Wave analysis of implant screw loosening using an air cylindrical cyclic loading device

Statement of Problem. The mechanics of implant screw loosening or fracture are well understood in the field of engineering. They have not been as widely explored in dentistry. Purpose. This study investigated the effects of simulated mastication on implant components and used wave analysis to docume...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of prosthetic dentistry 2002-10, Vol.88 (4), p.402-408
Hauptverfasser: Lee, Juneseok, Kim, Yung-Soo, Kim, Chang-Whe, Han, Jung-Suk
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 408
container_issue 4
container_start_page 402
container_title The Journal of prosthetic dentistry
container_volume 88
creator Lee, Juneseok
Kim, Yung-Soo
Kim, Chang-Whe
Han, Jung-Suk
description Statement of Problem. The mechanics of implant screw loosening or fracture are well understood in the field of engineering. They have not been as widely explored in dentistry. Purpose. This study investigated the effects of simulated mastication on implant components and used wave analysis to document the basic mechanisms of screw loosening in a simulated oral environment. Material and Methods. A pneumatic cylindrical cyclic loading device was fabricated to simulate masticatory movement. Thirteen standard abutments were connected on external hexagonal implants with titanium abutment screws tightened to 20 Newton centimeters (Ncm), and single crowns were retained with gold screws tightened to 10 Ncm on each abutment, respectively. Ten single-implant crowns were loaded with the use of a cyclic loading device with 100 N of force at 30° angles to the long-axis for 0.2 seconds of contact time with a frequency of 1 Hz. Three crowns were loaded vertically under the same conditions to serve as the control group. The effects of up to 1 million cyclic loads and various tightening torque forces (2, 4, 6, 8, 10, and 12 Ncm) on screw loosening were evaluated by wave analysis. A software program was written to record every wave mode and to stop the machine automatically if the amount of horizontal displacement of the crown was more than 0.5 mm, which was designated to represent perceptible loosened implant crown mobility clinically. The general wave patterns and characteristics of loosened and stable screws and the effect of various tightening torques were analyzed by comparing the differences in wave patterns. Results. The wave mode was divided into 4 stages for loosened gold screws: initial displacement, initial vibration, elastic deformation, and recovery stage. However, the initial displacement and initial vibration stages were not discernible for stable gold screws. Of the 10 gold occlusal screws, 4 loosened before the 1 million cyclic loads in the 10 single crowns tested. There was no screw loosening in the control group. There was no effect of screw loosening on the elastic deformation stage. Conclusion. Within the limitations of this study, tightening torque had a significant effect on screw loosening. It would appear that more than 10 Ncm of tightening torque should be recommended for the gold screws in this external hexagon implant system. (J Prosthet Dent 2002;88:402-8.)
doi_str_mv 10.1067/mpr.2002.128099
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_72722617</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S0022391302002329</els_id><sourcerecordid>72722617</sourcerecordid><originalsourceid>FETCH-LOGICAL-c440t-5fa8b1d542efb50771294375bcc8730e82fcdf9d0f8b3ec43f7957c56d8eb0c3</originalsourceid><addsrcrecordid>eNqFkMtLAzEQh4MoWh9nb7Inb1uT7GOSo4gvELwIHkN2MpHIPmrSVvrfm9KCJ_EyMzDf_GA-xi4Fnwvews2wiHPJuZwLqbjWB2wmuIayVbU4ZLO8kGWlRXXCTlP65JyrBsQxOxGyrkEKmLH3d7umwo6236SQiskXYVj0dlwWCSN9F_00JRrD-FGs0rbasbAhFrjpw-hiQNvnGfuAmbRuSzhaB6RzduRtn-hi38_Y28P9291T-fL6-Hx3-1JiXfNl2XirOuGaWpLvGg4gpK4raDpEBRUnJT06rx33qqsI68qDbgCb1inqOFZn7HoXu4jT14rS0gwhIfX5A5pWyYAEKVsB_4ISQEMrVQZvdiDGKaVI3ixiGGzcGMHN1rnJzs3Wudk5zxdX--hVN5D75feSM6B3AGUR60DRJAw0IrkQCZfGTeHP8B-vfJEt</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>27797628</pqid></control><display><type>article</type><title>Wave analysis of implant screw loosening using an air cylindrical cyclic loading device</title><source>MEDLINE</source><source>ScienceDirect Journals (5 years ago - present)</source><creator>Lee, Juneseok ; Kim, Yung-Soo ; Kim, Chang-Whe ; Han, Jung-Suk</creator><creatorcontrib>Lee, Juneseok ; Kim, Yung-Soo ; Kim, Chang-Whe ; Han, Jung-Suk</creatorcontrib><description>Statement of Problem. The mechanics of implant screw loosening or fracture