Accuracy of Methemoglobin Detection by Pulse CO-Oximetry During Hypoxia

Methemoglobin in the blood cannot be detected by conventional pulse oximetry, although it can bias the oximeter's estimate (Spo2) of the true arterial functional oxygen saturation (Sao2). A recently introduced "Pulse CO-Oximeter" (Masimo Rainbow SET(R) Radical-7 Pulse CO-Oximeter, Mas...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Anesthesia and analgesia 2010-07, Vol.111 (1), p.143-148
Hauptverfasser: Feiner, John R., Bickler, Philip E., Mannheimer, Paul D.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 148
container_issue 1
container_start_page 143
container_title Anesthesia and analgesia
container_volume 111
creator Feiner, John R.
Bickler, Philip E.
Mannheimer, Paul D.
description Methemoglobin in the blood cannot be detected by conventional pulse oximetry, although it can bias the oximeter's estimate (Spo2) of the true arterial functional oxygen saturation (Sao2). A recently introduced "Pulse CO-Oximeter" (Masimo Rainbow SET(R) Radical-7 Pulse CO-Oximeter, Masimo Corp., Irvine, CA) is intended to additionally monitor noninvasively the fractional carboxyhemoglobin and methemoglobin content in blood. The purpose of our study was to determine whether hypoxia affects the new device's estimated methemoglobin reading accuracy, and whether the presence of methemoglobin impairs the ability of the Radical-7 and a conventional pulse oximeter (Nonin 9700, Nonin Medical Inc., Plymouth, MN) to detect decreases in Sao2. Eight and 6 healthy adults were included in 2 study groups, respectively, each fitted with multiple sensors and a radial arterial catheter for blood sampling. In the first group, IV administration of approximately 300 mg sodium nitrite increased subjects' methemoglobin level to a 7% to 8% target and hypoxia was induced to different levels of Sao2 (70%-100%) by varying fractional inspired oxygen. In the second group, 15% methemoglobin at room air and 80% Sao2 were targeted. Pulse CO-oximeter readings were compared with arterial blood values measured using a Radiometer multiwavelength hemoximeter. Pulse CO-oximeter methemoglobin reading performance was analyzed by observing the incidence of meaningful reading errors at the various hypoxia levels. This was used to determine the impact on predictive values for detecting methemoglobinemia. Spo2 reading bias, precision, and root mean square error were evaluated during conditions of elevated methemoglobin. Observations spanned 66.2% to 99% Sao2 and 0.6% to 14.4% methemoglobin over the 2 groups (170 blood draws). Masimo methemoglobin reading bias and precision over the full Sao2 span was 7.7% +/- 13.0%. Best accuracy was found in the 95% to 100% Sao2 range (1.9% +/- 2.5%), progressing to its worst in the 70% to 80% range (24.8% +/- 15.6%). Occurrence of methemoglobin readings in error >5% increased over each 5-point decrease in Sao2 (P < 0.05). Masimo Spo2 readings were biased -6.3% +/- 3.0% in the 95% to 100% Sao2 range with 4% to 8.3% methemoglobin. Both the Radical-7 and Nonin 9700 pulse oximeters accurately detected decreases in Sao(2) 95%. The Radical-7's methemoglobin readings become progressivel
doi_str_mv 10.1213/ANE.0b013e3181c91bb6
