Wavelet based automatic seizure detection in intracerebral electroencephalogram
Background: Automatic seizure detection is often used during long-term monitoring, and is particularly important during intracerebral investigations. Existing methods make many false detections, particularly in intracerebral electroencephalogram (EEG) because of frequent large amplitude rhythmic act...
Gespeichert in:
Veröffentlicht in: | Clinical neurophysiology 2003-05, Vol.114 (5), p.898-908 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 908 |
---|---|
container_issue | 5 |
container_start_page | 898 |
container_title | Clinical neurophysiology |
container_volume | 114 |
creator | Khan, Y.U. Gotman, J. |
description | Background: Automatic seizure detection is often used during long-term monitoring, and is particularly important during intracerebral investigations. Existing methods make many false detections, particularly in intracerebral electroencephalogram (EEG) because of frequent large amplitude rhythmic activity bursts that are non-epileptiform.
Objective: To develop a seizure detection method for intracerebral monitoring that is as sensitive as existing methods but has fewer false detections.
Methods: To capture the rhythmic nature of seizure discharges, we developed a wavelet-based method, examining how different frequency ranges fluctuate compared to the background. In particular, the system remembers rhythmic bursts occurring commonly in the background to avoid detecting them as seizures.
Results: The method was evaluated on test data from 11 patients, including 229 h and 66 seizures, and its performance compared to the method of Gotman (Electroencephalogr clin Neurophysiol 76 (1990) 317). Detection sensitivity was unchanged at close to 90%, but false detections were reduced from 2.4 to 0.3/h.
Conclusions: Perfect sensitivity is unlikely because the morphology of seizure discharges is so variable. Nevertheless, the 87% sensitivity obtained in the combined training and testing data is quite high. We reduced the average false alarm rate to one per 3 h of recording, or 6 per 24-h period. Given how rapidly one can decide visually that a detection is erroneous, false detections should not cause any burden to the reviewer.
Significance: In intracerebral EEG it is possible to detect seizures automatically with high sensitivity and high specificity. |
doi_str_mv | 10.1016/S1388-2457(03)00035-X |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_73243999</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S138824570300035X</els_id><sourcerecordid>73243999</sourcerecordid><originalsourceid>FETCH-LOGICAL-c391t-a63703d78ca14184ac4487c859a4b4156f10dc0eb9579cca420bf9dd611551c23</originalsourceid><addsrcrecordid>eNqFkM9rFTEQx4NYbK3-CcpeFD1sTTbJJjkVKf4oFHpQsbcwO5nVyO7bZ5It6F9vXt-THgsDM4fPd2b4MPZC8DPBRf_ui5DWtp3S5g2XbznnUrc3j9iJsKZrrdPd4zr_R47Z05x_Vchw1T1hx6Iz0ippTtj1d7iliUozQKbQwFqWGUrEJlP8uyZqAhXCEpdNE3dVEiAlGhJMTc1hSQttkLY_YVp-JJifsaMRpkzPD_2Uffv44evF5_bq-tPlxfurFqUTpYVeGi6DsQhCCasAlbIGrXagBiV0PwoekNPgtHGIoDo-jC6EXgitBXbylL3e792m5fdKufg5ZqRpgg0ta_ZGdko65yqo9yCmJedEo9-mOEP64wX3O5P-zqTfafJc-juT_qbmXh4OrMNM4T51UFeBVwcAMsI0JthgzPecMspZ01fufM9R1XEbKfmMcecsxFT9-bDEB175Byn9kPE</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>73243999</pqid></control><display><type>article</type><title>Wavelet based automatic seizure detection in intracerebral electroencephalogram</title><source>MEDLINE</source><source>ScienceDirect Journals (5 years ago - present)</source><creator>Khan, Y.U. ; Gotman, J.</creator><creatorcontrib>Khan, Y.U. ; Gotman, J.</creatorcontrib><description>Background: Automatic seizure detection is often used during long-term monitoring, and is particularly important during intracerebral investigations. Existing methods make many false detections, particularly in intracerebral electroencephalogram (EEG) because of frequent large amplitude rhythmic activity bursts that are non-epileptiform.
Objective: To develop a seizure detection method for intracerebral monitoring that is as sensitive as existing methods but has fewer false detections.
Methods: To capture the rhythmic nature of seizure discharges, we developed a wavelet-based method, examining how different frequency ranges fluctuate compared to the background. In particular, the system remembers rhythmic bursts occurring commonly in the background to avoid detecting them as seizures.
Results: The method was evaluated on test data from 11 patients, including 229 h and 66 seizures, and its performance compared to the method of Gotman (Electroencephalogr clin Neurophysiol 76 (1990) 317). Detection sensitivity was unchanged at close to 90%, but false detections were reduced from 2.4 to 0.3/h.
