Evaluation of a soft-handling computerized pneumatic tube specimen delivery system : effects on analytical results and turnaround time
A computerized pneumatic tube specimen delivery system with system-wide air cushion soft handling features was evaluated. There were no significant differences in values (largely normal) for components of a standard chemical profile or complete blood count in specimens delivered from the outpatient...
Gespeichert in:
Veröffentlicht in: | American journal of clinical pathology 1992-04, Vol.97 (4), p.535-540 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 540 |
---|---|
container_issue | 4 |
container_start_page | 535 |
container_title | American journal of clinical pathology |
container_volume | 97 |
creator | KESHGEGIAN, A. A BULL, G. E |
description | A computerized pneumatic tube specimen delivery system with system-wide air cushion soft handling features was evaluated. There were no significant differences in values (largely normal) for components of a standard chemical profile or complete blood count in specimens delivered from the outpatient center or neonatal intensive care unit by pneumatic tube compared to couriers. The pneumatic tube system also did not affect values for pO2, pCO2, and pH over a wide range (pO2, 25 to 438 mmHg) in specimens sent from the operating room during cardiac surgery. The pneumatic tube system decreased the median turnaround time for potassium and hemoglobin results on specimens from the emergency department by 25%. The system evaluated is a rapid, efficient mechanism for sending specimens to the clinical laboratory that produces no significant effects on analytical results and has the ability to decrease turnaround time. |
doi_str_mv | 10.1093/ajcp/97.4.535 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_72871005</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>72871005</sourcerecordid><originalsourceid>FETCH-LOGICAL-c317t-f1b738d5220298d62d6e2736efd543b43b484856b85f022fe88a526077f2041f3</originalsourceid><addsrcrecordid>eNpFkE1LHTEUQENp0VfbpUshi-JunvmcZNyJaFsQumnXIZO50ZFMZkwmwvMH9Hebx3u0EEi49-QsDkLnlGwp6fiVfXbLVae2Yiu5_IA2tBO8UYqxj2hDCGFNRxU_RZ9zfiaEMk3ECTqhUvKOdhv09-7VhmLXcY549tjiPPu1ebJxCGN8xG6elrJCGt9gwEuEMlXU4bX0gPMCbpwg4gHC-Apph_MurzDhawzeg1szrlIbbdjVPzbgBLmEOq3yakjRprnsn1XyBX3yNmT4erzP0J_7u9-3P5qHX99_3t48NI5TtTae9orrQTJGWKeHlg0tMMVb8IMUvN8fLbRsey09YcyD1laylijlGRHU8zN0efAuaX4pkFczjdlBCDbCXLJRTCtKiKxgcwBdmnNO4M2SxsmmnaHE7LubfXfTKSNM7V75i6O49BMM_-lD6Lr_dtzbXFv4ZKMb8z9MVEVLBX8H73-Njg</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>72871005</pqid></control><display><type>article</type><title>Evaluation of a soft-handling computerized pneumatic tube specimen delivery system : effects on analytical results and turnaround time</title><source>MEDLINE</source><source>Oxford University Press Journals Digital Archive Legacy</source><creator>KESHGEGIAN, A. A ; BULL, G. E</creator><creatorcontrib>KESHGEGIAN, A. A ; BULL, G. E</creatorcontrib><description>A computerized pneumatic tube specimen delivery system with system-wide air cushion soft handling features was evaluated. There were no significant differences in values (largely normal) for components of a standard chemical profile or complete blood count in specimens delivered from the outpatient center or neonatal intensive care unit by pneumatic tube compared to couriers. The pneumatic tube system also did not affect values for pO2, pCO2, and pH over a wide range (pO2, 25 to 438 mmHg) in specimens sent from the operating room during cardiac surgery. The pneumatic tube system decreased the median turnaround time for potassium and hemoglobin results on specimens from the emergency department by 25%. The system evaluated is a rapid, efficient mechanism for sending specimens to the clinical laboratory that produces no significant effects on analytical results and has the ability to decrease turnaround time.