are well understood in the field of engineering. They have not been as widely explored in dentistry. Purpose. This study investigated the effects of simulated mastication on implant components and used wave analysis to document the basic mechanisms of screw loosening in a simulated oral environment. Material and Methods. A pneumatic cylindrical cyclic loading device was fabricated to simulate masticatory movement. Thirteen standard abutments were connected on external hexagonal implants with titanium abutment screws tightened to 20 Newton centimeters (Ncm), and single crowns were retained with gold screws tightened to 10 Ncm on each abutment, respectively. Ten single-implant crowns were loaded with the use of a cyclic loading device with 100 N of force at 30° angles to the long-axis for 0.2 seconds of contact time with a frequency of 1 Hz. Three crowns were loaded vertically under the same conditions to serve as the control group. The effects of up to 1 million cyclic loads and various tightening torque forces (2, 4, 6, 8, 10, and 12 Ncm) on screw loosening were evaluated by wave analysis. A software program was written to record every wave mode and to stop the machine automatically if the amount of horizontal displacement of the crown was more than 0.5 mm, which was designated to represent perceptible loosened implant crown mobility clinically. The general wave patterns and characteristics of loosened and stable screws and the effect of various tightening torques were analyzed by comparing the differences in wave patterns. Results. The wave mode was divided into 4 stages for loosened gold screws: initial displacement, initial vibration, elastic deformation, and recovery stage. However, the initial displacement and initial vibration stages were not discernible for stable gold screws. Of the 10 gold occlusal screws, 4 loosened before the 1 million cyclic loads in the 10 single crowns tested. There was no screw loosening in the control group. There was no effect of screw loosening on the elastic deformation stage. Conclusion. Within the limitations of this study, tightening torque had a significant effect on screw loosening. It would appear that more than 10 Ncm of tightening torque should be recommended for the gold screws in this external hexagon implant system. (J Prosthet Dent 2002;88:402-8.)</description><identifier>ISSN: 0022-3913</identifier><identifier>EISSN: 1097-6841</identifier><identifier>DOI: 10.1067/mpr.2002.128099</identifier><identifier>PMID: 12447217</identifier><language>eng</language><publisher>United States: Elsevier Inc</publisher><subject>Crowns ; Dental Abutments ; Dental Implants ; Dental Prosthesis Retention - instrumentation ; Dental Prosthesis, Implant-Supported ; Dental Restoration Failure ; Dental Stress Analysis - methods ; Dentistry ; Electric Capacitance ; Gold Alloys ; Signal Processing, Computer-Assisted ; Torque ; Vibration</subject><ispartof>The Journal of prosthetic dentistry, 2002-10, Vol.88 (4), p.402-408</ispartof><rights>2002 Editorial Council of The Journal of Prosthetic Dentistry</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c440t-5fa8b1d542efb50771294375bcc8730e82fcdf9d0f8b3ec43f7957c56d8eb0c3</citedby><cites>FETCH-LOGICAL-c440t-5fa8b1d542efb50771294375bcc8730e82fcdf9d0f8b3ec43f7957c56d8eb0c3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://dx.doi.org/10.1067/mpr.2002.128099$$EHTML$$P50$$Gelsevier$$H</linktohtml><link.rule.ids>314,780,784,3550,27924,27925,45995</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/12447217$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Lee, Juneseok</creatorcontrib><creatorcontrib>Kim, Yung-Soo</creatorcontrib><creatorcontrib>Kim, Chang-Whe</creatorcontrib><creatorcontrib>Han, Jung-Suk</creatorcontrib><title>Wave analysis of implant screw loosening using an air cylindrical cyclic loading device</title><title>The Journal of prosthetic dentistry</title><addtitle>J Prosthet Dent</addtitle><description>Statement of Problem. The mechanics of implant screw loosening or fracture are well understood in the field of engineering. They have not been as widely explored in dentistry. Purpose. This study investigated the effects of simulated mastication on implant components and used wave analysis to document the basic mechanisms of screw loosening in a simulated oral environment. Material and Methods. A pneumatic cylindrical cyclic loading device was fabricated to simulate masticatory movement. Thirteen standard abutments were connected on external hexagonal implants with titanium abutment screws tightened to 20 Newton