Conclusions: Perfect sensitivity is unlikely because the morphology of seizure discharges is so variable. Nevertheless, the 87% sensitivity obtained in the combined training and testing data is quite high. We reduced the average false alarm rate to one per 3 h of recording, or 6 per 24-h period. Given how rapidly one can decide visually that a detection is erroneous, false detections should not cause any burden to the reviewer.
Significance: In intracerebral EEG it is possible to detect seizures automatically with high sensitivity and high specificity.</description><identifier>ISSN: 1388-2457</identifier><identifier>EISSN: 1872-8952</identifier><identifier>DOI: 10.1016/S1388-2457(03)00035-X</identifier><identifier>PMID: 12738437</identifier><language>eng</language><publisher>Shannon: Elsevier Ireland Ltd</publisher><subject>Automatic seizure detection ; Biological and medical sciences ; Discrete wavelet transform ; Electrodiagnosis. Electric activity recording ; Electroencephalogram ; Electroencephalography - methods ; Epilepsy ; Humans ; Investigative techniques, diagnostic techniques (general aspects) ; Medical sciences ; Nervous system ; Seizures - diagnosis ; Seizures - physiopathology</subject><ispartof>Clinical neurophysiology, 2003-05, Vol.114 (5), p.898-908</ispartof><rights>2003 International Federation of Clinical Neurophysiology</rights><rights>2003 INIST-CNRS</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c391t-a63703d78ca14184ac4487c859a4b4156f10dc0eb9579cca420bf9dd611551c23</citedby><cites>FETCH-LOGICAL-c391t-a63703d78ca14184ac4487c859a4b4156f10dc0eb9579cca420bf9dd611551c23</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://dx.doi.org/10.1016/S1388-2457(03)00035-X$$EHTML$$P50$$Gelsevier$$H</linktohtml><link.rule.ids>314,780,784,3550,27924,27925,45995</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&idt=14749876$$DView record in Pascal Francis$$Hfree_for_read</backlink><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/12738437$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Khan, Y.U.</creatorcontrib><creatorcontrib>Gotman, J.</creatorcontrib><title>Wavelet based automatic seizure detection in intracerebral electroencephalogram</title><title>Clinical neurophysiology</title><addtitle>Clin Neurophysiol</addtitle><description>Background: Automatic seizure detection is often used during long-term monitoring, and is particularly important during intracerebral investigations. Existing methods make many false detections, particularly in intracerebral electroencephalogram (EEG) because of frequent large amplitude rhythmic activity bursts that are non-epileptiform.
Objective: To develop a seizure detection method for intracerebral monitoring that is as sensitive as existing methods but has fewer false detections.
Methods: To capture the rhythmic nature of seizure discharges, we developed a wavelet-based method, examining how different frequency ranges fluctuate compared to the background. In particular, the system remembers rhythmic bursts occurring commonly in the background to avoid detecting them as seizures.
Results: The method was evaluated on test data from 11 patients, including 229 h and 66 seizures, and its performance compared to the method of Gotman (Electroencephalogr clin Neurophysiol 76 (1990) 317). Detection sensitivity was unchanged at close to 90%, but false detections were reduced from 2.4 to 0.3/h.
Conclusions: Perfect sensitivity is unlikely because the morphology of seizure discharges is so variable. Nevertheless, the 87% sensitivity obtained in the combined training and testing data is quite high. We reduced the average false alarm rate to one per 3 h of recording, or 6 per 24-h period. Given how rapidly one can decide visually that a detection is erroneous, false detections should not cause any burden to the reviewer.
Significance: In intracerebral EEG it is possible to detect seizures automatically with high sensitivity and high specificity.</description><subject>Automatic seizure detection</subject><subject>Biological and medical sciences</subject><subject>Discrete wavelet transform</subject><subject>Electrodiagnosis. Electric activity recording</subject><subject>Electroencephalogram</subject><subject>Electroencephalography - methods</subject><subject>Epilepsy</subject><subject>Humans</subject><subject>Investigative techniques, diagnostic techniques (general aspects)</subject><subject>Medical sciences</subject><subject>Nervous system</subject><subject>Seizures - diagnosis</subject><subject>Seizures - physiopathology</subject><issn>1388-2457</issn><issn>1872-8952</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2003</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNqFkM9rFTEQx4NYbK3-CcpeFD1sTTbJJjkVKf4oFHpQsbcwO5nVyO7bZ5It6F9vXt-THgsDM4fPd2b4MPZC8DPBRf_ui5DWtp3S5g2XbznnUrc3j9iJsKZrrdPd4zr_R47Z05x_Vchw1T1hx6Iz0ippTtj1d7iliUozQKbQwFqWGUrEJlP8uyZqAhXCEpdNE3dVEiAlGhJMTc1hSQttkLY_YVp-JJifsaMRpkzPD_2Uffv44evF5_bq-tPlxfurFqUTpYVeGi6DsQhCCasAlbIGrXagBiV0PwoekNPgtHGIoDo-jC6EXgitBXbylL3e792m5fdKufg5ZqRpgg0ta_ZGdko65yqo9yCmJedEo9-mOEP64wX3O5P-zqTfafJc-juT_qbmXh4OrMNM4T51UFeBVwcAMsI0JthgzPecMspZ01fufM9R1XEbKfmMcecsxFT9-bDEB175Byn9kPE</recordid><startdate>20030501</startdate><enddate>20030501</enddate><creator>Khan, Y.U.