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_733441257</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>733441257</sourcerecordid><originalsourceid>FETCH-LOGICAL-c4936-3abd65d2d5baa4a9f9c35b8dc7ddc5b2ba0c43340c41f5db420c9c28c7675eef3</originalsourceid><addsrcrecordid>eNpdkVtP3DAQha2qqGxp_0FV5aXiKeBLHMePq-UqAcsDfbbsyYRNSeLFTgT59zViAQlL9nik78yRzhDyi9Ejxpk4Xt6cHlFHmUDBKgaaOVd-IQsmeZkrqauvZEEpFTnXWu-T7zH-Sy2jVfmN7PP0VUqwBTlfAkzBwpz5JrvGcYO9v--8a4fsBEeEsfVD5ubsduoiZqt1vn5uexzDnJ1MoR3us4t5659b-4PsNTYhP3f1gPw9O71bXeRX6_PL1fIqh0KLMhfW1aWseS2dtYXVjQYhXVWDqmuQjjtLoRCiSC9rZO0KTkEDr0CVSiI24oAcvs7dBv84YRxN30bArrMD-ikalcQF41IlsnglIfgYAzZmG9rehtkwal4SNClB8znBJPu9M5hcj_W76C2yBPzZATaC7ZpgB2jjB8d1lVZQfPg_-W7EEB-66QmD2aDtxo2hL0cKnfO0E6pSk6fLS_EfYSyKFg</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>733441257</pqid></control><display><type>article</type><title>Accuracy of Methemoglobin Detection by Pulse CO-Oximetry During Hypoxia</title><source>Ovid Lippincott Williams and Wilkins Journal Legacy Archive</source><source>MEDLINE</source><source>EZB-FREE-00999 freely available EZB journals</source><creator>Feiner, John R. ; Bickler, Philip E. ; Mannheimer, Paul D.</creator><creatorcontrib>Feiner, John R. ; Bickler, Philip E. ; Mannheimer, Paul D.</creatorcontrib><description>Methemoglobin in the blood cannot be detected by conventional pulse oximetry, although it can bias the oximeter's estimate (Spo2) of the true arterial functional oxygen saturation (Sao2). A recently introduced "Pulse CO-Oximeter" (Masimo Rainbow SET(R) Radical-7 Pulse CO-Oximeter, Masimo Corp., Irvine, CA) is intended to additionally monitor noninvasively the fractional carboxyhemoglobin and methemoglobin content in blood. The purpose of our study was to determine whether hypoxia affects the new device's estimated methemoglobin reading accuracy, and whether the presence of methemoglobin impairs the ability of the Radical-7 and a conventional pulse oximeter (Nonin 9700, Nonin Medical Inc., Plymouth, MN) to detect decreases in Sao2. Eight and 6 healthy adults were included in 2 study groups, respectively, each fitted with multiple sensors and a radial arterial catheter for blood sampling. In the first group, IV administration of approximately 300 mg sodium nitrite increased subjects' methemoglobin level to a 7% to 8% target and hypoxia was induced to different levels of Sao2 (70%-100%) by varying fractional inspired oxygen. In the second group, 15% methemoglobin at room air and 80% Sao2 were targeted. Pulse CO-oximeter readings were compared with arterial blood values measured using a Radiometer multiwavelength hemoximeter. Pulse CO-oximeter methemoglobin reading performance was analyzed by observing the incidence of meaningful reading errors at the various hypoxia levels. This was used to determine the impact on predictive values for detecting methemoglobinemia. Spo2 reading bias, precision, and root mean square error were evaluated during conditions of elevated methemoglobin. Observations spanned 66.2% to 99% Sao2 and 0.6% to 14.4% methemoglobin over the 2 groups (170 blood draws). Masimo methemoglobin reading bias and precision over the full Sao2 span was 7.7% +/- 13.0%. Best accuracy was found in the 95% to 100% Sao2 range (1.9% +/- 2.5%), progressing to its worst in the 70% to 80% range (24.8% +/- 15.6%). Occurrence of methemoglobin readings in error &gt;5% increased over each 5-point decrease in Sao2 (P &lt; 0.05). Masimo Spo2 readings were biased -6.3% +/- 3.0% in the 95% to 100% Sao2 range with 4% to 8.3% methemoglobin. Both the Radical-7 and Nonin 9700 pulse oximeters accurately detected decreases in Sao(2) &lt;90% with 4% to 15% methemoglobin, despite displaying low Spo2 readings when Sao2 was &gt;95%. The Radical-7's methemoglobin readings become progressively more inaccurate as Sao2 decreases &lt;95%, at times overestimating true values by 10% to 40%. Elevated methemoglobin causes the Spo2 readings to underestimate Sao2 similar to conventional 2-wavelength pulse oximeters at high saturation. Spo2 readings from both types of instruments continue to trend downward during the development of hypoxemia (Sao2 &lt;90%) with methemoglobin levels up to 15%.