Accuracy and reliability of the self-inflating bulb to verify tracheal intubation in out-of-hospital cardiac arrest patients
To determine the sensitivity and specificity of the self-inflating bulb (SIB) to verify tracheal intubation in out-of-hospital cardiac arrest patients. Sixty-five consecutive adult patients with out-of-hospital cardiac arrest were enrolled. Patients were provided chest compression and ventilation by...
Gespeichert in:
Veröffentlicht in: | Anesthesiology (Philadelphia) 2000-12, Vol.93 (6), p.1432-1436 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 1436 |
---|---|
container_issue | 6 |
container_start_page | 1432 |
container_title | Anesthesiology (Philadelphia) |
container_volume | 93 |
creator | TANIGAWA, Koichi TAKEDA, Taku GOTO, Eiichi TANAKA, Keiichi |
description | To determine the sensitivity and specificity of the self-inflating bulb (SIB) to verify tracheal intubation in out-of-hospital cardiac arrest patients.
Sixty-five consecutive adult patients with out-of-hospital cardiac arrest were enrolled. Patients were provided chest compression and ventilation by either ba-valve-mask or the esophageal tracheal double-lumen airway by ambulance crews when they arrived at the authors' department. Immediately after intubation in the emergency department, the endotracheal tube position was tested by the SIB and end-tidal carbon dioxide (ETCO2) monitor using an infrared carbon dioxide analyzer. We observed the SIB reinflating for 10 s, and full reinflation within 4 s was defined as a positive result (tracheal intubation).
Five esophageal intubations occurred, and the SIB correctly identified all esophageal intubations. Of the 65 tracheal intubations, the SIB correctly identified 47 tubes placed in the trachea (72.3%). Delayed but full reinflation occurred in one tracheal intubation during the 10-s observation period. Fifteen tracheal intubations had incomplete reinflation during the observation period, and two tracheal intubations did not achieve any reinflation. Thirty-nine tracheal intubations were identified by ETCO2 (60%). When the SIB test is combined with the ETCO2 detection, 59 tracheal intubations were identified with a 90.8% sensitivity.
The authors found a high incidence of false-negative results of the SIB in out-of-hospital cardiac arrest patients. Because no single test for verifying endotracheal tube position is reliable, all available modalities should be tested and used in conjunction with proper clinical judgment to verify tracheal intubation in cases of out-of-hospital cardiac arrest. |
doi_str_mv | 10.1097/00000542-200012000-00015 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_72519454</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>72519454</sourcerecordid><originalsourceid>FETCH-LOGICAL-c389t-8e8aa2f7ce2dae996165a1eb049f79bf2f42c01c8c2f892cd2c5f502d77119013</originalsourceid><addsrcrecordid>eNpFkF1rHCEUhqU0dDcff6EIhd6ZepxxRy-X0HxAIDfN9XDG0azFHbfqFBb64-Mm240XnqM-r8pDCAV-DVx3P_hhyFYwUSscJnZo5CeyBCkUA-jkZ7Ksew1ruBALcp7z77rsZKO-kAUAtLpt1JL8WxszJzR7itNIkw0eBx982dPoaNlYmm1wzE8uYPHTCx3mMNAS6V-bvNvTUqMbi4H6qcxDReJUWxrnwqJjm5h3vtRTg2n0aCimZHOhuwraqeRLcuYwZHt1rBfk-fbnr5t79vh093CzfmSmUbowZRWicJ2xYkSr9QpWEsEOvNWu04MTrhWGg1FGOKWFGYWRTnIxdh2A5tBckO_v9-5S_DPXH_Rbn40NAScb59x3QoJuZVtB9Q6aFHNO1vW75LeY9j3w_mC-_2--P5nv38zX6NfjG_OwteNH8Ki6At-OAGaDwSWcjM8nTgndamheAbrhjPg</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>72519454</pqid></control><display><type>article</type><title>Accuracy and reliability of the self-inflating bulb to verify tracheal intubation in out-of-hospital cardiac arrest patients</title><source>MEDLINE</source><source>Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals</source><source>Journals@Ovid Ovid Autoload</source><creator>TANIGAWA, Koichi ; TAKEDA, Taku ; GOTO, Eiichi ; TANAKA, Keiichi</creator><creatorcontrib>TANIGAWA, Koichi ; TAKEDA, Taku ; GOTO, Eiichi ; TANAKA, Keiichi</creatorcontrib><description>To determine the sensitivity and specificity of the self-inflating bulb (SIB) to verify tracheal intubation in out-of-hospital cardiac arrest patients.
