Suture-Endobutton Fixation of Ankle Tibio-Fibular Diastasis: A Cadaver Study
Suture-Endobutton fixation is proposed as a minimally invasive, flexible fixation of ankle tibio-fibular diastasis, which would not require routine removal. This study tested the Suture-Endobutton construct in a cadaver syndesmosis injury model and compared this against A.O. syndesmosis screw fixati...
Gespeichert in:
Veröffentlicht in: | Foot & ankle international 2003-02, Vol.24 (2), p.142-146 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Suture-Endobutton fixation is proposed as a minimally invasive, flexible fixation of ankle tibio-fibular diastasis, which would not require routine removal. This study tested the Suture-Endobutton construct in a cadaver syndesmosis injury model and compared this against A.O. syndesmosis screw fixation. Sixteen embalmed cadaver legs were used. Phase one consisted of placing the leg in a jig, generating an external rotation torque and measuring diastasis with increasing intraosseous membrane division. Phase two then compared the Suture-Endobutton construct vs. single four-cortex 4.5 mm A.O. screw fixation. Diastasis increased significantly with increasing intraosseous membrane division (p |
---|---|
ISSN: | 1071-1007 1944-7876 |
DOI: | 10.1177/107110070302400208 |