</description><identifier>ISSN: 0003-2999</identifier><identifier>EISSN: 1526-7598</identifier><identifier>DOI: 10.1213/ANE.0b013e3181c91bb6</identifier><identifier>PMID: 20007731</identifier><identifier>CODEN: AACRAT</identifier><language>eng</language><publisher>Hagerstown, MD: International Anesthesia Research Society</publisher><subject>Adult ; Anesthesia ; Anesthesia. Intensive care medicine. Transfusions. Cell therapy and gene therapy ; Biological and medical sciences ; Blood Gas Analysis ; Carbon Monoxide - blood ; Carboxyhemoglobin - analysis ; Data Interpretation, Statistical ; Female ; Hemoglobinometry - methods ; Humans ; Hypoxia - blood ; Male ; Medical sciences ; Methemoglobin - analysis ; Methemoglobinemia - blood ; Methemoglobinemia - diagnosis ; Oximetry - instrumentation ; Oximetry - methods ; Oxygen - blood ; Predictive Value of Tests ; Reproducibility of Results ; Sodium Nitrite</subject><ispartof>Anesthesia and analgesia, 2010-07, Vol.111 (1), p.143-148</ispartof><rights>International Anesthesia Research Society</rights><rights>2015 INIST-CNRS</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c4936-3abd65d2d5baa4a9f9c35b8dc7ddc5b2ba0c43340c41f5db420c9c28c7675eef3</citedby><cites>FETCH-LOGICAL-c4936-3abd65d2d5baa4a9f9c35b8dc7ddc5b2ba0c43340c41f5db420c9c28c7675eef3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf><![CDATA[$$Uhttp://ovidsp.ovid.com/ovidweb.cgi?T=JS&CSC=Y&NEWS=N&PAGE=fulltext&PDF=y&D=ovft&AN=00000539-201007000-00026$$EPDF$$P50$$Gwolterskluwer$$H]]></linktopdf><linktohtml>$$Uhttp://ovidsp.ovid.com/ovidweb.cgi?T=JS&amp;NEWS=n&amp;CSC=Y&amp;PAGE=fulltext&amp;D=ovft&amp;AN=00000539-201007000-00026$$EHTML$$P50$$Gwolterskluwer$$H</linktohtml><link.rule.ids>314,780,784,4609,27924,27925,64666,65461</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&amp;idt=22985264$$DView record in Pascal Francis$$Hfree_for_read</backlink><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/20007731$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Feiner, John R.</creatorcontrib><creatorcontrib>Bickler, Philip E.</creatorcontrib><creatorcontrib>Mannheimer, Paul D.</creatorcontrib><title>Accuracy of Methemoglobin Detection by Pulse CO-Oximetry During Hypoxia</title><title>Anesthesia and analgesia</title><addtitle>Anesth Analg</addtitle><description>Methemoglobin in the blood cannot be detected by conventional pulse oximetry, although it can bias the oximeter's estimate (Spo2) of the true arterial functional oxygen saturation (Sao2). A recently introduced "Pulse CO-Oximeter" (Masimo Rainbow SET(R) Radical-7 Pulse CO-Oximeter, Masimo Corp., Irvine, CA) is intended to additionally monitor noninvasively the fractional carboxyhemoglobin and methemoglobin content in blood. The purpose of our study was to determine whether hypoxia affects the new device's estimated methemoglobin reading accuracy, and whether the presence of methemoglobin impairs the ability of the Radical-7 and a conventional pulse oximeter (Nonin 9700, Nonin Medical Inc., Plymouth, MN) to detect decreases in Sao2. Eight and 6 healthy adults were included in 2 study groups, respectively, each fitted with multiple sensors and a radial arterial catheter for blood