Sixty-five consecutive adult patients with out-of-hospital cardiac arrest were enrolled. Patients were provided chest compression and ventilation by either ba-valve-mask or the esophageal tracheal double-lumen airway by ambulance crews when they arrived at the authors' department. Immediately after intubation in the emergency department, the endotracheal tube position was tested by the SIB and end-tidal carbon dioxide (ETCO2) monitor using an infrared carbon dioxide analyzer. We observed the SIB reinflating for 10 s, and full reinflation within 4 s was defined as a positive result (tracheal intubation).
Five esophageal intubations occurred, and the SIB correctly identified all esophageal intubations. Of the 65 tracheal intubations, the SIB correctly identified 47 tubes placed in the trachea (72.3%). Delayed but full reinflation occurred in one tracheal intubation during the 10-s observation period. Fifteen tracheal intubations had incomplete reinflation during the observation period, and two tracheal intubations did not achieve any reinflation. Thirty-nine tracheal intubations were identified by ETCO2 (60%). When the SIB test is combined with the ETCO2 detection, 59 tracheal intubations were identified with a 90.8% sensitivity.
The authors found a high incidence of false-negative results of the SIB in out-of-hospital cardiac arrest patients. Because no single test for verifying endotracheal tube position is reliable, all available modalities should be tested and used in conjunction with proper clinical judgment to verify tracheal intubation in cases of out-of-hospital cardiac arrest.</description><identifier>ISSN: 0003-3022</identifier><identifier>EISSN: 1528-1175</identifier><identifier>DOI: 10.1097/00000542-200012000-00015</identifier><identifier>PMID: 11149438</identifier><identifier>CODEN: ANESAV</identifier><language>eng</language><publisher>Hagerstown, MD: Lippincott</publisher><subject>Anesthesia. Intensive care medicine. Transfusions. Cell therapy and gene therapy ; Biological and medical sciences ; Carbon Dioxide - analysis ; Emergency and intensive respiratory care ; Emergency Medical Services - methods ; Esophagus ; False Negative Reactions ; False Positive Reactions ; Female ; Heart Arrest - therapy ; Humans ; Intensive care medicine ; Intubation ; Intubation, Intratracheal - instrumentation ; Male ; Medical sciences ; Middle Aged ; Prospective Studies ; Reproducibility of Results ; Sensitivity and Specificity</subject><ispartof>Anesthesiology (Philadelphia), 2000-12, Vol.93 (6), p.1432-1436</ispartof><rights>2001 INIST-CNRS</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c389t-8e8aa2f7ce2dae996165a1eb049f79bf2f42c01c8c2f892cd2c5f502d77119013</citedby></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,776,780,27903,27904</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&idt=829491$$DView record in Pascal Francis$$Hfree_for_read</backlink><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/11149438$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>TANIGAWA, Koichi</creatorcontrib><creatorcontrib>TAKEDA, Taku</creatorcontrib><creatorcontrib>GOTO, Eiichi</creatorcontrib><creatorcontrib>TANAKA, Keiichi</creatorcontrib><title>Accuracy and reliability of the self-inflating bulb to verify tracheal intubation in out-of-hospital cardiac arrest patients</title><title>Anesthesiology (Philadelphia)</title><addtitle>Anesthesiology</addtitle><description>To determine the sensitivity and specificity of the self-inflating bulb (SIB) to verify tracheal intubation in out-of-hospital cardiac arrest patients.
Sixty-five consecutive adult patients with out-of-hospital cardiac arrest were enrolled. Patients were provided chest compression and ventilation by either ba-valve-mask or the esophageal tracheal double-lumen airway by ambulance crews when they arrived at the authors' department. Immediately after intubation in the emergency department, the endotracheal tube position was tested by the SIB and end-tidal carbon dioxide (ETCO2) monitor using an infrared carbon dioxide analyzer. We observed the SIB reinflating for 10 s, and full reinflation within 4 s was defined as a positive result (tracheal intubation).