</description><identifier>ISSN: 0002-9173</identifier><identifier>EISSN: 1943-7722</identifier><identifier>DOI: 10.1093/ajcp/97.4.535</identifier><identifier>PMID: 1553919</identifier><identifier>CODEN: AJCPAI</identifier><language>eng</language><publisher>Chicago, IL: American Society of Clinical Pathologists</publisher><subject>Biological and medical sciences ; Blood Cell Count ; Blood Chemical Analysis ; Blood Gas Analysis ; Evaluation Studies as Topic ; Hospital Distribution Systems ; Humans ; Investigative techniques, diagnostic techniques (general aspects) ; Medical sciences ; Miscellaneous. Technology ; Pathology. Cytology. Biochemistry. Spectrometry. Miscellaneous investigative techniques ; Specimen Handling - instrumentation ; Specimen Handling - methods ; Time Factors</subject><ispartof>American journal of clinical pathology, 1992-04, Vol.97 (4), p.535-540</ispartof><rights>1993 INIST-CNRS</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c317t-f1b738d5220298d62d6e2736efd543b43b484856b85f022fe88a526077f2041f3</citedby></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,777,781,27905,27906</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&idt=4353614$$DView record in Pascal Francis$$Hfree_for_read</backlink><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/1553919$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>KESHGEGIAN, A. A</creatorcontrib><creatorcontrib>BULL, G. E</creatorcontrib><title>Evaluation of a soft-handling computerized pneumatic tube specimen delivery system : effects on analytical results and turnaround time</title><title>American journal of clinical pathology</title><addtitle>Am J Clin Pathol</addtitle><description>A computerized pneumatic tube specimen delivery system with system-wide air cushion soft handling features was evaluated. There were no significant differences in values (largely normal) for components of a standard chemical profile or complete blood count in specimens delivered from the outpatient center or neonatal intensive care unit by pneumatic tube compared to couriers. The pneumatic tube system also did not affect values for pO2, pCO2, and pH over a wide range (pO2, 25 to 438 mmHg) in specimens sent from the operating room during cardiac surgery. The pneumatic tube system decreased the median turnaround time for potassium and hemoglobin results on specimens from the emergency department by 25%. The system evaluated is a rapid, efficient mechanism for sending specimens to the clinical laboratory that produces no significant effects on analytical results and has the ability to decrease turnaround time.</description><subject>Biological and medical sciences</subject><subject>Blood Cell Count</subject><subject>Blood Chemical Analysis</subject><subject>Blood Gas Analysis</subject><subject>Evaluation Studies as Topic</subject><subject>Hospital Distribution Systems</subject><subject>Humans</subject><subject>Investigative techniques, diagnostic techniques (general aspects)</subject><subject>Medical sciences</subject><subject>Miscellaneous. Technology</subject><subject>Pathology. Cytology. Biochemistry. Spectrometry. Miscellaneous investigative techniques</subject><subject>Specimen Handling - instrumentation</subject><subject>Specimen Handling - methods</subject><subject>Time Factors</subject><issn>0002-9173</issn><issn>1943-7722</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>1992</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNpFkE1LHTEUQENp0VfbpUshi-JunvmcZNyJaFsQumnXIZO50ZFMZkwmwvMH9Hebx3u0EEi49-QsDkLnlGwp6fiVfXbLVae2Yiu5_IA2tBO8UYqxj2hDCGFNRxU_RZ9zfiaEMk3ECTqhUvKOdhv09-7VhmLXcY549tjiPPu1ebJxCGN8xG6elrJCGt9gwEuEMlXU4bX0gPMCbpwg4gHC-Apph_MurzDhawzeg1szrlIbbdjVPzbgBLmEOq3yakjRprnsn1XyBX3yNmT4erzP0J_7u9-3P5qHX99_3t48NI5TtTae9orrQTJGWKeHlg0tMMVb8IMUvN8fLbRsey09YcyD1laylijlGRHU8zN0efAuaX4pkFczjdlBCDbCXLJRTCtKiKxgcwBdmnNO4M2SxsmmnaHE7LubfXfTKSNM7V75i6O49BMM_-lD6Lr_dtzbXFv4ZKMb8z9MVEVLBX8H73-Njg</recordid><startdate>19920401</startdate><enddate>19920401</enddate><creator>KESHGEGIAN, A. A</creator><creator>BULL, G. E</creator><general>American Society of Clinical Pathologists</general><scope>IQODW</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>19920401</creationdate><title>Evaluation of a soft-handling computerized pneumatic tube specimen delivery system : effects on analytical results and turnaround time</title><author>KESHGEGIAN, A. A ; BULL, G. E</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c317t-f1b738d5220298d62d6e2736efd543b43b484856b85f022fe88a526077f2041f3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>1992</creationdate><topic>Biological