sampling. In the first group, IV administration of approximately 300 mg sodium nitrite increased subjects' methemoglobin level to a 7% to 8% target and hypoxia was induced to different levels of Sao2 (70%-100%) by varying fractional inspired oxygen. In the second group, 15% methemoglobin at room air and 80% Sao2 were targeted. Pulse CO-oximeter readings were compared with arterial blood values measured using a Radiometer multiwavelength hemoximeter. Pulse CO-oximeter methemoglobin reading performance was analyzed by observing the incidence of meaningful reading errors at the various hypoxia levels. This was used to determine the impact on predictive values for detecting methemoglobinemia. Spo2 reading bias, precision, and root mean square error were evaluated during conditions of elevated methemoglobin. Observations spanned 66.2% to 99% Sao2 and 0.6% to 14.4% methemoglobin over the 2 groups (170 blood draws). Masimo methemoglobin reading bias and precision over the full Sao2 span was 7.7% +/- 13.0%. Best accuracy was found in the 95% to 100% Sao2 range (1.9% +/- 2.5%), progressing to its worst in the 70% to 80% range (24.8% +/- 15.6%). Occurrence of methemoglobin readings in error &gt;5% increased over each 5-point decrease in Sao2 (P &lt; 0.05). Masimo Spo2 readings were biased -6.3% +/- 3.0% in the 95% to 100% Sao2 range with 4% to 8.3% methemoglobin. Both the Radical-7 and Nonin 9700 pulse oximeters accurately detected decreases in Sao(2) &lt;90% with 4% to 15% methemoglobin, despite displaying low Spo2 readings when Sao2 was &gt;95%. The Radical-7's methemoglobin readings become progressively more inaccurate as Sao2 decreases &lt;95%, at times overestimating true values by 10% to 40%. Elevated methemoglobin causes the Spo2 readings to underestimate Sao2 similar to conventional 2-wavelength pulse oximeters at high saturation. Spo2 readings from both types of instruments continue to trend downward during the development of hypoxemia (Sao2 &lt;90%) with methemoglobin levels up to 15%.</description><subject>Adult</subject><subject>Anesthesia</subject><subject>Anesthesia. Intensive care medicine. Transfusions. Cell therapy and gene therapy</subject><subject>Biological and medical sciences</subject><subject>Blood Gas Analysis</subject><subject>Carbon Monoxide - blood</subject><subject>Carboxyhemoglobin - analysis</subject><subject>Data Interpretation, Statistical</subject><subject>Female</subject><subject>Hemoglobinometry - methods</subject><subject>Humans</subject><subject>Hypoxia - blood</subject><subject>Male</subject><subject>Medical sciences</subject><subject>Methemoglobin - analysis</subject><subject>Methemoglobinemia - blood</subject><subject>Methemoglobinemia - diagnosis</subject><subject>Oximetry - instrumentation</subject><subject>Oximetry - methods</subject><subject>Oxygen - blood</subject><subject>Predictive Value of Tests</subject><subject>Reproducibility of Results</subject><subject>Sodium Nitrite</subject><issn>0003-2999</issn><issn>1526-7598</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2010</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNpdkVtP3DAQha2qqGxp_0FV5aXiKeBLHMePq-UqAcsDfbbsyYRNSeLFTgT59zViAQlL9nik78yRzhDyi9Ejxpk4Xt6cHlFHmUDBKgaaOVd-IQsmeZkrqauvZEEpFTnXWu-T7zH-Sy2jVfmN7PP0VUqwBTlfAkzBwpz5JrvGcYO9v--8a4fsBEeEsfVD5ubsduoiZqt1vn5uexzDnJ1MoR3us4t5659b-4PsNTYhP3f1gPw9O71bXeRX6_PL1fIqh0KLMhfW1aWseS2dtYXVjQYhXVWDqmuQjjtLoRCiSC9rZO0KTkEDr0CVSiI24oAcvs7dBv84YRxN30bArrMD-ikalcQF41IlsnglIfgYAzZmG9rehtkwal4SNClB8znBJPu9M5hcj_W76C2yBPzZATaC7ZpgB2jjB8d1lVZQfPg_-W7EEB-66QmD2aDtxo2hL0cKnfO0E6pSk6fLS_EfYSyKFg</recordid><startdate>20100701</startdate><enddate>20100701</enddate><creator>Feiner, John R.