Five esophageal intubations occurred, and the SIB correctly identified all esophageal intubations. Of the 65 tracheal intubations, the SIB correctly identified 47 tubes placed in the trachea (72.3%). Delayed but full reinflation occurred in one tracheal intubation during the 10-s observation period. Fifteen tracheal intubations had incomplete reinflation during the observation period, and two tracheal intubations did not achieve any reinflation. Thirty-nine tracheal intubations were identified by ETCO2 (60%). When the SIB test is combined with the ETCO2 detection, 59 tracheal intubations were identified with a 90.8% sensitivity.
The authors found a high incidence of false-negative results of the SIB in out-of-hospital cardiac arrest patients. Because no single test for verifying endotracheal tube position is reliable, all available modalities should be tested and used in conjunction with proper clinical judgment to verify tracheal intubation in cases of out-of-hospital cardiac arrest.</description><subject>Anesthesia. Intensive care medicine. Transfusions. Cell therapy and gene therapy</subject><subject>Biological and medical sciences</subject><subject>Carbon Dioxide - analysis</subject><subject>Emergency and intensive respiratory care</subject><subject>Emergency Medical Services - methods</subject><subject>Esophagus</subject><subject>False Negative Reactions</subject><subject>False Positive Reactions</subject><subject>Female</subject><subject>Heart Arrest - therapy</subject><subject>Humans</subject><subject>Intensive care medicine</subject><subject>Intubation</subject><subject>Intubation, Intratracheal - instrumentation</subject><subject>Male</subject><subject>Medical sciences</subject><subject>Middle Aged</subject><subject>Prospective Studies</subject><subject>Reproducibility of Results</subject><subject>Sensitivity and Specificity</subject><issn>0003-3022</issn><issn>1528-1175</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2000</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNpFkF1rHCEUhqU0dDcff6EIhd6ZepxxRy-X0HxAIDfN9XDG0azFHbfqFBb64-Mm240XnqM-r8pDCAV-DVx3P_hhyFYwUSscJnZo5CeyBCkUA-jkZ7Ksew1ruBALcp7z77rsZKO-kAUAtLpt1JL8WxszJzR7itNIkw0eBx982dPoaNlYmm1wzE8uYPHTCx3mMNAS6V-bvNvTUqMbi4H6qcxDReJUWxrnwqJjm5h3vtRTg2n0aCimZHOhuwraqeRLcuYwZHt1rBfk-fbnr5t79vh093CzfmSmUbowZRWicJ2xYkSr9QpWEsEOvNWu04MTrhWGg1FGOKWFGYWRTnIxdh2A5tBckO_v9-5S_DPXH_Rbn40NAScb59x3QoJuZVtB9Q6aFHNO1vW75LeY9j3w_mC-_2--P5nv38zX6NfjG_OwteNH8Ki6At-OAGaDwSWcjM8nTgndamheAbrhjPg</recordid><startdate>20001201</startdate><enddate>20001201</enddate><creator>TANIGAWA, Koichi</creator><creator>TAKEDA, Taku</creator><creator>GOTO, Eiichi</creator><creator>TANAKA, Keiichi</creator><general>Lippincott</general><scope>IQODW</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>20001201</creationdate><title>Accuracy and reliability of the self-inflating bulb to verify tracheal intubation in out-of-hospital cardiac arrest patients</title><author>TANIGAWA, Koichi ; TAKEDA, Taku ; GOTO, Eiichi ; TANAKA, Keiichi</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c389t-8e8aa2f7ce2dae996165a1eb049f79bf2f42c01c8c2f892cd2c5f502d77119013</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2000</creationdate><topic>Anesthesia. Intensive care medicine. Transfusions. Cell therapy and gene therapy</topic><topic>Biological and medical sciences</topic><topic>Carbon Dioxide - analysis</topic><topic>Emergency and intensive respiratory care</topic><topic>Emergency Medical Services - methods</topic><topic>Esophagus</topic><topic>False Negative Reactions</topic><topic>False Positive Reactions</topic><topic>Female</topic><topic>Heart Arrest - therapy</topic><topic>Humans</topic><topic>Intensive care medicine</topic><topic>Intubation</topic><topic>Intubation, Intratracheal - instrumentation</topic><topic>Male</topic><topic>Medical sciences</topic><topic>Middle Aged</topic><topic>Prospective Studies</topic><topic>Reproducibility of Results</topic><topic>Sensitivity and Specificity</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>TANIGAWA, Koichi</creatorcontrib><creatorcontrib>TAKEDA, Taku</creatorcontrib><creatorcontrib>GOTO, Eiichi</creatorcontrib><creatorcontrib>TANAKA, Keiichi</creatorcontrib><collection>Pascal-Francis</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Anesthesiology (Philadelphia)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>TANIGAWA, Koichi</au><au>TAKEDA, Taku</au><au>GOTO, Eiichi</au><au>TANAKA, Keiichi</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Accuracy and reliability of the self-inflating bulb to verify tracheal intubation in out-of-hospital cardiac arrest patients</atitle><jtitle>Anesthesiology (Philadelphia)</jtitle><addtitle>Anesthesiology</addtitle><date>2000-12-01</date><risdate>2000</risdate><volume>93</volume><issue>6</issue><spage>1432</spage><epage>1436</epage><pages>1432-1436</pages><issn>0003-3022</issn><eissn>1528-1175</eissn><coden>ANESAV</coden><abstract>To determine the sensitivity and specificity of the self-inflating bulb (SIB) to verify tracheal intubation in out-of-hospital cardiac arrest patients.
