radiation and isotopes, 2010-02, Vol.68 (2), p.275-279 |
issn | 0969-8043 1872-9800 |
language | eng |
recordid | cdi_proquest_miscellaneous_733705873 |
source | MEDLINE; ScienceDirect Journals (5 years ago - present) |
subjects | Calibration Efficiency calibration Equipment Design Equipment Failure Analysis Experimental procedure accuracy Geometrical effect Liquid waste monitoring Measurement geometry Medical Waste - analysis Nuclear medicine Radiation Dosage Radiation Monitoring - instrumentation Radiation Monitoring - standards Radioactive Waste - analysis Radiopharmaceuticals - analysis Reference Standards Reproducibility of Results Rome Sensitivity and Specificity Sodium Iodide - radiation effects Solutions Traceability |
title | A new experimental procedure for determination of photoelectric efficiency of a NaI(Tl) detector used for nuclear medicine liquid waste monitoring with traceability to a reference standard radionuclide calibrator |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-06T02%3A17%3A54IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=A%20new%20experimental%20procedure%20for%20determination%20of%20photoelectric%20efficiency%20of%20a%20NaI(Tl)%20detector%20used%20for%20nuclear%20medicine%20liquid%20waste%20monitoring%20with%20traceability%20to%20a%20reference%20standard%20radionuclide%20calibrator&rft.jtitle=Applied%20radiation%20and%20isotopes&rft.au=Ceccatelli,%20A.&rft.date=2010-02-01&rft.volume=68&rft.issue=2&rft.spage=275&rft.epage=279&rft.pages=275-279&rft.issn=0969-8043&rft.eissn=1872-9800&rft_id=info:doi/10.1016/j.apradiso.2009.10.041&rft_dat=%3Cproquest_cross%3E733705873%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=733705873&rft_id=info:pmid/19914080&rft_els_id=S0969804309006587&rfr_iscdi=true |