</creator><creator>Bickler, Philip E.</creator><creator>Mannheimer, Paul D.</creator><general>International Anesthesia Research Society</general><general>Lippincott Williams &amp; Wilkins</general><scope>IQODW</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>20100701</creationdate><title>Accuracy of Methemoglobin Detection by Pulse CO-Oximetry During Hypoxia</title><author>Feiner, John R. ; Bickler, Philip E. ; Mannheimer, Paul D.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c4936-3abd65d2d5baa4a9f9c35b8dc7ddc5b2ba0c43340c41f5db420c9c28c7675eef3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2010</creationdate><topic>Adult</topic><topic>Anesthesia</topic><topic>Anesthesia. Intensive care medicine. Transfusions. Cell therapy and gene therapy</topic><topic>Biological and medical sciences</topic><topic>Blood Gas Analysis</topic><topic>Carbon Monoxide - blood</topic><topic>Carboxyhemoglobin - analysis</topic><topic>Data Interpretation, Statistical</topic><topic>Female</topic><topic>Hemoglobinometry - methods</topic><topic>Humans</topic><topic>Hypoxia - blood</topic><topic>Male</topic><topic>Medical sciences</topic><topic>Methemoglobin - analysis</topic><topic>Methemoglobinemia - blood</topic><topic>Methemoglobinemia - diagnosis</topic><topic>Oximetry - instrumentation</topic><topic>Oximetry - methods</topic><topic>Oxygen - blood</topic><topic>Predictive Value of Tests</topic><topic>Reproducibility of Results</topic><topic>Sodium Nitrite</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Feiner, John R.</creatorcontrib><creatorcontrib>Bickler, Philip E.</creatorcontrib><creatorcontrib>Mannheimer, Paul D.</creatorcontrib><collection>Pascal-Francis</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Anesthesia and analgesia</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Feiner, John R.</au><au>Bickler, Philip E.</au><au>Mannheimer, Paul D.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Accuracy of Methemoglobin Detection by Pulse CO-Oximetry During Hypoxia</atitle><jtitle>Anesthesia and analgesia</jtitle><addtitle>Anesth Analg</addtitle><date>2010-07-01</date><risdate>2010</risdate><volume>111</volume><issue>1</issue><spage>143</spage><epage>148</epage><pages>143-148</pages><issn>0003-2999</issn><eissn>1526-7598</eissn><coden>AACRAT</coden><abstract>Methemoglobin in the blood cannot be detected by conventional pulse oximetry, although it can bias the oximeter's estimate (Spo2) of the true arterial functional oxygen saturation (Sao2). A recently introduced "Pulse CO-Oximeter" (Masimo Rainbow SET(R) Radical-7 Pulse CO-Oximeter, Masimo Corp., Irvine, CA) is intended to additionally monitor noninvasively the fractional carboxyhemoglobin and methemoglobin content in blood. The purpose of our study was to determine whether hypoxia affects the new device's estimated methemoglobin reading accuracy, and whether the presence of methemoglobin impairs the ability of the Radical-7 and a conventional pulse oximeter (Nonin 9700, Nonin Medical Inc., Plymouth, MN) to detect decreases in Sao2. Eight and 6 healthy adults were included in 2 study groups, respectively, each fitted with multiple sensors and a radial arterial catheter for blood sampling. In the first group, IV administration of approximately 300 mg sodium nitrite increased subjects' methemoglobin level to a 7% to 8% target and hypoxia was induced to different levels of Sao2 (70%-100%) by varying fractional inspired oxygen. In the second group, 15% methemoglobin at room air and 80% Sao2 were targeted. Pulse CO-oximeter readings were compared with arterial blood values measured using a Radiometer multiwavelength hemoximeter. Pulse CO-oximeter methemoglobin reading performance was analyzed by observing the incidence of meaningful reading errors at the various hypoxia levels. This was used to determine the impact on predictive values for detecting methemoglobinemia. Spo2 reading bias, precision, and root mean square error were evaluated during conditions of elevated methemoglobin. Observations spanned 66.2% to 99% Sao2 and 0.6% to 14.4% methemoglobin over the 2 groups (170 blood draws). Masimo methemoglobin reading bias and precision over the full Sao2 span was 7.7% +/- 13.0%. Best accuracy was found in the 95% to 100% Sao2 range (1.9% +/- 2.5%), progressing to its worst in the 70% to 80% range (24.8% +/- 15.6%). Occurrence of methemoglobin readings in error &gt;5% increased over each 5-point decrease in Sao2 (P &lt; 0.05). Masimo Spo2 readings were biased -6.3% +/- 3.0% in the 95% to 100% Sao2 range with 4% to 8.3% methemoglobin. Both the Radical-7 and Nonin 9700 pulse oximeters accurately detected decreases in Sao(2) &lt;90% with 4% to 15% methemoglobin, despite displaying low Spo2 readings when Sao2 was &gt;95%. The Radical-7's methemoglobin readings become progressively more inaccurate as Sao2 decreases &lt;95%, at times overestimating true values by 10% to 40%. Elevated methemoglobin causes the Spo2 readings to underestimate Sao2 similar to conventional 2-wavelength pulse oximeters at high saturation. Spo2 readings from both types of instruments continue to trend downward during the development of hypoxemia (Sao2 &lt;90%) with methemoglobin levels up to 15%.</abstract><cop>Hagerstown, MD</cop><pub>International Anesthesia Research Society</pub><pmid>20007731</pmid><doi>10.1213/ANE.0b013e3181c91bb6</doi><tpages>6</tpages><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 0003-2999
ispartof Anesthesia and analgesia, 2010-07, Vol.111 (1), p.143-148
issn 0003-2999
1526-7598
language eng
recordid cdi_proquest_miscellaneous_733441257
source Ovid Lippincott Williams and Wilkins Journal Legacy Archive; MEDLINE; EZB-FREE-00999 freely available EZB journals
subjects Adult
Anesthesia
Anesthesia. Intensive care medicine. Transfusions. Cell therapy and gene therapy
Biological and medical sciences
Blood Gas Analysis
Carbon Monoxide - blood
Carboxyhemoglobin - analysis
Data Interpretation, Statistical
Female
Hemoglobinometry - methods
Humans
Hypoxia - blood
Male
Medical sciences
Methemoglobin - analysis
Methemoglobinemia - blood
Methemoglobinemia - diagnosis
Oximetry - instrumentation
Oximetry - methods
Oxygen - blood
Predictive Value of Tests
Reproducibility of Results
Sodium Nitrite
title Accuracy of Methemoglobin Detection by Pulse CO-Oximetry During Hypoxia
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-01T03%3A07%3A06IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Accuracy%20of%20Methemoglobin%20Detection%20by%20Pulse%20CO-Oximetry%20During%20Hypoxia&rft.jtitle=Anesthesia%20and%20analgesia&rft.au=Feiner,%20John%20R.&rft.date=2010-07-01&rft.volume=111&rft.issue=1&rft.spage=143&rft.epage=148&rft.pages=143-148&rft.issn=0003-2999&rft.eissn=1526-7598&rft.coden=AACRAT&rft_id=info:doi/10.1213/ANE.0b013e3181c91bb6&rft_dat=%3Cproquest_cross%3E733441257%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=733441257&rft_id=info:pmid/20007731&rfr_iscdi=true