Direct and indirect catheter-to-catheter carbonization caused by high power radiofrequency delivery via a large tip ablation electrode

We report two cases of directly and indirectly induced catheter carbonizations by radiofrequency application via a large tip ablation catheter. We assume that during a large tip high power delivery of more than 50 watts, an adjacent, smaller-sized, low resistance electrode produces a local increase...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of interventional cardiology 2002-10, Vol.15 (5), p.407-410
Hauptverfasser: Schneider, Michael A E, Zrenner, Bernhard, Karch, Martin R, Weyerbrock, Sonja, Schreieck, Jürgen, Ndrepepa, Gjin, Schömig, Albert, Schmitt, Claus
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We report two cases of directly and indirectly induced catheter carbonizations by radiofrequency application via a large tip ablation catheter. We assume that during a large tip high power delivery of more than 50 watts, an adjacent, smaller-sized, low resistance electrode produces a local increase in the intensity of the current field that is sufficient to elevate the tissue temperature above 100 degrees C. Due to the potential risk of embolism, this may have an impact on ablation procedures in the left atrium and ventricle using similar mapping configurations.