and medical sciences</topic><topic>Blood Cell Count</topic><topic>Blood Chemical Analysis</topic><topic>Blood Gas Analysis</topic><topic>Evaluation Studies as Topic</topic><topic>Hospital Distribution Systems</topic><topic>Humans</topic><topic>Investigative techniques, diagnostic techniques (general aspects)</topic><topic>Medical sciences</topic><topic>Miscellaneous. Technology</topic><topic>Pathology. Cytology. Biochemistry. Spectrometry. Miscellaneous investigative techniques</topic><topic>Specimen Handling - instrumentation</topic><topic>Specimen Handling - methods</topic><topic>Time Factors</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>KESHGEGIAN, A. A</creatorcontrib><creatorcontrib>BULL, G. E</creatorcontrib><collection>Pascal-Francis</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>American journal of clinical pathology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>KESHGEGIAN, A. A</au><au>BULL, G. E</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Evaluation of a soft-handling computerized pneumatic tube specimen delivery system : effects on analytical results and turnaround time</atitle><jtitle>American journal of clinical pathology</jtitle><addtitle>Am J Clin Pathol</addtitle><date>1992-04-01</date><risdate>1992</risdate><volume>97</volume><issue>4</issue><spage>535</spage><epage>540</epage><pages>535-540</pages><issn>0002-9173</issn><eissn>1943-7722</eissn><coden>AJCPAI</coden><abstract>A computerized pneumatic tube specimen delivery system with system-wide air cushion soft handling features was evaluated. There were no significant differences in values (largely normal) for components of a standard chemical profile or complete blood count in specimens delivered from the outpatient center or neonatal intensive care unit by pneumatic tube compared to couriers. The pneumatic tube system also did not affect values for pO2, pCO2, and pH over a wide range (pO2, 25 to 438 mmHg) in specimens sent from the operating room during cardiac surgery. The pneumatic tube system decreased the median turnaround time for potassium and hemoglobin results on specimens from the emergency department by 25%. The system evaluated is a rapid, efficient mechanism for sending specimens to the clinical laboratory that produces no significant effects on analytical results and has the ability to decrease turnaround time.</abstract><cop>Chicago, IL</cop><pub>American Society of Clinical Pathologists</pub><pmid>1553919</pmid><doi>10.1093/ajcp/97.4.535</doi><tpages>6</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0002-9173 |
ispartof | American journal of clinical pathology, 1992-04, Vol.97 (4), p.535-540 |
issn | 0002-9173 1943-7722 |
language | eng |
recordid | cdi_proquest_miscellaneous_72871005 |
source | MEDLINE; Oxford University Press Journals Digital Archive Legacy |
subjects | Biological and medical sciences Blood Cell Count Blood Chemical Analysis Blood Gas Analysis Evaluation Studies as Topic Hospital Distribution Systems Humans Investigative techniques, diagnostic techniques (general aspects) Medical sciences Miscellaneous. Technology Pathology. Cytology. Biochemistry. Spectrometry. Miscellaneous investigative techniques Specimen Handling - instrumentation Specimen Handling - methods Time Factors |
title | Evaluation of a soft-handling computerized pneumatic tube specimen delivery system : effects on analytical results and turnaround time |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-19T22%3A02%3A47IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Evaluation%20of%20a%20soft-handling%20computerized%20pneumatic%20tube%20specimen%20delivery%20system%20:%20effects%20on%20analytical%20results%20and%20turnaround%20time&rft.jtitle=American%20journal%20of%20clinical%20pathology&rft.au=KESHGEGIAN,%20A.%20A&rft.date=1992-04-01&rft.volume=97&rft.issue=4&rft.spage=535&rft.epage=540&rft.pages=535-540&rft.issn=0002-9173&rft.eissn=1943-7722&rft.coden=AJCPAI&rft_id=info:doi/10.1093/ajcp/97.4.535&rft_dat=%3Cproquest_cross%3E72871005%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=72871005&rft_id=info:pmid/1553919&rfr_iscdi=true |