centimeters (Ncm), and single crowns were retained with gold screws tightened to 10 Ncm on each abutment, respectively. Ten single-implant crowns were loaded with the use of a cyclic loading device with 100 N of force at 30° angles to the long-axis for 0.2 seconds of contact time with a frequency of 1 Hz. Three crowns were loaded vertically under the same conditions to serve as the control group. The effects of up to 1 million cyclic loads and various tightening torque forces (2, 4, 6, 8, 10, and 12 Ncm) on screw loosening were evaluated by wave analysis. A software program was written to record every wave mode and to stop the machine automatically if the amount of horizontal displacement of the crown was more than 0.5 mm, which was designated to represent perceptible loosened implant crown mobility clinically. The general wave patterns and characteristics of loosened and stable screws and the effect of various tightening torques were analyzed by comparing the differences in wave patterns. Results. The wave mode was divided into 4 stages for loosened gold screws: initial displacement, initial vibration, elastic deformation, and recovery stage. However, the initial displacement and initial vibration stages were not discernible for stable gold screws. Of the 10 gold occlusal screws, 4 loosened before the 1 million cyclic loads in the 10 single crowns tested. There was no screw loosening in the control group. There was no effect of screw loosening on the elastic deformation stage. Conclusion. Within the limitations of this study, tightening torque had a significant effect on screw loosening. It would appear that more than 10 Ncm of tightening torque should be recommended for the gold screws in this external hexagon implant system. (J Prosthet Dent 2002;88:402-8.)</description><subject>Crowns</subject><subject>Dental Abutments</subject><subject>Dental Implants</subject><subject>Dental Prosthesis Retention - instrumentation</subject><subject>Dental Prosthesis, Implant-Supported</subject><subject>Dental Restoration Failure</subject><subject>Dental Stress Analysis - methods</subject><subject>Dentistry</subject><subject>Electric Capacitance</subject><subject>Gold Alloys</subject><subject>Signal Processing, Computer-Assisted</subject><subject>Torque</subject><subject>Vibration</subject><issn>0022-3913</issn><issn>1097-6841</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2002</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNqFkMtLAzEQh4MoWh9nb7Inb1uT7GOSo4gvELwIHkN2MpHIPmrSVvrfm9KCJ_EyMzDf_GA-xi4Fnwvews2wiHPJuZwLqbjWB2wmuIayVbU4ZLO8kGWlRXXCTlP65JyrBsQxOxGyrkEKmLH3d7umwo6236SQiskXYVj0dlwWCSN9F_00JRrD-FGs0rbasbAhFrjpw-hiQNvnGfuAmbRuSzhaB6RzduRtn-hi38_Y28P9291T-fL6-Hx3-1JiXfNl2XirOuGaWpLvGg4gpK4raDpEBRUnJT06rx33qqsI68qDbgCb1inqOFZn7HoXu4jT14rS0gwhIfX5A5pWyYAEKVsB_4ISQEMrVQZvdiDGKaVI3ixiGGzcGMHN1rnJzs3Wudk5zxdX--hVN5D75feSM6B3AGUR60DRJAw0IrkQCZfGTeHP8B-vfJEt</recordid><startdate>20021001</startdate><enddate>20021001</enddate><creator>Lee, Juneseok</creator><creator>Kim, Yung-Soo</creator><creator>Kim, Chang-Whe</creator><creator>Han, Jung-Suk</creator><general>Elsevier Inc</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7TB</scope><scope>8BQ</scope><scope>8FD</scope><scope>FR3</scope><scope>JG9</scope><scope>7X8</scope></search><sort><creationdate>20021001</creationdate><title>Wave analysis of implant screw loosening using an air cylindrical cyclic loading device</title><author>Lee, Juneseok ; Kim, Yung-Soo ; Kim, Chang-Whe ; Han, Jung-Suk</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c440t-5fa8b1d542efb50771294375bcc8730e82fcdf9d0f8b3ec43f7957c56d8eb0c3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2002</creationdate><topic>Crowns</topic><topic>Dental Abutments</topic><topic>Dental Implants</topic><topic>Dental Prosthesis Retention - instrumentation</topic><topic>Dental Prosthesis, Implant-Supported</topic><topic>Dental Restoration Failure</topic><topic>Dental Stress Analysis - methods</topic><topic>Dentistry</topic><topic>Electric Capacitance</topic><topic>Gold Alloys</topic><topic>Signal Processing, Computer-Assisted</topic><topic>Torque</topic><topic>Vibration</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Lee, Juneseok</creatorcontrib><creatorcontrib>Kim, Yung-Soo</creatorcontrib><creatorcontrib>Kim, Chang-Whe</creatorcontrib><creatorcontrib>Han, Jung-Suk</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>Mechanical &amp; Transportation Engineering Abstracts</collection><collection>METADEX</collection><collection>Technology Research Database</collection><collection>Engineering