Comparison of 99mTc-MIBI uptakes on planar images with those in excised rats organs

The precision with which images reflect tracer uptake in the myocardium has been studied. Additionally, the degree to which Tc methoxyisobutylisonitrile (Tc-MIBI) in the liver gave the effect to a myocardial image has been examined. After administering Tc-MIBI to normal male rats, we compared the my...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Nuclear medicine communications 2003-01, Vol.24 (1), p.47-54
Hauptverfasser: ONOGUCHI, M, TAKAYAMA, T, MIYATI, T, TONAMI, N, SUMI, Y, MAEHARA, T
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 54
container_issue 1
container_start_page 47
container_title Nuclear medicine communications
container_volume 24
creator ONOGUCHI, M
TAKAYAMA, T
MIYATI, T
TONAMI, N
SUMI, Y
MAEHARA, T
description The precision with which images reflect tracer uptake in the myocardium has been studied. Additionally, the degree to which Tc methoxyisobutylisonitrile (Tc-MIBI) in the liver gave the effect to a myocardial image has been examined. After administering Tc-MIBI to normal male rats, we compared the myocardial uptakes obtained using a gamma camera with the actual uptakes in the excised organs. Twenty-nine rats were used. Following imaging, the anterior view at 5, 10, 15, 30, 45, 60, 90 and 120 min after administration of the tracer, uptakes in the heart, lung, liver and blood were estimated with a well-type scintillation counter (WC) and represented as the percentage of the injected dose per gram of tissue (%ID/g). The regions of interest (ROIs) were placed on planar images (PI) and the uptake in each organ was estimated as the percentage of the injected dose per pixel (%ID/pixel). The ratios of PI-to-WC and heart-to-organ were also evaluated. Cardiac uptake with WC was maximum (1.581%±1.893%) at 10 min post-injection. On the other hand, that with PI was maximum (1.493%±0.598%) at 45 min post-injection, but there were significant differences between both measurements (PI/WC ratioabout 1.0 time). Pulmonary uptake with WC was the maximum at 5 min (0.808%±0.015%) post-injection, and decreased gradually. PI measurement showed the maximum value at 45 min (0.760%±0.012%). Hepatic uptake with WC was the maximum at 30 min (0.594%±0.254%). On the other hand, PI measurement showed the same pattern with WC, but these values were higher value than WC as the whole. PI measurement showed higher uptakes in each organ than WC measurement. It was concluded that uptakes or the heart-to-organ ratio obtained clinically with PI might not represent a value that is always accurate.
doi_str_mv 10.1097/00006231-200301000-00014
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_72889264</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>72889264</sourcerecordid><originalsourceid>FETCH-LOGICAL-c3034-eae6616ddb86c74a2d2af4e686f2fe56c50f80275609551945c85391c66ee9523</originalsourceid><addsrcrecordid>eNp1kUtPAyEQgInRaK3-BcNFb6u8F47a-Gii8WA9E2Rn7eq-hN1U_71oq54kIYThm2H4QAhTckqJyc9IGopxmjFCOKFpl6VJxRaaUJHzTCqmt9EkRXjGFVd7aD_Gl4RorvJdtEeZTFnUTNDDrGt6F6rYtbgrsTHNwmd384s5HvvBvULE6aCvXesCrhr3nAKraljiYdlFwFWL4d1XEQoc3JDY8OzaeIB2SldHONysU_R4dbmY3WS399fz2flt5jnhIgMHSlFVFE9a-Vw4VjBXClBalawEqbwkpSYsl4oYKakR0mvJDfVKARjJ-BSdrOv2oXsbIQ62qaKHOnUL3RhtzrQ2TIkE6jXoQxdjgNL2IT0mfFhK7JdQ-yPU_gq130JT6tHmjvGpgeIvcWMwAccbwEXv6jK4Ngn544TkUqe-p0isuVVXDxDiaz2uINgluHpY2v8-lH8ChTmLJw</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>72889264</pqid></control><display><type>article</type><title>Comparison of 99mTc-MIBI uptakes on planar images with those in excised rats organs</title><source>MEDLINE</source><source>Journals@Ovid Complete</source><creator>ONOGUCHI, M ; TAKAYAMA, T ; MIYATI, T ; TONAMI, N ; SUMI, Y ; MAEHARA, T</creator><creatorcontrib>ONOGUCHI, M ; TAKAYAMA, T ; MIYATI, T ; TONAMI, N ; SUMI, Y ; MAEHARA, T</creatorcontrib><description>The precision with which images reflect tracer uptake in the myocardium has been studied. Additionally, the degree to which Tc methoxyisobutylisonitrile (Tc-MIBI) in the liver gave the effect to a myocardial image has been examined. After administering Tc-MIBI to normal male rats, we compared the myocardial uptakes obtained using a gamma camera with the actual uptakes in the excised organs. Twenty-nine rats were used. Following imaging, the anterior view at 5, 10, 15, 30, 45, 60, 90 and 120 min after administration of the tracer, uptakes