Sixty-five consecutive adult patients with out-of-hospital cardiac arrest were enrolled. Patients were provided chest compression and ventilation by either ba-valve-mask or the esophageal tracheal double-lumen airway by ambulance crews when they arrived at the authors' department. Immediately after intubation in the emergency department, the endotracheal tube position was tested by the SIB and end-tidal carbon dioxide (ETCO2) monitor using an infrared carbon dioxide analyzer. We observed the SIB reinflating for 10 s, and full reinflation within 4 s was defined as a positive result (tracheal intubation).
Five esophageal intubations occurred, and the SIB correctly identified all esophageal intubations. Of the 65 tracheal intubations, the SIB correctly identified 47 tubes placed in the trachea (72.3%). Delayed but full reinflation occurred in one tracheal intubation during the 10-s observation period. Fifteen tracheal intubations had incomplete reinflation during the observation period, and two tracheal intubations did not achieve any reinflation. Thirty-nine tracheal intubations were identified by ETCO2 (60%). When the SIB test is combined with the ETCO2 detection, 59 tracheal intubations were identified with a 90.8% sensitivity.
The authors found a high incidence of false-negative results of the SIB in out-of-hospital cardiac arrest patients. Because no single test for verifying endotracheal tube position is reliable, all available modalities should be tested and used in conjunction with proper clinical judgment to verify tracheal intubation in cases of out-of-hospital cardiac arrest.</abstract><cop>Hagerstown, MD</cop><pub>Lippincott</pub><pmid>11149438</pmid><doi>10.1097/00000542-200012000-00015</doi><tpages>5</tpages><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0003-3022 |
ispartof | Anesthesiology (Philadelphia), 2000-12, Vol.93 (6), p.1432-1436 |
issn | 0003-3022 1528-1175 |
language | eng |
recordid | cdi_proquest_miscellaneous_72519454 |
source | MEDLINE; Elektronische Zeitschriftenbibliothek - Frei zugängliche E-Journals; Journals@Ovid Ovid Autoload |
subjects | Anesthesia. Intensive care medicine. Transfusions. Cell therapy and gene therapy Biological and medical sciences Carbon Dioxide - analysis Emergency and intensive respiratory care Emergency Medical Services - methods Esophagus False Negative Reactions False Positive Reactions Female Heart Arrest - therapy Humans Intensive care medicine Intubation Intubation, Intratracheal - instrumentation Male Medical sciences Middle Aged Prospective Studies Reproducibility of Results Sensitivity and Specificity |
title | Accuracy and reliability of the self-inflating bulb to verify tracheal intubation in out-of-hospital cardiac arrest patients |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-25T16%3A24%3A09IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Accuracy%20and%20reliability%20of%20the%20self-inflating%20bulb%20to%20verify%20tracheal%20intubation%20in%20out-of-hospital%20cardiac%20arrest%20patients&rft.jtitle=Anesthesiology%20(Philadelphia)&rft.au=TANIGAWA,%20Koichi&rft.date=2000-12-01&rft.volume=93&rft.issue=6&rft.spage=1432&rft.epage=1436&rft.pages=1432-1436&rft.issn=0003-3022&rft.eissn=1528-1175&rft.coden=ANESAV&rft_id=info:doi/10.1097/00000542-200012000-00015&rft_dat=%3Cproquest_cross%3E72519454%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=72519454&rft_id=info:pmid/11149438&rfr_iscdi=true |