Research Database</collection><collection>Materials Research Database</collection><collection>MEDLINE - Academic</collection><jtitle>The Journal of prosthetic dentistry</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Lee, Juneseok</au><au>Kim, Yung-Soo</au><au>Kim, Chang-Whe</au><au>Han, Jung-Suk</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Wave analysis of implant screw loosening using an air cylindrical cyclic loading device</atitle><jtitle>The Journal of prosthetic dentistry</jtitle><addtitle>J Prosthet Dent</addtitle><date>2002-10-01</date><risdate>2002</risdate><volume>88</volume><issue>4</issue><spage>402</spage><epage>408</epage><pages>402-408</pages><issn>0022-3913</issn><eissn>1097-6841</eissn><abstract>Statement of Problem. The mechanics of implant screw loosening or fracture are well understood in the field of engineering. They have not been as widely explored in dentistry. Purpose. This study investigated the effects of simulated mastication on implant components and used wave analysis to document the basic mechanisms of screw loosening in a simulated oral environment. Material and Methods. A pneumatic cylindrical cyclic loading device was fabricated to simulate masticatory movement. Thirteen standard abutments were connected on external hexagonal implants with titanium abutment screws tightened to 20 Newton centimeters (Ncm), and single crowns were retained with gold screws tightened to 10 Ncm on each abutment, respectively. Ten single-implant crowns were loaded with the use of a cyclic loading device with 100 N of force at 30° angles to the long-axis for 0.2 seconds of contact time with a frequency of 1 Hz. Three crowns were loaded vertically under the same conditions to serve as the control group. The effects of up to 1 million cyclic loads and various tightening torque forces (2, 4, 6, 8, 10, and 12 Ncm) on screw loosening were evaluated by wave analysis. A software program was written to record every wave mode and to stop the machine automatically if the amount of horizontal displacement of the crown was more than 0.5 mm, which was designated to represent perceptible loosened implant crown mobility clinically. The general wave patterns and characteristics of loosened and stable screws and the effect of various tightening torques were analyzed by comparing the differences in wave patterns. Results. The wave mode was divided into 4 stages for loosened gold screws: initial displacement, initial vibration, elastic deformation, and recovery stage. However, the initial displacement and initial vibration stages were not discernible for stable gold screws. Of the 10 gold occlusal screws, 4 loosened before the 1 million cyclic loads in the 10 single crowns tested. There was no screw loosening in the control group. There was no effect of screw loosening on the elastic deformation stage. Conclusion. Within the limitations of this study, tightening torque had a significant effect on screw loosening. It would appear that more than 10 Ncm of tightening torque should be recommended for the gold screws in this external hexagon implant system. (J Prosthet Dent 2002;88:402-8.)</abstract><cop>United States</cop><pub>Elsevier Inc</pub><pmid>12447217</pmid><doi>10.1067/mpr.2002.128099</doi><tpages>7</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0022-3913
ispartof The Journal of prosthetic dentistry, 2002-10, Vol.88 (4), p.402-408
issn 0022-3913
1097-6841
language eng
recordid cdi_proquest_miscellaneous_72722617
source MEDLINE; ScienceDirect Journals (5 years ago - present)
subjects Crowns
Dental Abutments
Dental Implants
Dental Prosthesis Retention - instrumentation
Dental Prosthesis, Implant-Supported
Dental Restoration Failure
Dental Stress Analysis - methods
Dentistry
Electric Capacitance
Gold Alloys
Signal Processing, Computer-Assisted
Torque
Vibration
title Wave analysis of implant screw loosening using an air cylindrical cyclic loading device
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-08T02%3A09%3A59IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Wave%20analysis%20of%20implant%20screw%20loosening%20using%20an%20air%20cylindrical%20cyclic%20loading%20device&rft.jtitle=The%20Journal%20of%20prosthetic%20dentistry&rft.au=Lee,%20Juneseok&rft.date=2002-10-01&rft.volume=88&rft.issue=4&rft.spage=402&rft.epage=408&rft.pages=402-408&rft.issn=0022-3913&rft.eissn=1097-6841&rft_id=info:doi/10.1067/mpr.2002.128099&rft_dat=%3Cproquest_cross%3E72722617%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=27797628&rft_id=info:pmid/12447217&rft_els_id=S0022391302002329&rfr_iscdi=true