Noninvasive carbon dioxide monitoring during neurosurgical procedures in adults : End-tidal versus transcutaneous techniques
We prospectively compared transcutaneous (TC) versus end-tidal (ET) carbon dioxide monitoring during neurosurgical procedures in adults. After calibration and an equilibration time for the TC-CO2 monitor, arterial blood gas (ABG) values were obtained as clinically indicated. The PaCO2 values were co...
Gespeichert in:
Veröffentlicht in: | Southern medical journal (Birmingham, Ala.) Ala.), 2002-08, Vol.95 (8), p.870-874 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We prospectively compared transcutaneous (TC) versus end-tidal (ET) carbon dioxide monitoring during neurosurgical procedures in adults.
After calibration and an equilibration time for the TC-CO2 monitor, arterial blood gas (ABG) values were obtained as clinically indicated. The PaCO2 values were compared with the values recorded by the noninvasive monitors (TC and ET).
The ET-CO2 to PaCO2 difference was 6.1 +/- 5.6 mm Hg, and the TC-CO2 to PaCO2 difference was 3.7 +/- 2.9 mm Hg. The difference between the PaCO2 and ET-CO2 was 3 mm Hg or less in 17 of 57 values, while the difference between the PaCO2 and TC-CO2 was 3 mm Hg or less in 35 of 57 values. Linear regression analysis of ET-CO2 versus PaCO2 revealed a slope of 0.381 +/- 0.007. Linear regression analysis of TC-CO2 versus PaCO2 revealed a slope of 1.17 +/- 0.008.
Transcutaneous CO2 monitoring provides a more accurate estimate of PaCO2 than ET-CO2 monitoring during neurosurgical procedures. |
---|---|
ISSN: | 0038-4348 1541-8243 |
DOI: | 10.1097/00007611-200208000-00017 |