in the heart, lung, liver and blood were estimated with a well-type scintillation counter (WC) and represented as the percentage of the injected dose per gram of tissue (%ID/g). The regions of interest (ROIs) were placed on planar images (PI) and the uptake in each organ was estimated as the percentage of the injected dose per pixel (%ID/pixel). The ratios of PI-to-WC and heart-to-organ were also evaluated. Cardiac uptake with WC was maximum (1.581%±1.893%) at 10 min post-injection. On the other hand, that with PI was maximum (1.493%±0.598%) at 45 min post-injection, but there were significant differences between both measurements (PI/WC ratioabout 1.0 time). Pulmonary uptake with WC was the maximum at 5 min (0.808%±0.015%) post-injection, and decreased gradually. PI measurement showed the maximum value at 45 min (0.760%±0.012%). Hepatic uptake with WC was the maximum at 30 min (0.594%±0.254%). On the other hand, PI measurement showed the same pattern with WC, but these values were higher value than WC as the whole. PI measurement showed higher uptakes in each organ than WC measurement. It was concluded that uptakes or the heart-to-organ ratio obtained clinically with PI might not represent a value that is always accurate.</description><identifier>ISSN: 0143-3636</identifier><identifier>EISSN: 1473-5628</identifier><identifier>DOI: 10.1097/00006231-200301000-00014</identifier><identifier>PMID: 12501019</identifier><language>eng</language><publisher>Hagerstown, MD: Lippincott Williams &amp; Wilkins, Inc</publisher><subject>Animals ; Biological and medical sciences ; Heart - diagnostic imaging ; Investigative techniques, diagnostic techniques (general aspects) ; Liver - diagnostic imaging ; Liver - metabolism ; Lung - diagnostic imaging ; Lung - metabolism ; Male ; Medical sciences ; Miscellaneous. Technology ; Myocardium - metabolism ; Organ Specificity ; Radionuclide investigations ; Radiopharmaceuticals - blood ; Radiopharmaceuticals - pharmacokinetics ; Rats ; Reproducibility of Results ; Scintillation Counting - methods ; Sensitivity and Specificity ; Technetium Tc 99m Sestamibi - blood ; Technetium Tc 99m Sestamibi - pharmacokinetics ; Tissue Distribution ; Tomography, Emission-Computed, Single-Photon - methods</subject><ispartof>Nuclear medicine communications, 2003-01, Vol.24 (1), p.47-54</ispartof><rights>2003 Lippincott Williams &amp; Wilkins, Inc.</rights><rights>2003 INIST-CNRS</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c3034-eae6616ddb86c74a2d2af4e686f2fe56c50f80275609551945c85391c66ee9523</citedby><cites>FETCH-LOGICAL-c3034-eae6616ddb86c74a2d2af4e686f2fe56c50f80275609551945c85391c66ee9523</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,778,782,4012,27906,27907,27908</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&amp;idt=14535885$$DView record in Pascal Francis$$Hfree_for_read</backlink><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/12501019$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>ONOGUCHI, M</creatorcontrib><creatorcontrib>TAKAYAMA, T</creatorcontrib><creatorcontrib>MIYATI, T</creatorcontrib><creatorcontrib>TONAMI, N</creatorcontrib><creatorcontrib>SUMI, Y</creatorcontrib><creatorcontrib>MAEHARA, T</creatorcontrib><title>Comparison of 99mTc-MIBI uptakes on planar images with those in excised rats organs</title><title>Nuclear medicine communications</title><addtitle>Nucl Med Commun</addtitle><description>The precision with which images reflect tracer uptake in the myocardium has been studied. Additionally, the degree to which Tc methoxyisobutylisonitrile (Tc-MIBI) in the liver gave the effect to a myocardial image has been examined. After administering Tc-MIBI to normal male rats, we compared the myocardial uptakes obtained using a gamma camera with the actual uptakes in the excised organs. Twenty-nine rats were used. Following imaging, the anterior view at 5, 10, 15, 30, 45, 60, 90 and 120 min after administration of the tracer, uptakes in the heart, lung, liver and blood were estimated with a well-type scintillation counter (WC) and represented as the percentage of the injected dose per gram of tissue (%ID/g). The regions of interest (ROIs) were placed on planar images (PI) and the uptake in each organ was estimated as the percentage of the injected dose per pixel (%ID/pixel). The ratios of PI-to-WC and heart-to-organ were also evaluated. Cardiac uptake with WC was maximum (1.581%±1.893%) at 