</creator><creator>Gotman, J.</creator><general>Elsevier Ireland Ltd</general><general>Elsevier Science</general><scope>IQODW</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>20030501</creationdate><title>Wavelet based automatic seizure detection in intracerebral electroencephalogram</title><author>Khan, Y.U. ; Gotman, J.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c391t-a63703d78ca14184ac4487c859a4b4156f10dc0eb9579cca420bf9dd611551c23</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2003</creationdate><topic>Automatic seizure detection</topic><topic>Biological and medical sciences</topic><topic>Discrete wavelet transform</topic><topic>Electrodiagnosis. Electric activity recording</topic><topic>Electroencephalogram</topic><topic>Electroencephalography - methods</topic><topic>Epilepsy</topic><topic>Humans</topic><topic>Investigative techniques, diagnostic techniques (general aspects)</topic><topic>Medical sciences</topic><topic>Nervous system</topic><topic>Seizures - diagnosis</topic><topic>Seizures - physiopathology</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Khan, Y.U.</creatorcontrib><creatorcontrib>Gotman, J.</creatorcontrib><collection>Pascal-Francis</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Clinical neurophysiology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Khan, Y.U.</au><au>Gotman, J.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Wavelet based automatic seizure detection in intracerebral electroencephalogram</atitle><jtitle>Clinical neurophysiology</jtitle><addtitle>Clin Neurophysiol</addtitle><date>2003-05-01</date><risdate>2003</risdate><volume>114</volume><issue>5</issue><spage>898</spage><epage>908</epage><pages>898-908</pages><issn>1388-2457</issn><eissn>1872-8952</eissn><abstract>Background: Automatic seizure detection is often used during long-term monitoring, and is particularly important during intracerebral investigations. Existing methods make many false detections, particularly in intracerebral electroencephalogram (EEG) because of frequent large amplitude rhythmic activity bursts that are non-epileptiform.
Objective: To develop a seizure detection method for intracerebral monitoring that is as sensitive as existing methods but has fewer false detections.
Methods: To capture the rhythmic nature of seizure discharges, we developed a wavelet-based method, examining how different frequency ranges fluctuate compared to the background. In particular, the system remembers rhythmic bursts occurring commonly in the background to avoid detecting them as seizures.
Results: The method was evaluated on test data from 11 patients, including 229 h and 66 seizures, and its performance compared to the method of Gotman (Electroencephalogr clin Neurophysiol 76 (1990) 317). Detection sensitivity was unchanged at close to 90%, but false detections were reduced from 2.4 to 0.3/h.
Conclusions: Perfect sensitivity is unlikely because the morphology of seizure discharges is so variable. Nevertheless, the 87% sensitivity obtained in the combined training and testing data is quite high. We reduced the average false alarm rate to one per 3 h of recording, or 6 per 24-h period. Given how rapidly one can decide visually that a detection is erroneous, false detections should not cause any burden to the reviewer.
Significance: In intracerebral EEG it is possible to detect seizures automatically with high sensitivity and high specificity.</abstract><cop>Shannon</cop><pub>Elsevier Ireland Ltd</pub><pmid>12738437</pmid><doi>10.1016/S1388-2457(03)00035-X</doi><tpages>11</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1388-2457 |
ispartof | Clinical neurophysiology, 2003-05, Vol.114 (5), p.898-908 |
issn | 1388-2457 1872-8952 |
language | eng |
recordid | cdi_proquest_miscellaneous_73243999 |
source | MEDLINE; ScienceDirect Journals (5 years ago - present) |
subjects | Automatic seizure detection Biological and medical sciences Discrete wavelet transform Electrodiagnosis. Electric activity recording Electroencephalogram Electroencephalography - methods Epilepsy Humans Investigative techniques, diagnostic techniques (general aspects) Medical sciences Nervous system Seizures - diagnosis Seizures - physiopathology |
title | Wavelet based automatic seizure detection in intracerebral electroencephalogram |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-05T20%3A31%3A27IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Wavelet%20based%20automatic%20seizure%20detection%20in%20intracerebral%20electroencephalogram&rft.jtitle=Clinical%20neurophysiology&rft.au=Khan,%20Y.U.&rft.date=2003-05-01&rft.volume=114&rft.issue=5&rft.spage=898&rft.epage=908&rft.pages=898-908&rft.issn=1388-2457&rft.eissn=1872-8952&rft_id=info:doi/10.1016/S1388-2457(03)00035-X&rft_dat=%3Cproquest_cross%3E73243999%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=73243999&rft_id=info:pmid/12738437&rft_els_id=S138824570300035X&rfr_iscdi=true |