Amperometric glucose biosensor with extended concentration range utilizing complexation effect of borate

Borate buffer strongly decreases amperometric response of a glucose oxidase linked pO 2 or H 2O 2 sensing electrode, extending substantially its linear calibration range. With increasing pH and concentration of the buffer the upper limit for glucose can be varied between 1 and 30 mmol 1 −1 glucose....

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Biosensors & bioelectronics 1992, Vol.7 (8), p.593-598
Hauptverfasser: MACHOLAN, L, SKLADAL, P, BOHACKOVA, I, KREJCI, J
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Borate buffer strongly decreases amperometric response of a glucose oxidase linked pO 2 or H 2O 2 sensing electrode, extending substantially its linear calibration range. With increasing pH and concentration of the buffer the upper limit for glucose can be varied between 1 and 30 mmol 1 −1 glucose. The effect of borate ion is explained by the rapid complexation of glucose decreasing the equilibrium concentration of free β-anomer, the specific substrate of glucose oxidase. The high loading of cross-linked enzyme inside the sensor membrane is necessary for the measurement to ensure an almost constant response factor (Δ i per 1 mmol 1 −1) between pH 5 and 10. Analysis is stirred solution and in a flow-through system has been employed for the measurement of elevated glucose levels in heparinized human blood or plasma samples.
ISSN:0956-5663
1873-4235
DOI:10.1016/0956-5663(92)85011-X