10 min post-injection. On the other hand, that with PI was maximum (1.493%±0.598%) at 45 min post-injection, but there were significant differences between both measurements (PI/WC ratioabout 1.0 time). Pulmonary uptake with WC was the maximum at 5 min (0.808%±0.015%) post-injection, and decreased gradually. PI measurement showed the maximum value at 45 min (0.760%±0.012%). Hepatic uptake with WC was the maximum at 30 min (0.594%±0.254%). On the other hand, PI measurement showed the same pattern with WC, but these values were higher value than WC as the whole. PI measurement showed higher uptakes in each organ than WC measurement. It was concluded that uptakes or the heart-to-organ ratio obtained clinically with PI might not represent a value that is always accurate.</description><subject>Animals</subject><subject>Biological and medical sciences</subject><subject>Heart - diagnostic imaging</subject><subject>Investigative techniques, diagnostic techniques (general aspects)</subject><subject>Liver - diagnostic imaging</subject><subject>Liver - metabolism</subject><subject>Lung - diagnostic imaging</subject><subject>Lung - metabolism</subject><subject>Male</subject><subject>Medical sciences</subject><subject>Miscellaneous. Technology</subject><subject>Myocardium - metabolism</subject><subject>Organ Specificity</subject><subject>Radionuclide investigations</subject><subject>Radiopharmaceuticals - blood</subject><subject>Radiopharmaceuticals - pharmacokinetics</subject><subject>Rats</subject><subject>Reproducibility of Results</subject><subject>Scintillation Counting - methods</subject><subject>Sensitivity and Specificity</subject><subject>Technetium Tc 99m Sestamibi - blood</subject><subject>Technetium Tc 99m Sestamibi - pharmacokinetics</subject><subject>Tissue Distribution</subject><subject>Tomography, Emission-Computed, Single-Photon - methods</subject><issn>0143-3636</issn><issn>1473-5628</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2003</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp1kUtPAyEQgInRaK3-BcNFb6u8F47a-Gii8WA9E2Rn7eq-hN1U_71oq54kIYThm2H4QAhTckqJyc9IGopxmjFCOKFpl6VJxRaaUJHzTCqmt9EkRXjGFVd7aD_Gl4RorvJdtEeZTFnUTNDDrGt6F6rYtbgrsTHNwmd384s5HvvBvULE6aCvXesCrhr3nAKraljiYdlFwFWL4d1XEQoc3JDY8OzaeIB2SldHONysU_R4dbmY3WS399fz2flt5jnhIgMHSlFVFE9a-Vw4VjBXClBalawEqbwkpSYsl4oYKakR0mvJDfVKARjJ-BSdrOv2oXsbIQ62qaKHOnUL3RhtzrQ2TIkE6jXoQxdjgNL2IT0mfFhK7JdQ-yPU_gq130JT6tHmjvGpgeIvcWMwAccbwEXv6jK4Ngn544TkUqe-p0isuVVXDxDiaz2uINgluHpY2v8-lH8ChTmLJw</recordid><startdate>200301</startdate><enddate>200301</enddate><creator>ONOGUCHI, M</creator><creator>TAKAYAMA, T</creator><creator>MIYATI, T</creator><creator>TONAMI, N</creator><creator>SUMI, Y</creator><creator>MAEHARA, T</creator><general>Lippincott Williams &amp; Wilkins, Inc</general><general>Lippincott Williams &amp; Wilkins</general><scope>IQODW</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>200301</creationdate><title>Comparison of 99mTc-MIBI uptakes on planar images with those in excised rats organs</title><author>ONOGUCHI, M ; TAKAYAMA, T ; MIYATI, T ; TONAMI, N ; SUMI, Y ; MAEHARA, T</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c3034-eae6616ddb86c74a2d2af4e686f2fe56c50f80275609551945c85391c66ee9523</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2003</creationdate><topic>Animals</topic><topic>Biological and medical sciences</topic><topic>Heart - diagnostic imaging</topic><topic>Investigative techniques, diagnostic techniques (general aspects)</topic><topic>Liver - diagnostic imaging</topic><topic>Liver - metabolism</topic><topic>Lung - diagnostic imaging</topic><topic>Lung - metabolism</topic><topic>Male</topic><topic>Medical sciences</topic><topic>Miscellaneous. Technology</topic><topic>Myocardium - metabolism</topic><topic>Organ Specificity</topic><topic>Radionuclide investigations</topic><topic>Radiopharmaceuticals - blood</topic><topic>Radiopharmaceuticals - pharmacokinetics</topic><topic>Rats</topic><topic>Reproducibility of Results</topic><topic>Scintillation Counting - methods</topic><topic>Sensitivity and Specificity</topic><topic>Technetium Tc 99m Sestamibi - blood</topic><topic>Technetium Tc 99m Sestamibi - pharmacokinetics</topic><topic>Tissue Distribution</topic><topic>Tomography, Emission-Computed, Single-Photon - methods</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>ONOGUCHI, M</creatorcontrib><creatorcontrib>TAKAYAMA, T</creatorcontrib><creatorcontrib>MIYATI, T</creatorcontrib><creatorcontrib>TONAMI, N</creatorcontrib><creatorcontrib>SUMI, Y</creatorcontrib><creatorcontrib>MAEHARA, T</creatorcontrib><collection>Pascal-Francis</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Nuclear medicine communications</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>ONOGUCHI, M</au><au>TAKAYAMA, T</au><au>MIYATI, T</au><au>TONAMI, N</au><au>SUMI, Y</au><au>MAEHARA, T</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Comparison of 99mTc-MIBI uptakes on planar images with those in excised rats organs</atitle><jtitle>Nuclear medicine communications</jtitle><addtitle>Nucl Med Commun</addtitle><date>2003-01</date><risdate>2003</risdate><volume>24</volume><issue>1</issue><spage>47</spage><epage>54</epage><pages>47-54</pages><issn>0143-3636</issn><eissn>1473-5628</eissn><abstract>The precision with which images reflect tracer uptake in the myocardium has been studied. Additionally, the degree to which Tc methoxyisobutylisonitrile (Tc-MIBI) in the liver gave the effect to a myocardial image has been examined. After administering Tc-MIBI to normal male rats, we compared the myocardial uptakes obtained using a gamma camera with the actual uptakes in the excised organs. Twenty-nine rats were used. Following imaging, the anterior view at 5, 10, 15, 30, 45, 60, 90 and 120 min after administration of the tracer, uptakes in the heart, lung, liver and blood were estimated with a well-type scintillation counter (WC) and represented as the percentage of the injected dose per gram of tissue (%ID/g). The regions of interest (ROIs) were placed on planar images (PI) and the uptake in each organ was estimated as the percentage of the injected dose per pixel (%ID/pixel). The ratios of PI-to-WC and heart-to-organ were also evaluated. Cardiac uptake with WC was maximum (1.581%±1.893%) at 10 min post-injection. On the other hand, that with PI was maximum (1.493%±0.598%) at 45 min post-injection, but there were significant differences between both measurements (PI/WC ratioabout 1.0 time). Pulmonary uptake with WC was the maximum at 5 min (0.808%±0.015%) post-injection, and decreased gradually. PI measurement showed the maximum value at 45 min (0.760%±0.012%). Hepatic uptake with WC was the maximum at 30 min (0.594%±0.254%). On the other hand, PI measurement showed the same pattern with WC, but these values were higher value than WC as the whole. PI measurement showed higher uptakes in each organ than WC measurement. It was concluded that uptakes or the heart-to-organ ratio obtained clinically with PI might not represent a value that is always accurate.</abstract><cop>Hagerstown, MD</cop><pub>Lippincott Williams &amp; Wilkins, Inc</pub><pmid>12501019</pmid><doi>10.1097/00006231-200301000-00014</doi><tpages>8</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0143-3636
ispartof Nuclear medicine communications, 2003-01, Vol.24 (1), p.47-54
issn 0143-3636
1473-5628
language eng
recordid cdi_proquest_miscellaneous_72889264
source MEDLINE; Journals@Ovid Complete
subjects Animals
Biological and medical sciences
Heart - diagnostic imaging
Investigative techniques, diagnostic techniques (general aspects)
Liver - diagnostic imaging
Liver - metabolism
Lung - diagnostic imaging
Lung - metabolism
Male
Medical sciences
Miscellaneous. Technology
Myocardium - metabolism
Organ Specificity
Radionuclide investigations
Radiopharmaceuticals - blood
Radiopharmaceuticals - pharmacokinetics
Rats
Reproducibility of Results
Scintillation Counting - methods
Sensitivity and Specificity
Technetium Tc 99m Sestamibi - blood
Technetium Tc 99m Sestamibi - pharmacokinetics
Tissue Distribution
Tomography, Emission-Computed, Single-Photon - methods
title Comparison of 99mTc-MIBI uptakes on planar images with those in excised rats organs
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-16T06%3A39%3A33IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Comparison%20of%2099mTc-MIBI%20uptakes%20on%20planar%20images%20with%20those%20in%20excised%20rats%20organs&rft.jtitle=Nuclear%20medicine%20communications&rft.au=ONOGUCHI,%20M&rft.date=2003-01&rft.volume=24&rft.issue=1&rft.spage=47&rft.epage=54&rft.pages=47-54&rft.issn=0143-3636&rft.eissn=1473-5628&rft_id=info:doi/10.1097/00006231-200301000-00014&rft_dat=%3Cproquest_cross%3E72889264%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=72889264&rft_id=info:pmid/12501019&rfr_iscdi=true